BrandAvesta
OriginRussia
ModelASP-IR
Instrument TypeLaboratory Benchtop Scanning Spectrometer
Spectral Range (nm)500–3450 (model-dependent: ASP-IR-1.7: 500–1700
ASP-IR-2.6900–2600
ASP-IR-3.51200–3450)
Resolution (free-space input)0.5 nm / 0.8 nm / 1.2 nm
Resolution (400 µm fiber input)0.7 nm / 1.0 nm / 1.5 nm
Scan Speedup to 400 nm/s
Signal-to-Noise Ratio5000:1
Optical DesignCzerny-Turner monochromator (f = 150 mm, f/# = 1:13)
Input Slit15 µm
Output Slit20 µm
DetectorThermoelectrically cooled InGaAs (ASP-IR-1.7) or二级-cooled InAs (ASP-IR-2.6 & ASP-IR-3.5)
NEP0.6 nW (0.15 pJ) / 42 nW (2.5 pJ) / 145 nW (7 pJ)
Dynamic Range3×10⁵ / 7×10⁵ / 6×10⁴
ADC16-bit
InterfaceUSB 2.0
Dimensions (mm)225×136×81 (ASP-IR-1.7), 295×136×81 (ASP-IR-2.6/3.5)
Weight2.9 kg / 3.6 kg
Trigger InputYes (TTL-compatible, for pulsed laser synchronization)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSpectral Products
OriginUSA
ModelSM301 / SM301-EX
Instrument TypePortable
Spectral Range1.0–3.0 µm (PbS) or 1.5–5.0 µm (PbSe)
Spectral Resolution10–30 nm (dependent on slit width and grating)
Detector256-element PbS or PbSe linear array
Pixel Size45 µm × 450 µm
Pixel Clockup to 2 MHz
Data Output Rateup to 4 MHz
Integration Time0.01–200 ms (onboard)
Slit Options10, 25, 50, 100 µm (default: 25 µm)
f/#3.5
CoolingTE-cooled to –15 °C ARO (down to –20 °C ARO optional)
Dark Noise~20 RMS counts (16-bit, balanced)
SNR4000:1
Stray Light< 0.1%
InterfaceUSB 1.1/2.0, 16-bit
SoftwareSM32Pro & SMProMX (Windows XP–10, 32/64-bit), DLL & SDK included
Dimensions173 × 120 × 98.1 mm
Weight2.5 kg
Optical CouplingSMA 905 or FC connector
Trigger ModesFree-running or external TTL
Added to wishlistRemoved from wishlist 0
Add to compare
BrandSPL
OriginSwitzerland
ModelARCoptix
TypeFourier Transform (FTIR)
ConfigurationOnline/Portable
Wavelength Range2–12 µm
Wavenumber Resolution4 cm⁻¹
Wavelength Resolution<2 nm (at 2 µm), <10 nm (at 4 µm), <12 nm (at 6 µm)
Signal-to-Noise Ratio>1000:1
Scan Rate1 spectrum/sec (interferogram averaging mode)
IR SourceMiniature blackbody emitter (1150 K)
Reference LaserSolid-state 850 nm diode laser
Interference CompensationBuilt-in dynamic self-compensation
ComplianceCE, RoHS, ISO/IEC 17025-compatible operation
Added to wishlistRemoved from wishlist 0
Add to compare
BrandARC
OriginSwitzerland
ModelFT-NIR Rocket
Instrument TypePortable
Spectral Range900–2600 nm (11000–3800 cm⁻¹)
Resolution4 cm⁻¹ (≤0.8 nm @ 1100 nm, ≤2 nm @ 1700 nm, ≤5 nm @ 2600 nm)
Wavenumber Accuracy<1 cm⁻¹ @ 4000 cm⁻¹ (25±5°C)
Signal-to-Noise Ratio>10,000:1 (RMS, full spectrum, single scan)
Dynamic Range>3000:1
Scan Speed1 spectrum/sec (averaging mode)
A/D Conversion24-bit
DetectorInGaAs photodiode array
Reference LaserIntegrated solid-state 850 nm laser
Optical InterfaceSMA905, fiber input ≥1 mm core, NA = 0.2
Power Supply5 V DC via USB 2.0
Operating Temperature10–30 °C
Dimensions180 × 160 × 80 mm
Weight850 g
ComplianceCE, RoHS, USB-IF certified