Sabio
Filter
Showing all 12 results
| Brand | Sabio |
|---|---|
| Origin | Beijing, China |
| Model | 1001 |
| Output Pressure | 25 psi |
| Standard Flow Rate | 22 L/min (SLP) |
| Condensate Tank Capacity | 3.4 L |
| Air Purity | NO/NO₂/O₃/H₂S/SO₂ < 0.5 ppb, CO < 25 ppb, HC < 200 ppb |
| Dew Point | ≤ −20 °C at 15 L/min |
| Dimensions | 222 mm (H) × 412 mm (W) × 565 mm (D) |
| Power Supply | 220 VAC, 50 Hz |
| Compressor Type | Brushless, Oil-Free |
| Condensate Drain | Automatic or Manual |
| Optional | Built-in CO oxidation catalyst, HC oxidation catalyst, pressure switch, flow meter, rack-mount kit, LED dew point display |
| Brand | Sabio |
|---|---|
| Origin | Beijing, China |
| Model | 2020 |
| Flow Rate | 10 SLPM at 20 psig |
| Purification Targets | NO, NO₂, O₃, SO₂, H₂S < 0.5 ppb (standard) |
| Dew Point | < –40 °C (via integrated permeation dryer) |
| Power Supply | 196–264 VAC, 50–60 Hz |
| Analog Output | 4–20 mA (max. load 750 Ω) |
| Dimensions | 165 mm (H) × 353 mm (W) × 381 mm (D) |
| Weight | 10.4 kg |
| Compliance | Designed for ISO 8573-1 Class 1 compressed air purity |
| Brand | Sabio |
|---|---|
| Model | 2030 |
| Measured Species | O₃ |
| Measurement Principle | UV Absorption Spectrophotometry |
| Range Options | 0–100 ppb, 0–200 ppb, 0–500 ppb, 0–1 ppm, 0–2 ppm |
| Zero Noise | <0.6 ppb |
| Limit of Detection | <1 ppb |
| Zero Drift | <1 ppb/24 h or <2 ppb/7 days |
| Span Drift | <1% per month |
| Response Time (T90) | 20 s (10 s lag) |
| Accuracy | ±1 ppb |
| Linearity | <1% FS |
| Sample Flow Rate | 1.0 LPM |
| Operating Temperature | 5–45 °C |
| Analog Outputs | 0–0.1 V, 0–1 V, 0–2 V, 0–5 V, 0–10 V (user-configurable, 4 channels) |
| Power Supply | 196–264 VAC, 50–60 Hz |
| Integrated High-Precision Ozone Generator | Output Range 0.05–1.5 ppm at 5 SLM, Accuracy ±2% of setpoint or ±3.0 ppb at 5 SLM |
| Environmental Compensation | Automatic temperature and pressure compensation |
| Connectivity | Ethernet (TCP/IP), USB, RS-232/RS-485 |
| Data Logging | Onboard non-volatile memory with timestamped records |
| Display | Large backlit TFT LCD with touch interface |
| Physical Design | Portable chassis with integrated side handles, ruggedized instrument case, built-in zero-air source |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Model | 4010 |
| Flow Accuracy | ±1.0% of Full Scale |
| Flow Repeatability | ±0.15% of Full Scale |
| Flow Linearity | ±0.5% of Full Scale |
| Diluent MFC Range | 0–10 SLPM (optional 0–20 SLPM) |
| Source MFC Range | 0–100 SCCM (optional 0–50 / 0–200 / 0–500 SCCM / 0–1 SLM) |
| Input Pressure | 20–35 psig |
| Ozone Output Range | 0.05–1.5 ppm at 5 SLPM (extendable to 0.05–1.0 ppm at 15 SLPM) |
| Ozone Accuracy | ±2% of setpoint or ±3 ppb at 5 SLPM |
| Ozone Flow Control Precision | ±1 SCCM at 100 SCCM |
| 95% Setpoint Response Time | <60 s |
| Wetted Materials | 316 Stainless Steel, PTFE, PEEK, Halar |
| Seals | Viton, Kalrez |
| Optional Internal Permeation Oven | 94 cm³ Volume, 13.2 cm Length, 3.02 cm ID, Temperature Stability: ±0.1°C (setpoint +60°C above ambient), Preheat Time: <30 min |
| Brand | Sabio |
|---|---|
| Origin | Beijing, China |
| Model | 4010L |
| Diluent Flow Range | 0–10 SLM (optional up to 20 SLM) |
| Source Gas Flow Range | 0–100 SCCM / 0–10,000 SCCM |
| Flow Accuracy | ≤ ±1% FS |
| Flow Repeatability | ≤ ±0.15% FS |
| Linearity | ≤ ±0.5% FS |
| Inlet Pressure | 20–35 PSIG |
| Ozone Generator Output | 0.05–1.5 ppm at 5 SLPM |
| Ozone Generator Accuracy | ±1% of setpoint or ±2 ppb at 5 SLPM |
| UV Absorption Photometer Range | 0–0.5 / 1.0 / 2.0 ppm |
| Photometer Zero Drift | < ±1 ppb per 24 h (or 30 days) |
| Photometer Flow Rate | 1.0 LPM |
| Dilution Furnace Temperature Range | Ambient +5 °C to 65 °C |
| Temperature Stability | < ±0.1 °C |
| Calibration Programs | 20 user-configurable, time- and event-driven (7-day cycle support) |
| Input/Output | 24-bit digital I/O, 2× serial ports (RS232 standard), parallel printer port, external keyboard interface, air source control port |
| Display | High-brightness color LCD |
| Keypad | 55-key membrane panel |
| Gas Ports | 4 source gas inlets, 1 purge inlet, 1 diluent inlet, 5 calibrated outputs, 1 exhaust port |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | 1001 |
| Output Flow | 22 SLPM at 25 psi |
| Condensate Tank Capacity | 3.4 L |
| Compression Technology | Brushless, Oil-Free Compressor |
| Purification Media | Catalytic beds for NO, NO₂, SO₂, H₂S, and O₃ removal |
| Zero Air Purity | NO, NO₂, O₃, H₂S, SO₂ < 0.5 ppb |
| Optional Methane Catalytic Furnace (Model 1000M) | Microprocessor-controlled proportional heating |
| Dimensions | 13.3 cm H × 48.3 cm W × 38.1 cm D |
| Input Power | 98–132 VAC (optional 196–264 VAC), 360 VA, 50/60 Hz |
| Post-Catalysis Purity | CO < 5 ppb, HC < 0.20 ppb |
| Model 2020 Variant | 10 SLPM at 20 psig |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Model 2010 |
| Pricing | Upon Request |
| Diluent Mass Flow Range | 0–10 SLM (optional up to 20 SLM) |
| Source Gas Mass Flow Range | 0–100 SCCM (optional up to 2 SLM) |
| Flow Accuracy | ≤1% FS |
| Flow Repeatability | ≤±0.15% FS |
| Linearity | ≤±0.5% FS |
| Response Time | <60 s |
| Control Interfaces | Keypad, Keyboard, RS-232 Serial, USB, Ethernet |
| Predefined Calibration Paths | 20 factory-configured flow paths, each supporting up to 20 linear calibration points |
| Supported Calibration Types | Diluent gas, Ozone (O₃), Gas Phase Titration (GPT) |
| Temperature Control System | Microprocessor-based proportional PID controller |
| Temperature Sensor | Platinum RTD (Pt100) |
| Heating Zone Dimensions | Internal Diameter 3.09 cm, Length 12.7 cm |
| Temperature Range | Ambient +2 °C to 65 °C |
| Front Panel Integration | Adjustable pressure regulator, analog pressure gauge, digital temperature controller, Teflon-lined dilution furnace inlet |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Model 2030 |
| Price | Upon Request |
| Measurement Principle | UV Absorption at 254 nm (Beer–Lambert Law) |
| Range Options | 0–100 / 0–200 / 0–500 ppb or 0–1 / 0–2 ppm |
| Detection Limit | 1 ppb |
| Zero Noise | 0.6 ppb RMS |
| Zero Drift | <1.0 ppb/24 hr, <2.0 ppb/7 days |
| Span Drift | <1% per month |
| Response Time | 20 s (10 s delay) |
| Accuracy | ±1 ppb |
| Linearity | ±1% of full scale |
| Sample Flow Rate | 1 L/min |
| Operating Temperature | 5–45 °C |
| Power Supply | 98–132 VAC / 196–264 VAC, 50/60 Hz |
| Analog Outputs | Four independent, scalable (0.1 V / 1 V / 2 V / 5 V / 10 V) |
| Digital Interfaces | USB, RS-232/RS-485 (DB-9) |
| Display | Color LCD with touchscreen, front-panel keys, external keyboard/mouse support |
| Data Storage | Internal logging with timestamped records |
| Built-in Ozone Generator | 0.05–1.5 ppm @ 5 SLM, ±2% of setpoint or ±3 ppb @ 5 SLM |
| Auto Zero Compensation | Yes |
| Temperature & Pressure Compensation | Yes |
| Dimensions | 15.6 cm H × 36.4 cm W × 31.6 cm D |
| Weight | 9.1 kg |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Import Status | Imported |
| Model | Model 6020 |
| Price | Upon Request |
| Measurement Principle | Pulsed UV Fluorescence (PUVF) |
| Range | 0–50 ppb to 0–20 ppm (user-selectable) |
| Detection Limit | < 0.4 ppb |
| Noise | < 0.2 ppb |
| Zero Drift | < 0.5 ppb/24 hr |
| Accuracy | < 1% Full Scale |
| Sample Flow Rate | 0.4–0.8 L/min |
| Operating Temperature | 5–40 °C |
| Power Consumption | 200 W |
| Dimensions (H×W×D) | 133 × 432 × 571.5 mm |
| Weight | 11.3 kg |
| Display | High-resolution color TFT LCD with touchscreen support |
| Communication Interfaces | Ethernet (TCP/IP), USB 2.0, RS-232/485 |
| Analog I/O | 4 configurable voltage channels (0.1 V, 1 V, 2 V, 5 V, 10 V output ranges) |
| Digital I/O | 8 standard TTL-compatible channels |
| Data Logging | Internal non-volatile memory with timestamped records |
| Compliance | EPA Reference Method EQOA-0808-191 (equivalent to EPA Method 40 CFR Part 53, Class I) |
| Software | Menu-driven embedded OS with diagnostic suite, barcode-enabled time-series visualization, Modbus RTU/TCP support |
| Environmental Compensation | Automatic temperature and barometric pressure correction |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Import Status | Imported |
| Model | 6040 |
| Price | Upon Request |
| Measurement Principle | Chemiluminescence Detection |
| Range | 0–50 ppb to 0–20 ppm (user-selectable) |
| Units | ppm, ppb |
| Display | High-resolution color TFT LCD |
| Interface Options | Touchscreen, front-panel keypad, external keyboard/mouse |
| Data Logging | Internal non-volatile memory with timestamping |
| Communication | Ethernet (TCP/IP), USB 2.0 (host & device), RS-232/485 |
| Analog I/O | 4 channels, configurable voltage output (0.1 V, 1 V, 2 V, 5 V, 10 V) |
| Digital I/O | 8 standard TTL-compatible inputs/outputs |
| Environmental Compensation | Automatic temperature and pressure correction |
| EPA Certification | RFNA-0418-250 |
| Detection Limit | < 0.4 ppb |
| Noise | < 0.2 ppb |
| Zero Drift | < 0.5 ppb/24 hr |
| Accuracy | < 0.5% of full scale |
| Sample Flow Rate | 0.4–0.8 L/min |
| Operating Temperature | 5–45 °C (EPA-specified range) |
| Operating Humidity | 0–90% RH (non-condensing) |
| Power Requirement | 200 W, 100–240 VAC, 50/60 Hz |
| Dimensions (H×W×D) | 133 × 432 × 571.5 mm |
| Weight | 11.3 kg (25 lbs) |
| Protocols | Modbus RTU/TCP, SCPI-compliant command set |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Model | Model 6050 |
| Detection Principle | Non-Dispersive Infrared (NDIR) with Gas Filter Correlation (GFC) |
| Measurement Range | 0–200 ppm (user-selectable) |
| Resolution | 0.01 ppm |
| Units | ppm, ppb |
| Display | 7-inch high-resolution color TFT LCD (touch-enabled) |
| Communication Interfaces | Ethernet (TCP/IP), USB 2.0 (front-panel & rear), RS-232/485 |
| Analog I/O | 4-channel configurable voltage/current inputs & outputs |
| Digital I/O | 8-channel opto-isolated inputs/outputs |
| Environmental Compensation | Automatic temperature and barometric pressure compensation |
| Data Logging | Internal non-volatile memory (≥30 days at 1-min intervals) |
| Compliance | Meets EPA Method 40 CFR Part 53 reference equivalent performance criteria for CO analyzers |
| Software | Embedded menu-driven firmware with diagnostic suite, barcode-enabled time-series visualization, and remote configuration via web interface |
| Brand | Sabio |
|---|---|
| Origin | Beijing, China |
| Model | Sabio2010 |
| Diluent Flow Range | 0–10 SLM (up to 20 SLM optional) |
| Source Gas Flow Range | 0–100 SCCM (up to 2 SLM optional) |
| Flow Accuracy | <1% FS |
| Flow Repeatability | ±0.15% FS |
| Linearity | ±0.5% FS |
| Inlet Pressure | 12–35 PSIG |
| Operating Temperature | 5–40°C |
| Dimensions | 156 mm (H) × 364 mm (W) × 316 mm (D) |
| Weight | 8.6 kg |
| Power Supply | 98–264 VAC, 100 VA, 50/60 Hz |
| Communication Interfaces | RS-232, Ethernet (TCP/IP optional) |
| Digital I/O | 8-bit TTL/contact-triggered |
| Calibration Paths | 20 preconfigured, each supporting up to 20 linear points |
| Gas Configuration | 1 diluent + 20 source gases |
| Output Ports | 3 calibrated outputs + 1 exhaust port (1 permeation port optional) |
| Input Interface | 42-key membrane keypad or external keyboard |