| Brand | ADVANCE RIKO |
|---|---|
| Origin | Japan |
| Model | F-CAL |
| Measurement Principle | Heat Flow Method |
| Thermal Resistance Range | 1×10⁻⁵ m²·K/W |
| Repeatability | <2% fluctuation |
| Test Atmosphere | Ambient air |
| Hot-Side Temperature Setting | 50°C |
| Sample Geometry | Square, 25 mm or 40 mm edge length, thickness 1–5 mm |
| Instrument Type | Steady-State Thermal Resistance Analyzer |
| Brand | Advance Riko |
|---|---|
| Origin | Japan |
| Model | TC-1200RH |
| Measurement Principle | Laser Flash Method (LFM) |
| Temperature Range | RT to 1150 °C (up to 1200 °C short-term) |
| Thermal Conductivity Range | 0.1–2000 W/m·K |
| Accuracy | ±5% |
| Repeatability | ±3% |
| Sample Dimensions | Ø10 mm × 1–3 mm (thickness), measured through-thickness |
| Atmosphere | Vacuum (standard) |
| Heating Rate | 10 °C/min (to 100 °C), 20 °C/min (to 300 °C), 50 °C/min (to 1150 °C) |
| Dimensions (W×D×H) | 900 × 1050 × 1700 mm |
| Weight | ~350 kg |
| Power Supply | AC200 V, single-phase, 8 kVA (main unit) |
| Cooling Water | ≥5 L/min, pressure ≥0.15 MPa |
| Compliance | JIS R 1611, JIS R 1650-3, JIS H 7801, ISO 13826, ASTM E1461 |