BrandHitachi
ModelTMA7100 / TMA7300
Temperature RangeTMA7100: −170 °C to 600 °C
TMA7300Ambient to 1500 °C
Displacement Range±5 mm
RMS Noise / Sensitivity0.005 μm / 0.01 μm
Load Range (Resolution)±5.8 N (9.8 μN)
Heating Rate0.01–100 °C/min
Sample Max DimensionsØ10 mm × 25 mm (TMA7100/TMA7300)
Probe MountingCantilever-type
Atmosphere ControlAmbient air, inert gas, vacuum (~1.3 Pa)
Load Control ModesFixed load, constant-rate loading (9.8×10⁻²–9.8×10⁶ mN/min), frequency sweep (0.001–1 Hz), multi-step program (up to 40 steps)
Displacement Control ModesFixed displacement (±5000 μm), constant-rate displacement (0.01–10⁶ μm/min), frequency sweep (0.001–1 Hz), multi-step program (up to 40 steps)
Optional AccessoriesAutomated liquid nitrogen cooling system (−150 °C to 600 °C), electronic cooler (−60 °C to 450 °C), fan-assisted ambient cooling (RT to 600 °C or 1500 °C), LN₂ Dewar kit (−170 °C to 600 °C), mass flow controller, humidity control module, expansion/compression/penetration/tensile/bending probes (quartz or alumina), sapphire dilatometry fixture
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHitachi
OriginImported
ManufacturerHitachi High-Tech Corporation
Instrument TypeDynamic Mechanical Analyzer (DMA/DMTA)
Temperature Range−150 °C to 600 °C
Frequency RangeSinusoidal mode: 0.01–200 Hz
Composite waveform modesimultaneous 5 frequencies
ComplianceASTM D4065, ISO 6721, ISO 11357-4, USP <1031>, FDA 21 CFR Part 11 ready (with optional audit trail module)