| Brand | Shenkai |
|---|---|
| Origin | Shanghai, China |
| Model | SKY1002-II |
| Operation | Fully Automatic |
| Compliant Standards | ASTM D93, GB/T 261, ISO 2719, IP 34 |
| Flash Point Range | 40–370 °C |
| Temperature Display Resolution | 0.1 °C |
| Heating Rates | Step A: 5–6 °C/min |
| Step B | 1.0–1.5 °C/min |
| Step C | 2.5–3.5 °C/min |
| Stirring Speeds | Step A: 90–120 r/min |
| Step B | 250 ±10 r/min |
| Step C | 90–120 r/min |
| Ignition Modes | Electronic or Gas (LPG, butane, natural gas <10 kPa) |
| Safety Cut-off | EFP+20 °C (flash mode), absolute limit 400 °C |
| Cooling | High-efficiency forced-air system |
| Data Interface | RJ-45 for LIMS integration |
| Power Supply | AC 220 V ±10%, 50/60 Hz |
| Dimensions (W×D×H) | 265 × 525 × 365 mm |
| Net Weight | 20 kg |
| Brand | Shenkai |
|---|---|
| Origin | Shanghai, China |
| Model | SYP1002B-IV |
| Test Method | Closed-Cup (Pensky-Martens) |
| Standards Compliance | GB/T 261, ISO 2719, ASTM D93, IP 34 |
| Heating Power | 500 W |
| Temperature Range | −5°C to 370°C (with three calibrated thermometer ranges) |
| Stirring Speeds | 90–120 rpm and 250 ±10 rpm |
| Ignition Motor Speed | 24 rpm |
| Cooling Rate | 12–14 °C/min |
| Cup Type | Standard Pensky-Martens closed cup (GB/T 261 compliant) |
| Ignition Orifice Diameter | 0.7 mm |
| Voltage Control Range | 0–220 V (digitally set, 3½-digit LCD display) |
| Flame Adjustment | Via adjustable gas flow valve for butane, propane, or town gas |
| Brand | Shenkai |
|---|---|
| Origin | Shanghai, China |
| Model | SYP1001-V-J |
| Test Method | Cleveland Open Cup (COC) |
| Standards Compliance | GB/T 3536, ASTM D92, ISO 2592, SY/T 5645 |
| Cleveland Cup Material | H62 Brass |
| Cup Inner Diameter | Φ63.5 ± 0.5 mm |
| Ignition Source | Liquefied Petroleum Gas (LPG), Butane, or Town Gas |
| Igniter Orifice Diameter | 0.8 mm |
| Flame Sphere Diameter | φ4 mm |
| Heating Element Power | 400 W |
| Temperature Sensor | PT100, Range: 0–410 °C |
| Thermometer | GB/T 514 compliant, -6 to 400 °C, 2 °C graduation |
| Remote Control Distance | ≥10 m |
| Auto-Stop Conditions | 21 °C above expected flash point or sample temperature >400 °C |
| Voltage Compensation | Yes (maintains consistent heating rate under fluctuating line voltage) |
| Brand | Shenkai |
|---|---|
| Model | SKY1001-II |
| Test Method | Cleveland Open Cup (COC) |
| Standards Compliance | ASTM D92, GB/T 3536, ISO 2592, IP 36 |
| Temperature Range | 79–400 °C |
| Display Resolution | 0.1 °C |
| Repeatability | ≤6 °C |
| Reproducibility | ≤8 °C |
| Safety Cutoff Temperatures | EFP+20 °C (Flash), EFP+50 °C (Fire), Absolute Limit: 400 °C |
| Heating Power | 800 W |
| Total Power | 900 W |
| Supply | AC 220 V ±10%, 50/60 Hz |
| Dimensions | 233 × 530 × 427 mm (W×D×H) |
| Weight | 19 kg |
| Cooling | Forced-air |
| Ignition Detection | Ionization flame sensing |
| Pressure Requirement for Fuel Gas | <10 kPa (LPG, butane, natural gas, or town gas) |