Empowering Scientific Discovery

Shanghai Hogon Scientific Instrument Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHOGON
OriginShanghai, China
Manufacturer TypeDirect Manufacturer
Instrument TypeFlame–Graphite Furnace Hybrid AAS
MonochromatorFixed-grating with selectable slits
Optical SystemDouble-beam
DetectorSolid-state CCD array
Background CorrectionSimultaneous D₂-lamp and Self-reversal (SR) correction
Spectral Bandwidth0.1, 0.2, 0.4, 0.7, 1.0, 2.0 nm (auto-switching)
Wavelength Range190–900 nm
Wavelength Accuracy±0.2 nm
Wavelength Repeatability0.1 nm
Resolution0.2 nm
Cu Detection Limit (Flame)≤0.004 µg/mL
Cu Characteristic Concentration (Flame)≤0.025 µg/mL per 1% absorption
Precision (RSD, Cu, A > 0.8)≤0.5%
Baseline Drift≤0.002 A / 30 min
Graphite Furnace Max Ramp Rate≥3000 °C/s
Graphite Furnace Temp RangeAmbient to 3000 °C
Cd Characteristic Mass≤1 pg
Cu Characteristic Mass≤10 pg
Graphite Furnace Precision (RSD)≤3% (Cd, Cu)
Lamp Turret6-position automatic
Auto-sampler Options45- or 85-position
Gas ControlDual-path (in-tube / out-tube) inert gas regulation
Safety SystemsReal-time flame status monitoring, gas pressure sensors (air, acetylene, N₂O), auto-shutdown on flame failure/flashback, power-interruption cutoff, overtemperature/coolant-flow interlock
Dimensions700 × 550 × 440 mm
Weight80 kg
Power Requirement220 V AC, 6000 W nominal, transient peak up to 8 kW
Show next
InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0