| Brand | ESCO |
|---|---|
| Origin | Singapore |
| Model | CCL-170B-8 |
| Heating Method | Direct-heat air-jacketed |
| Temperature Range | Ambient +3 to 60 °C |
| Temp. Uniformity | <±0.2 °C |
| Temp. Accuracy | <±0.1 °C |
| Temp. Recovery (after 1-min door opening) | 6 min |
| CO₂ Range | 0–20% |
| CO₂ Accuracy | ±0.1% |
| CO₂ Sensor | NDIR single-beam dual-wavelength infrared |
| CO₂ Recovery | 4 min |
| Humidity Range | Ambient to 97% RH |
| Humidity Recovery (±5% from setpoint) | 15 min |
| Chamber Volume | 170 L |
| Internal Dimensions (W×D×H) | 505 × 530 × 635 mm |
| External Dimensions (W×D×H) | 660 × 660 × 900 mm |
| Power Supply | 230 V, 50/60 Hz, 3.4 A |
| Max. Power Consumption | 800 W |
| Steady-State Power @ 37 °C | 80 W |
| Net Weight | 120 kg |
| Chamber Material | 304 stainless steel, seamless one-piece construction with large-radius internal corners |
| Exterior Coating | ISOCIDE™ antimicrobial-coated galvanized steel |
| Air Filtration | ULPA filter (≥99.999% @ 0.12 µm), ISO 5 (Class 100) cleanroom-grade chamber air |
| Gas Inlet Filtration | 0.2 µm inline sterilizing filter |
| Sterilization | SwiftCon™ 90 °C wet-heat cycle with automated drying and cooling phases |
| Control System | Smartsense™ microprocessor-based PID controller with 16 MB embedded flash memory |
| Data Logging | Full-parameter event logging with timestamped records accessible via LCD interface |
| Interface | RS-485 standard output |
| Compliance | Designed for GLP/GMP environments |