- All
- Favorite
- Popular
- Most rated
| Brand | Nittoseiko Analytech |
|---|---|
| Origin | Japan |
| Model | CA-310 |
| Measurement Principle | Coulometric Karl Fischer Titration |
| Moisture Range | 1 ppm to 100% (5 µg – 999.9999 mg H₂O) |
| Accuracy | ≤ 0.3% RSD (for samples > 1 mg H₂O) |
| Precision | ±3 µg (for 10–1000 µg H₂O) |
| Sensitivity | 0.1 µg H₂O |
| Max Electrolysis Rate | 2620 µg H₂O/min |
| Electrolysis Current | Up to 516 mA |
| Endpoint Detection | Continuous current polarization potential difference |
| Background Correction | Automatic, real-time display |
| Sample Introduction | Manual or via optional vaporizer (VA-300/VA-200 series, etc.) |
| Channels | Up to 4 simultaneous coulometric cells (optional multi-channel control unit) |
| Display | 8.4-inch color TFT-LCD touchscreen (Chinese & English UI) |
| Data Storage | 9999 test records |
| Compliance | ASTM D1533, D3401, D4017, D4928, D6304, D7375, E1064 |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Model | GT-200 |
| Measurement Range | 0–14 pH, −2000 to +2000 mV, 0–20 µA |
| Titrant Dispensing Accuracy | ±0.01 mL (20 mL burette) |
| Resolution | 0.002 mL |
| Titration Resolution | 1/10000 |
| Display | 5.7″ color LCD |
| Operating Temperature | 5–40 °C |
| Humidity | ≤85% RH (non-condensing) |
| Power Supply | 100–240 V AC, 100 VA |
| Dimensions (main unit) | 175 × 408 × 176 mm (W×D×H) |
| Weight | 4 kg |
| GLP Compliance | Full audit trail for calibration, electrode validation, titrant standardization, and user access logs |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | VA-121 |
| Titration Method | Fully Automatic Coulometric Karl Fischer Titration |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Measurement Resolution | ±0.1 µA |
| Water Content Range | 0.5 µg to 999.9999 mg |
| Accuracy | <0.3% RSD |
| Minimum Dispense Volume | 0 µL |
| Titration Control Precision | N/A |
| Heating Temperature Range | 250–1000 °C |
| Heating Module | Fe-Cr-Al Resistance Wire Electronic Furnace (800 W) |
| Max Sample Mass | ≤10 g |
| Detection Limit | ≥5 ppm (as H₂O) |
| Carrier Gas | Nitrogen, Flow Rate: 0.1–0.5 L/min |
| Power Supply | AC 100–120 V / 220–240 V, 50/60 Hz, 1100 VA |
| Dimensions (W×D×H) | ~420 × 260 × 420 mm |
| Weight | ~15.5 kg |
| System Configuration | VA-121 coupled with CA-310 Coulometric Titrator |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Model | CA-200 |
| Measurement Principle | Coulometric Karl Fischer Titration |
| Channel Configuration | Single-channel |
| Water Content Range | 10 µg – 100 mg H₂O |
| Precision | ≤0.3% RSD (for samples ≥1 mg H₂O, n=10) |
| Minimum Dispensing Increment | 1/2000 of burette volume |
| Electrolysis Current | 430 mA |
| Average Titration Rate | 2.2 mg H₂O/min (36 µg H₂O/sec) |
| Display | 5.7-inch color LCD |
| Data Storage | 20 single files + 50 composite files |
| Interface Options | RS-232C (optional), USB host port (dual), balance auto-input via serial/USB, internal I/O for titration cell/vaporizer control |
| Power Supply | AC 100–120 V / 230–240 V, 50/60 Hz |
| Power Consumption | 80 VA (100–120 V), 310 VA (230–240 V) |
| Dimensions (main unit) | 220 × 350 × 360 mm (W×D×H) |
| Weight | ~5 kg |
| Stirrer Dimensions | 120 × 180 × 148 mm |
| Weight | ~1 kg |
| Compliance | GLP/GMP-ready architecture, 21 CFR Part 11-compatible software available (optional) |
| Origin | Japan |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | HR250A |
| Instrument Type | Analytical Balance |
| Capacity Range | 200–500 g |
| Readability | 0.1 mg |
| Calibration Method | Internal Automatic Calibration |
| Maximum Capacity | 252 g |
| Accuracy | ±0.3 mg |
| Repeatability | 0.2 mg |
| Stabilization Time | 2 s |
| Sensitivity Drift | ±2 ppm/°C (10–30 °C) |
| Operating Environment | 5–40 °C, RH <85% non-condensing |
| Display Update Rate | 5 or 10 Hz |
| Counting Mode Minimum Unit | 0.1 mg |
| Sample Count Options | 10 / 25 / 50 / 100 pcs |
| % Weighing Minimum Reference Weight | 10.0 mg |
| % Display Resolution | 0.01% / 0.1% / 1% |
| Interface | RS-232C |
| Software Compatibility | WinCT Data Acquisition Suite |
| External Calibration Weights | 250 g / 200 g / 100 g / 50 g |
| Pan Diameter | Ø90 mm |
| Draft Shield Dimensions | 193(W) × 262.5(D) × 84.5(H) mm |
| Power Supply | AC adapter (~11 VA), region-specific voltage compliance |
| Net Weight | ~3 kg |
| Brand | MITSUBISHI CHEMICAL |
|---|---|
| Origin | Japan |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer Accessory |
| Sample Introduction Method | Boat-type automated vaporization unit |
| Compatible Titrator | CA-310 |
| Moisture Measurement Range | 0.5 µg to 999.9999 mg |
| Measurement Resolution | ±0.1 µA |
| Accuracy | <0.3% RSD |
| Minimum Dispensing Volume | 0 µL |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | VA-236S |
| Titration Method | Fully Automatic Coulometric Titration |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Measurement Resolution | ±0.1 µA |
| Water Content Range | 0.5 µg to 999.9999 mg |
| Accuracy | <0.3% RSD |
| Minimum Dispensing Volume | 0 µL |
| Titration Control Precision | N/A (Coulometric mode, current-controlled electrolysis) |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Model | CA200C |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Water Measurement Range | 10 µg – 100 mg H₂O |
| Accuracy | ≤ 0.3% RSD (for samples ≥ 1 mg H₂O) |
| Minimum Dispensing Volume | 1/2000 of burette volume |
| Titration Speed | 2.2 mg H₂O/min (36 µg H₂O/sec) |
| Electrolysis Current | 430 mA |
| Precision | CV ≤ 0.3% (n = 10, ≥1 mg H₂O) |
| Display | 5.7-inch color LCD |
| Data Storage | 20 single files + 50 composite files |
| Interfaces | Dual USB ports, RS-232C (optional), balance auto-input support |
| Power Supply | AC 100–120 V / 230–240 V, 50/60 Hz |
| Power Consumption | 80 VA (100–120 V), 310 VA (230–240 V) |
| Dimensions (main unit) | 220 × 350 × 360 mm (W×D×H) |
| Weight | ~5 kg |
| Brand | Nittoseiko Analytech |
|---|---|
| Origin | Japan |
| Instrument Type | Volumetric Karl Fischer Titrator |
| Titration Mode | Fully Automatic |
| Measurement Resolution | 0.1 µg |
| Water Content Range | 0.1–999.999 mg |
| Measurement Accuracy | ±0.2 mV, ±0.02 µA |
| Minimum Dispensing Volume | 1/1000 of burette volume |
| Titration Control Precision | 0.1% |
| Display | 8.4-inch color touchscreen |
| Channel Configuration | Single-channel dual-pump system |
| Max. Expandable Channels | 4 (via master console) |
| Sample Heating | Stepwise Temperature Control (Patent No. 3284783) |
| Data Integrity Compliance | GLP/GMP-ready |
| Interface | USB host/device ports for data export/import and method transfer |
| SOP Functionality | Embedded standard operating procedure guidance engine |
| Brand | Nittoseiko Analytech |
|---|---|
| Origin | Japan |
| Model | CA-310EO |
| Titration Method | Fully Automated Coulometric Titration |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Measurement Resolution | ±0.1 μA |
| Water Content Range | 5 μg to 999.9999 mg |
| Accuracy | < 0.3% RSD (for samples > 1 mg H₂O) |
| Sensitivity | 0.1 μg H₂O |
| Maximum Electrolysis Current | 516 mA |
| Average Titration Speed | 2.2 mg H₂O/min (up to 2.64 mg H₂O/min in High-Speed Mode) |
| Endpoint Detection | Continuous Polarization Potential Difference |
| Sample Heating Control | Stepwise Temperature Control (Patent No. 3284783) |
| Display | 8.4-inch Color TFT Touchscreen LCD |
| Data Storage Capacity | 9999 Test Records |
| GLP/GMP Compliance | Full Audit Trail, Electronic Signatures, and Data Integrity Support |
| Connectivity | 4× USB ports, 1× LAN port, Optional Wireless Channel Control |
| Brand | MITSUBISHI CHEMICAL |
|---|---|
| Origin | Japan |
| Model | CA-200 |
| Measurement Principle | Coulometric Karl Fischer Titration |
| Detection Range | 10 µg – 100 mg H₂O |
| Accuracy | CV ≤ 0.3% for samples ≥ 1 mg H₂O (n = 10) |
| Electrolysis Current | 430 mA |
| Average Titration Speed | 2.2 mg H₂O/min (36 µg H₂O/sec) |
| Display | 5.7-inch color LCD |
| Data Storage | 20 user-defined methods × 50 sub-files each |
| Interface | RS-232C (optional), USB storage port, analytical balance input |
| Power Supply | AC 100–120 V / 230–240 V, 50/60 Hz |
| Host Dimensions & Weight | 220 × 350 × 360 mm, ~5 kg |
| Stirrer Dimensions & Weight | 120 × 180 × 148 mm, ~1 kg |
| VA-200 Sample Heater | Temp. Range 70–300 °C, N₂ carrier gas flow 0.1–0.5 L/min, max. sample mass 10 g, weight ~5.5 kg |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Imported |
| Model | CVS-300 |
| Titration Method | Fully Automated Coulometric Karl Fischer Titration |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Current Resolution | ±0.1 µA |
| Water Content Range | 5 µg to 999.9999 mg |
| Measurement Accuracy | <0.3% RSD |
| Minimum Dispense Volume | 0 µL |
| Titration Control Precision | N/A |
| Origin | Japan |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | SV-10 / SV-100 |
| Instrument Type | Oscillating Viscometer |
| Viscosity Range | 0.3–10,000 mPa·s (SV-10), 1,000–10,000 mPa·s (SV-100) |
| Repeatability | ±1% of reading (full scale, S.D., 20–30 °C, no condensation) |
| Sample Volume | ≥35 mL |
| Operating Frequency | 30 Hz sinusoidal fork oscillation |
| Display Units | mPa·s, Pa·s, cP, P (user-selectable) |
| Interface | RS-232C |
| Sensor Material | Stainless steel |
| Temperature Equilibration Time | <5 s |
| Brand | MITSUBISHI CHEMICAL |
|---|---|
| Origin | Japan |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Moisture Range | 10 µg – 100 mg H₂O |
| Accuracy | ≤0.3% RSD (≥1 mg H₂O, n=10) |
| Minimum Dispensing Volume | 1/2000 of burette volume |
| Electrolysis Current | 430 mA |
| Average Titration Speed | 2.2 mg H₂O/min |
| Display | 5.7-inch color LCD |
| Power Supply | AC 100–120 V / 230–240 V, 50/60 Hz |
| Dimensions (Analyzer) | 220 × 350 × 360 mm (W×D×H) |
| Weight | ~5 kg |
| Brand | MITSUBISHI CHEMICAL |
|---|---|
| Origin | Japan |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Moisture Range | 1 ppm – 100% (10 µg – 100 mg H₂O) |
| Accuracy | ≤0.3% RSD (n=10, ≥1 mg H₂O) |
| Minimum Dispensing Volume | 1/2000 of Titrator Volume |
| Electrolysis Current | 430 mA |
| Average Titration Speed | 2.2 mg H₂O/min |
| Display | 5.7-inch Color LCD |
| Power Supply | AC 100–120 V / 230–240 V, 50/60 Hz |
| Dimensions (Analyzer) | 220 × 350 × 360 mm (W×D×H) |
| Weight | ~5 kg |
| VA-210 Vaporizer Temp. Range | 70–199 °C |
| Carrier Gas | N₂, 0.1–0.5 L/min |
| Power | AC 100 V, 160 VA |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Model | VA-122 |
| Heating Zones | Low-Temp Furnace (70–300 °C) & High-Temp Furnace (250–1000 °C) |
| Heating Element | Fe-Cr-Al resistance wire, 1000 W |
| Sample Capacity | ≤10 g |
| Detection Limit | ≥10 ppm (as H₂O) |
| Carrier Gas | N₂, flow rate 0.1–0.5 L/min |
| Power Supply | AC 100–120 V or 220–240 V, 50/60 Hz, 1900 VA |
| Dimensions | 690 × 290 × 410 mm (W×D×H) |
| Weight | ~20 kg |
| Compatible Analyzer | CA-310 Coulometric Karl Fischer Titrator |
| Measurement Principle | Thermal liberation of surface-bound and adsorbed water prior to coulometric KF titration |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Model | PT-200 |
| Form Factor | Benchtop Laboratory Instrument |
| Turbidity Measurement Range | 0–1000 ppm (kaolin), 0–100 ppm (polystyrene) with 10 mm cell |
| Colorimetric Range | 0–10 degrees |
| Resolution | 0.1 NTU |
| Accuracy | ±3% RSD (1–10 ppm, 10 mm cell) |
| Operating Temperature | 0–25 °C |
| Light Source | 12 V / 50 W Halogen Lamp |
| Detector | Silicon Photodiode |
| Integrating Sphere Diameter | Φ80 mm |
| Display | 5.7-inch Full-Color LCD (supporting alphanumeric, kanji, and katakana) |
| Interface | RS-232C, USB |
| Power Supply | AC 100 ±10 V, 50/60 Hz, 118 VA |
| Dimensions | 300(W) × 430(D) × 220(H) mm |
| Weight | ~11 kg |
| Compliance | JIS K 0101, Japanese Waterworks Testing Methods |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | KF-31 |
| Titration Method | Fully Automatic Volumetric Titration |
| Instrument Type | Volumetric Karl Fischer Moisture Analyzer |
| Measurement Resolution | 0.1 µA |
| Water Content Range | 0.01 mg – 999.999 mg H₂O (equivalent to ~10 µg – 999.999 mg absolute water) |
| Accuracy | ≤0.3% RSD |
| Minimum Dispensing Volume | 1/2000 of burette capacity |
| Titration Control Precision | 0.1 µL |
| Burette Capacity | 5 mL |
| Electrode Detection | Continuous Current Polarization (Dual-Platinum Electrode System) |
| Display | 5.7-inch Full-Color LCD Touchscreen |
| Data Storage | Internal memory for ≥1 titration factor calibration, ≥3 user-defined methods |
| Connectivity | USB × 2 (PC, keyboard, USB flash drive, auto fluid handling), RS-232C × 2 (printer, analytical balance, vaporizer units) |
| Compatible Vaporizers | VA-200, VA-300, VA-230, VG-200 (all with stepwise temperature control) |
| Operating Environment | 15–40 °C, ≤85% RH (non-condensing) |
| Power Supply | AC 100/120/220/240 V, 50/60 Hz, 60 VA |
| Dimensions | 220 × 350 × 360 mm (W×D×H) |
| Weight | ~3.8 kg (excluding titration cell and power adapter) |
| Brand | Nittoseiko Analytech |
|---|---|
| Origin | Japan |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Titration Method | Fully Automated Coulometric Titration |
| Measurement Resolution | 0.1 µg H₂O |
| Moisture Range | 5 µg – 999.9999 mg H₂O |
| Accuracy | < 0.3% RSD (for samples > 1 mg H₂O) |
| Minimum Dispensing Increment | 1/2000 of burette volume |
| Titration Control Precision | 0.1 µg H₂O |
| Electrolysis Current | Up to 516 mA |
| Average Titration Speed | 2.2 mg H₂O/min (2.64 mg H₂O/min in High-Speed Mode) |
| Background Correction | Automatic, Real-Time Display |
| Sample Heating Control | Stepwise Temperature Optimization (Patent No. 3284783) |
| Electrode Options | Membrane-Free or Membrane-Type Generator Electrodes |
| Compliance | GLP/GMP Data Integrity Support |
| Interface | 4× USB, 1× LAN, RS-232 for Balance Integration |
| Display | 8.4″ Color TFT Touchscreen |
| Data Storage | 9999 Test Records |
| Scheduled Power-On/Off | 99 Presets |
| Dimensions (Main Unit) | 244 × 161 × 215 mm (W×D×H) |
| Weight (Main Unit) | 2.0 kg |
| Brand | MITSUBISHI CHEMICAL |
|---|---|
| Origin | Japan |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Moisture Range | 1 ppm – 100% w/w (10 µg – 100 mg H₂O) |
| Accuracy | ≤0.3% RSD (for ≥1 mg H₂O, n=10) |
| Minimum Dispensing Volume | 1/2000 of burette volume |
| Electrolysis Current | 430 mA |
| Titration Speed | 36 µg H₂O/sec (avg. 2.2 mg H₂O/min) |
| Display | 5.7-inch color LCD |
| Data Storage | 20 user-defined methods × 50 sub-files |
| Interface | RS-232C (optional), USB storage port, balance input (auto-weighing), internal/external communication ports |
| Power | AC 100–120 V / 230–240 V, 50/60 Hz |
| Dimensions (Analyzer) | 220 × 350 × 360 mm (W×D×H) |
| Weight | ~5 kg |
| VA-200 Vaporizer | Temp. range 70–300 °C, N₂ carrier gas 0.1–0.5 L/min, max. sample mass 10 g, weight ~5.5 kg |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Model | CVS-210 |
| Titration Method | Fully Automated Coulometric Karl Fischer Titration |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Measurement Resolution | ±0.1 μA |
| Water Content Range | 0.5 μg to 999.9999 mg |
| Accuracy | <0.3% RSD |
| Minimum Dispensing Volume | 0 μL |
| Titration Control Precision | 0 (coulometric, current-controlled, no reagent dispensing) |
| Brand | Mitsubishi Chemical |
|---|---|
| Origin | Japan |
| Model | CA-200 |
| Measurement Principle | Coulometric Karl Fischer Titration |
| Method Type | Single-Channel Coulometric |
| Detection Range | 10 µg – 100 mg H₂O |
| Average Titration Speed | 2.2 mg H₂O/min (36 µg H₂O/sec) |
| Electrolysis Current | 430 mA |
| Precision | ≤0.3% RSD (for samples ≥1 mg H₂O, n=10) |
| Data Storage | 20 standalone methods + 50 composite methods |
| Display | 5.7-inch full-color LCD |
| Interfaces | RS-232C (optional), dual USB ports (data export & printer), internal balance auto-input support |
| Power Supply | AC 100–120 V / 230–240 V, 50/60 Hz |
| Power Consumption | 80 VA (100–120 V), 310 VA (230–240 V) |
| Dimensions (Main Unit) | 220 × 350 × 360 mm (W×D×H) |
| Weight | ~5 kg |
| Stirrer Dimensions | 120 × 180 × 148 mm |
| Weight | ~1 kg |
| Compliance | GLP/GMP-ready architecture, 21 CFR Part 11-compatible software available (optional) |
| Origin | Japan |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | GR200 |
| Instrument Type | Analytical Balance |
| Capacity | 210 g |
| Readability | 0.1 mg |
| Calibration | Internal Automatic Calibration with Built-in Weights + External Calibration Support (50 g / 100 g / 200 g / 250 g) |
| Repeatability (SD) | 0.1 g |
| Linearity Error | ±0.2 g |
| Stabilization Time | 2 s |
| Sensitivity Drift | ±2 ppm/°C (10–30 °C) |
| Operating Environment | 5–40 °C, RH <85% non-condensing |
| Display Update Rate | 5 or 10 Hz |
| Counting Mode Minimum Unit | 0.1 mg |
| Sample Count Options | 10 / 25 / 50 / 100 units |
| Interface | RS-232C |
| Software Compatibility | WinCT Data Acquisition Suite |
| Pan Diameter | Ø85 mm |
| Draft Shield Dimensions | 249 × 330 × 328 mm (W×D×H) |
| Power Supply | AC Adapter (~11 VA), Region-Specific Input Voltage |
| Net Weight | ~6 kg |
| Brand | Nittoseiko Analytech |
|---|---|
| Origin | Japan |
| Manufacturer Type | Authorized Distributor |
| Product Category | Imported Instrument |
| Model | CA-310SJ |
| Titration Method | Fully Automatic Coulometric Titration |
| Instrument Type | Coulometric Karl Fischer Moisture Analyzer |
| Measurement Resolution | ±0.1 µA |
| Water Content Range | 5 µg to 999.9999 mg |
| Accuracy | < 0.3% RSD (for samples > 1 mg H₂O) |
| Minimum Dispensing Volume | 1/2000 of burette volume |
| Titration Control Precision | 0.1 µL |
| Electrolysis Current | Up to 516 mA |
| Average Titration Speed | 2.2 mg H₂O/min (2.64 mg H₂O/min in High-Speed Mode) |
| Detection Sensitivity | 0.1 µg H₂O |
| Repeatability | ±3 µg (for 10–1000 µg H₂O) |
| Display | 8.4-inch Color TFT Touchscreen LCD |
| Data Storage Capacity | 9999 test records |
| Environmental Operating Range | 15–40 °C, ≤85% RH (non-condensing) |
| Power Supply | AC 230–240 V, 310 VA, 50/60 Hz |
| Dimensions (Main Unit) | 244 × 161 × 215 mm (W×D×H) |
| Weight (Main Unit) | 2.0 kg |
Show next