BrandJWGB
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product CategoryDomestic
ModelJW-M100A True Density Analyzer (Open/Closed Cell Porosity)
Instrument TypeTrue Density Analyzer
Form FactorBenchtop Laboratory Instrument
Sample TypeSolid Materials
Density Range0 g/mL — No Known Upper Limit
Density Accuracy≤ 0.03%
Resolution0.00001 g/mL
Temperature Control Range0–50 °C
Measurement PrincipleGas Displacement (Helium/Nitrogen/Air)
Analysis Stations1
Vacuum SystemIntegrated Oil-Free Diaphragm Pump
Pressure SensorImported High-Stability Sensor, ±0.04% FS Accuracy, Long-Term Drift ≤ 0.025% FS
Thermal ManagementIntegrated Active Temperature Stabilization Module for Sample & Reference Chambers
Sample Chamber Volume Options10–120 mL (6 Standard Aluminum Cells Supplied)
Cycle Time≤ 2 Minutes per Measurement
Data InterfaceFull-Touch LCD Console + USB Export
ComplianceCE Certified, CNAS-Validated Calibration Traceability, Meets ISO 12154, ASTM D5550, and USP <699> Requirements for Solid Density Determination
Added to wishlistRemoved from wishlist 0
Add to compare
BrandJWGB
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product CategoryDomestic
ModelJW-M100A Automated True Density Analyzer
Instrument TypeTrue Density Analyzer
Form FactorBenchtop Laboratory Instrument
Sample TypeSolid Materials
Density Measurement Range0 g/mL to no known upper limit
Density Accuracy≤ 0.03%
Resolution0.00001 g/mL
Temperature Control Range0–50 °C
Minimum Sample Mass5 g
PrincipleGas Displacement (Helium, Nitrogen, or Air)
Analysis Stations1
Pressure Sensor Accuracy±0.04% FS
Long-Term Stability±0.025% FS
Vacuum SourceIntegrated Oil-Free Diaphragm Pump
Sample Cell Volumes10–120 mL (6 standard aluminum cells included)
Cycle Time≤ 2 minutes per analysis
Data InterfaceFull-touch LCD panel with USB export
ComplianceCE certified
Added to wishlistRemoved from wishlist 0
Add to compare
BrandJWGB
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product CategoryDomestic
ModelJW-M100A Foam-Specific True Density and Porosity Analyzer
Instrument TypeTrue Density Analyzer
Form FactorBenchtop Laboratory Instrument
Sample TypeSolid Materials (Including Rigid & Flexible Foams)
Measurement PrincipleGas Displacement Pycnometry (Helium/Nitrogen/Air)
Density Range0 g/mL to No Known Upper Limit
Density Accuracy≤ 0.03% (RSD)
Resolution0.00001 g/mL
Temperature Control Range0–50 °C
Analysis Station1
Vacuum SystemIntegrated Oil-Free Diaphragm Pump
Pressure SensorImported High-Stability Transducer, 0.04% FS Accuracy, Long-Term Drift ≤ 0.025% FS
Pressure RegulationFully Automated, Max Pressure ≤ Atmospheric
Sample Chamber Volume Options10–120 mL (6 Standard Aluminum Cells Supplied)
Cycle Time≤ 2 Minutes per Measurement
Data InterfaceFull-Touch LCD Console + USB Export
ComplianceCE Certified, Meets JJG 99–2006 (Chinese National Metrology Verification Regulation), Supports GLP Audit Trail via Exported Timestamped Logs
Added to wishlistRemoved from wishlist 0
Add to compare
BrandJWGB
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product CategoryDomestic
ModelJW-M100A True Density Analyzer for Ceramic Proppants and Quartz Sand
Instrument TypeTrue Density Analyzer
Form FactorBenchtop Laboratory Instrument
Sample TypeSolid Materials
Density Measurement Range0 g/mL — No Upper Limit
Density Accuracy≤ 0.03% (RSD)
Resolution0.00001 g/mL
Temperature Control Range0–50 °C
Measurement PrincipleGas Displacement (Helium/Nitrogen/Air)
Analysis Stations1
Vacuum SystemIntegrated Oil-Free Diaphragm Pump
Pressure SensorImported High-Stability Sensor, ±0.04% FS Accuracy, Long-Term Drift ≤ 0.025% FS
Thermal ManagementIntegrated Precision Isothermal Module for Sample & Reference Chambers
Sample Chamber Volume Options10–120 mL (6 Standard Aluminum Cells Supplied)
Data InterfaceFull-Touch LCD Console + USB Export
ComplianceCE Certified
CalibrationIncludes Certified Reference Standards
Testing Time per Sample≤ 2 Minutes
Porosity ModesOpen-Cell & Closed-Cell Porosity Calculation
SoftwareEmbedded Firmware with Audit-Trail Capable Data Logging (GLP-Compatible)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandJWGB
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Product CategoryDomestic
ModelJW-M100A Fracturing Proppant-Specific True Density Analyzer
Instrument TypeTrue Density Analyzer
Form FactorBenchtop Laboratory Instrument
Sample TypeSolid Materials
Measurement PrincipleGas Displacement (Helium Pycnometry)
Density Range0 g/mL to no known upper limit
Density Accuracy≤ 0.03% RSD
Resolution0.00001 g/mL
Temperature Control Range0–50 °C
Analysis Stations1
Sample Chamber Volumes10–120 mL (6 standard aluminum cells included)
Vacuum SystemIntegrated oil-free diaphragm pump
Pressure SensorImported high-stability sensor, 0.04% FS accuracy, long-term stability ≤ 0.025% FS
Pressure ControlFully automated, peak pressure ≤ atmospheric
Data InterfaceBuilt-in full-touch LCD control panel with USB export capability
ComplianceCE certified