- All
- Favorite
- Popular
- Most rated
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Model | CP610 |
| Temperature Range | –70 °C to +30 °C |
| Temperature Control Accuracy | ±0.1 °C |
| Cooling Capacity | 600 W at 0 °C, 300 W at –40 °C, 100 W at –70 °C |
| Cooling Rate | ~30 °C/h (average) |
| Sample Capacity | 4-position rotating bath |
| Bath Volume | 3.8 L |
| Dimensions | 42 × 53 × 66 cm |
| Weight | ~68 kg |
| Power Supply | 220 V, 10 A, 50/60 Hz, single-phase |
| Compliance | ASTM D97, ASTM D2500, ISO 3016, ISO 2084, IP 15, GB/T 6986, GB/T 3536 |
| Brand | Falex |
|---|---|
| Origin | USA |
| Model | 002‐001‐003 |
| Testing Principle | Timken EP Test (ASTM D2509, ASTM D2782, SH/T 0203) |
| Maximum Applied Load | 92,300 psi (on 0.022 in. wide Timken test block) |
| Rotational Speed | 800 rpm ± 5 rpm (adjustable up to 850 rpm) |
| Test Duration | 10 min ± 15 s |
| Temperature Control Range | Ambient to 120 °C (digital PID controller with fluid circulation system) |
| Oil Reservoir Capacity | 3 L (heated, pumped, valve-controlled |
| Grease Auto-Feeder Rate | 45 ± 9 g/min |
| Drive Motor | 2 HP, unidirectional, with RPM display |
| Load Application | Static load via 10:1 lever arm |
| Sample Types | Liquid lubricants and greases |
| Compliance | ASTM D2509, ASTM D2782, SH/T 0203 |
| Brand | Falex |
|---|---|
| Origin | USA |
| Model | 001-001-331 |
| Test Type | Friction and Wear |
| Load Capacity (Standard) | 805 lbf |
| Load Capacity (Optional) | 1300 lbf |
| Speed Range | Programmable RPM Control |
| Temperature Range (Standard Chamber) | Ambient to 200 °C |
| Temperature Range (Pressurized Chamber, Optional) | Ambient to 250 °C |
| Sub-ambient Testing | Available with External Chiller |
| Friction Force Measurement Range (Standard) | 0–250 lbf |
| Friction Force Measurement Range (Optional Low-Load Sensor) | 0–25 lbf |
| Compliant Standards | ASTM D2714, D2981, D3704, D2509, D2782, G77 |
| Test Geometries | Block-on-Ring, Conforming Block-on-Ring, Timken Canted Cylinder, Chip-on-Ring, Ball-on-Ring |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | TBS2100E-F / TBS3000 |
| Price | Upon Request |
| Dimensions (Viscometer) | 18 × 30.5 × 35.5 cm |
| Dimensions (Control Unit) | 46 × 46 × 15 cm |
| Dimensions (Auto-sampler, optional) | 46 × 51 × 61 cm |
| Weight (Viscometer) | 15.5 kg |
| Weight (Control Unit) | 9 kg |
| Weight (Auto-sampler) | 25.4 kg |
| Voltage | 220–240 V AC |
| Frequency | 50/60 Hz |
| Sample Volume | 50 mL (recommended) |
| Test Duration per Sample | ~5–7 min |
| Temperature Range | 40 °C to 200 °C (±0.1 °C stability) |
| Shear Rate Range | 5×10⁴ s⁻¹ to 7×10⁶ s⁻¹ |
| Speed Steps | 12 discrete settings (800–8000 rpm) |
| Overtemperature Protection | Programmable safety limit |
| Compliance | CE-marked |
| Shipping Weight (Manual System) | 68 kg |
| Shipping Weight (Auto System) | 72.5 kg |
| Shipping Dimensions (Manual) | 61 × 81 × 89 cm |
| Shipping Dimensions (Auto) | 61 × 81 × 127 cm |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | BLB701 / BLB702 / BLB DIN |
| Product Type | Laboratory Instrument |
| Test Parameter | Low-Temperature Apparent Viscosity (ASTM D2983, IP 295, GB/T 6538) |
| Compliance | ASTM D2983, IP 295, GB/T 6538 |
| Dimensions (W×D×H) | 48 × 53 × 66 cm |
| Weight | ~68 kg |
| Voltage | 220 VAC, 50/60 Hz, Single Phase |
| Fuse Rating | BLB701: 10 A |
| BLB702 | 2 × 10 A |
| BLB-DIN | 12 A max |
| Cooling Capacity | 600 W at 0 °C, 300 W at −40 °C, 100 W at −70 °C |
| Heating Capacity | 600 W |
| Temperature Range | BLB701: 30 °C to −40 °C |
| BLB702 | 30 °C to −70 °C |
| BLB-DIN | 30 °C to −55 °C |
| Viewing Window | ~41 × 23 cm |
| Sample Capacity | 12-position bath |
| Bath Volume | 3.8 L |
| Safety Features | High-pressure cutoff, over-temperature limit, low-fluid-level detection and auto-shutdown |
| Certification | CE |
| Shipping Weight | 122–136 kg |
| Shipping Dimensions | 81 × 76 × 104 cm |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | TFAB |
| Product Type | Laboratory Instrument |
| Test Parameter | Foam Characteristics |
| Compliance Standards | ASTM D892, IP 146, ISO 6247 |
| Dimensions (W×D×H) | 61 × 61 × 53 cm |
| Weight | 72.5 kg |
| Power Supply | 220 V, Single-phase, 10 A, 50/60 Hz |
| Heating Medium | Forced Air |
| Motor | DC-driven, maintenance-free |
| Aeration Head | Stainless steel cylindrical head with pre-heated air supply |
| Air Flow System | Dual integrated mass flow meters |
| Air Flow Rate | 94–200 mL/min (±5 mL/min) |
| Temperature Range | 20–175 °C (±0.1 °C) |
| Test Programs | Four standardized sequences (I: 24 °C, II: 93.5 °C, III: 24 °C, IV: 150 °C) |
| Sample Volume | 200 mL |
| Test Capacity | Six 1000-mL graduated cylinders |
| Display | Touchscreen controller with real-time digital readouts, programmable alarms, and elapsed time tracking |
| Safety | Dual thermal protection (fuse + current-limiting fuse), CE certified |
| Shipping Weight/Dimensions | ~90 kg / 71 × 71 × 122 cm |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Model | KV801 / KV802 / KV803 |
| Temperature Range | KV801: +30 °C to –40 °C |
| KV802 | +30 °C to –70 °C |
| KV803 | +40 °C to 150 °C |
| Temperature Stability | ±0.1 °C (digital) / ±0.01 °C (analog) for KV801/KV802 |
| Cooling Capacity | 600 W at 0 °C, 300 W at –40 °C, 100 W at –70 °C |
| Heating Capacity | 1500 W (KV803 only) |
| Bath Volume | ~11.5 L |
| Sample Capacity | 4 capillary viscometers (KV801/KV802), 6 capillary viscometers (KV803) |
| Dimensions (W×D×H) | 42 cm × 53 cm × 66 cm |
| Net Weight | ~68 kg |
| Shipping Weight | 122–136 kg |
| Shipping Dimensions (W×D×H) | 81 cm × 76 cm × 104 cm |
| Power Supply | KV801/KV802: 208–220 V, 10 A, 50/60 Hz |
| KV803 | 220 V, 8 A, 50/60 Hz |
| Refrigerants | R410A (KV801), R507/R508B (KV802) |
| Compressor | 1/3 HP |
| Viewing Window | 41 cm × 23 cm |
| Compliance | ASTM D445, GB/T 265 |
| Brand | Falex |
|---|---|
| Origin | USA |
| Model | FAL400 |
| Compliance | ASTM D3241, IP 323, GB/T 9169 |
| Heating Tube Certification | ASTM D3241 Table 2 Note c, IP 323 Table A.1 |
| Automation | Fully automated air purge (1.5 L/min), pre-test pump priming (60 s), flow verification at start/end |
| Pump System | Dual HPLC-grade precision pumps, pulse-free flow delivery |
| Control System | Embedded microprocessor (non-Windows), corrosion-resistant color touchscreen |
| Safety | Interlocked safety shield with real-time alarm |
| Filtration | Integrated activated carbon fume scrubber |
| Optional Accessory | FAL430 Deposit Rating Ellipsometer (ellipsometric thickness & 3D topography measurement) |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | TBS 3000 |
| Product Type | Laboratory Instrument |
| Test Parameter | HTHS Viscosity |
| Compliance Standards | ASTM D4683, ASTM D6616, SAE J300, GB/T 6538 |
| Dimensions (W × D × H) | 66 × 66 × 48 cm |
| Weight | 38.5 kg |
| Power Supply | 100–240 VAC, ≤6 A, 50/60 Hz |
| Viscosity Range | 1–30 mPa·s (cP) |
| Sample Volume | 40 mL (recommended) |
| Test Duration per Sample | ~5–7 min |
| Auto-sampler Capacity | 1–40 samples per sequence |
| Temperature Control Setpoints | 80 °C, 100 °C, 150 °C (±0.1 °C) |
| Shear Rate Range | 5×10⁵ s⁻¹ to 3×10⁶ s⁻¹ (dependent on fluid viscosity and temperature) |
| Display | Digital touchscreen interface |
| Safety Features | Automatic torque motor and stator heater shutdown upon power interruption |
| Brand | PCS Instruments |
|---|---|
| Origin | United Kingdom |
| Model | ABS |
| Power Supply | 100–230 V AC, 750 VA |
| Operating Temperature | 15–35 °C |
| Operating Humidity | 20–80 % RH (non-condensing) |
| Weight | 34 kg |
| Dimensions | Not specified |
| Compressed Air Requirement | ASTM D5001-compliant (≤0.1 ppm hydrocarbons), dry and oil-free |
| Deionized Water Reservoir Capacity | ~1000 tests per fill |
| Control | Microprocessor-based, fully automated |
| Environmental Compliance | Designed for ISO/IEC 17025-aligned laboratories |
| Software Interface | PC-based control and data logging with audit-trail capability |
| Brand | Falex |
|---|---|
| Origin | USA |
| Model | 450-001-002 |
| Heating Tube Temperature Range | 300–450 °C |
| Temperature Control Accuracy | ±2 °C |
| Flow Rate Range | 0.55–10.0 mL/min |
| Pressure Range | 0–200 psi |
| Test Duration | 2.5–500 hours |
| Sample Reservoir Temperature | 100–210 °C |
| Sample Volume | 100 mL–1 L |
| Compliance | SAE ARP 5996B, SAE AS5780 |
| Brand | Tannas&King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | TanEV WCT |
| Product Type | Laboratory Instrument |
| Compliance | SAE J3200 (in development), ASTM method under preparation |
| Brand | Falex |
|---|---|
| Origin | USA |
| Model | 000-001-305 |
| Test Method Compliance | ASTM D2670, D2625, D3233, D5620 |
| Load Range | 200–3000 lbf (±2% FS) |
| Rotational Speed | 290 ± 10 rpm |
| Torque Measurement | 0–100 in-lb (±2% FS) |
| Wear Resolution | 0.001 inch (0.025 mm) per 14.4 ratchet clicks |
| Operating Temperature | Ambient to 100 °C |
| Brand | PCS Instruments |
|---|---|
| Origin | United Kingdom |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | ETM |
| Product Type | Laboratory Instrument |
| Test Parameter | Friction and Wear |
| Load Range | 100–1650 N |
| Maximum Contact Pressure | 3.5 GPa (standard steel test specimens) |
| Sliding Velocity | up to 3.5 m/s |
| Temperature Range | 10–150 °C |
| Sample Volume | 30 mL (10 mL with optional filler insert) |
| Operating System | Windows 10 |
| Safety Features | Dual PT100 RTD temperature monitoring, emergency stop button, interlocked lubricant reservoir lid |
| Power Supply | 100–240 V, 50/60 Hz, 1600 VA |
| Weight (main unit) | 52 kg |
| Dimensions (H × W × D) | 425 mm × 400 mm × 605 mm |
| Brand | Tannas&King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Product Category | Imported Laboratory Instrument |
| Model | SB+2 |
| Application Scope | Lubricating Oil & Low-Temperature Fluid Characterization |
| Compliance Standards | ASTM D5133, D7110, D2983, D4684, D445, D97, D2500 |
| Temperature Range | +30 °C to –70 °C |
| Temperature Control Accuracy | ±0.1 °C (programmed), ±0.03 °C (manual) |
| Cooling Rate | 1 °C/h (for Brookfield Scan/Gel Index), Isothermal for Brotz–Feldt Viscosity |
| Bath Medium | Methanol (temperature-dependent) |
| Dimensions (W×D×H) | 48 × 61 × 69 cm |
| Net Weight | ~86 kg |
| Power Supply | 220 VAC, 50/60 Hz |
| Safety Certifications | CE |
| Transport Weight (Bath Unit) | ~123 kg |
| Transport Dimensions (Bath Unit) | 66 × 59 × 122 cm |
| Brand | Tannas&King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | TanEV CDT |
| Product Type | Laboratory Instrument |
| Compliance Standards | ASTM D8544, SAE J3200 |
| Brand | Falex |
|---|---|
| Origin | USA |
| Model | 081-001-005 |
| Test Principle | Oscillating Ball-on-Cylinder Tribometry |
| Applied Load | 2450 N (550 lbf) |
| Test Duration | 22 h |
| Oscillation Frequency | 30.0 Hz (1800 cpm) |
| Compliance Standards | ASTM D4170, SH/T 0716 |
| Optional Feature | Load Uniformity Adjustment System |
| Brand | PCS Instruments |
|---|---|
| Origin | United Kingdom |
| Model | HFRR |
| Frequency Range | 10–200 Hz |
| Stroke Range | 20 µm–2.0 mm |
| Load Capacity | 0–1.0 kg (via calibrated weights) |
| Maximum Friction Force | up to 10.0 N (amplitude-dependent) |
| Operating Temperature | ambient to 150 °C (up to 400 °C with optional high-power heating module) |
| Standard Counterface | 10.0 mm diameter × 3.0 mm thick steel disc |
| Standard Indenter | 6.0 mm diameter hardened steel ball |
| Compliance | ASTM D6079, ASTM D7688, ISO 12156-1, SH/T 0765–2021 |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | Quantum |
| Pricing | Upon Request |
| Key | Brand: CFR |
|---|---|
| Origin | USA |
| Model | F1/F2 |
| Application | Gasoline Octane Number (RON/MON) Determination |
| Configuration | Benchtop |
| Standards Compliance | ASTM D2699, ASTM D2700, ISO 5164, ISO 5163, GB/T 5487, GB/T 503 |
| Compression Ratio Range | 4:1 to 18:1 (continuously adjustable) |
| Ignition Timing | 13° BTDC (capacitor discharge system) |
| Knock Detection | Factory-calibrated analog knock meter (0–100 scale), magnetic reluctance sensor mounted in combustion chamber |
| Lubrication | Pressure-fed, external oil pump with filtration, cooling, and flow monitoring |
| Cooling System | Thermosiphon jacketed cylinder with boiling/condensation recirculation |
| Air Intake Humidity Control | 25–50 g H₂O per lb dry air (ASTM-compliant) |
| Safety Interlocks | Low oil pressure, low coolant pressure, low voltage, motor overload |
| Drive Motor | Three-phase synchronous/reluctance motor (constant-speed operation) |
| Carburetor | Single-venturi, level-viewable fuel bowl |
| Cylinder Head | Monolithic, integrated design for thermal stability and measurement repeatability |
| Crankcase | CFR 48D cast iron box-type, service-accessible side and gear covers |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Manufacturer Status | Authorized Distributor |
| Origin Category | Imported |
| Model | TBR |
| Pricing | Upon Request |
| Brand | Tannas & King |
|---|---|
| Origin | USA |
| Model | MRV TP-1 |
| Standards Compliance | ASTM D4684 (Method A), SH/T 0562 |
| Temperature Range (BLB702) | 30 °C to −70 °C |
| Cooling Power | 600 W at 0 °C, 100 W at −40 °C, 300 W at −70 °C |
| Heating Power | 600 W |
| Sample Capacity | 12-position bath |
| Bath Volume | 3.8 L |
| Dimensions (W×D×H) | 48 cm × 53 cm × 66 cm |
| Net Weight | ~68 kg |
| Transport Weight | 122–136 kg |
| Transport Dimensions | 81 cm × 76 cm × 104 cm |
| Electrical Supply | 220 VAC, 50/60 Hz, Single Phase |
| Safety Certifications | CE, High-Temp Cut-off, Low-Level Fluid Detection, Over-Pressure Breaker |
| Viewing Window | 41 cm × 23 cm |
| Brand | Tannas&King |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | SBT DC-II |
| Pricing | Upon Request |
| Dimensions (W×D×H) | 32 × 51 × 74 cm |
| Weight | ~36.3 kg |
| Power Supply | 220–240 VAC, 50/60 Hz |
| Cooling Range | +90 °C to –45 °C (±0.1 °C stability) |
| Max. Cooling Rate | 60 °C/h |
| Heating Capability | Up to 90 °C |
| Sample Capacity | 2 simultaneous tests |
| Control | Single PC can manage up to 4 units |
| Safety Certifications | CE compliant |
| Transport Dimensions | ~81 × 76 × 104 cm |
| Transport Weight | ~125 kg |
| Compliant Standards | ASTM D5133, ASTM D7110, ILSAC GF-2 through GF-6 (A&B), API SL/SM/SN, GB 11121, SH/T 0732, JPI-5S-56-99 |
| Brand | CFR |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | CFR F-5 |
| Configuration | Benchtop |
| Measurement Parameter | Diesel Cetane Number |
| Compliance Standard | ASTM D613, GB/T 386–1991, ASTM D975 (Arbitration Method) |
Show next