| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | LISA SAC 254 |
| Measurement Principle | UV absorbance at 254 nm (SAC₂₅₄) and visible reference at 530 nm |
| Light Source | Dual LED (254 nm UV-C, 530 nm green) |
| Detector | Filtered photodiode |
| Path Length Options | 1, 2, 5, 10, or 50 mm |
| Output Interfaces | Ethernet (TCP/IP), RS-232/RS-485 (Modbus RTU, TriOS protocol), 4–20 mA analog |
| Turbidity Compensation | Internal dual-wavelength correction (530 nm channel) |
| Accuracy | ±0.2% FS |
| Response Time | 4 s |
| Measurement Interval | 2 s |
| Data Logging Capacity | 2 MB (T100 format) |
| Operating Temperature | 2–40 °C (sample & ambient) |
| Storage Temperature | −20–80 °C |
| Flow Velocity Range | 0.1–10 m/s |
| Max. Pressure (SubConn) | 30 bar |
| Max. Pressure (Fixed Cable) | 3 bar |
| Max. Pressure (Flow Cell) | 1 bar (2–4 L/min) |
| Enclosure Material | Stainless Steel 1.4571 / 1.4404 or Titanium Grade 3.7035 |
| IP Rating | IP68 |
| Power Supply | 12–24 VDC (±10%) |
| Power Consumption | ≤1 W |
| Dimensions (10 mm path) | 300 mm × 48 mm |
| Weight (10 mm, stainless steel) | 1.6 kg |
| Compliant With | ISO 7027 (turbidity correlation), ASTM D5540 (SAC-based organic carbon estimation), EN ISO 8466-1 (calibration linearity), FDA 21 CFR Part 11-ready data logging (via TriOS software) |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | microFlu CDOM |
| Light Source | High-efficiency blue LEDs (370 nm) with internal reference photodiode |
| Detector | Photodiode (460 nm, 20 nm FWHM) |
| Measurement Range | 0–20 μg/L and 0–200 μg/L (dual-range auto-adaptive) |
| Detection Limit | 0.02 μg/L |
| Resolution | 0.02 μg/L |
| Accuracy | ±3% of reading |
| Housing | Marine-grade stainless steel (corrosion-resistant to seawater) |
| Dimensions | Ø48 × 200 mm |
| Weight | 0.5 kg |
| Pressure Rating | 500 m water depth |
| Electrical Interface | RS-232 and 0–5 V analog output |
| Connector | SUB-CONN Micro 5-pin male |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | microFlu V2 |
| Excitation/Emission Wavelengths | Chl-a (470/685 nm), Blue-Green Algae (620/655 nm), CDOM (375/460 nm), TRP (275/360 nm) |
| Measurement Range | 0–500 ppb (varies by analyte) |
| Detection Limit | 0.05–3 ppb (analyte-dependent) |
| Accuracy | ±(5% + detection limit) |
| T90 Response Time | 6 s |
| Minimum Measurement Interval | 3 s |
| Digital Interface | RS-485, Modbus RTU |
| Analog Outputs | 4–20 mA (default), 0–5 V, 0–10 V |
| Power Supply | 12–24 VDC (±10%) |
| Typical Power Consumption | 0.6 W |
| Housing Material | Stainless Steel 1.4571/1.4404 or Titanium Grade 3.7035 |
| Dimensions | 162 mm × 48 mm |
| Weight | ~650 g (stainless), ~510 g (titanium) |
| Pressure Rating | SubConn version: 30 bar |
| Fixed-cable version | 3 bar |
| Flow-cell version | 1 bar |
| IP Rating | IP68 |
| Operating Temperature | +2 to +40 °C |
| Storage Temperature | −20 to +80 °C |
| Flow Rate Compatibility | 0.1–10 m/s |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | microFlu-HC |
| Measurement Principle | UV-induced fluorescence (255 nm excitation, 360 nm emission) |
| Target Analytes | Polycyclic Aromatic Hydrocarbons (PAHs) and total petroleum hydrocarbons (TPH) in water |
| Range | PAHs 0–5000 ppb, TPH 0–150 ppm |
| LOD | PAHs 5 ppb, TPH 0.15 ppm |
| Accuracy | ±10 % FS |
| Response Time (T90) | 6 s |
| Min. Measurement Interval | 3 s |
| Digital Interface | RS-485 / Modbus RTU |
| Analog Outputs | 4–20 mA (default), 0–5 V, 0–10 V |
| Power Supply | 12–24 VDC (±10 %) |
| Housing Material | Stainless Steel 1.4571/1.4404 or Titanium Grade 3.7035 |
| Dimensions | 162 mm × 48 mm |
| Weight | ~650 g (SS), ~510 g (Ti) |
| Max. Pressure Rating | 30 bar (SubConn), 3 bar (fixed cable), 1 bar (flow cell) |
| IP Rating | IP68 |
| Operating Temp. | +2 to +40 °C (sample & ambient) |
| Storage Temp. | −20 to +80 °C |
| Flow Rate Compatibility | 0.1–10 m/s |
| Optical Window | Nano-coated quartz |
| Turbidity Compensation | None |
| Data Logging | Internal storage not supported |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | NICO plus |
| Measurement Principle | UV Absorption Spectrophotometry (254 nm) |
| Optical Path Length | 10 mm |
| Detection Method | Dual-wavelength (254 nm + reference) |
| Calibration | Built-in temperature compensation |
| Interface | G2 Web-enabled Ethernet interface |
| Compliance | DIN EN ISO 7027 (turbidity), ASTM D5540 (UVT), USP <643> (TOC estimation), GLP/GMP-ready data logging |
| Sample Handling | In-situ, flow-through or immersion mode |
| Reagent-Free | Yes |
| Optical Window | Nano-coated quartz |
| Data Storage | Internal SD card (optional), real-time streaming via Modbus TCP/RTU, HTTP API |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | NICO |
| Measurement Principle | UV Absorption Spectrophotometry (Multi-Wavelength Compensation) |
| Detection Channels | 4 |
| Wavelengths | 212 nm (NO₃⁻), 254 nm (DOC correction), 360 nm (turbidity correction) |
| Optical Path Lengths | 0.3, 1, 2, 5, 10, 50 mm |
| NO₃-N Range | 0–100 mg/L |
| Accuracy | ±(5% + 0.1 mg/L) |
| Turbidity & DOC Compensation | Yes |
| Response Time | 2 min |
| Minimum Measurement Interval | 1 min |
| Data Storage | 2 GB internal |
| Digital Interfaces | Ethernet (TCP/IP), RS-232/RS-485 (Modbus RTU) |
| Analog Output | 4–20 mA |
| Power Supply | 12–24 VDC (±10%) |
| Max. Power Consumption | ≤8 W |
| Housing Materials | Stainless Steel 1.4571 / 1.4404 or Titanium Grade 2 (3.7035) |
| Dimensions (L × Ø) | 470 mm × 48 mm (for 10 mm path length) |
| Weight | ~3 kg (stainless), ~2 kg (titanium) |
| Pressure Rating | SubConn: 30 bar |
| Fixed Cable | 3 bar |
| Flow Cell | 1 bar |
| Flow Rate | 2–4 LPM |
| Inlet Velocity | 0.1–10 m/s |
| IP Rating | IP68 |
| Operating Temp. | +2 to +40 °C (sample & ambient) |
| Storage Temp. | −20 to +80 °C |
| Calibration/Maintenance Interval | 24 months |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | oFlu Miniature Fluorometer |
| Measurement Principle | Dual-wavelength fluorescence excitation & emission detection |
| Target Analytes | CDOM (colored dissolved organic matter), Chlorophyll a, Phycocyanin, Rhodamine WT, Tryptophan |
| Detection Range | 0–200 µg/L |
| Accuracy | ±3 % FS |
| Response Time | <2 s |
| Sampling Interval | <1 s |
| Housing Material | Stainless Steel 1.4571/1.4404 or Titanium Grade 2 (3.7035) |
| Dimensions | 186 mm × 36 mm (L × Ø) |
| Weight | ~0.6 kg (stainless steel), ~0.5 kg (titanium) |
| Pressure Rating | SubConn connector: 30 bar |
| Fixed cable version | 3 bar |
| Flow cell version | 1 bar (2–4 LPM) |
| IP Rating | IP68 |
| Operating Temperature | +2 °C to +40 °C |
| Storage Temperature | −20 °C to +80 °C |
| Flow Velocity Range | 0.1–10 m/s |
| Power Supply | 12–24 VDC (±10 %) |
| Power Consumption | ≤1 W |
| Digital Interfaces | Ethernet (TCP/IP), RS-232 / RS-485 (Modbus RTU) |
| Analog Outputs | 4–20 mA or 0–5 V |
| Data Storage | ~10 MB internal flash memory |
| Calibration Interval | 24 months |
| Turbidity Compensation | Not supported |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | OPUS aero |
| Optical Path Lengths | 0.3 mm, 1 mm, 2 mm |
| Nitrate-N (NO₃-N) Range | 2.4–120 mg/L (0.3 mm), 0.7–36 mg/L (1 mm), 0.35–18 mg/L (2 mm) |
| Nitrite-N (NO₂-N) Range | 4.4–220 mg/L (0.3 mm), 1.3–67 mg/L (1 mm), 0.65–33.5 mg/L (2 mm) |
| Wavelength Range | 200–360 nm |
| Resolution | 0.8 nm/pixel |
| Detector | 256-channel Miniature Spectrometer |
| Light Source | Xenon Flash Lamp |
| Turbidity Compensation | Yes |
| Accuracy | ±(5% + 0.1 mg/L) |
| Response Time | 2 min |
| Minimum Measurement Interval | 1 min |
| Housing Material | Stainless Steel 1.4571 / 1.4404 |
| Dimensions | 470 × 48 mm |
| Weight | ~3 kg |
| Digital Interfaces | Ethernet (TCP/IP), RS-232/RS-485 (Modbus RTU) |
| Power Supply | 12–24 VDC (±10%) |
| Max. Power Consumption | ≤8 W |
| Pressure Rating | 3 bar (fixed cable), 1 bar (flow cell, 2–4 LPM) |
| IP Rating | IP68 |
| Operating Temperature | +2 to +40 °C (sample & ambient), -20 to +80 °C (storage) |
| Flow Velocity | 0.1–10 m/s |
| Calibration Interval | 24 months |
| Data Storage | 2 GB internal |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | OPUS Deep-Sea Edition |
| Housing Material | Titanium Alloy (Grade 3.7035) |
| Optical Path Length | 10 mm |
| Wavelength Range | 200–360 nm |
| Spectral Resolution | 0.8 nm/pixel |
| Detector | High-Performance Miniature Spectrometer (256 channels) |
| Light Source | Pulsed Xenon Lamp |
| Measurement Principle | Linear Spectral Analysis (LSA) based on Beer–Lambert absorption spectroscopy |
| Turbidity Compensation | Yes |
| Depth Rating | 600 bar / 6000 m |
| IP Rating | IP68 |
| Operating Temperature | +2 °C to +40 °C |
| Storage Temperature | −20 °C to +80 °C |
| Flow Rate Compatibility | 0.1–10 m/s |
| Digital Interface | RS-232 / RS-485 (Modbus RTU) |
| Connector | SubConn 8-pin |
| Power Supply | 12–24 VDC (±10 %) |
| Power Consumption | ≤ 8 W |
| Response Time | 2 min |
| Minimum Measurement Interval | ≥ 1 min |
| Internal Data Storage | 2 GB |
| Calibration Interval | 24 months |
| Seawater-Specific Calibration Modules | OPUS cal seawater series |
| Detection Limits (10 mm path) | NO₃⁻: 0.1 mg/L |
| COD-SACeq | 2.2 mg/L |
| DOC | 0.5 mg/L |
| Quantification Ranges (10 mm path) | NO₃⁻: 0–44 mg/L |
| COD-SACeq | 0–320 mg/L |
| DOC | 0–100 mg/L |
| Accuracy (NO₃⁻) | ±0.73 mg/L (based on certified seawater reference standards) |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | OPUS UV |
| Light Source | Xenon lamp |
| Detector | 256-channel miniature spectrometer (200–360 nm, 0.8 nm/pixel) |
| Measurement Principle | Absorption spectroscopy (attenuation-based) |
| Optical Path Lengths | 0.3 mm, 1 mm, 2 mm, 5 mm, 10 mm, 50 mm |
| Turbidity Compensation | Yes |
| Data Memory | 2 GB internal |
| Measurement Interval | ≥1 minute |
| Output Interfaces | RS232 or RS485 |
| Power Consumption | ≤8 W |
| Housing Material | Stainless steel or titanium alloy (seawater-resistant) |
| Dimensions | Ø48 × 470 mm |
| Weight | 3 kg (stainless steel, 10 mm path) |
| Pressure Rating | 30 bar (with SubConn connector), 3 bar (fixed cable), 1 bar (flow cell, 2–4 L/min) |
| Ingress Protection | IP68 |
| Sample Temperature Range | 2–40 °C |
| Ambient Temperature Range | 2–40 °C |
| Storage Temperature Range | −20–80 °C |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | OPUS |
| Spectral Range | 200–360 nm |
| Optical Path Length Options | 0.3, 1, 2, 5, 10, 50 mm |
| Detection Principle | Linear Spectral Analysis (LSA) based on UV absorption |
| Turbidity Compensation | Yes |
| Data Storage | 2 GB internal |
| Interface | Ethernet (TCP/IP), RS-232/RS-485 (Modbus RTU), Wi-Fi |
| Housing Material | Stainless Steel 1.4571/1.4404 or Titanium 3.7035 |
| IP Rating | IP68 |
| Operating Temperature | +2 °C to +40 °C |
| Power Supply | 12–24 VDC (±10 %) |
| Max. Pressure Rating | 30 bar (SubConn), 3 bar (fixed cable), 1 bar (flow cell) |
| Response Time | ≤ 2 min |
| Measurement Interval | ≥ 1 min |
| Calibration Interval | 24 months |
| Accuracy (NO₃-N, 10 mm path) | ±(5 % + 0.01) mg/L |
| LOD (NO₃-N, 10 mm path) | 0.03 mg/L |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Tcon |
| Price | Upon Request |
| Brand | TriOS |
|---|---|
| Origin | Germany |
| Model | TpH-D |
| Measurement Principle | Potentiometric Analysis |
| pH Range | 0–14 pH |
| Temperature Range | 0–65 °C |
| Resolution | 0.01 pH / 0.1 °C |
| Accuracy | ±0.06 pH / ±0.5 °C |
| Repeatability | ±0.05 pH (pH 7), ±0.1 pH (pH 1 & 13) |
| Drift (24 h) | <0.03 pH |
| Response Time T90 | ≤5 s |
| Temperature Compensation | Pt1000 |
| Output Interface | RS-485, Modbus RTU |
| Power Supply | 12–24 VDC (±10%) |
| Power Consumption | 0.2 W |
| Housing Material | PPS / PET / NBR |
| Dimensions (L×D) | 225 × 32 mm |
| Weight | 180 g |
| IP Rating | IP68 |
| Max Operating Pressure | 3 bar (typical), 1 bar (in flow cell) |
| Flow Rate Compatibility | 0–3 m/s |
| Cable Options | 0.5 m, 2 m, 10 m fixed |
| Maintenance Interval | ≤0.5 h/month |
| Calibration Interval | Every 4 weeks |
| Sample Temp. | +2–+40 °C |
| Ambient Temp. | −5–+55 °C |
| Storage Temp. | +5–+15 °C |
| Brand | Trojan |
|---|---|
| Origin | Imported |
| Manufacturer Type | Authorized Distributor |
| Model | TrojanUVTelos |
| Pricing | Available Upon Request |
| Brand | Trojan |
|---|---|
| Origin | Imported |
| Manufacturer Type | Authorized Distributor |
| Model | TrojanUV3000Plus |
| Pricing | Available Upon Request |
| Brand | Trojan |
|---|---|
| Origin | Imported |
| Manufacturer Type | Authorized Distributor |
| Model | TrojanUVFit |
| Pricing | Available Upon Request |
| Brand | Trojan |
|---|---|
| Origin | Imported |
| Manufacturer Type | Authorized Distributor |
| Model | TrojanUVSigna |
| Pricing | Available Upon Request |
| Brand | Trojan |
|---|---|
| Origin | Imported |
| Manufacturer Type | Authorized Distributor |
| Model | TrojanUVSwiftECT |
| Pricing | Available Upon Request |
| Brand | Trojan |
|---|---|
| Origin | Imported |
| Manufacturer Type | Authorized Distributor |
| Model | TrojanUVSwift™ |
| Pricing | Available Upon Request |
| Brand | Truelab |
|---|---|
| Origin | Shanghai, China |
| Model | WGZ-2-XJ |
| Minimum Readout | 0.01 MCF |
| Measuring Range | 0–6 MCF |
| Basic Error | ±5% F.S. (McFarland Units) |
| Repeatability | ≤1.5% |
| Display | Backlit LCD with Microprocessor Control |
| Interface | RS232 Serial Port |
| Light Source | High-Intensity Stable LED (Rated Lifetime: 100,000 hours) |
| Data Storage | Internal Clock-Synchronized Memory for Calibration Values and Measurement Logs |
| Measurement Modes | Auto-Balanced Reading, Timed-Acquisition, Averaging Mode for Unstable Samples |
| Optical System | Precision Aligned Path with Chromatic Compensation |
| Compliance | Designed to support GLP-compliant workflows in microbiological water testing |
| Brand | TTL |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Category | Domestic |
| Model | TTL-HS |
| Instrument Type | Laboratory Benchtop System |
| Sample Capacity | 4 positions |
| Heating Method | Constant-Temperature Water Bath (Uniform Thermal Distribution) |
| Heating Power | 1000 W |
| Display | Digital LED Interface |
| Compliance Standard | HJ 1226–2021 (PRC National Ecological and Environmental Standard) |
| Brand | TTL |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Category | Domestic |
| Model | TTL-QZS |
| Instrument Type | Laboratory Benchtop System |
| Price Range | USD 7,000 – 14,000 (FOB Beijing) |
| Certifications | Chinese Utility Model Patent ZL202221875739.3 |
| Brand | TTL |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Country of Origin | China |
| Model | TTL-QHS |
| Price Range | USD 7,000 – 14,000 |
| Sample Capacity | 6 positions |
| Heating Method | PID-controlled recirculating water bath |
| Temperature Range | Ambient to 99 °C |
| Temperature Accuracy | ±1 °C |
| Nitrogen Flow Range | 60–800 mL/min per sample |
| Total Mass Flow Range | 0–10 L/min |
| Reaction Vessel | 500 mL × 6 |
| Absorption Vial | 100 mL × 6 |
| Display | 7-inch color touchscreen with graphical interface |
| Power Rating | 1000 W |
| Dimensions (W×D×H) | 1110 × 500 × 800 mm |
| Brand | TUNE VOLT |
|---|---|
| Model | TW-5235 |
| Origin | Beijing, China |
| Measurement Range | 0.05–100 mg/L (three sub-ranges: 0–1, 1–10, 10–100 mg/L) |
| Accuracy | ±5% or ±0.2 mg/L (whichever is greater) |
| Repeatability | ≤±3% |
| Chloride Interference Tolerance | ≤2000 mg/L |
| Digestion Temperature | 125 °C (adjustable from ambient to 180 °C) |
| Digestion Time | 30 min |
| Optical Drift | <0.002 A over 20 min |
| Dimensions | 400 × 310 × 158 mm |
| Weight | 5 kg |
| Display | Backlit LCD with Chinese menu interface |
| Data Storage | 10 calibration curves + 99 measurement records (non-volatile memory) |
| Connectivity | USB port for data export |
| Output | Integrated thermal printer |
| Light Source | High-brightness, long-life (100,000 h) imported LED |
| Brand | TUNE VOLT |
|---|---|
| Model | TW-5258 |
| Origin | Beijing, China |
| Instrument Type | Benchtop Laboratory Analyzer |
| Measurement Range | 0–1000 mg/L |
| Minimum Resolution | 0.01 mg/L |
| Accuracy | ≤ ±2% F.S. |
| Repeatability | ≤ 3% |
| Optical Drift | < 0.002 A in 20 min |
| Dimensions (with printer) | 340 mm × 250 mm × 130 mm |
| Dimensions (without printer) | 266 mm × 200 mm × 130 mm |
| Operating Temperature | 5–45 °C |
| Relative Humidity | ≤ 85% RH |
| Power Supply | AC 220 V ± 10%, 50 Hz ± 0.5 Hz |
| Weight | 4 kg |
| Brand | Turner Designs |
|---|---|
| Origin | USA |
| Model | 10-AU |
| Detection Principle | Dual-beam UV-Vis fluorescence spectroscopy |
| Sensitivity | 10 ppt Rhodamine WT |
| Dynamic Range | 4 decades (0–9999.999 fluorescence units) |
| Optical Path Options | 1 mm, 3 mm, 10 mm, 25 mm flow cells |
| Environmental Rating | IP67-rated enclosure with removable waterproof optical filter chamber |
| Calibration | Single-point auto-ranging calibration |
| Compliance | Designed for ASTM D5867, ISO 10260, and EPA Method 445.0-compatible workflows |
| Brand | Turner Designs |
|---|---|
| Origin | USA |
| Model | AquaFlash |
| Detection Principle | Pulse-Amplitude-Modulated (PAM) Fluorescence |
| Measurement Targets | Total Chlorophyll a Concentration (µg/L) & Maximum Quantum Yield of PSII (Fv/Fm) |
| Detection Limit | 0.30 µg/L |
| Range | 0–100 µg/L |
| Linearity | R² > 0.99 |
| Dimensions | 4.45 cm × 8.9 cm × 18.4 cm |
| Weight | 0.4 kg |
| Operating Temperature | 5–40 °C |
| Warm-up Time | 5 s |
| Detector | Silicon Photodiode (300–1000 nm) |
| Data Storage | 1000 measurements |
| Power | 4 × AAA batteries (>1000 measurements per set) |
| Enclosure Rating | IP67 |
| Sample Cell | 10 × 10 mm square quartz or glass cuvette |
| Calibration | Solid calibration block + liquid calibration standard |
| Data Output | ASCII via USB interface |
| Brand | Turner Designs |
|---|---|
| Model | Aquafluor |
| Origin | USA |
| Type | Portable Dual-Channel Fluorometer and Turbidimeter |
| Detection Channels | Live Chlorophyll a, Extracted Chlorophyll a, Phycocyanin (Cyanobacteria), Phycoerythrin (Red Algae), Rhodamine WT, Fluorescein, Ammonium, CDOM, Optical Brighteners, Turbidity |
| LODs | Chlorophyll a (live): 0.3 µg/L |
| Chlorophyll a (extracted) | 0.5 µg/L |
| Ammonium | 0.1 µmol/L |
| CDOM | 0.1 ppb |
| Cyanobacteria | 150 cells/mL |
| Turbidity | 0.5 NTU |
| Range | Chlorophyll a up to 300 µg/L |
| Resolution | 12-bit |
| Dynamic Range | 3-digit |
| IP Rating | IP67 |
| Operating Temperature | 5–40 °C |
| Dimensions | 4.45 × 8.9 × 18.4 cm |
| Weight | 0.4 kg |
| Power | Two AAA batteries |
| Warm-up Time | <5 seconds |
| Brand | Turner Designs |
|---|---|
| Origin | USA |
| Model | C3 |
| Dimensions | 23 cm L × 10 cm Ø |
| Weight | 1.64 kg (3.6 lbs) |
| Housing Material | Acetal resin |
| Operating Temperature | −2 to 50 °C |
| Max Depth Rating | 600 m |
| Power Supply | 8–30 VDC |
| Interface | RS-232 digital output |
| Minimum Sampling Interval | 1 s |
| Data Storage Capacity | >480,000 records |
| Integrated Sensors | Factory-installed temperature and pressure (depth) sensors |
| Compatible Probes | Up to three Cyclops-7 optical sensors (selectable from: live chlorophyll a, phycocyanin, phycoerythrin, rhodamine WT, CDOM/UV, hydrocarbon oil, turbidity) |
| Brand | Turner Designs |
|---|---|
| Origin | USA |
| Model | C6p |
| Instrument Type | Online Water Quality Analyzer |
| Detection Parameters | Chlorophyll-a, Phycocyanin (Cyanobacteria), Oil-in-Water (Crude & Refined), Turbidity, Rhodamine WT, Fluorescein |
| Housing Material | Polyoxymethylene (POM) |
| Max Operating Depth | 600 m |
| Temperature Range | –2 to 50 °C |
| Power Supply | 8–30 VDC |
| Power Consumption | ≤5 W |
| Data Storage Capacity | >480,000 records |
| Minimum Sampling Interval | 1 s |
| Output | ASCII via RS-232 |
| Ingress Protection | IP68 |
| Dimensions | 685.75 mm (L) × 346 mm (Ø) |
| Weight (air) | 2.54 kg |
| Built-in Temperature Sensor | Yes |
| Optional Depth Sensor | Yes |
| Anti-fouling System | Copper anti-fouling band + motorized mechanical cleaning brush |
| Calibration Standards Available | Yes (for all standard fluorophores and turbidity) |