- All
- Favorite
- Popular
- Most rated
| Brand | NeuronBC |
|---|---|
| Model | V6.5 |
| Origin | Beijing, China |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (mm) | 400 (D) × 240 (W) × 270/220 (H1/H2) |
| Weight | 8 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusive Flow Accuracy | ±4% |
| Water Intrusion Accuracy | ±0.01 mL |
| Net Volume Test Accuracy | ±4% |
| Bubble Point Range | 100–8,000 mbar |
| Diffusive Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Pre-stored Protocols | 80 |
| Event Log Capacity | ≥16 auditable events with timestamped traceability |
| User Access Levels | 4-tier role-based authentication with electronic signature support |
| Data Storage | ≥5,000 test records, USB backup of full datasets including real-time curves |
| Display | 7" high-resolution color touchscreen |
| Interface Options | RS232, USB, optional industrial fieldbus (e.g., Modbus TCP, Profibus DP) and analog I/O |
| Language Support | English / Chinese |
| Compliance | FDA 21 CFR Part 11, ALCOA+ principles, Annex 11 (EU GMP), ISO 13408-2, ASTM F838, USP <1207> |
| Brand | NeuronBC |
|---|---|
| Model | BGT-100 |
| Origin | Beijing, China |
| Power Supply | AC 220 V, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 0–5000 Pa |
| Pressure Resolution | 0.1 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Dimensions | 370 × 220 × 210 mm |
| Weight | 4.5 kg |
| Data Storage Capacity | ≥10,000 test records |
| External Gas Source | Compressed air, nitrogen, or inert gas (≤0.5 MPa inlet pressure) |
| Audit Trail | Optional |
| Printer | Integrated thermal printer |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | NeuronBC GT-2.0 Online |
| Power Supply | 100–220 VAC, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Pressure Resolution | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Operating Mode | Continuous |
| External Gas Source | Compressed air, nitrogen, or inert gas (max. inlet pressure ≤0.5 MPa) |
| Seal Pressure Capacity | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 400 × 260 × 250 mm |
| Weight | 7.5 kg |
| Origin | USA |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | ME2 |
| Price | USD 198,000 (FOB) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V8.0 |
| Operating Pressure Range | 100–10,000 mbar (150 psi) |
| Pressure Units | mbar, kPa, psi, kgf/cm² |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Enclosure Rating | IP54 |
| Dimensions (L×W×H) | 480 × 300 × 190 mm |
| Weight | 10 kg |
| Compliance | CE Marked, FDA 21 CFR Part 11, GMP-Aligned Audit Trail & Electronic Signature Support |
| Display | 10-inch true-color touchscreen |
| Data Export | USB (raw test data, configuration files, audit logs) |
| Connectivity | RS232, USB |
| Sensor Accuracy | High-precision pressure transducers with low drift and extended pneumatic line compensation up to 100 m |
| Integrated Thermal Printer | Non-ink, particle-free, archival-grade print retention (>10 years) |
| User Management | Four-tier role-based access control with customizable permissions, mandatory electronic signatures, and password-protected login |
| Brand | NeuronBC |
|---|---|
| Model | GT-2.0 |
| Type | Online & Offline Glove Integrity Tester |
| Compliance | GB/T 25915.7–2010 / ISO 14644-7:2004 |
| Power Supply | 100–220 VAC, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Pressure Resolution | ≤50 Pa |
| Operating Noise | ≤50 dB(A) |
| Max Seal Pressure | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 400 × 260 × 250 mm |
| Weight | 7.5 kg |
| External Gas Source | Compressed air, nitrogen, or inert gas |
| Mobility | Portable with rapid-connect fittings |
| Interface | Color touchscreen with time/date stamping |
| Regulatory Conformance | 21 CFR Part 11 (electronic records & signatures), audit trail, role-based access control |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | GT-2.0 |
| Power Supply | 100–220 VAC, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Operating Mode | Continuous |
| External Gas Source | Compressed Air, Nitrogen, or Inert Gas |
| Max Seal Pressure | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 400 × 260 × 250 mm |
| Weight | 7.5 kg |
| Brand | NeuronBC |
|---|---|
| Model | BGT-120 |
| Principle | Pressure Decay Method (ASTM F2095-01) |
| Power Supply | AC 220 V, 50 Hz with Li-ion backup |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 0–5000 Pa |
| Pressure Resolution | 0.1 Pa |
| Response Threshold | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions | 370 × 220 × 210 mm |
| Weight | 4.5 kg |
| Audit Trail | Optional |
| Compliance | ASTM F2095-01, FDA 21 CFR Part 11 (electronic records & signatures), GMP-aligned validation support (IQ/OQ/PQ/SOP) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | Integtest™ V4.0 Portable |
| Power Supply | 100–240 V AC, 50/60 Hz, 100 W |
| Operating Pressure Range | 100–6900 mbar |
| Pressure Resolution | ±1.0 mbar |
| Bubble Point Accuracy | ±75 mbar |
| Test Duration (Rapid Bubble Point) | 8 ± 2 min |
| Test Duration (Bubble Point Scan) | 20 ± 2 min |
| Ambient Operating Conditions | 5–35 °C, 10–80% RH |
| Display | Monochrome LCD, 73 × 39 mm with backlight |
| Data Storage | 20 test records |
| Print Output | Integrated Chinese-language thermal printer |
| Dimensions (W × D × H₁ × H₂) | 400 × 380 × 100 × 235 mm |
| Weight | Approx. 8.2 kg |
| Compliance | Designed to support FDA 21 CFR Part 11 readiness, USP <1207>, ISO 13408-2, and Annex 1 (2022) requirements for sterile filtration validation |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic (China-manufactured) |
| Model | WGT-1200 |
| Price | Upon Request |
| Power Supply | DC 12 V, 20 W |
| Built-in Battery | 12 V Li-ion |
| Detection Accuracy | Minimum Detectable Hole Size: 300 µm |
| Test Duration | 3–8 min |
| Pressure Range | 500–3000 Pa |
| Pressure Resolution | 0.1 Pa |
| Noise Level | ≤50 dB(A) |
| Communication | Wi-Fi |
| Air Source | Integrated Sanitary-Grade Diaphragm Pump |
| Maximum Seal Pressure | ≤200 kPa |
| Data Storage Capacity | >12,000 test records |
| Data Storage Medium | Onboard Flash Memory |
| Glove Identification | RFID or Manual Entry |
| Unit Weight | 4.2 kg |
| Max. Controlled Test Units per Controller | 10 |
| Operating Time | Up to 8 h continuous / >48 h standby |
| User Accounts | 120 max |
| Predefined Test Protocols | 120 max |
| Access Control | Four-Tier Role-Based Permissions |
| Software Compliance | FDA 21 CFR Part 11 (Electronic Records & Signatures), GMP Annex 11, ALCOA+ Principles |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | FIT-V4.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (115 psi) |
| Test Methods | Bubble Point, Diffusion Flow, Pressure Hold, Water Intrusion |
| Accuracy | Bubble Point ±50 mbar |
| Test Range | Bubble Point 590–6900 mbar |
| Display | 5.7″ TFT monochrome LCD with backlight |
| Data Storage | 500 test records |
| Connectivity | RS232 serial interface |
| Language | English / Chinese |
| Operating Environment | ISO Class B cleanroom compliant |
| Dimensions (W×D×H) | 240 × 400 × 270 mm (base) + 220 mm (top module) |
| Weight | 7.5 kg |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (PRC) |
| Model | V1.2 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (max. 115 psi) |
| Bubble Point Accuracy | ±75 mbar |
| Diffusion Flow Accuracy | ±4% |
| Bubble Point Range | 590–6900 mbar |
| Diffusion Flow Range | 1–180 mL/min |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (base) + 220 mm (height with display module) |
| Weight | 7.5 kg |
| Display | 3.2″ monochrome LCD |
| Data Storage | 20 test records |
| Language | Chinese only |
| Test Modes | Bubble Point (fully automated), Diffusion Flow (semi-automated), Hold Pressure (pressure decay) |
| Sample Compatibility | Symmetric & asymmetric membranes |
| Cleanroom Suitability | ISO Class B and above |
| Compliance | Designed to support execution of ASTM F838-22, USP <1207>, EP 2.6.27, and ISO 13408-2 integrity test protocols |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | MFT-1000 |
| Detection Principles | Vacuum Decay & Pressure Decay |
| Compliance | USP <1207>, ASTM F2338-22, FDA 21 CFR Part 11, GMP Annex 1 (2022), ISO 15378:2017 |
| Test Range | −0.1 to 1.0 MPa (selectable ranges) |
| Differential Pressure Sensor | ±2 kPa, accuracy ≤0.5% FS, resolution 0.1 Pa |
| Minimum Detectable Equivalent Orifice Diameter | 1–2 µm |
| Display | 10-inch color capacitive touchscreen |
| Power | AC 100–240 V, 50/60 Hz, ≤200 W |
| Enclosure Rating | IP54 |
| Data Retention | ≥10 years (audit trail + event logs) |
| User Management | 4-tier role-based access control with customizable permissions |
| Output Interfaces | 4–20 mA, RS485, Ethernet, USB, 12 V alarm |
| Data Export | Raw test data, metadata, configuration files via USB |
| Optional Connectivity | Wireless module (Wi-Fi/Bluetooth), OPC UA, MQTT support |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V6.5 Audit-Trail |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar |
| Temperature Range | +5°C to +40°C |
| Relative Humidity | 10–80% |
| Dimensions (mm) | 400(D) × 240(W) × 270(H1) × 220(H2) |
| Weight | 8 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusion Flow Accuracy | ±4% |
| Water Intrusion Accuracy | ±0.01 mL |
| Bubble Point Range | 100–8,000 mbar |
| Diffusion Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Pre-stored Protocols | 80 |
| Event Log Capacity | ≥16 auditable parameters with timestamp |
| User Access Levels | 4-tier role-based authentication with electronic signature support |
| Compliance | Fully aligned with FDA 21 CFR Part 11, ALCOA+ data integrity principles, and ISO 13485 / ASTM F838-22 / USP <1207> guidance for filter integrity testing |
| Brand | Neuronbc |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | NBC-1002 |
| Price | USD 28,000 (FOB Beijing) |
| Excitation Wavelength | 785 nm |
| Laser Power | ≤500 mW (adjustable) |
| Spectral Range | 200–3200 cm⁻¹ |
| Spectral Resolution | ≤8 cm⁻¹ |
| Detector | TE-cooled CCD |
| Battery Life | ≥6 hours (continuous operation) |
| Connectivity | Wi-Fi 5 (802.11ac), Bluetooth 5.0, USB-C |
| Data Storage | Internal 64 GB + cloud-sync capable |
| Compliance | Meets ISO/IEC 17025 calibration traceability requirements |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | FPOSM-V2.0 |
| Measurement Range | 0–2000 mOsm/kg H₂O |
| Sample Volume | 100 µL |
| Analysis Time | ≤3 min |
| Pre-cooling Time | ≤3 min |
| Accuracy | ≤±1% |
| Repeatability | ≤1% |
| Linearity | ≤1% |
| Operating Temperature | 0–30 °C (recommended 15–25 °C) |
| Relative Humidity | ≤60% RH |
| Power Supply | 220 V AC, 50 Hz, 150 VA |
| Dimensions | 225 × 275 × 345 mm |
| Weight | 15 kg |
| Cooling System | Dual-stage thermoelectric (no heat-transfer fluid) |
| Display | 7-inch color capacitive touchscreen |
| Data Storage | 3-year audit-trail compliant archive with timestamped records |
| Calibration | Up to 31-point linear calibration |
| Regulatory Support | CFR 21 Part 11–compatible electronic signatures, user-level access control, GLP/GMP-ready data management |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | FPOSM2.0 |
| Measurement Principle | Freezing Point Depression (based on Raoult’s Law) |
| Measurement Range | 0–2000 mOsmol/kg |
| Sample Volume | 100 µL |
| Test Time | ≤3 min |
| Pre-cooling Time | ≤3 min |
| Accuracy | ≤1% |
| Repeatability | ≤1% |
| Linearity | ≤1% |
| Operating Temperature | 0–30 °C (recommended 15–25 °C) |
| Relative Humidity | ≤60% |
| Power Supply | 220 V AC, 50 Hz, 150 VA |
| Dimensions | 225 × 275 × 345 mm |
| Weight | 15 kg |
| Compliance | Designed to support 21 CFR Part 11-compliant data integrity requirements (electronic signatures, user-level access control, audit-trail-capable data storage) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V8.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (145 psi) |
| Display | 10-inch true-color capacitive touchscreen |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Enclosure Rating | IP54 |
| Dimensions | 480 × 300 × 190 mm |
| Weight | 12 kg |
| Compliance | CE certified, FDA 21 CFR Part 11 compliant, GMP-aligned audit trail & electronic signature support |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | MFT-900 |
| Detection Principle | Vacuum Decay / Differential Pressure Method |
| Compliance | USP <1207>, ASTM F2338-22, FDA Guidance on Container Closure Integrity Testing (CCIT) |
| Automation Level | Semi-Automatic |
| Detection Sensitivity | Equivalent to ≥2 µm defect aperture |
| Pressure Range | −0.1 to 1.0 MPa (selectable ranges: −0.1–0.2 MPa / −0.1–0.7 MPa / −0.1–1.0 MPa) |
| Differential Pressure Sensor | ±2 kPa, Accuracy ≤0.5% FS, Resolution: 0.1 Pa |
| Test Pressure Sensor Accuracy | ≤±1% FS |
| Display | 10-inch color capacitive touchscreen |
| Data Storage | ≥5 years audit-trail-enabled event logs & test records |
| User Management | 4-tier role-based access control (RBAC), password-authenticated login |
| Regulatory Support | Full 21 CFR Part 11 compliance (electronic signatures, audit trail, data integrity) |
| Interface Options | Ethernet, USB, RS485, 4–20 mA analog output, 12 V alarm signal |
| Power Supply | AC 100–240 V, 50/60 Hz, ≤200 W |
| Data Export | USB flash drive backup |
| Optional Peripherals | External thermal printer |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China) |
| Model | 6.0 |
| Price | USD 11,800 (approx.) |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–10,000 mbar (14.5–145 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (mm) | 400 D × 240 W × 270 H₁ × 220 H₂ |
| Weight | 8 kg |
| Bubble Point Accuracy | ±50 mbar |
| Diffusion Flow Accuracy | ±4% |
| Water Intrusion Accuracy | ±0.01 mL |
| Net Volume Test Accuracy | ±4% |
| Bubble Point Range | 100–8,000 mbar |
| Diffusion Flow Range | 1–600 mL/min |
| Water Intrusion Range | 0.01–100 mL/min |
| Pre-stored Protocols | 80 |
| Audit Trail Events | 16+ logged parameters (user, timestamp, method, pressure, flow, pass/fail status, etc.) |
| Data Storage Capacity | >5,000 test records (unlimited with USB backup) |
| Display | 7" high-resolution color touchscreen |
| Interface Options | RS232, USB, optional industrial bus (e.g., Modbus TCP, Profibus DP) and analog I/O |
| Language Support | English & Chinese |
| Compliance | Fully aligned with USP <1207>, EP 2.6.27, ISO 13408-2, and FDA 21 CFR Part 11 requirements for electronic records and signatures |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | V1.2 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (115 psi) |
| Bubble Point Accuracy | ±75 mbar |
| Diffusion Flow Accuracy | ±4% |
| Bubble Point Range | 590–6900 mbar |
| Diffusion Flow Range | 1–180 mL/min |
| Display | 3.2″ monochrome LCD |
| Data Storage | 20 test records |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (main unit) + 220 mm (height with printer) |
| Weight | 7.5 kg |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Compliance | FDA, USP <1207>, ISO 13408-2, Annex 1 (2022), and Chinese Pharmacopoeia (ChP) 2020 Edition |
| Brand | NeuronBC |
|---|---|
| Model | GT-2.0 |
| Origin | Beijing, China |
| Power Supply | AC 220 V, 50 Hz |
| Rated Power | 120 W |
| Test Duration | 2–8 min |
| Pressure Range | 500–5000 Pa |
| Pressure Resolution | ≤50 Pa |
| Noise Level | ≤50 dB(A) |
| Operating Mode | Continuous |
| External Gas Source | Compressed Air, Nitrogen, or Inert Gas |
| Max Sealing Pressure | ≤0.8 MPa |
| Data Storage Capacity | ≥10,000 test records |
| Dimensions (L×W×H₁×H₂) | 400 × 260 × 270 × 250 mm |
| Glove Tray Configuration | Up to 6 trays per unit (customizable) |
| Optional Features | Audit Trail, USB Data Export, Printer Integration |
| Compliance | Designed per GB/T 25915.7–2010 / ISO 14644-7:2004, FDA 21 CFR Part 11–ready (with electronic signature and role-based access control) |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | WGT-1000 |
| Power Supply | DC 12 V, 20 W |
| Test Accuracy | Detects leaks equivalent to ≥300 µm aperture |
| Test Duration | 3–8 min |
| Pressure Range | 500–3000 Pa |
| Pressure Resolution | 0.1 Pa |
| Noise Level | ≤50 dB(A) |
| Communication Method | Wi-Fi |
| Gas Source | Built-in sanitary air pump |
| Maximum Sealing Pressure | ≤200 kPa |
| Data Storage Capacity | >12,000 test records |
| Glove Identification | RFID or manual input |
| Enclosure Rating | IP65 |
| Weight | 4.2 kg |
| Display | Color touchscreen interface |
| Compliance | FDA 21 CFR Part 11, GMP, ISO 14644-7:2004, GB/T 25915.7–2010 |
| Brand | NeuronBC |
|---|---|
| Origin | Beijing, China |
| Model | Integtest V4.0 |
| Power Supply | 100–240 VAC, 50 Hz, 120 W |
| Operating Pressure Range | 100–8000 mbar (115 psi) |
| Environmental Operating Conditions | 5–40 °C, 10–80% RH |
| Dimensions (L×W×H) | 400 × 240 × 270 mm (base) + 220 mm (height extension) |
| Weight | 7.5 kg |
| Display | 5.7″ monochrome TFT LCD with backlight |
| Data Storage | 500 test records |
| Interface | RS232 serial port |
| Language Support | English & Chinese |
| Test Modes | Manual Bubble Point, Standard Bubble Point, Enhanced Bubble Point, Diffusion Flow, Water Intrusion |
| Accuracy | Net Volume ±4%, Bubble Point ±50 mbar, Diffusion Flow ±4%, Water Intrusion ±0.01 mL |
| Measurement Ranges | Bubble Point 590–6900 mbar, Diffusion Flow 1–180 mL/min, Water Intrusion 0.01–30 mL/min |
| Print Output | Integrated thermal micro-printer |
| Filter Compatibility | Symmetric & asymmetric membranes, syringe filters, capsule filters, flat-sheet filters, cartridge filters (≤3 elements, ≤20 inches) |
| Compliance | Designed for ISO 13408-2, ASTM F838-22, and USP <1207> water intrusion method validation |
| Application Context | Offline and online integrity verification in Grade B cleanroom environments |
Show next