Empowering Scientific Discovery

Xiamen Longlide Environmental Technology Development Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAPI
OriginUSA
ModelT300
Instrument TypeOnline Gas Analyzer
Detection PrincipleGas Filter Correlation Infrared Absorption (GFC-IR)
Measurement Range0–1 ppm or 0–1000 ppm (selectable dual range)
Repeatability±1% of Full Scale (F.S.)
Resolution0.1 ppm
Response Time (T95)<60 s
Zero Drift<0.1 ppm/24 h, <0.2 ppm/7 days
Span Drift<0.5% of reading/24 h, <1% of reading/7 days
Linearity Error±1% of full scale
Zero Noise (RMS)<0.02 ppm
Span Noise<0.5% of reading
LOD (Limit of Detection)<0.04 ppm
Optical Path Length14 m
Sample Flow Rate800 cm³/min ±10%
Operating Temperature5–40 °C
Relative Humidity0–95% RH (non-condensing)
Pressure & Temperature CompensationYes
Output Signals10 V, 5 V, 1 V, 100 mV, 4–20 mA (isolated), dual RS-232 (bidirectional)
Power Supply100–240 V AC, 50/60 Hz, 250 W
Dimensions (H×W×D)178 × 432 × 597 mm
Weight18 kg
CertificationsU.S. EPA Designated Method, Eignungsgeprüft (German Performance Verification), CE