Hangzhou Chromatogram Chromatography Technology Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandChrom
ModelGC-80PDD on-line
Detection PrinciplePulsed Discharge Helium Ionization Detection (PDHID)
Column Temperature RangeUp to 420°C
Pressure ControlUp to 100 psi
Split RatioUp to 999.9:1
Baseline Noise≤0.02 mV
Baseline Drift≤0.5 mV/30 min
Repeatability (CH₄)RSD ≤3.0%
Detection Limit (CH₄)≤3.5×10⁻¹³ g/mL
Power Consumption<2.0 kW
Temperature Control Accuracy±0.1°C
Output Interfaces4–20 mA (up to 8 channels), RS232/RS485, Ethernet
Enclosure19″ rack-mount, 6U standard chassis
Carrier GasUltra-high-purity helium
ComplianceDesigned for continuous industrial process monitoring per ISO/IEC 17025-aligned operation and compatible with DCS integration under IEC 61511 functional safety context
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAgilent Technologies
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Product OriginDomestic (China)
ModelAgilent 8890
PricingUpon Request
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAgilent Technologies
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Regional ClassificationDomestic (China)
ModelAgilent 8890
QuotationUpon Request
DetectorDual-Pulse Helium Ionization Detector (PDHID)
Column Oven Temp Range40 °C above ambient to 450 °C
Temp Setpoint Resolution1 °C
Ramp Rate0.1–120 °C/min
Temp Stability<0.01 °C per 1 °C ambient fluctuation
Programmable Temperature Steps20 ramps / 21 platforms
Max Run Time999.99 min
Cool-down≤6 min (450 °C → 50 °C at 22 °C ambient)
Detection Limits5–10 ppb for H₂, O₂/Ar, N₂, CH₄, CO, CO₂, C₂–C₄ hydrocarbons, SiH₃Cl, Si₂H₆, and other silicon-containing impurities in electronic-grade silane (≥99.9995% purity)
ComplianceGB/T 28726–2012, JJG 700–2016, GB/T 15909–2017
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAgilent Technologies
OriginZhejiang, China
ModelAgilent 8860
Detector TypePulsed Discharge Helium Ionization Detector (PDHID)
Column Oven Temp Range8 °C above ambient to 425 °C
Max Ramp Rate75 °C/min
Cooling Rate300 °C → 50 °C in ≤5.7 min (at 25 °C ambient)
Temperature Resolution0.1 °C
Max Run Time999.99 min
PDHID Detection Limit≤10 ppb (as CH₄)
Linear Dynamic Range10⁵
Measurable ImpuritiesH₂, O₂/Ar, N₂, CH₄, CO, CO₂, C₂H₆
ComplianceGB/T 28726–2012, JJG 700–2016, GB/T 8979–2008, GB/T 4842–2006, GB/T 4844.3–2005, GB/T 3634.2–2011, GB/T 6052–2011
Added to wishlistRemoved from wishlist 0
Add to compare
BrandChrom
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Regional ClassificationDomestic (PRC)
ModelGC-126N PDD
Instrument TypeLaboratory Gas Chromatograph
ApplicationUltra-High-Purity (UHP) and Electronic Specialty Gases Analysis
Oven Temperature Range+5 °C above ambient to 400 °C
Max. Ramp Rate60 °C/min
Cooling Rate350 °C → 50 °C in ≤9 min
Carrier Gas Flow Control (He)0–500 mL/min
Carrier Gas Pressure Control (He)0–0.8 MPa
Injector Max. Temp.400 °C
Injector Pressure Setting Range0–1 MPa
Injector Total Flow Setting Range0–200 mL/min
Added to wishlistRemoved from wishlist 0
Add to compare
[BrandChrom
OriginZhejiang, China
ModelGC-5190PDD
Instrument TypeLaboratory Gas Chromatograph
ApplicationUltra-High-Purity Gas Analysis
Oven Temperature Range4°C above ambient to 450°C
Max. Ramp Rate60°C/min
Carrier Gas Flow Control Range30–35 mL/min
Carrier Gas Pressure Control Range0.35–0.55 MPa
Injector Max. Temp.450°C
Injector Pressure Setting Range0–100 psi
Injector Total Flow Setting Range0–1000 mL/min
PDHID Detection Limit (CH₄)≤3.5×10⁻¹³ g/mL
PDHID Dynamic Linear Range≥10⁷
PDHID Noise<0.1 pA
PDHID Drift≤1.0 pA/30 min
PDHID Repeatability (CH₄)≤3.0% RSD
Column Oven Temp. Stability±0.02°C
Programmable Temperature Steps9 ramps / 10 plateaus
Max. Program Duration9999.9 min]
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAgilent Technologies
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Regional ClassificationDomestic (China)
ModelAgilent 8860
Instrument TypeLaboratory Gas Chromatograph
Detection PrinciplePulsed Discharge Helium Ionization Detection (PDHID)
Column Oven Dimensions28.0 × 30.5 × 16.5 cm
Operating Temperature Range8 °C above ambient to 425 °C
Temperature Setpoint Resolution0.1 °C
Max Heating Rate75 °C/min
Cooling Rate300 °C → 50 °C in ≤5.7 min (at 25 °C ambient)
Max Run Time999.99 min
Temperature ProgrammingUp to 20 ramps (including negative ramps)
Ambient Temperature Rejection<0.01 °C per 1 °C ambient change
Analog OutputsTwo channels (1 V and 10 V standard)
Remote ControlStart/Stop via digital I/O
ConnectivityEthernet (LAN)
Flow Selection ValveBCD-encoded input
Operating Environment15–35 °C, 5–90% RH (non-condensing)
Storage Range−40 to 70 °C
Max Operating Altitude4600 m
PDHID Detection Limit≤10 ppb (as CH₄)
Linear Dynamic Range10⁵
Measurable Impurities in HBrH₂, O₂, N₂, CH₄, CO, CO₂, C₂H₆, H₂O, HCl
Compliance StandardsGB/T 28726–2012, JJG 700–2016
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAgilent Technologies
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Regional ClassificationDomestic (PRC)
ModelAgilent 8860
Instrument TypeLaboratory Gas Chromatograph
Application DomainUltra-High-Purity (UHP) Electronic Specialty Gas Analysis
Oven Temperature Range8 °C above ambient to 425 °C
Maximum Ramp Rate75 °C/min
Cooling Rate300 °C → 50 °C in ≤5.7 min (at 25 °C ambient)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandChrom
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Product CategoryDomestic
ModelPSANA 6850
Instrument TypeOnline Gas Chromatograph
Application FieldHigh-Purity Gas Analysis
Oven Temperature Range15 °C to 400 °C
Oven Ramp Rate0.1–35 °C/min
Oven Cool-Down2500 °C to 35 °C in ~10 min
Carrier Gas Flow Range & Control0–350 mL/min
Carrier Gas Pressure Range & Control0–500 kPa
Injector Maximum Operating Temperature400 °C
Injector Pressure Setpoint Range0–500 kPa
Injector Total Flow Setpoint Range0–500 mL/min
DetectorPulsed Discharge Helium Ionization Detector (PDHID)
PDHID Temperature Stability±0.1 °C
Baseline Noise≤0.5 pA
Baseline Drift≤1.0 pA/30 min
Repeatability (CH₄)RSD ≤3.0%
Detection Limit (CH₄)≤3.5×10⁻¹³ g/mL
EPC Pressure Accuracy±0.01 psi
EPC Flow Sensor Accuracy<±5%
Detector EPC Module Accuracy±8% of setpoint
Analog Output4–20 mA (8–12 channels)
Digital InterfaceRS232 / RS485 / Ethernet
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAgilent Technologies
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Regional CategoryDomestic (China)
ModelAgilent 8890 PDD
PricingAvailable Upon Request
Detection PrinciplePulsed Discharge Helium Ionization Detection (PDHID)
Column Oven Range40 °C above ambient to 450 °C
Temperature Stability< 0.01 °C per 1 °C ambient fluctuation
Max Ramp Rate120 °C/min
Cooling Time≤6 min (450 °C → 50 °C at 22 °C ambient)
LOD5–25 ppb (gas-dependent)
ComplianceGB/T 28726-2012, JJG 700-2016, GB/T 8979-2008, GB/T 4842-2006, GB/T 4844.3-2005, GB/T 3634.2-2011, GB/T 6052-2011
Added to wishlistRemoved from wishlist 0
Add to compare
BrandAgilent Technologies
OriginZhejiang, China
Manufacturer TypeAuthorized Distributor
Instrument TypeLaboratory Gas Chromatograph
ApplicationHigh-Purity Specialty Gas Analysis
Oven Temperature RangeAmbient + 5 °C to 400 °C
Maximum Ramp Rate60 °C/min
Cooling Rate≤9 min (350 °C → 50 °C)
Carrier Gas (He) Flow Range0–500 mL/min
Carrier Gas (He) Pressure Range0–500 kPa
Injector Max Temp400 °C
Injector Pressure Setting Range (He)0–100 kPa
Injector Total Flow Setting Range (He)0–500 mL/min
Show next