Biological Safety Cabinets
Filter
Showing 1–30 of 80 results
| Brand | AIRTECH |
|---|---|
| Origin | Jiangsu, China |
| Model Variants | BSC-1004IIA2 / BSC-1304IIA2 / BSC-1604IIA2 / BSC-1804IIA2 |
| External Dimensions (L×W×H, mm) | 1200/1500/1800/2000 × 795 × 2050 |
| Internal Work Area Dimensions (L×W×H, mm) | 1004/1304/1604/1804 × 630 × 630 |
| ULPA Filter Efficiency | ≥99.9995% @ 0.12 µm |
| Cabinet Classification | NSF/ANSI 49–2022 Class II, Type A2 |
| Airflow Velocity (Inflow/Downflow) | 0.50–0.55 m/s / 0.25–0.30 m/s (typical, per model and certification) |
| Filter Type | Dual ULPA (Supply & Exhaust) with Leak-Proof Sealing Technology |
| Power Supply | 220 V AC, 50 Hz |
| Motor Type | Brushless DC Constant-Flow Fan |
| Brand | AIRTECH |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Regional Category | Domestic (China) |
| Model Variants | BSC-0605IIA2 / BSC-1005IIA2 / BSC-1305IIA2 / BSC-1605IIA2 / BSC-1805IIA2 |
| Cabinet Type | Class II, Type A2 |
| External Dimensions (L×W×H, mm) | 1200/1500/1800/2000 × 795 × 2050 |
| Internal Work Area Dimensions (L×W×H, mm) | 1004/1304/1604/1804 × 630 × 630 |
| Filter Efficiency | ≥99.995% @ 0.3 µm (EN1822-1 H14/U15 certified) |
| Brand | AIRTECH |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model Variants | BSC-0605IIB2 / BSC-1005IIB2 / BSC-1305IIB2 / BSC-1605IIB2 / BSC-1805IIB2 |
| Instrument Class | Class II, Type B2 (Total Exhaust, Non-Recirculating) |
| External Dimensions (L×W×H, mm) | 1200/1500/1800/2000 × 795 × 2050 |
| Internal Work Area Dimensions (L×W×H, mm) | 1004/1304/1604/1804 × 630 × 630 |
| Filter Efficiency | ≥99.995% @ 0.3 µm (HEPA H14 per EN 1822-1) |
| Brand | AIRTECH |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic |
| Model | BSC-1000III |
| Instrument Type | Class III Biological Safety Cabinet |
| External Dimensions (L×W×H, mm) | 1200×750×2200 |
| Internal Dimensions (L×W×H, mm) | 800×600×650 |
| Filter Efficiency | ≥99.999% at 0.3 µm (HEPA/ULPA dual-stage filtration) |
| Compliance Standard | NSF/ANSI 49-2016 |
| Work Chamber Negative Pressure | ≥120 Pa |
| Airflow Rate | ≥0.05 m³/s |
| Noise Level | ≤65 dB(A) |
| Power Supply | AC 220–240 V, 1φ, 50 Hz |
| Rated Power (excl. outlet) | 1.0 kVA |
| Net Weight | 350 kg |
| Glove Quantity | 2 pairs |
| Lighting | LED, 24.5 W × 1, Illuminance ≥650 lx |
| UV Lamp | 18 W × 1 + 8 W × 1 |
| Cabinet Material | Powder-coated cold-rolled steel (ivory white) |
| Work Area | Seamless 304 stainless steel |
| Exhaust Configuration | Top discharge |
| Cleanliness Class | ISO 5 (Class 100) |
| Brand | AIRTECH |
|---|---|
| Origin | Jiangsu, China |
| Cabinet Type | Class II, Type B2 (Total Exhaust, Non-Recirculating) |
| Models | BSC-1004IIB2, BSC-1304IIB2, BSC-1604IIB2, BSC-1804IIB2 |
| External Dimensions (L×W×H, mm) | 1200/1500/1800/2000 × 795 × 2050 |
| Internal Work Area Dimensions (L×W×H, mm) | 1004/1304/1604/1804 × 630 × 630 |
| ULPA Filter Efficiency | ≥99.9995% @ 0.12 µm |
| Airflow Control | Brushless DC Constant-Flow Fan with Intelligent Compensation |
| Filter Leak Integrity | Dual “Leak-Resistant” Sealing Technology for Supply & Exhaust Filters |
| Cleanliness Class | ISO 4 (equivalent to Federal Standard 209E Class 10) |
| Power Management | Power-Fail Memory & Standby Mode with Sash-Closure Trigger |
| Brand | AIRTECH (Suzhou Antai) |
|---|---|
| Origin | Jiangsu, China |
| Instrument Type | Class II, Type B2 (Total Exhaust) Biosafety Cabinet |
| External Dimensions (L×W×H) | 1500 × 790 × 2050 mm |
| Internal Work Area Dimensions (L×W×H) | 1270 × 600 × 620 mm |
| HEPA Filter Efficiency | ≥99.995% @ 0.3 µm |
| ULPA Filter Efficiency | ≥99.9995% @ 0.12 µm |
| Downflow Velocity | 0.35 m/s |
| Inflow Velocity | 0.55 m/s |
| Noise Level | ≤65 dB(A) |
| Power Supply | AC 220 V, 1φ, 50 Hz |
| Rated Power | 1.65 kW |
| Weight | 270 kg |
| Air Cleanliness Class | ISO 4 (equivalent to Federal Standard 209E Class 10) |
| Personnel Protection | ≤5 CFU/plate (slit sampler) |
| Product Protection | ≤2 CFU/plate |
| Cross-Contamination Protection | ≤5 CFU/plate (impact sampler), ≤5 CFU/plate (slit sampler) |
| Lighting | 24.5 W LED, ≥900 lx uniformity |
| UV Lamp | 25 W, germicidal |
| Exhaust Direction | Top-mounted ducted exhaust |
| Brand | AIRTECH (Suzhou Jintan Airtech) |
|---|---|
| Origin | Jiangsu, China |
| Model | BSC-1600IIA2 |
| Instrument Type | Class II, Type A2 Biological Safety Cabinet |
| External Dimensions (L×W×H) | 1800 × 790 × 2050 mm |
| Internal Work Area Dimensions (L×W×H) | 1600 × 600 × 620 mm |
| Filtration Efficiency | HEPA ≥99.995% @ 0.3 µm |
| Downflow Velocity | 0.35 m/s |
| Inflow Velocity | 0.55 m/s |
| Noise Level | ≤62 dB(A) |
| Lighting Intensity | ≥800 lx |
| Power Consumption | 2.0 kW |
| Weight | 360 kg |
| Compliance | YY 0569–2011 (China), NSF/ANSI 49–2022 (USA) |
| Origin | Jiangsu, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Models | BSC-1600III / BSC-1300III / BSC-1000III |
| Price Range | USD 14,000 – 28,000 |
| Cabinet Class | Class III (Total Enclosure, Glove-Box Type) |
| External Dimensions (L×W×H) | 2150 × 620 × 1980 mm |
| Internal Work Area (L×W×H) | 1600 × 600 × 650 mm |
| Filter Efficiency (HEPA/ULPA) | ≥99.999% at 0.12 µm (EN 1822 H14 equivalent) |
| Origin | Zhejiang, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | BHC-1000IIA2 |
| Price | USD 4,200 (FOB) |
| Cabinet Class | Class II, Type A2 |
| External Dimensions (L×W×H) | 1200 × 780 × 1950 mm |
| Internal Work Area (L×W×H) | 1000 × 700 × 680 mm |
| HEPA Filter Efficiency | ≥99.99% @ 0.3 µm (EN 1822-1 |
| Inflow Velocity | 0.35 ± 0.025 m/s |
| Downflow Velocity (Center) | 0.26 ± 0.025 m/s |
| Downflow Velocity (Rear) | 0.27 ± 0.025 m/s |
| Side Airflow Velocity | 0.38 ± 0.025 m/s |
| Air Leakage Rate | ≤1×10⁻⁶ m/s @ 500 Pa |
| Sound Pressure Level | ≤62 dB(A) |
| Vibration Amplitude (Half-Peak) | ≤5 µm |
| Illumination | ≥300 lux |
| Power Supply | AC 220 V, 50 Hz, Single Phase |
| Power Consumption | 0.6 kW |
| Net Weight | <200 kg |
| HEPA Filter Size & Qty | 995 × 600 × 38 mm × 1 |
| Fluorescent Lamp / UV Lamp | 20 W × 1 / 20 W × 1 |
| Origin | Jiangsu, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | BHC-1300IIA/B3 |
| Instrument Class | Class II A2 Biological Safety Cabinet (30% Exhaust) |
| External Dimensions (L×W×H) | 1500 × 815 × 2270 mm |
| Internal Work Area (L×W×H) | 1300 × 600 × 640 mm |
| Filter Efficiency | HEPA ≥99.995% @ 0.3 µm |
| Downflow Velocity (Center) | 0.27 m/s |
| Inflow Velocity | ≥0.5 m/s |
| Noise Level | ≤58 dB(A) |
| Power Consumption | 2.1 kW |
| Weight | 340 kg |
| Lighting | 30 W × 2 Fluorescent Lamps + 30 W UV Lamp |
| Illumination | ≥650 lx |
| Airflow Configuration | Left/Right/Top Exhaust Options |
| Certified Protection Levels | Personnel Protection (≤10 CFU/plate, impact sampler), Product Protection (≤5 CFU/plate), Cross-Contamination Protection (≤2 CFU/plate) |
| Compliance | Meets ISO 14644-1 Class 5 (ISO Class 5, formerly “100”), EN 12469:2000, and GB/T 16292–16294 (Chinese Pharmacopoeia & GMP requirements) |
| Origin | Jiangsu, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic (China-Made) |
| Model | BHC-1300IIA2 |
| Cabinet Type | Class II, Type A2 (30% Exhaust / 70% Recirculated) |
| External Dimensions (L×W×H) | 1500 × 815 × 2270 mm |
| Internal Work Area Dimensions (L×W×H) | 1300 × 600 × 640 mm |
| Filter Efficiency | HEPA ≥99.995% @ 0.3 µm |
| Downflow Velocity | 0.33 m/s |
| Inflow Velocity | 0.55 m/s |
| Noise Level | ≤58 dB(A) |
| Lighting Intensity | ≥650 lux |
| Power Supply | AC 220 V, 50 Hz |
| Total Power Consumption | 2.0 kW (including dedicated GFCI outlets) |
| Net Weight | 340 kg |
| Lighting | Two 30 W fluorescent lamps + one 20 W UV lamp |
| Exhaust Options | Left, right, or top discharge |
| User Capacity | Dual-operator configuration |
| Origin | Jiangsu, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Product Category | Class II, Type B2 (Total Exhaust) Biosafety Cabinet |
| Model | BHC-1300IIB2 |
| External Dimensions (L×W×H) | 1525 × 830 × 2300 mm |
| Internal Work Area Dimensions (L×W×H) | 1300 × 600 × 640 mm |
| Filter Efficiency | HEPA ≥99.995% @ 0.3 µm |
| Downflow Velocity | 0.30 m/s |
| Inflow Velocity | 0.55 m/s |
| Noise Level | ≤58 dB(A) |
| Power Supply | AC 220 V, 1Φ, 50 Hz |
| Total Power Consumption | 2.0 kW (including auxiliary socket) |
| Weight | 340 kg |
| Lighting | 2 × 36 W fluorescent lamps |
| UV Lamp | 1 × 30 W |
| Illuminance | ≥650 lx |
| Duct Port Diameter | Ø250 mm |
| Safety Certification Compliance | EN 12469:2000, NSF/ANSI 49-2022, GB 50346-2011 |
| Origin | Jiangsu, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | BSC-1300IIA2 |
| Price | USD 4,500 (FOB Jiangsu) |
| Certifications | EN 12469:2000, NSF/ANSI 49-2022 compliant |
| Airflow | Inflow velocity ≥ 0.65 m/s, Downflow velocity ≥ 0.35 m/s |
| Filtration | Dual HEPA (≥99.9995% @ 0.3 µm), Pre-filter + Exhaust HEPA + Supply HEPA |
| Noise Level | ≤58 dB(A) |
| Illumination | ≥1000 lux |
| UV Intensity | ≥80 µW/cm² |
| Work Surface Load Capacity | 25 kg |
| Electrical | 220 V ±20 V, 50 Hz, Ground Resistance ≤0.1 Ω |
| Leak Integrity | ≤10% pressure decay at 500 Pa for 30 min |
| Cleanliness Class | ISO Class 5 (100-grade) |
| Containment Performance | Personnel protection ≤10 CFU/plate (impact sampling), Product protection ≤4 CFU/plate, Cross-contamination ≤2 CFU/plate |
| Display | Full-function LCD with real-time monitoring of airflow velocities, filter differential pressure, temperature, humidity, UV status, door position, fan operation, timer, and service life tracking |
| Brand | BOXUN |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic |
| Model | BSC-1000A2 |
| Instrument Type | Class II Biological Safety Cabinet |
| External Dimensions (L×W×H) | 1220 × 860 × 2225 mm |
| Internal Dimensions (L×W×H) | 1100 × 500 × 640 mm |
| Filter Efficiency (for 0.3 µm particles) | ≥99.99% |
| Inflow Velocity | 0.55 m/s ± 0.02 m/s |
| Downflow Velocity | 0.35 m/s ± 0.02 m/s |
| Airflow Pattern | 70% recirculated, 30% exhausted |
| Work Area Classification | ISO 5 (equivalent to US FED STD 209E Class 100) |
| Personnel Protection Factor (KI test) | ≥1 × 10⁵ |
| Product Protection (colony count) | ≤5 CFU/test |
| Cross-Contamination Protection (impactor sampling) | ≤2 CFU/test |
| Noise Level | ≤67 dB(A) |
| Vibration Peak-to-Peak | ≤5 µm |
| Illumination | ≥650 lux |
| Operating Temperature | 10–30 °C |
| Relative Humidity | ≤70% at 25 °C |
| Ambient Pressure Range | 80–105 kPa |
| Power Supply | AC 220 V ±10%, 50/60 Hz |
| Total Power Consumption | 1.25 kW (including integrated AC outlets) |
| Brand | BOXUN |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | BSC-1000B2 |
| Instrument Type | Class II Biological Safety Cabinet |
| External Dimensions (L×W×H) | 1120 × 895 × 2340 mm |
| Internal Work Area Dimensions (L×W×H) | 1000 × 500 × 640 mm |
| HEPA Filter Efficiency | ≥99.99% @ 0.3 µm |
| Inflow Velocity | 0.55 m/s |
| Downflow Velocity | 0.35 m/s |
| Downflow Control Accuracy | ±0.015 m/s |
| Pressure Differential Range | 0–500 Pa |
| Noise Level | ≤67 dB(A) |
| Vibration Peak-to-Peak | ≤5 µm |
| Average Illuminance | ≥650 lx |
| Operating Ambient Temperature | 10–30 °C |
| Relative Humidity | ≤70% (at 25 °C) |
| Atmospheric Pressure Range | 80–105 kPa |
| Lighting | 2 × 30 W fluorescent lamps + 1 × 18 W UV lamp |
| Power Consumption | 2.1 kW |
| Electrical Supply | AC 220 V ±10%, 50/60 Hz |
| Brand | BOXUN |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | BSC-1360A2 |
| Instrument Type | Class II A2 Biological Safety Cabinet |
| External Dimensions (L×W×H) | 1420 × 860 × 2225 mm |
| Internal Work Area Dimensions (L×W×H) | 1300 × 500 × 640 mm |
| HEPA Filter Efficiency | ≥99.99% @ 0.3 µm |
| Inflow Velocity | 0.55 m/s ±0.02 m/s |
| Downflow Velocity | 0.35 m/s ±0.02 m/s |
| Air Recirculation Ratio | 70% |
| Exhaust Ratio | 30% |
| Noise Level | ≤67 dB(A) |
| Vibration Peak-to-Peak | ≤5 µm |
| Average Illuminance | ≥650 lux |
| Operating Temperature Range | 10–30 °C |
| Relative Humidity Limit | ≤70% @ 25 °C |
| Atmospheric Pressure Range | 80–105 kPa |
| Personnel Protection Factor (KI Test) | ≥1×10⁵ |
| Product Protection (Settle Plate) | ≤5 CFU/plate/test |
| Cross-Contamination Protection (Impactor Sampler) | ≤2 CFU/plate/test |
| Lighting & UV Lamp | 30 W × 1 each |
| Total Power Consumption | 1.4 kW |
| Electrical Supply | AC 220 V ±10%, 50/60 Hz |
| Brand | BOXUN |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | BSC-1360B2 |
| Instrument Type | Class II Biological Safety Cabinet |
| External Dimensions (L×W×H) | 1420 × 895 × 2340 mm |
| Internal Work Area Dimensions (L×W×H) | 1300 × 500 × 640 mm |
| HEPA Filter Efficiency | ≥99.99% @ 0.3 µm |
| Inflow Velocity | 0.55 m/s |
| Downflow Velocity | 0.33 m/s |
| Downflow Control Accuracy | ±0.015 m/s |
| Pressure Differential Range | 0–500 Pa |
| Noise Level | ≤67 dB(A) |
| Vibration Half-Peak | ≤5 µm |
| Average Illuminance | ≥650 lx |
| Operating Ambient Temperature | 10–30 °C |
| Relative Humidity | ≤70% @ 25 °C |
| Atmospheric Pressure Range | 80–105 kPa |
| Power Supply | AC 220 V ±10%, 50/60 Hz |
| Total Power Consumption | 2.2 kW |
| Lighting | 2 × 40 W fluorescent lamps + 1 × 30 W UV lamp |
| Supply Filter Size | 1326 × 606 × 70 mm |
| Exhaust Filter Size | 782 × 515 × 90 mm |
| Work Surface | Removable SUS304 stainless steel with integrated liquid collection trough and drain valve |
| Airflow Sensors | Imported U.S.-made constant-velocity anemometers |
| Fan System | EBMPAPST (Germany) centrifugal blowers |
| Safety Interlocks | UV/light interlock, sash position alarm, fan shutdown at sash closure |
| Brand | Brookfield |
|---|---|
| Origin | Shanghai, China |
| Model | BSC-1360A2 |
| Cabinet Class | Class II, Type A2 |
| External Dimensions (L×W×H) | 1500 × 770 × 2135 mm |
| Internal Work Area (L×W×H) | 1360 × 580 × 595 mm |
| Airflow Mode | 70% recirculated, 30% exhausted |
| Inflow Velocity | 0.55 m/s ± 0.02 m/s |
| Downflow Velocity | 0.35 m/s ± 0.02 m/s |
| HEPA Filter Efficiency | ≥99.998% @ 0.3 µm |
| Noise Level | ≤67 dB(A) |
| Illumination | ≥650 lux |
| Power Supply | AC 220 V ±10%, 50/60 Hz |
| Total Power Consumption | 1.4 kW |
| Personnel Protection Factor (KI test) | ≥1 × 10⁵ |
| Product Protection (CFU/test) | ≤5 CFU |
| Cross-Contamination Protection (impactor sampling) | ≤2 CFU |
| Operating Temperature | 10–30 °C |
| Relative Humidity | ≤70% @ 25 °C |
| Ambient Pressure Range | 80–105 kPa |
| Origin | Jiangsu, China |
|---|---|
| Cabinet Type | Class II, Type B2 (Total Exhaust) |
| External Dimensions (L×W×H) | 1500 × 790 × 2050 mm |
| Internal Work Area (L×W×H) | 1300 × 600 × 620 mm |
| Filter Efficiency | HEPA ≥99.995% @ 0.3 µm |
| Downflow Velocity | 0.35 m/s |
| Inflow Velocity | 0.35 m/s |
| Noise Level | ≤58 dB(A) |
| Power Supply | AC 220 V, 1Φ, 50 Hz |
| Max. Rated Power | 2.1 kVA |
| Weight | 300 kg |
| Lighting | 36 W × 2 fluorescent lamps |
| UV Lamp | 20 W × 1 |
| Illuminance | ≥800 lx |
| Biological Protection Performance | ≤10 CFU/plate (impact sampler), ≤5 CFU/plate (slit sampler) |
| Product Protection | ≤5 CFU/plate |
| Cross-contamination Protection | ≤2 CFU/plate |
| Brand | Chuanyu Instruments |
|---|---|
| Origin | Shanghai, China |
| Model | BHC-1300IIA2 |
| Cabinet Class | NSF/EN 12469 Class II, Type A2 |
| External Dimensions (L×W×H) | 1500 × 640 × 1950 mm |
| Internal Work Area (L×W×H) | 1300 × 580 × 680 mm |
| HEPA Filter Efficiency | ≥99.99% @ 0.3 µm (tested per ISO 14644-3 & IEST-RP-CC001.3) |
| Inflow Velocity | 0.35 ± 0.025 m/s |
| Downflow Velocity | 0.26–0.27 ± 0.025 m/s |
| Noise Level | ≤62 dB(A) |
| Illumination | ≥300 lx |
| Power Supply | AC 220 V / 50 Hz |
| Max. Power Consumption | 0.9 kW |
| Weight | <250 kg |
| Filter Configuration | Single integrated HEPA filter (1295 × 600 × 38 mm), liquid-sealed, pre-filtered intake |
| Brand | Dynamica |
|---|---|
| Model | EHS-530 |
| Cabinet Class | Class II, Type A2 |
| Construction Material | 304 Stainless Steel Interior (One-Piece Seamless Welded) |
| External Dimensions (L×W×H) | 1950 × 750 × 2100 mm |
| Internal Work Area Dimensions (L×W×H) | 1862 × 574 × 628 mm |
| HEPA/ULPA Filtration Efficiency | ≥99.9995% @ 0.12 µm |
| Airflow System | Recirculating (70% re-circulated, 30% exhausted via dedicated duct or thimble connection) |
| Motor Type | Brushless DC ECM Motor with Variable Speed Control |
| Control Interface | DynaAir Color Touchscreen Controller with Real-Time Airflow Monitoring |
| Sensor | High-Sensitivity Thermal Anemometer for Inflow & Downflow Velocity Detection |
| Compliance | EN 12469:2000, NSF/ANSI 49-2022, ISO 14644-1 Class 5 (ISO Class 5 Equivalent), GLP-Ready Configuration |
| Brand | Dynamica |
|---|---|
| Model | NU-437-400S / NU-437-600S |
| Type | Class II A2 Biological Safety Cabinet |
| External Dimensions (W×D×H) | 1972 × 835 × 1600 mm (NU-437-600S) / 1362 × 835 × 1600 mm (NU-437-400S) |
| Internal Work Area (W×D×H) | 1788 × 597 × 724 mm (NU-437-600S) / 1178 × 597 × 724 mm (NU-437-400S) |
| Inflow Velocity | 0.53 m/s |
| Downflow Velocity | 0.30 m/s |
| HEPA Filtration Efficiency | ≥99.999% for particles 0.1–0.3 µm (H14 grade) |
| Airflow Configuration | 30% exhaust, 70% recirculation |
| Construction | 16-gauge 304 stainless steel, fully welded double-wall chamber |
| Pressure Integrity | Withstands 500 Pa static pressure test |
| Compliance | YY 0569–2011 (China), aligned with NSF/ANSI 49 and EN 12469 principles |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Manufacturer | Esco Micro Pte Ltd |
| Product Type | Class II Type A2 Biological Safety Cabinet |
| Model | AC2-S-CN |
| Filter Efficiency | ≥99.999% for Most Penetrating Particle Size (MPPS) or ≥0.12 µm particles |
| Compliance | EN 12469, NSF/ANSI 49, YY 0569 |
| Airflow System | Constant-velocity DC ECM blower with voltage fluctuation compensation and resistance-sensing dynamic compensation |
| Surface Coating | Esco ISOCIDE antimicrobial coating |
| Construction | Seamless stainless steel work surface |
| Interface | RS-232 serial port for operational status monitoring |
| Cleanliness Class | ISO 14644-1 Class 3 (≤1,000 particles ≥0.1 µm/m³) |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Manufacturer | Esco Micro Pte Ltd |
| Product Type | Class II, Type B2 Biosafety Cabinet |
| Filter Efficiency | ≥99.999% for MPPS (Most Penetrating Particle Size) or ≥0.12 µm particles |
| Standards Compliance | NSF/ANSI 49-2023, EN 12469:2020, YY 0569-2011 |
| Control System | Sentinel Gold™ Microprocessor-Based Controller |
| Filtration | ULPA (Ultra-Low Penetration Air) Filter |
| Airflow Design | 100% Exhaust, Non-Recirculating |
| Construction | Seamless Stainless Steel Work Surface with ISOCIDE® Antimicrobial Coating |
| Interface | RS-232 Serial Port for Operational Data Logging |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Origin | Imported |
| Cabinet Class | Class III |
| External Dimensions (L×W×H) | Refer to model-specific specifications |
| Internal Dimensions (L×W×H) | Refer to model-specific specifications |
| Filter Efficiency | ≥99.999% for particles ≥0.1–0.3 µm |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Instrument Type | Class II, Type B2 Biosafety Cabinet |
| External Dimensions (L×W×H) | 1340 × 811 × 2171 mm |
| Internal Dimensions (L×W×H) | 1270 × 585 × 670 mm |
| Filter Efficiency | ≥99.999% (ULPA/HEPA) |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Cabinet Type | Class II, Type A2 |
| Certification | EN 12469 compliant |
| Filter Efficiency | ≥99.999% @ 0.3 µm (ULPA) |
| Inflow Velocity | 0.45 m/s (90 fpm) |
| Downflow Velocity | 0.30 m/s (60 fpm) |
| External Dimensions (AC2-4D) | 1340 × 753 × 1400 mm (w/o armrest), 1340 × 810 × 1400 mm (w/ armrest) |
| Internal Dimensions (AC2-4D) | 1220 × 580 × 660 mm |
| Power Supply | 220 V, 50 Hz |
| Work Surface | Seamless stainless steel, integral arch spill containment |
| Antimicrobial Coating | Esco ISOCIDE® on all external and internal surfaces |
| Control System | EscoSentinel Gold microprocessor-based display with real-time airflow monitoring |
| Communication Interface | RS-232 for BMS integration |
| Ergonomic Features | Adjustable armrests, sloped front grille, footrest option, height-adjustable base options |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Manufacturer | Esco Micro Pte Ltd |
| Product Category | Imported Class II Biological Safety Cabinet |
| Model | AC2-D |
| Cabinet Type | Type A2 (Recirculating/Exhaust) |
| External Dimensions (L×W×H, mm) | 1340×753×1400 (AC2-4D) / 1950×753×1400 (AC2-6D) |
| Internal Dimensions (L×W×H, mm) | 1220×580×660 (AC2-4D) / 1830×580×660 (AC2-6D) |
| Inflow Velocity | 0.45 m/s (90 fpm) |
| Downflow Velocity | 0.30 m/s (60 fpm) |
| Filtration Efficiency | ≥99.999% @ 0.3 µm (ULPA), ≥99.995% @ 0.1 µm |
| Power Supply | 220 V, 50 Hz |
| Compliance | EN 12469:2000, NSF/ANSI 49-2022 (Class II, Type A2), ISO 14644-1 Class 5 (ISO 5) work environment |
| Surface Coating | Esco ISOCIDE® antimicrobial coating |
| Control System | EscoSentinel Gold microprocessor-based interface with real-time airflow monitoring and alarm logic |
| Communication Interface | RS-232 for BMS integration |
| Work Surface | Seamless one-piece stainless steel with integral spill containment arch |
| Structural Design | Weld-free, single-piece formed inner chamber with no seams or fasteners |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Instrument Type | Class II Type A2 Biological Safety Cabinet |
| Filter Efficiency | ≥99.999% for 0.1 µm particles (ULPA-grade) |
| Inflow Velocity | 0.45 m/s (90 fpm) |
| Downflow Velocity | 0.30 m/s (60 fpm) |
| Power Supply | 230 V, 50/60 Hz |
| Compliance | EN 12469:2000, NSF/ANSI 49-2022, IEST-RP-CC001-2022 (equivalent to EN 1822 H14) |
| Brand | Esco |
|---|---|
| Origin | Singapore |
| Instrument Type | Class II Type A2 Biological Safety Cabinet |
| External Dimensions (L×W×H) | 1340 × 753 × 1400 mm |
| Internal Dimensions (L×W×H) | 1220 × 580 × 660 mm |
| Filter Efficiency | ≥99.999% for 0.1–0.3 µm particles |