| Brand | Dynamica |
|---|---|
| Model | HALO MPR-96 |
| Origin | Switzerland |
| Wavelength Range | 400–750 nm (MPR-96UV: 340–750 nm) |
| Absorbance Range | 0–4.0 OD |
| Resolution | 0.001 OD |
| Stability | ±0.001 OD (after 15 min warm-up) |
| Accuracy | ±1.0% ±0.01 OD (405 nm, 0.000–2.500 OD) |
| Repeatability | ±0.5% ±0.005 OD (405 nm, 0.000–2.500 OD) |
| Linearity | ±1.0% (405 nm, 0.000–2.500 OD) |
| Filter Positions | 5 (4 fixed + 1 user-selectable, 400–750 nm) |
| Read Time | ≤10 s (single wavelength, 96-well) |
| Plate Formats | 96-well flat, U-bottom, V-bottom |
| Optical Channels | 8 measurement + 1 reference |
| Detector | Silicon photodiode |
| Light Source | Tungsten-halogen lamp |
| Display | Backlit LCD, 320×240 px |
| Incubation (optional) | Ambient +4°C to +50°C, ±0.5°C accuracy |
| Dimensions (W×D×H) | 290×425×200 mm |
| Weight | 15 kg (net), 20 kg (gross) |
| Power | 110–220 V auto-switching, 50/60 Hz, 50 W (standby), 100 W (operational) |