| Brand | Setaram |
|---|---|
| Origin | France |
| Model | Sensys Evo TG-DSC |
| Temperature Range | –120 to 830 °C |
| Heating/Cooling Rate | 0.01–30 °C/min |
| Temperature Stability | ±0.02 °C |
| Max Sample Mass | 3 g |
| Pressure Range (HP-DSC) | up to 500 bar (controlled up to 400 bar) |
| Atmosphere Compatibility | inert, oxidizing, reducing, corrosive (e.g., H₂S, NH₃), humidified, vacuum, steam |
| Balance Resolution | 0.002 µg |
| DSC Sensitivity | < 0.035 µW |
| TG Baseline Drift | < 10 µg (full-scale scan) |
| TG Baseline Repeatability | ±1 µg |
| Gas Control | integrated mass flow controllers, 3 carrier + 1 reactive gas channels |
| Sample Holder Options | TG-DSC and TG-DTA dual-mode crucibles |
| Optional Features | ASC auto-sampler (48 positions), humidity control module, FT-IR/MS/GC coupling interfaces |
| Brand | SETARAM |
|---|---|
| Origin | France |
| Model | THEMYS H2 |
| Temperature Range | Ambient to 1000/1600/1750 °C |
| Heating Rate | 0.01–100 °C/min |
| Max Sample Mass | 35 g |
| Atmosphere | 3 Gas Lines (including dedicated H₂ line) |
| Safety | Dual H₂/O₂ Sensors, Interlocked Furnace & Balance, Emergency H₂ Shutoff, Vacuum- and High-Temp-Interlocked Access |
| Vacuum Level | < 5×10⁻² mbar |
| Resolution | 0.4 μW (DTA), 1 μW (DSC), 0.002 μg (TGA), 0.02 μg (microbalance), 0.2 nm (TMA) |
| Quantitative Accuracy | ±1.4% (DTA), ±0.9% (DSC), ±0.06% (TGA), ±0.08×10⁻⁶/°C (TMA expansion) |
| Brand | SETARAM |
|---|---|
| Model | Setline STA / Setline STA+ |
| Temperature Range | Ambient to 1100 °C |
| Heating Rate | 0.01–50 °C/min |
| Temperature Stability | ±0.05 °C |
| Max Sample Mass | 20 g |
| Atmosphere Options | Air, O₂, Inert (N₂, Ar) |
| Balance Resolution | 0.02 µg |
| TG Dynamic Range | ±200 mg / ±1000 mg |
| DSC Sensitivity | 1 µW |
| DSC Enthalpy Accuracy | ±1% |
| Optional Autosampler | 49-position |
| Cooling Method | Water-cooled furnace liner |
| Software | Calisto 2.0 (Data Acquisition & Processing) |
| Brand | SETARAM |
|---|---|
| Origin | France |
| Model | Labsys Evo STA |
| Heating Rate | 0–100 K/min |
| Temperature Range | Ambient to 1600 °C |
| Temperature Stability | ±0.05 °C |
| Maximum Sample Mass | 1000 mg |
| Atmosphere Compatibility | Inert, Oxidizing, Reducing, Steam, Corrosive Gases |
| Balance Resolution | 0.02 µg |
| DSC Sensitivity | 0.4 µW (sensor-dependent) |
| Cp Accuracy | <2% error |
| Gas Control | 3 mass-flow-controlled carrier gases + 1 auxiliary/reactive gas |
| EGA Interface | Standard MS/FT-IR/GC coupling ports |
| Optional Autosampler | 25-position ASC |
| Brand | SETARAM |
|---|---|
| Origin | France |
| Model | 96 Line |
| Temperature Range | RT to 1750 °C (upgradable to 2100 °C) |
| Heating Rate | 0–100 K/min |
| TG Sample Capacity | 100 g |
| Max. TG Sample Dimensions | Ø20 mm × H80 mm |
| TG Resolution | 0.3 µg |
| Calorimetric Sample Volume | 5700 µL (Ø14.5 mm × H35 mm) |
| Specific Heat Accuracy | ±1% (relative) |
| TMA Stroke Range | ±6 mm |
| TMA Resolution | 1.6 nm |
| TMA Max. Sample Dimensions | Ø18 mm × H50 mm |
| Sensor Architecture | 3D microcalorimetric sensor with Drop method implementation |
| Modular Configurations | TG, TG-DSC, TG-DTA, TMA, 3D calorimetry |
| Brand | SETARAM |
|---|---|
| Origin | France |
| Model | Setsys Evo16 / Evo18 / Evo24 |
| Temperature Range | –150 °C to 1750 °C (upgradable to 2400 °C) |
| Heating Rate | 0–100 K/min (full range) |
| Max. TG Sample Capacity | 35 g / 100 g |
| TG Resolution | 0.002 µg / 0.02 µg |
| TG Baseline Repeatability | <10 µg (RT to 1750 °C) |
| DSC Resolution | 1 µW |
| DTA Resolution | 0.4 µW |
| Gas Channels | 3 carrier + 1 auxiliary/reactive gas, all MFC-controlled |
| Atmosphere Options | Inert, oxidative, reductive, steam, corrosive gases |
| Vacuum Level | ≤1×10⁻⁴ mbar |
| EGA Compatibility | MS, FT-IR, GC interfaces standard |
| Brand | SETARAM |
|---|---|
| Origin | France |
| Model | Themys One |
| Temperature Range | RT–1150°C / RT–1600°C |
| Heating Rate | 0.01–100 °C/min |
| Atmosphere | Inert or reactive gas |
| Compliance | ASTM E1131, ISO 11358, USP <1227>, ISO 17972 (Cp), GLP/GMP-ready software architecture |
| Brand | SETARAM |
|---|---|
| Origin | France |
| Model | Setsys STA 2400°C |
| Temperature Range | –150 to 2400 °C |
| Heating Rate | 0.01–100 K/min |
| Max. Sample Mass | 35 g (standard), 100 g (high-capacity option) |
| TG Resolution | 0.002 µg / 0.02 µg |
| TG Baseline Repeatability | <10 µg (RT to 1750 °C) |
| DSC Resolution | 1 µW |
| DTA Resolution | 0.4 µW |
| Atmosphere Control | Inert, oxidizing, reducing, steam, corrosive gases (static/dynamic) |
| Gas Configuration | 3 carrier + 1 auxiliary/reactive gas lines with MFCs and solenoid valves |
| Vacuum Level | ≤10⁻⁴ mbar |
| EGA Compatibility | Direct coupling ports for MS, FT-IR, GC |