Infrared Spectrometer
Filter
Showing 61–90 of 186 results
| Brand | GBPI |
|---|---|
| Origin | Guangdong, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory FTIR Spectrometer |
| Spectral Range | 7800–350 cm⁻¹ |
| Resolution | <1.0 cm⁻¹ |
| Scan Speed | 4 cm⁻¹ |
| Signal-to-Noise Ratio | 15,000:1 (at 4 cm⁻¹, 1 min scan) / 30,000:1 (at 4 cm⁻¹, 5 min scan) |
| Beamsplitter | KBr substrate with Ge coating |
| Detector | Imported high-sensitivity DLaTGS |
| Source | Imported long-life, air-cooled high-intensity IR source |
| Interferometer | Sealed 30°-incidence Michelson interferometer with automatic optical alignment |
| Laser | VCSEL-based He–Ne reference laser |
| A/D Conversion | 24-bit, 500 kHz |
| Communication Interface | USB 2.0 (backward compatible with USB 3.0) |
| Dimensions | 450 × 350 × 235 mm (L×W×H) |
| Weight | 14 kg |
| Operating Temperature | 15–35 °C |
| Power Consumption | <30 W |
| Brand | GBPI |
|---|---|
| Origin | Guangdong, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory FTIR Spectrometer |
| Wavenumber Range | 7800–350 cm⁻¹ |
| Spectral Resolution | <1.0 cm⁻¹ |
| Scan Speed | 4 cm⁻¹ (optical path difference rate) |
| Signal-to-Noise Ratio | 15,000:1 (at 4 cm⁻¹, 1 min scan) / 30,000:1 (at 4 cm⁻¹, 1 min scan, DLaTGS detector) |
| Beamsplitter | KBr substrate with Ge coating |
| Detector | Imported high-sensitivity deuterated triglycine sulfate (DLaTGS) |
| Source | Imported long-life, air-cooled high-intensity IR source |
| Interferometer | Sealed 30°-incidence Michelson interferometer with integrated auto-alignment optics |
| A/D Conversion | 24-bit, 500 kHz |
| Communication Interface | USB 2.0 (backward compatible with USB 3.0) |
| Dimensions | 450 × 350 × 235 mm (L×W×H) |
| Weight | 14 kg |
| Operating Temperature | 15–35 °C |
| Power Consumption | <30 W |
| Brand | GBPI |
|---|---|
| Origin | Guangdong, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory FTIR Spectrometer |
| Spectral Range | 7800–350 cm⁻¹ |
| Resolution | <1.0 cm⁻¹ |
| Scan Speed | 4 cm⁻¹ |
| Signal-to-Noise Ratio | 15,000:1 / 30,000:1 (at 4 cm⁻¹, 1 min scan) |
| Beamsplitter | KBr substrate with Ge coating |
| Detector | Imported high-sensitivity deuterated L-alanine triglycine sulfate (DLaTGS) |
| Source | Imported long-life, air-cooled high-intensity IR source |
| Interferometer | Sealed 30°-incidence Michelson interferometer with built-in auto-alignment optics |
| A/D Conversion | 24-bit, 500 kHz |
| Communication Interface | USB 2.0 (backward compatible with USB 3.0) |
| Dimensions | 450 × 350 × 235 mm (L×W×H) |
| Weight | 14 kg |
| Operating Temperature | 15–35 °C |
| Power Consumption | <30 W |
| Brand | Hamamatsu |
|---|---|
| Origin | Japan |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Imported Instrument |
| Model | C15511-01 |
| Instrument Type | Portable Fourier Transform Infrared (FTIR) Spectrometer Engine |
| Spectral Range | 1100–2500 nm |
| Spectral Resolution | 5.7 nm at λ = 1533 nm |
| Signal-to-Noise Ratio (SNR) | 10,000:1 |
| Input Interface | Fiber-optic coupled |
| Form Factor | Compact OEM module (palm-sized) |
| Origin | Japan |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Import Status | Imported |
| Pricing | Upon Request |
| Instrument Category | Spectral Software & Reference Database |
| Brand | Innoteg |
|---|---|
| Origin | Guangdong, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China-made) |
| Model | InnoFTIR10 |
| Instrument Type | Laboratory FTIR Spectrometer |
| Wavenumber Range | 7800–350 cm⁻¹ |
| Spectral Resolution | Better than 1.0 cm⁻¹ |
| Signal-to-Noise Ratio (S/N) | 45,000:1 (peak-to-peak, 2100 cm⁻¹, 4 cm⁻¹ resolution, 1-minute acquisition) |
| Beam Diameter | 10 mm |
| Beamsplitter | Imported multi-layer germanium-coated KBr, Ø60 mm with anti-humidity coating |
| Source | Imported high-intensity air-cooled ceramic IR source |
| Detector | Imported low-noise deuterated triglycine sulfate (DLATGS) detector |
| Interferometer | Oscillating-mirror interferometer (space-derived optical architecture) |
| Baseline Flatness | ≤0.1% T |
| Dimensions (W×D×H) | ≥590 × 355 × 185 mm |
| Sample Compartment Size (W×D×H) | ≥200 × 255 × 150 mm |
| Weight | 26 kg |
| Interface | USB |
| Power Supply | AC 220 V / DC 12 V, 40 W |
| Operating System | Windows 10 (64-bit or later) |
| Ambient Temperature | 15–28 °C |
| Relative Humidity | <65% RH (non-condensing) |
| Optional Features | N₂ purge port, built-in desiccant chamber with visual indicator, heater-assisted humidity control, adjustable interferometer energy output |
| Brand | Innoteg |
|---|---|
| Origin | Guangdong, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China-made) |
| Model | InnoFTIR15 |
| Instrument Type | Laboratory FTIR Spectrometer |
| Wavenumber Range | 7800–350 cm⁻¹ |
| Spectral Resolution | <1.0 cm⁻¹ |
| Signal-to-Noise Ratio (S/N) | 30,000:1 (peak-to-peak, 2100 cm⁻¹, 4 cm⁻¹ resolution, 60 s acquisition) |
| Beam Diameter | 15 mm |
| Beamsplitter | Multi-layer Ge-coated KBr (imported) |
| Source | High-intensity air-cooled ceramic IR source (imported) |
| Interferometer | Michelson-type, self-compensating, automated alignment |
| Detector | Low-noise deuterated triglycine sulfate (DLATGS) detector (imported) |
| Baseline Flatness | ≤0.1% T |
| Measurement Modes | Transmittance, Absorbance, Radiant Energy |
| Dimensions (W×D×H) | 486 × 406 × 230 mm |
| Sample Compartment Size | 200 × 170 × 145 mm |
| Weight | 23 kg |
| Interface | USB 2.0 |
| Power Supply | AC 220 V / DC 12 V, 40 W |
| OS Compatibility | Windows 10 (64-bit or later) |
| Operating Environment | 15–28 °C, RH <65% (non-condensing) |
| Brand | Innoteg |
|---|---|
| Origin | Guangdong, China |
| Manufacturer Type | Authorized Distributor |
| Regional Classification | Domestic (China) |
| Model | InnoFTIR15C |
| Instrument Type | Laboratory FTIR Spectrometer |
| Wavenumber Range | 4000–400 cm⁻¹ |
| Signal-to-Noise Ratio (S/N) | 15,000:1 (at 4 cm⁻¹ resolution, 1 min scan) |
| Wavenumber Accuracy | ±0.01 cm⁻¹ |
| Optical System | Michelson Interferometer with KBr Multilayer Beam Splitter (Ge-coated, anti-humidity) |
| Detector | High-Sensitivity DLATGS |
| Source | High-Intensity Air-Cooled Ceramic IR Source |
| Enclosure | Electromagnetically Shielded Metal Housing |
| Included Accessory | Integrated Desiccant Dry Box |
| Software | Building Glass Compliance Suite (GB/T 2680–2021, JGJ/T 151–2008 compliant) |
| Brand | INSA Optics |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Instrument Type | Laboratory FTIR Spectrometer |
| Spectral Range | 8000–350 cm⁻¹ (extendable to 12800–4000 cm⁻¹ in NIR |
| Spectral Resolution | 0.2 cm⁻¹ |
| Scan Speed | 2.5–60 kHz |
| Signal-to-Noise Ratio (SNR) | 40,000:1 (at 4 cm⁻¹, 1 min scan, MCT detector) |
| Wavenumber Accuracy | < ±0.1 cm⁻¹ |
| Vacuum Level | ≤ 0.2 mbar |
| Optical Design | All-metal vacuum-sealed cast-aluminum interferometer housing with independent optical chamber |
| Brand | Spectrolight |
|---|---|
| Origin | Switzerland |
| Model | IRsweep IRis-F1 |
| Instrument Type | Laboratory Benchtop Laser-Based Infrared Spectrometer |
| Light Source Technology | Dual Quantum Cascade Laser (QCL) Frequency Comb |
| Time Resolution | 1 µs |
| Spectral Resolution | 0.25–0.5 cm⁻¹ |
| Spectral Range | Mid-Infrared (Typical 1,200–3,000 cm⁻¹, configurable) |
| Compliance | CE, RoHS, ISO 9001-certified manufacturing |
| Software Environment | IRsweep Control Suite v4.x (Windows-based, FDA 21 CFR Part 11-ready audit trail option) |
| Brand | IRsweep |
|---|---|
| Origin | Switzerland |
| Model | IRis-F1 |
| Technology | Dual-Comb Quantum Cascade Laser (QCL) Infrared Spectroscopy |
| Time Resolution | 1 µs |
| Spectral Resolution | 0.25–0.5 cm⁻¹ |
| Spectral Range | Mid-IR (typically 1200–2200 cm⁻¹, configurable) |
| SNR | High (optimized for trace-level detection) |
| Sample Compatibility | Solid, Liquid, Gas |
| Compliance | CE, RoHS, ISO 9001-certified manufacturing |
| Brand | ISPEC |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Manufacturer |
| Regional Classification | Domestic |
| Model | Analytical System |
| Quotation | Upon Request |
| Instrument Category | Spectral Software & Database Suite |
| Instrument Type | Laboratory-Based |
| Wavenumber Range | 4000–400 cm⁻¹ |
| Resolution | 1 cm⁻¹ |
| Brand | JASCO |
|---|---|
| Origin | Japan |
| Model | FT/IR-4000, FT/IR-6000 Series |
| Instrument Type | Laboratory Fourier Transform Infrared (FTIR) Spectrometer |
| Wavenumber Range | 7800–400 cm⁻¹ |
| Spectral Resolution | 0.4 cm⁻¹ |
| Signal-to-Noise Ratio (S/N) | 55,000:1 (at 4 cm⁻¹, 1 min scan, MCT detector) |
| Sample Compatibility | Up to 300 mm (12″) silicon wafers |
| Measurement Modes | Specular reflectance, transmission, and attenuated total reflectance (ATR) with wafer-specific accessories |
| Compliance | JEITA RC-2001, JEIDA-6301, SEMI F39, ASTM F1597, ISO 14644-1 (cleanroom-compatible operation) |
| Brand | JASCO |
|---|---|
| Origin | Japan |
| Model | FT/IR-4000, FT/IR-6000 |
| Instrument Type | Laboratory Benchtop FTIR |
| Wavenumber Range | 25,000–10 cm⁻¹ (extended options available) |
| Spectral Resolution | 0.7–0.07 cm⁻¹ |
| Scan Speed | 20 Hz |
| Signal-to-Noise Ratio | 25,000:1 to 55,000:1 |
| A/D Converter | 24-bit |
| Interferometer | Corner-Cube Retroreflector with Real-Time DSP Mirror Control |
| Optical Stability | Active Beam Alignment Compensation |
| Compatibility | Integrated with JASCO IR Microscopes (e.g., IRT-5200, IRT-7200) |
| Brand | JASCO |
|---|---|
| Origin | Japan |
| Instrument Type | Laboratory FTIR Spectrometer |
| Wavenumber Range | 11,500–50 cm⁻¹ |
| Spectral Resolution | 0.4 cm⁻¹ |
| Signal-to-Noise Ratio (S/N) | 35,000:1 |
| Dimensions | 386 mm (W) × 479 mm (D) × 290 mm (H) |
| Interferometer | Sealed, Thermally & Hygroscopically Stabilized Michelson |
| Scan Speed | Up to 80 spectra/sec |
| Detector Compatibility | MCT, InGaAs, DTGS |
| Sample Compartment Width | 200 mm |
| Software | Spectrum Manager Ver. 2.5 with Auto-Diagnostic & Guided Workflow |
| Brand | JASCO |
|---|---|
| Origin | Japan |
| Model | FT/IR-6X / FT/IR-8X |
| Instrument Type | Laboratory Benchtop FTIR Spectrometer |
| Spectral Range | 25,000–20 cm⁻¹ |
| Spectral Resolution | 0.07 cm⁻¹ |
| Signal-to-Noise Ratio (S/N) | 55,000:1 (at 4 cm⁻¹, 1 min scan, MCT detector, peak-to-peak) |
| Interferometer | 28° Michelson with Corner Cube Retroreflectors |
| Light Source | High-intensity Ceramic Source |
| Detector | Temperature-Controlled DLATGS or Optional MCT |
| Beam Splitter Exchange | Automated Beam Splitter (BS) Switching Unit |
| Software | Spectrum Manager™ Ver. 2.5 |
| Compliance | Fully Compatible with ASTM E1421, ISO 18382, USP <1119>, GLP/GMP Data Integrity Requirements (21 CFR Part 11 Ready) |
| Brand | JASCO |
|---|---|
| Origin | Japan |
| Instrument Type | Laboratory Benchtop |
| FTIR Architecture | Interferometer-Based Michelson Design |
| Wavenumber Range | 7800–400 cm⁻¹ |
| Spectral Resolution | 0.07 cm⁻¹ (unapodized, with He–Ne laser referencing) |
| Signal-to-Noise Ratio | 55,000:1 (at 4 cm⁻¹ resolution, 1 min scan, MCT detector, peak-to-peak) |
| Vacuum Performance | <10⁻³ Pa base pressure, <5×10⁻⁴ Pa during spectral acquisition |
| Optical Path Environment | Fully evacuated sample compartment and interferometer housing |
| Brand | JASCO |
|---|---|
| Origin | Japan |
| Model | IRT-7200 / IRT-5200 / IRT-1000 |
| Instrument Type | Laboratory FTIR Microscope |
| Wavenumber Range | 25,000–10 cm⁻¹ |
| Spectral Resolution | 0.7–0.07 cm⁻¹ |
| Scan Speed | 20 Hz |
| Signal-to-Noise Ratio | 25,000:1 to 55,000:1 |
| Detector Configuration | 16-channel MCT (IRT-7200), single-element MCT + CMOS visual camera (IRT-5200), compact ATR-integrated module (IRT-1000) |
| Optical Interface | Cassegrain objectives (×16, ×32 standard), ATR imaging capability |
| Software | JASCO Spectra Manager™ with mapping, time-resolved imaging, and co-registered spectral–spatial analysis modules |
| Brand | JASCO |
|---|---|
| Origin | Japan |
| Model | VIR-100 / VIR-200 / VIR-300 |
| Instrument Type | Laboratory FTIR Spectrometer |
| Wavenumber Range | 7,800–350 cm⁻¹ (standard) |
| Spectral Resolution | 0.4–16 cm⁻¹ |
| Interferometer | Sealed, DSP-driven corner-cube retroreflector with auto-alignment and learning-based active stabilization (VIR-200/VIR-300 only) |
| Detectors | Standard DLATGS |
| Scan Speed | Up to 40 Hz (optional) |
| Power Supply | AC adapter or rechargeable battery operation |
| Compliance | Designed for GLP/GMP environments |
| Brand | Jingke Ruida |
|---|---|
| Model | WQF-520 |
| Type | Laboratory FTIR Spectrometer |
| Spectral Range | 400–7800 cm⁻¹ |
| Resolution | <0.85 cm⁻¹ |
| Wavenumber Accuracy | ±0.01 cm⁻¹ |
| Scan Speed | 0.2–2.5 cm/s (motor-controlled, continuously adjustable) |
| SNR | >15,000:1 (RMS, at 2100 cm⁻¹, 4 cm⁻¹ resolution, 60 s acquisition, DTGS detector) |
| Beamsplitter | Ge-coated KBr |
| Detector | Standard DTGS, optional MCT |
| Source | Air-cooled infrared source |
| Brand | Jingpu Technology |
|---|---|
| Origin | Henan, China |
| Model | JP-AFL25B |
| Instrument Type | Portable |
| FTIR Architecture | Interferometric (Michelson-type) |
| Wavenumber Range | 7800–350 cm⁻¹ |
| Spectral Resolution | ≤1 cm⁻¹ |
| Wavenumber Accuracy | ≤0.01 cm⁻¹ |
| Signal-to-Noise Ratio (SNR) | >45,000:1 (at 4 cm⁻¹, 1 min scan, MCT detector) |
| Beam Splitter | Multi-layer coated KBr with hydrophobic coating |
| Detector | Dual-mode DTGS / MCT (user-selectable) |
| IR Source | Long-life air-cooled high-intensity broadband source |
| Scan Speed | Software-selectable, continuously variable (0.1–2.5 cm/s) |
| Optical Platform | Monolithic hardened aluminum base with integrated interferometer and detector chambers |
| Mirror System | Gold-coated corner-cube retroreflectors (monolithic, adhesive-free) |
| Environmental Protection | Industrial-grade resistive humidity control + physical desiccant barrier |
| Data Interface | USB 2.0/3.0 |
| OS Compatibility | Windows 7/10/11 (64-bit) |
| Software | IRPro v5.x (database-driven, ASTM E2412 & EN 14078 compliant modules included) |
| Certified Compliance | ISO 9001, AAA Measurement Management System, EMC-compliant per GB/T 18268.1–2010 (CNAS-verified) |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | LCT-700 |
| Instrument Class | IR Auxiliary Coupling Interface |
| Instrument Type | Benchtop Laboratory System |
| Wavenumber Range | 8000–500 cm⁻¹ |
| Spectral Resolution | 0.5 cm⁻¹ |
| Scan Speed | 20 scans/sec |
| Signal-to-Noise Ratio | 15,000:1 |
| Brand | Leshi |
|---|---|
| Origin | Beijing, China |
| Model | 9100FIR |
| Instrument Type | Portable |
| Spectral Range | 485–8500 cm⁻¹ |
| Resolution Options | 1, 2, 4, or 8 cm⁻¹ |
| Scan Speed | Up to 1 scan/sec (resolution-dependent) |
| Signal-to-Noise Ratio | >20,000:1 (peak-to-peak, 60-sec acquisition), >120,000:1 (RMS, 60-sec acquisition) |
| Wavenumber Accuracy | ±0.01 cm⁻¹ |
| Operating Sample Temperature | Up to 180 °C (hot-wet sampling) |
| Detector | Ambient-temperature DTGS |
| Brand | Leshi |
|---|---|
| Origin | Beijing, China |
| Model | 9100FIR |
| Instrument Type | Portable FTIR Spectrometer |
| Resolution Options | 0.7 cm⁻¹ (unapodized), 1 cm⁻¹, 2 cm⁻¹, 4 cm⁻¹, 8 cm⁻¹ |
| Scan Speed | Adjustable up to 1 scan/sec (dependent on resolution) |
| Signal-to-Noise Ratio (SNR) | >20,000:1 (peak-to-peak, 60-sec acquisition), >120,000:1 (RMS, 60-sec acquisition) |
| Wavenumber Accuracy | ±0.01 cm⁻¹ |
| Operating Sample Temperature | Up to 180 °C (hot-wet gas analysis) |
| Detector | Room-temperature DTGS |
| Reference Laser | VCSEL-based stabilization |
| Internal O₂ Sensor | Zirconia-based electrochemical sensor |
| Particulate Filtration | Dual-stage, <2 µm |
| Nitrogen Purge Interfaces | Dual-path (sample cell reverse purge + optical path purge) |
| Data Output | Real-time spectral & concentration data, Excel-compatible export |
| Compliance | HJ 919–2017, HJ 920–2017, HJ 1011–2018, HJ 1240–2021, HJ 1330–2023 |
| Brand | LISICO |
|---|---|
| Model | Online FTIR Spectrometer |
| Type | Fourier Transform Infrared (FTIR), Online Process Analyser |
| Wavenumber Range | 8000–350 cm⁻¹ (extendable to 12,500–4000 cm⁻¹) |
| Spectral Resolution | < 0.25 cm⁻¹ |
| Wavenumber Accuracy | ≤ 0.01 cm⁻¹ |
| Interferometer | Domestic high-stability立体角镜 interferometer (permanent alignment, >10-year service life) |
| Beam Splitter | Standard KBr-coated mid-IR beam splitter (moisture-resistant option available) |
| Detector | Standard room-temperature DLaTGS |
| IR Source | Long-life ceramic source (1550 K) |
| Probe Material | Hastelloy C-22 |
| Probe Dimensions | Ø12 mm × 300 mm |
| Fiber Type | PIR-900/1000 silver halide fiber |
| Operating Temperature | −100 °C to +140 °C (custom ranges available) |
| Pressure Rating | Up to 200 bar |
| Interface | SMA connector |
| Fiber Length | 1.5 m |
| Minimum Bend Radius | 130 mm |
| Brand | LISICO |
|---|---|
| Model | LS-100 |
| Instrument Type | Laboratory FT-NIR Spectrometer |
| Configuration | Modular Benchtop System |
| Optical Architecture | Michelson Interferometer with Solid-State Laser Reference |
| Spectral Range | 10,000–4,000 cm⁻¹ (1,000–2,500 nm) |
| Resolution | Adjustable down to 4 cm⁻¹ |
| Detector | Thermoelectrically Cooled InGaAs |
| Sampling Modes | Transmission (liquids), Diffuse Reflectance (solids), Non-contact Reflectance (Dual-Eye probe) |
| Connectivity | USB 3.0, Ethernet, Optional Fiber-Optic Interface (up to 8 channels) |
| Software Platform | S-Cal (PLS/QPLS modeling), S-Seq (automated sequence control), S-Online (real-time DCS integration) |
| Brand | LISICO |
|---|---|
| Model | LS-100-Pro |
| Type | Online FT-NIR Spectrometer |
| Channels | 8-channel multiplexed |
| Form Factor | Integrated cabinet system with embedded industrial PC, temperature control module, and PLC for DCS integration |
| Dimensions | 540 mm × 350 mm × 650 mm (L×W×H) |
| Weight | 10 kg (excluding spectrometer optics core) |
| Origin | Shanghai, China |
| Instrument Category | Domestic manufactured |
| Distribution Model | Authorized distributor supply |
| Brand | LUMEX |
|---|---|
| Origin | Russia |
| Model | FT-08 |
| Instrument Type | Laboratory FTIR Spectrometer |
| Wavenumber Range | 8000–350 cm⁻¹ |
| Resolution | 0.25–64 cm⁻¹ (selectable) |
| Scan Speed | 0.8 s at 16 cm⁻¹ |
| SNR | >75,000 (4 cm⁻¹, 1 min, 2150±50 cm⁻¹) |
| Wavenumber Accuracy | ±0.05 cm⁻¹ |
| Detector | High-Sensitivity DLATGS |
| Beamsplitter | KBr-coated Ge (8000–350 cm⁻¹), hermetically sealed and desiccated |
| Interferometer | Patented Suspended Double Cat’s Eye Design |
| Light Source | High-Stability Ceramic IR Emitter |
| Power Consumption | 65 W |
| Dimensions | 580 × 550 × 340 mm |
| Weight | 32 kg |
| Origin | Imported |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Model | M500 |
| Price | Upon Request |
| Wavelength Range | 4000–600 cm⁻¹ |
| Scan Speed Options | 3 / 6 / 12 / 24 min per scan |
| Data Points per Scan | 4,250 (3 min) to 34,000 (24 min) |
| Detector | 1×2 mm DLaTGS (Deuterated L-Alanine doped Triglycine Sulfate), Sensitivity: 1 V/mW |
| Sample Compartment Area | 150 mm² |
| Display | Built-in LCD for wavenumber/absorbance readout |
| Interface | RS-232, analog recorder output |
| Heater Element | NiCr wire on sintered ceramic core |
| Dimensions | 66.7 × 38.7 × 18.4 cm (26.25″ × 15.25″ × 7.25″) |
| Weight | 17.2 kg (38 lbs) |
| Power Supply | 115/230 V AC, 50/60 Hz, 150 W |
| Brand | METTLER TOLEDO |
|---|---|
| Origin | USA |
| Model | ReactIR 701L |
| Instrument Type | In-line / Flow-through Process Spectrometer |
| FTIR Architecture | Michelson Interferometer-Based |
| Wavenumber Range | 4000–650 cm⁻¹ |
| Spectral Resolution | 4 cm⁻¹ |
| Scan Rate | 7 scans per second |
| Signal-to-Noise Ratio (SNR) | 2000:1 |
| Detector | High-Sensitivity, Liquid-Nitrogen-Cooled MCT (Mercury Cadmium Telluride) |
| Probe Compatibility | Fiber-Optic ATR, Transmission, and Reflectance Probes |
| Operating Environment | Lab-scale reactors (glass, stainless steel), flow cells, and microreactors |