Empowering Scientific Discovery

Wuhan Corrtest Instruments Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCorrTest
OriginHubei, China
Manufacturer TypeAuthorized Distributor
Product OriginDomestic (China)
ModelCS310X
Price RangeUSD 14,000–28,000
Instrument TypeElectrochemical Workstation
Current Range±200 mA (8-channel mode) / ±500 mA (4-channel mode)
Current Accuracy0.1% of full-scale reading
Potential Accuracy0.1% of full-scale reading ±1 mV
Potentiostatic Range±10 V
EIS Frequency Range10 mHz to 1 MHz
Input Impedance (Reference Electrode)10¹² Ω
Channel Isolation Resistance>100 MΩ
Data Acquisition16-bit @ 1 MHz, 20-bit @ 1 kHz
ADC Resolution16-bit DAC with 1 ms settling time
Low-Pass Filter8 programmable stages
Communication InterfaceEthernet
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCorrTest
ModelCS300M
Current Range±2 A
Current Accuracy±0.1% of full-scale reading
Potential Accuracy±0.1% of full-scale reading ±1 mV
Potentiostatic Range±10 V
Potentiodynamic Scan Rate (CV/LSV)0.001 mV/s to 10 V/s
Current Sensitivity1 pA
Potential Resolution3 mV (<10 Hz), 10 mV (>100 Hz)
EIS Frequency RangeNot specified
AD Acquisition16-bit @ 1 MHz, 20-bit @ 1 kHz
DA Resolution16-bit
Rise Time<1 ms (<10 mA), <10 ms (<2 A)
Reference Electrode Input Impedance10¹² Ω
Voltage Compliance±21 V
Communication InterfaceUSB 2.0
Low-Pass Filter8-stage programmable
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCorrTest
OriginHubei, China
ModelCS350M
Current Range±2.0 A
Current Accuracy0.1% of full-scale reading
Potential Accuracy0.1% of full-scale reading ±1 mV
Potentiostatic Range±10 V
EIS Frequency Range10 µHz to 1 MHz
ADC Resolution16-bit @ 1 MHz, 20-bit @ 1 kHz
DAC Resolution16-bit
Input Impedance (Reference Electrode)10¹² Ω
Minimum Current Sensitivity1 pA
CV/LSV Scan Rate0.001 mV/s to 10 V/s
Pulse Width (CA/CC)0.0001 s to 65,000 s
EIS Signal Amplitude1–2500 mV RMS
DC Bias Range–10 V to +10 V
Communication InterfaceUSB 2.0
Low-Pass Filter8 programmable bandwidth settings
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCorrTest
OriginHubei, China
ModelCS310M
Current Range±2.0 A
Current Accuracy0.1% of full-scale reading
Potential Accuracy0.1% of full-scale reading ±1 mV
Potentiostatic Range±10 V
EIS Frequency Range10 µHz to 1 MHz
Potential Resolution3 mV (<10 Hz), 10 mV (>100 Hz)
Current Sensitivity1 pA
Rise Time<1 ms (<10 mA), <10 ms (<2 A)
Reference Electrode Input Impedance10¹² Ω
Current Ranges2 nA–2 A (10 decades)
Compliance Voltage±21 V
CV/LSV Scan Rate0.001 mV/s to 10 V/s
CA/CC Pulse Width0.0001–65,000 s
AD Acquisition16-bit @ 1 MHz, 20-bit @ 1 kHz
DA Resolution16-bit, settling time: 1 µs
EIS Signal GeneratorSine/triangle/square wave
AC Amplitude1–2500 mV RMS
DC Bias−10 to +10 V
Bandwidth8 programmable settings
InterfaceUSB 2.0
SoftwareCorrTest CS Studio v4.x with GLP-compliant audit trail, method validation, and CFR Part 11–ready user access control
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCorrTest
OriginHubei, China
ModelCS2350M
Instrument TypeDual-Channel Electrochemical Workstation
Number of Channels2 Independent Potentiostat/Galvanostat Units
Current Range±1.0 A
Current Accuracy±0.1% of full-scale reading
Potential Accuracy±0.1% of full-scale reading ±1 mV
Potentiostatic Control Range±10 V
Electrochemical Impedance Spectroscopy (EIS) Frequency Range10 µHz to 1 MHz
Input Impedance (Reference Electrode)10¹² Ω
Current Sensitivity1 pA
Potential Sensitivity3 µV (<10 Hz), 10 µV (>100 Hz)
Potentiostat Rise Time<1 µs (<10 mA), <10 µs (<2 A)
Current Ranges2 A to 2 nA (10 decades)
DA/AD Resolution16-bit DAC / 16-bit @ 1 MHz, 20-bit @ 1 kHz ADC
Communication InterfaceEthernet
Dimensions (W×D×H)36.5 × 30.5 × 16 cm
Net Weight7.5 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCorrTest
OriginHubei, China
ModelCS300X
Channel Count2–8 (configurable)
Current Range±500 mA (Ch. 1–3), ±200 mA (Ch. 4–8)
Current Accuracy0.1% of full-scale reading
Potential Accuracy0.1% of full-scale reading ±1 mV
Potentiostatic Range±10 V
EIS Frequency Range10 µHz to 1 MHz
Data Acquisition16-bit @ 1 MHz, 20-bit @ 1 kHz
Input Impedance (Reference)10¹² Ω
Output Compliance Voltage±21 V (expandable to ±200 V)
Communication InterfaceUSB 2.0 / Ethernet
Added to wishlistRemoved from wishlist 0
Add to compare
BrandCorrTest
ModelCS150M
Current Range±2.0 A
Current Accuracy0.1% of full-scale reading
Potential Accuracy0.1% of full-scale reading ±1 mV
Potentiostatic Range±10 V
EIS Frequency Range10 µHz to 115 kHz
Potential Resolution3 mV (<10 Hz), 10 mV (>100 Hz)
Current Sensitivity1 pA
Rise Time<1 ms (<10 mA), <10 ms (<2 A)
Reference Electrode Input Impedance10¹² Ω
Current Ranges2 nA–2 A (10 decades)
Compliance Voltage±21 V
CV/LSV Scan Rate0.001 mV/s–10 V/s
CA/CC Pulse Width0.0001–65,000 s
AD Acquisition16-bit @ 1 MHz, 20-bit @ 1 kHz
DA Resolution16-bit, Settling Time: 1 ms
Minimum CV Potential Increment0.075 mV
SWV Frequency0.001–100 kHz
DPV/NPV Pulse Width0.0001–1000 s
Low-Pass Filter8-stage programmable
Communication InterfaceUSB 2.0
Show next
InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0