| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Fluoride Ion Online Monitor |
| Instrument Type | Fluoride Ion Analyzer |
| Measurement Principle | Potentiometric Ion-Selective Electrode (ISE) |
| Range | 0–20 ppm |
| Accuracy | ±5% of reading or full scale |
| Resolution | 0.001 ppm |
| Repeatability | ±5% of reading or full scale |
| Cycle Time | 2 minutes |
| Sample Flow Rate | 50–900 mL/min |
| Sample Interface | 1/4" compression fitting |
| Sample Temperature | 5–70 °C |
| Max. Suspended Solids | 2% w/w, max. particle size 1.27 mm |
| Sample Pressure | 0–50 psig |
| Waste Outlet | 3/4" NPT male, open drain |
| Alarm Outputs | User-programmable digital outputs |
| Analog Output | 4–20 mA, fully configurable |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Dimensions (W×H×D) | 610 × 640 × 330 mm |
| Weight | ~36 kg (fully assembled) |
| Mounting | Wall-mounted with included hardware |
| Certifications | CE-compliant, designed to meet general requirements for industrial environmental monitoring installations |
| Brand | Galvanic |
|---|---|
| Origin | USA |
| Model | Accuseries LPA Online Analyzer |
| Instrument Type | Online Chloride Analyzer |
| Measurement Range | 0–50 ppm |
| Detection Limit | 0.1 ppm |
| Response Time | 8 min |
| Accuracy | ±1% of reading |
| Operating Temperature Range | −20 °C to +60 °C |
| Communication Interfaces | 4–20 mA, Modbus TCP/IP, RS232, RS458, USB, GUI PC Direct Connection, Relay Outputs |
| Sample Streams | Up to 6 independent flow paths |
| Analytical Method | Potentiometric Titration |
| Enclosure Rating | IP65 (corrosion-resistant, H₂S/Cl₂ tolerant, moisture-proof) |
| Service Life | Designed for continuous 24/7 operation with automated reagent management |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Chromium Ion Online Monitor |
| Price | Upon Request |
| Instrument Type | Multi-Ion Analyzer |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Chromium Ion Online Monitor |
| Price | Upon Request |
| Instrument Type | Multi-Ion Analyzer |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentine Phosphate Online Monitor |
| Instrument Type | Phosphate (Orthophosphate) Analyzer |
| Accuracy | ±2% of reading or full scale |
| Resolution | 0.001 mg/L |
| Repeatability | ±2% of reading or full scale |
| Measurement Cycle | 5 minutes |
| Sample Flow Rate | 50–900 mL/min |
| Sample Interface | 1/4" compression fitting |
| Sample Temperature Range | 5–70 °C |
| Max. Suspended Solids | 2% w/w, max. particle size 1.27 mm |
| Sample Pressure Range | 0–345 kPa (0–50 psig) |
| Drain Outlet | 3/4" NPT male, atmospheric discharge |
| Alarm Outputs | User-programmable digital outputs |
| Analog Output | 4–20 mA, fully configurable |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting | Wall-mounted with included hardware |
| Dimensions (W×H×D) | 387 × 406 × 210 mm |
| Weight | ~36 kg (fully configured) |
| Certifications | CE compliant |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Zinc Ion Online Monitor |
| Pricing | Upon Request |
| Instrument Type | Multi-Ion Analyzer |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Nitrate Online Monitor |
| Instrument Type | Nitrate Ion (NO₃⁻) Analyzer |
| Measurement Principle | Ion-Selective Electrode (ISE) |
| Range Options | 0–10 ppm, 0–50 ppm, 0–100 ppm |
| Accuracy | ±5% of reading or full scale |
| Resolution | 0.001 ppm |
| Repeatability | ±5% of reading or full scale |
| Cycle Time | 2 minutes |
| Sample Flow Rate | 50–900 mL/min |
| Sample Inlet | 1/4″ compression fitting |
| Sample Temperature | 5–70 °C |
| Max Suspended Solids | 2% w/w, max particle size 1.27 mm (0.05″) |
| Sample Pressure | 0–50 psig |
| Waste Outlet | 3/4″ NPT male, atmospheric discharge |
| Alarm Outputs | User-programmable digital outputs |
| Analog Output | 4–20 mA, fully configurable |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Dimensions (W×H×D) | 610 × 640 × 330 mm |
| Weight | ~36 kg (fully assembled) |
| Mounting | Wall-mounted with included hardware |
| Certifications | CE compliant |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Total Manganese Online Monitor |
| Instrument Type | Multi-Ion Analyzer |
| Measurement Principle | Colorimetric Analysis (Formaldehyde Oxime Complexation) |
| Range | Low: 0–25 ppb, High: 0–200 ppb |
| Accuracy | ±2% of reading or full scale |
| Resolution | 0.001 ppb |
| Repeatability | ±2% of reading or full scale |
| Cycle Time | 10 minutes |
| Sample Flow Rate | 50–900 mL/min |
| Sample Inlet | 1/4" compression fitting |
| Sample Temperature | 5–70 °C |
| Suspended Solids | ≤2% w/w, max particle size 1.27 mm |
| Sample Pressure | 0–50 psig |
| Effluent Outlet | 3/4" NPT male, atmospheric discharge |
| Alarm Outputs | User-programmable |
| Analog Output | 4–20 mA, programmable |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting | Wall-mounted with included hardware |
| Dimensions (W×H×D) | 387 mm × 406 mm × 210 mm |
| Weight | ~36 kg |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Cu Ion Online Monitor |
| Instrument Type | Multi-Ion Analyzer |
| Measurement Principle | Colorimetric Spectrophotometry (Bathocuproine or Neocuproine Method) |
| Range | 0–10 ppm (Low), 0–20 ppm (High) |
| Accuracy | ±2% of reading or full scale |
| Resolution | 0.001 ppm |
| Repeatability | ±2% of reading or full scale |
| Cycle Time | 5–15 minutes |
| Sample Flow Rate | 50–900 mL/min |
| Sample Temperature | 5–70 °C |
| Max Suspended Solids | 2% w/w, particle size ≤1.27 mm |
| Sample Pressure | 0–50 psig |
| Sample Interface | 1/4" compression fitting |
| Waste Outlet | 3/4" NPT male, atmospheric discharge |
| Analog Output | 4–20 mA, user-configurable |
| Alarm Outputs | Programmable digital relays |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting | Wall-mounted with integrated hardware |
| Dimensions (W×H×D) | 387 × 406 × 210 mm |
| Weight | ~36 kg |
| Brand | Galvanic |
|---|---|
| Origin | USA |
| Model | Accuseries LPA Online Analyzer |
| Instrument Type | Online Sodium Ion Analyzer |
| Measurement Range | 0–10,000 ppb |
| Detection Limit | 0 ppb |
| Response Time | 10 min |
| Accuracy | ±1% of reading |
| Operating Temperature | −20 to +60 °C |
| Communication Interfaces | 4–20 mA, Modbus TCP/IP, RS232, RS485, USB, GUI PC Direct Connection |
| Sample Streams | Up to 6 modular flow paths |
| Compliance | Designed for continuous operation in industrial water monitoring per ASTM D4192, ISO 10523, and USP <643> guidance for sodium in purified water systems |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentine Sulfide Ion Online Monitor |
| Instrument Type | Multi-Ion Analyzer |
| Measurement Principle | Colorimetric (with DMPD Oxalate Reagent) |
| Range | 0–5 ppm (customizable for higher ranges) |
| Accuracy | ±2% of reading or full scale |
| Resolution | 0.001 ppm |
| Repeatability | ±2% of reading or full scale |
| Measurement Wavelength | 650 nm |
| Reference Wavelength | 810 nm |
| Cycle Time | 15 minutes |
| Sample Flow Rate | 50–900 mL/min |
| Sample Volume per Analysis | 30 mL |
| Sample Temperature Range | 5–70 °C |
| Max. Suspended Solids | 2% w/v, max. particle size 1.27 mm |
| Sample Pressure | 0–50 psig |
| Sample Interface | 1/4" compression fitting |
| Effluent Port | 3/4" NPT male, atmospheric vent |
| Alarm Outputs | User-programmable digital outputs |
| Analog Output | 4–20 mA, fully configurable |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting | Wall-mounted (hardware included) |
| Dimensions (W×H×D) | 387 × 406 × 210 mm |
| Weight | ~36 kg (fully configured) |
| Certifications | CE compliant |
| Brand | Galvanic |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Product Category | Imported Instrument |
| Model | Accuseries LPA Online Analyzer |
| Instrument Type | Online Multi-Ion Analyzer |
| Measurement Range | 0–500 mg/L |
| Detection Limit | 1 ppm |
| Response Time | 10 min |
| Accuracy | ±1% of reading |
| Operating Temperature Range | −20 °C to +60 °C |
| Measurement Principle | Potentiometric Titration |
| Communication Interfaces | 4–20 mA, Modbus TCP/IP, RS232, RS485, USB, GUI PC Direct Connection |
| Relay Outputs | Yes |
| Sample Streams | Up to 6 modular flow paths |
| Enclosure Rating | IP65 (corrosion-resistant, H₂S/Cl₂ tolerant, moisture-proof) |
| Automation Level | Fully Automated (auto-calibration, auto-cleaning, reagent replenishment, 24/7 unattended operation) |
| Brand | Galvanic |
|---|---|
| Origin | USA |
| Model | Accuseries |
| Type | Online Multi-Ion Analyzer |
| Measurement Accuracy | ±1% FS |
| Linearity | ±1% FS |
| Repeatability | ±1% FS |
| Response Time | 8–12 min |
| Operating Temperature Range | −20 °C to +60 °C |
| Dimensions | 1016 mm × 406 mm × 241 mm |
| Weight | 45.3 kg |
| Power Supply | 10–32 VDC or 90–240 VAC |
| Power Consumption | 300 W |
| Analog Outputs | up to eight 4–20 mA |
| Communication Protocols | Modbus RTU (RS-232/RS-485), Modbus TCP/IP |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Sentinel Silicon Online Analyzer |
| Instrument Type | Silica (Orthosilicic Acid) Analyzer |
| Measurement Principle | Colorimetric Spectrophotometry (Molybdenum Blue Method) |
| Detection Wavelengths | 810 nm (low-range), 312 nm (high-range) |
| Range Options | 0–100 ppb, 0–500 ppb, 0–1000 ppb, 0–2 ppm, 0–5 ppm, 0–10 ppm |
| Accuracy | ±2% of reading or full scale |
| Resolution | 0.001 ppb (at lowest range) |
| Repeatability | ±2% of reading or full scale |
| Sample Flow Rate | 50–900 mL/min |
| Sample Interface | 1/4" compression fitting |
| Sample Temperature | 5–70 °C |
| Max Suspended Solids | 2% w/w, particle size ≤ 1.27 mm |
| Sample Pressure | 0–50 psig |
| Drain Outlet | 3/4" NPT male, atmospheric vent |
| Alarm Outputs | User-programmable digital relays |
| Analog Output | 4–20 mA, fully configurable |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting | Wall-mounted with integrated hardware |
| Dimensions (W×H×D) | 387 × 406 × 210 mm |
| Weight | ~36 kg (fully assembled) |
| Brand | Galvanic |
|---|---|
| Origin | USA |
| Model | Accuseries-LPA Online Analyzer |
| Instrument Type | Online Multi-Ion Analyzer |
| Measurement Range | 0–200 ppb (as Ca²⁺/Mg²⁺) |
| Detection Limit | <1 ppb (typical, method-dependent) |
| Response Time | ≤10 min (T90) |
| Accuracy | ±1% of reading |
| Operating Temperature | −20 to +60 °C |
| Communication Interfaces | 4–20 mA, Modbus TCP/IP, RS232, RS485, USB, GUI PC Direct Connection, Relay Outputs |
| Sample Flow Paths | Up to 6 independent streams (modular configuration) |
| Enclosure Rating | IP65 (corrosion-resistant housing, H₂S/Cl₂ tolerant) |