| Origin | Jilin, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | WKW-300 |
| Pricing | Available Upon Request |
| Temperature Range | Ambient to 300 °C |
| Temperature Accuracy | ±0.5 °C |
| Heating Rates | 5 ±0.5 °C/6 min (Method A), 12 ±1.0 °C/6 min (Method B) |
| HDT Deformation Range | 0–0.5 mm |
| Vicat Penetration Range | 0–1 mm |
| Dial Indicator Accuracy | ±0.005 mm |
| Vicat Needle | 3 mm length, cross-sectional area 1.000 ±0.015 mm² |
| HDT Probe Fillet Radius | 3.0 ±0.2 mm |
| Max HDT Load | 5 kgf (≈49.0 N) |
| Vicat Loads | GA = 10 N ±0.2 N, GB = 50 N ±1 N |
| Max Heating Power | ≤4000 W |
| Electrical Supply | 220 V, 50 Hz, 30 A |
| HDT Span Lengths | 64 mm and 100 mm |
| Accuracy Class | Class 1 |
| Cooling Options | Natural convection or water-cooled |
| Over-Temperature Protection | Programmable upper limit |
| Compatible Heat Transfer Medium | Methyl silicone oil (flash point ≥300 °C, kinematic viscosity ≈200 cSt) |
| Brand | Changchun Intelligent |
|---|---|
| Origin | Jilin, China |
| Model | WKW-300 |
| Temperature Range | Ambient to 300 °C |
| Temperature Accuracy | ±0.5 °C |
| Heating Rates | 5 ±0.5 °C/6 min (Method A), 12 ±1.0 °C/6 min (Method B) |
| Deformation Measurement Range | 0–1 mm |
| Deformation Accuracy | ±0.005 mm |
| Dial Indicator Resolution | 0.001 mm (±0.003 mm accuracy) |
| Test Loads | 10 N ±0.2 N (VST Method A), 50 N ±1 N (VST Method B) |
| Bending Stress Levels | 0.45 MPa (HDT Method B), 1.80 MPa (HDT Method A), 8.00 MPa (HDT Method C) |
| Span Lengths | 64 mm and 100 mm |
| Heating Medium | Methyl silicone oil (kinematic viscosity ≤200 cSt, flash point ≥300 °C) |
| Cooling Options | Natural convection or water cooling |
| Power Supply | AC 220 V, 50 Hz |
| Rated Power | 4 kW |
| Accuracy Class | Class 1 (per GB/T 1633 & ISO 306) |
| Brand | Changchun Intelligent |
|---|---|
| Origin | Jilin, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | LDS-3000 |
| Load Cell Capacity | 3000 N |
| Load Accuracy | ±1% FS |
| Load Resolution | 0.5 N |
| Test Speed Range | 5–300 mm/min |
| Speed Accuracy | ±5% |
| Effective Stroke | 300 mm |
| Displacement Measurement Range | 0–300 mm |
| Displacement Accuracy | ±1% FS |
| Platen Dimensions | 310 × 310 mm |
Show next