Water Quality Analysis
Filter
Showing 991–1020 of 3714 results
| Brand | HACH |
|---|---|
| Model | Amtax™ inter2 |
| Measurement Principle | Indophenol Blue Method (DIN 38406-E5) |
| Measurement Ranges | 0.02–2.00 mg/L NH₄–N, 0.1–20.0 mg/L NH₄–N, 1.0–80 mg/L NH₄–N |
| Accuracy | ±2% of reading |
| Cycle Time | 5, 10, 15, 20, or 30 minutes (user-selectable) |
| Calibration | Manual or automated (scheduled or event-triggered) |
| Maintenance Interval | Every 6 months |
| Reagent Consumption | Reagents A/B: 4–8 weeks |
| Zero & Span Standards | 12 months |
| Cleaning Solution | 6–12 months (dependent on cleaning frequency) |
| Analog Outputs | Two isolated 0/4–20 mA outputs (max. load 500 Ω) |
| Alarm Outputs | Two relay outputs (24 VDC, 1 A) |
| Digital Communication | MODBUS RTU (RS-485) or PROFIBUS DP |
| Operating Ambient Temperature | 5–40 °C |
| Power Supply | 230 VAC ±10%, 50 Hz |
| Power Consumption | ~310 VA (including integrated refrigerator) |
| Sample Pretreatment | Compatible with HACH FILTRAX® sampling and filtration system |
| Optical System | Dual-beam photometer with dual interference filters |
| Internal Refrigeration | Yes (for reagent stabilization at 4–8 °C) |
| Brand | Hach |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Portable |
| Measurement Principle | Ion-Selective Electrode (ISE) |
| Sensor Architecture | Integrated CARTRICAL Cartridge with pH, Temperature, and Selective Electrodes |
| Calibration | Factory-Calibrated Cartridge |
| Maintenance Interval | ~2x per year (typical operation) |
| Sample Handling | No pretreatment required |
| Controller Compatibility | SC100 / SC200 / SC1000 Digital Controllers |
| Housing Material | Stainless Steel Body with Ryton End Caps |
| Immersion Rating | Fully Submersible |
| Mounting Options | Rod, Chain, or Wall-Mount Kits |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online TOC Analyzer |
| Detection Principle | Direct Conductivity Measurement |
| Oxidation Method | UV Photolysis (185/254 nm) |
| Measurement Range | 0.05–1999 ppb C |
| Accuracy | ±0.5 ppb or ±5% of reading, whichever is greater |
| Detection Limit | 0.05 ppb C |
| Repeatability | ±0.05 ppb (for TOC < 5 ppb) |
| Conductivity Range for Sample Inlet | 0.05–5 µS/cm (Auto Mode), 0.05–100 µS/cm (Clean Mode) |
| Maximum Inlet Pressure | < 100 psig (690 kPa) |
| Display Resolution | 0.01 ppb (0.00–19.99 ppb), 0.1 ppb (20.0–199.9 ppb), 1 ppb (200–1999 ppb) |
| Sensor Options | S10, S20, or S20P |
| Communication | RS-485 (Modbus RTU) |
| Data Output | Local print support, real-time analog output (4–20 mA), digital interface |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online TOC Analyzer |
| Model | A-1000XP |
| Detection Principle | Direct Conductivity Measurement |
| Oxidation Method | UV Photolysis (185/254 nm) |
| Measurement Range | 0.05–1999 ppb C |
| Accuracy | ±0.05 ppb (MDL to 0.999 ppb) |
| Detection Limit | 0.02 ppb C |
| Conductivity Cell Temperature Control | ±0.01 °C |
| Inlet Water Resistivity Requirement (Auto Mode) | 15.0–18.2 MΩ·cm |
| Inlet Conductivity Range (Clean Mode) | 0.05–100 µS/cm |
| Max Inlet Pressure | 100 psig (690 kPa) |
| Display Resolution | 0.001 ppb C |
| Communication Interfaces | RS-232 (local printing), RS-485 (modbus RTU, up to 1200 m) |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | ASX-7200 |
| Dimensions (H×W×D) | 45.7 cm × 56.9 cm × 53.7 cm |
| Weight | 19.5 kg (43 lbs) |
| Sample Tray Capacity | 36 sealed HACH vials |
| Communication Interface | RS232 |
| Power Supply | 100–240 VAC ±10%, 50/60 Hz, 1.9 A |
| Warranty | 2 years |
| Fluidic Path Materials | PTFE and PEEK (metal-free, inert) |
| Positioning System | Motorized X-Y-Z Cartesian coordinate arm with high-torque stepper motors |
| Environmental Protection | Conformal-coated PCBs, sealed electronics enclosure, plated lead screws |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Status | Manufacturer |
| Model | Anatel PAT700 |
| Measurement Range | 0.5–2,000 ppb C TOC |
| Repeatability | ±0.3 ppb or ±1% (whichever is greater) |
| Accuracy | ±1 ppb or ±5% (whichever is greater) |
| TOC Resolution | 0.1 ppb |
| Conductivity/Resistivity Resolution | 0.01 µS/cm or MΩ·cm |
| Temperature Resolution | 0.1 °C |
| Conductivity Compensation | Temperature-compensated and non-temperature-compensated modes |
| Inlet Pressure | 10–100 psig (69–690 kPa) |
| Flow Rate | 60–300 mL/min |
| Analog Outputs | Three 4–20 mA |
| Digital I/O | Two digital inputs, four digital outputs |
| Communication Interfaces | RS232, USB, Ethernet, Serial printer port |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | TOC600 |
| Instrument Type | Online TOC Analyzer |
| Detection Principle | Direct Conductivity Method |
| Oxidation Method | UV Photolysis |
| Measurement Range | 1–1000 ppb C |
| Accuracy | ±2% (10–40 °C), ±5% (40–50 °C) |
| Resolution | 1 ppb C |
| Repeatability | ±1% or ±1 ppb (whichever is greater) |
| Inlet Conductivity Range | 0.05–150 mS/cm |
| Inlet Pressure | 10–100 psig (69–690 kPa) |
| Flow Rate | 60–300 mL/min |
| Output Options | 4–20 mA, RS-232/RS-485 |
| Operating Environment | Ambient ≤70 °C (heat sink required above 70 °C) |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer | Anatel (a HACH company) |
| Model | UP100 |
| Detection Sizes | 0.1 µm and 0.2 µm (customizable from 0.1–5.0 µm) |
| Sample Flow Rate | 10 mL/min |
| System Flow Rate | >1 L/min |
| Operating Pressure Range | 30–100 psig |
| Light Source | Laser Diode |
| Output | 4–20 mA |
| Clean Recovery Time | <60 seconds |
| Brand | HACH |
|---|---|
| Model | APA 6000 |
| Measurement Principle | Modified Phenol-Hypochlorite Colorimetry (ASTM D1426, ISO 7150-1) |
| Range | 0.02–2 mg/L as N (0.1–10 mg/L as Cl₂) |
| Accuracy | ±5% of reading or ±0.02 mg/L as N (whichever is greater) |
| Repeatability | ±3% of reading or ±0.01 mg/L as N (whichever is greater) |
| Detection Limit | 0.01 mg/L as N (0.05 mg/L as Cl₂) |
| Response Time (90%) | <5 min per cycle |
| Cycle Time | 5.0 min (average) |
| Sample Pressure | 0.03–2.04 bar |
| Sample Temperature | 5–50 °C |
| Sample Flow Rate | 100–2000 mL/min |
| Power Supply | 95–240 VAC, 50/60 Hz ±2 Hz |
| Enclosure Rating | NEMA 4X / IP66 (with air purge interface) |
| Dimensions (W×H×D) | 522 × 627 × 526 mm |
| Weight | 25.5 kg |
| Mounting Options | Wall-mounted, benchtop, or panel-integrated |
| Analog Outputs | 2 × 4–20 mA (expandable to 14) |
| Relay Outputs | 2 alarm relays (expandable to 14) |
| Data Storage | 30-day internal logging with trend and graphical analysis capability |
| Automation | Fully autonomous operation for ≥30 days with auto-calibration, auto-cleaning, and reagent dosing control |
| Sample Channels | Dual independent sample inlets (optional) |
| Brand | HACH |
|---|---|
| Model | APA6000 |
| Measurement Principle | Automated titration pump coupled with colorimetric detection |
| Range (Low) | 0.05–2.0 mg/L Cu²⁺ |
| Range (High) | 1.0–10.0 mg/L Cu²⁺ |
| Accuracy (Low) | ±3% or ±0.03 mg/L (whichever is greater) |
| Accuracy (High) | ±5% or ±0.05 mg/L (whichever is greater) |
| Resolution | 0.001 ppm |
| Repeatability | ±3% of reading or <±0.03 mg/L (whichever is greater) |
| Cycle Time | ~4 minutes per analysis |
| Display | Digital and graphical output calibrated to Cu²⁺ concentration |
| Sample Flow Rate | 20–1000 mL/min |
| Operating Temperature | 5–50 °C |
| Relative Humidity | ≤95% RH, non-condensing |
| Inlet Pressure | 0.17–6.9 bar |
| Calibration Frequency | Configurable from twice daily to once weekly (calibration duration: ~35 min) |
| Analog Outputs | Two isolated 4–20 mA outputs |
| Alarm Outputs | Two SPDT relays |
| Power Supply | 95–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting Options | Panel-mount, benchtop, or wall-mount (hardware included) |
| Data Storage | 30-day onboard history with trend and graphical analysis capability |
| Maintenance Interval | Up to 30 days unattended operation |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | APA6000 |
| Measurement Principle | Flow Injection Analysis (FIA) with Photometric Detection at 600 nm |
| Range | Total Alkalinity: 0–500 mg/L as CaCO₃ |
| Phenolphthalein Alkalinity | 0–250 mg/L as CaCO₃ |
| Accuracy | ±5% of reading or ±1.0 mg/L (whichever is greater) |
| Repeatability | ±3% of reading or ±0.6 mg/L (whichever is greater) |
| Response Time (90%) | <10 min |
| Analysis Cycle | 8 min |
| Auto-Calibration Interval | 60 min |
| Sample Temperature | 5–50 °C |
| Analog Outputs | Two isolated 4–20 mA outputs, fully configurable across full scale |
| Alarm Outputs | Two SPDT relays (230 VAC, 5 A) |
| Power Supply | 95–240 VAC, 50/60 Hz ±2 Hz |
| Enclosure Rating | NEMA 4X (indoor) / IEC 529 IP66 with air-purge interface |
| Drain Connection | Vertical orientation recommended |
| Sample Flow Rate | 100–2000 mL/min |
| Inlet/Outlet Fittings | 1/4" NPT inlet, 3/4" NPT drain |
| Sample Pressure | 0.03–2.04 bar |
| Dimensions (W×H×D) | 522 × 627 × 526 mm |
| Mounting Options | Wall-mounted, benchtop, or panel-mounted |
| Weight | 25.5 kg |
| Data Storage | 30 days of timestamped results with trend/graphing capability |
| Key Features | Fully automated calibration, reagent replenishment, and system cleaning |
| Brand | HACH |
|---|---|
| Model | APA6000 |
| Measurement Principle | Colorimetric EDTA Complexation (USEPA-Approved Method) |
| Low-Range Range | 0.05–10 mg/L as CaCO₃ |
| High-Range Range | 10–1000 mg/L as CaCO₃ |
| Accuracy (Low) | ±5% or ±0.05 mg/L as CaCO₃, whichever is greater |
| Accuracy (High) | ±5% or ±2 mg/L as CaCO₃, whichever is greater |
| Repeatability (Low) | ±3% or ±0.03 mg/L as CaCO₃ |
| Repeatability (High) | ±5% or ±2 mg/L as CaCO₃ |
| T90 Response Time (Low) | ≤5 min |
| T90 Response Time (High) | ≤17 min |
| Cycle Time (Low) | 4 min |
| Cycle Time (High) | 8.2 min |
| Sample Temp. | 5–50 °C |
| Sample Flow Rate | 100–2000 mL/min |
| Inlet Pressure | 0.03–2.04 bar |
| Drain Fitting | 3/4″ NPT |
| Analog Outputs | Dual 4–20 mA (isolated, configurable for hardness or status) |
| Alarm Outputs | Two SPDT relays |
| Certifications | NRTL Listed (UL/CSA), CE Marked |
| Power Supply | 95–240 VAC, 50/60 Hz ±2 Hz |
| Data Storage | 30-day internal logging with timestamped values, trend plotting, and export via serial interface |
| Brand | Hach |
|---|---|
| Origin | USA |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Imported Instrument |
| Model | AS950 Series |
| Instrument Type | Portable / Refrigerated / All-Weather Automatic Water Sampler |
| Sampling Mode | Time-Proportional and Flow-Proportional Sampling |
| Sampling Configuration | Single-Bottle or Multi-Bottle (up to 24 bottles) |
| Bottle Capacity Options | 350 mL (standard base) or 575 mL (compact base) |
| Cooling Performance | Maintains sample temperature at 4 °C |
| Insulation | Dual-layer insulated base with ice retention capability |
| Enclosure Material | Chemical-resistant, UV-stabilized polymer (portable) |
| Environmental Rating | IP65-rated controls |
| Brand | HACH |
|---|---|
| Model | AS950 IO9000 |
| Type | Laboratory Water Sampler |
| Input | (2) 0/4–20 mA analog inputs |
| Output | (3) 0/4–20 mA analog outputs, (4) digital outputs, (1) high-voltage relay (IO9001) or (4) high-voltage relays (IO9004) |
| Operating Temperature | –40 to +50 °C |
| Storage Temperature | –40 to +80 °C |
| Enclosure Rating | NEMA 4X / IP66 |
| Certifications | cETLus |
| Weight | 2.5 kg |
| Dimensions (L×W×D) | 9.71 × 11.82 × 5.12 in |
| Cable Length | 2.74 m |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online TOC Analyzer |
| Model | Astro TOC UV TURBO |
| Measurement Principle | UV-Persulfate Oxidation + NDIR Detection |
| Oxidation Method | Dual-stage oxidation (acidic sparging for inorganic carbon removal followed by 185/254 nm UV photolysis with persulfate) |
| Measurement Range | 0–2.000, 0–5.000, 0–10.000, 0–25.000, 0–50.000 mg/L as C |
| Detection Limit | ≤5 µg/L (at 25°C, 0–5000 µg/L range) |
| Response Time | T₉₀ ≤ 5 min |
| Repeatability | < ±2% RSD (at ≥100 µg/L) |
| Sample Flow Rate | 25–200 mL/min |
| Sample Temperature Range | 2–70°C |
| Operating Ambient Temperature | 5–40°C |
| Inlet Pressure | 0.15–6 bar |
| Enclosure Rating | IP66 / NEMA 4 |
| Certifications | CE, UL 61010-1, CSA C22.2 No. 61010-1 |
| Dimensions (H×W×D) | 981 × 675 × 220 mm |
| Weight | 54 kg |
| Power Supply | 115/230 VAC, 50/60 Hz |
| Analog Outputs | Two isolated 4–20 mA outputs (user-configurable scaling) |
| Digital Communication | RS232/RS485 (Modbus RTU or CSV format) |
| Alarm Functions | Five configurable alarms (TOC concentration, system fault, maintenance due, flow loss, pressure deviation) |
| Carrier Gas | CO₂-free air or nitrogen (2.8–6.2 bar) |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online TOC Analyzer |
| Oxidation Method | UV/Persulfate (Acid + UV) |
| Detection Principle | NDIR (Non-Dispersive Infrared) |
| Measurement Range | 0–2.000 / 5.000 / 10.000 / 25.000 / 50.000 mg/L |
| Detection Limit | ≤5 µg/L (at 25°C, 0–5000 µg/L range) |
| Response Time | T₉₀ ≤ 5 min |
| Sample Flow Rate | 25–200 mL/min |
| Sample Temperature | 2–70°C |
| Operating Temperature | 5–40°C |
| Inlet Pressure | 0.15–6 bar |
| Output | Dual 4–20 mA (user-configurable) |
| Alarms | 5 programmable alarms (TOC concentration, system fault, shutdown) |
| Communication | RS232/RS485 (Modbus RTU, CSV) |
| Power Supply | 115/230 VAC, 50/60 Hz |
| Enclosure Rating | IP66 / NEMA 4 |
| Dimensions (W×H×D) | 981 × 675 × 220 mm |
| Weight | 54 kg |
| Carrier Gas | CO₂-free air or nitrogen (2.8–6.2 bar) |
| Certifications | CE, UL, CSA |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online Analyzer |
| Measurement Principle | UV-Persulfate Oxidation with NDIR Detection |
| Measurement Range | 0–5 to 20,000 mg/L TOC (selectable ranges) |
| Accuracy | ±2% of full scale at 25°C (non-diluted) |
| Repeatability | ±2% of reading at 25°C (non-diluted) |
| Detection Limit | <0.015 mg/L TOC (in 0–5 mg/L range, at 25°C) |
| Response Time | T90 < 8 minutes |
| Carrier Gas | CO₂-free air or nitrogen, 2.8–6.2 bar, 20–200 mL/min |
| Sample Temperature | 2–70°C |
| Operating Temperature | 5–40°C |
| Analog Outputs | Two isolated 4–20 mA outputs (user-configurable range) |
| Alarm Functions | Five configurable alarms (TOC concentration, system fault, shutdown, etc.) |
| Communication Interface | RS-232 or RS-485 (Modbus RTU / ASCII, CSV export) |
| Power Supply | 115/230 VAC, 50/60 Hz |
| Sample Inlet/Outlet | 1/4" OD compression fittings |
| Certifications | CE, UL, CSA |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online TOC Analyzer |
| Model | AstroTOC™ HT |
| Oxidation Method | High-Temperature Catalytic Combustion (Pt-catalyzed, 680 °C) |
| Measurement Range | 0–25 mg/L to 0–20,000 mg/L TOC |
| Accuracy | ±5% of reading (≤1000 mg/L, undiluted, 25 °C) |
| Detection Limit | <0.1 mg/L (at 0–25 mg/L range, 25 °C) |
| Repeatability | ±5% of reading (≤1000 mg/L) or ±2% (2000–20,000 mg/L), undiluted, 25 °C |
| Response Time (T₉₀) | <8 minutes |
| Sample Flow Rate | 20–200 mL/min |
| Sample Temperature Range | 2–70 °C |
| Operating Ambient Temperature | 5–40 °C |
| Inlet Pressure | 0.15–6 bar |
| Power Supply | 115/230 VAC, 50/60 Hz, max. 1500 W |
| Enclosure Rating | IP54 / NEMA 12 |
| Dimensions (H×W×D) | 983 × 973 × 244 mm |
| Weight | 97 kg |
| Certifications | CE, UL 61010-1, CSA C22.2 No. 61010-1 |
| Carrier Gas | CO₂-free air or nitrogen, 2.8–6.2 bar |
| Brand | HACH |
|---|---|
| Model | BIOTECTOR B3500dw |
| Instrument Type | Online TOC Analyzer |
| Oxidation Method | Two-Stage Advanced Oxidation (TSAO) with Hydroxyl Radicals |
| Measurement Principle | NDIR Detection of CO₂ after Acidification and Oxidation |
| Measurement Range (Standard) | 0–250 mgC/L |
| Extended Range (with Overrange Tracking) | 0–1,000 mgC/L |
| Accuracy | ±3% of reading or ±0.45 mgC/L (standard range) |
| Repeatability | Same as accuracy specification |
| Detection Limit | Not specified |
| Resolution | Not specified |
| Sample Particle Size Limit | ≤100 µm |
| Operating Temperature | 0–45 °C |
| Sample Inlet Temperature | 0–60 °C |
| Chloride Tolerance | Up to 30% w/w |
| Humidity Range | 5–85% RH, non-condensing |
| Power Supply | 115/220 VAC |
| Dimensions (H×W×D) | 750 × 500 × 320 mm |
| Weight | 46 kg |
| Data Storage | 9,999 reaction records + 99 error logs |
| Communication Protocols | Modbus®, PROFIBUS |
| Cycle Time | Typically 7.5 minutes |
| Display | 40×16 LED-backlit LCD |
| Languages | English, Spanish, French (others on request) |
| Maintenance Interval | Every 6 months |
| Compliance | EPA-approved oxidation method (U.S. Patent No. 6,623,974 B1) |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer | HACH Company |
| Model | BIOTECTOR B3500e TOC |
| Instrument Type | Online TOC Analyzer |
| Oxidation Method | Two-Stage Advanced Oxidation (TSAO) using hydroxyl radicals |
| Measurement Principle | NDIR detection of CO₂ after acidification and purge |
| Measurement Range (Standard) | 0–250 mg C/L |
| Extended Range (with Overrange Tracking) | 0–1,000 mg C/L |
| Accuracy | ±3% of reading or ±0.45 mg C/L (whichever is greater) for standard range |
| Repeatability | Same as accuracy specification |
| Detection Limit | Not specified |
| Resolution | Not specified |
| Sample Inlet Temperature | 0–60 °C |
| Operating Ambient Temperature | 0–45 °C |
| Relative Humidity | 5–85% non-condensing |
| Chloride Tolerance | Up to 30% w/w |
| Particle Size Limit | ≤100 µm |
| Cycle Time | Typically 7.5 minutes |
| Data Storage | 9,999 reaction records + 99 error logs |
| Display | 40×16 high-contrast LED-backlit LCD |
| Communication Protocols | Modbus®, PROFIBUS |
| Power Supply | 115/220 VAC |
| Dimensions (H×W×D) | 750 × 500 × 320 mm |
| Weight | 46 kg |
| Languages | English, Spanish, French (others available on request) |
| Auto-Cleaning | Integrated TSAO-based self-cleaning per cycle |
| Calibration | Fully automated |
| TIC Measurement | Acidification + sparging prior to NDIR detection |
| TC Measurement | Sum of TOC + TIC |
| Brand | Hach |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Product Category | Imported Instrument |
| Model | BODTrak II |
| Instrument Type | Laboratory BOD Analyzer |
| Measurement Principle | Manometric (Pressure-Difference) Method |
| Measurement Ranges | 0–35, 0–70, 0–350, 0–700 mg/L O₂ |
| Detection Limit | 0.01 mg/L O₂ |
| Accuracy | ±2% at 235 mg/L BOD |
| Repeatability | ±2% at 235 mg/L BOD |
| Incubation Temperature | 20 °C (±0.5 °C) |
| Sample Throughput | 6 samples per run (473 mL bottles) |
| Data Storage | 480 data points per sample |
| Display | Real-time graphical trace with cursor readout |
| Power Supply | AC adapter |
| Dimensions | 28.9 × 26.0 × 9.8 cm |
| Weight | 3.6 kg |
| Brand | Hach |
|---|---|
| Origin | Shanghai, China |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (China-manufactured under Hach licensing) |
| Model | BODTrak II |
| Instrument Type | Laboratory BOD Analyzer |
| Measurement Principle | Pressure Differential Method |
| Measurement Ranges | 0–35, 0–70, 0–350, 0–700 mg/L |
| Detection Limit | 1 mg/L |
| Accuracy | ±1 mg/L |
| Incubation Duration | 5–10 days (user-selectable) |
| Incubation Temperature | 20 °C ±0.5 °C |
| Repeatability | 3–5% RSD |
| Sample Throughput | Up to 6 samples simultaneously |
| Data Logging Interval | Every 15 minutes |
| Total Data Points Storage | 480 |
| Software Interface | HachLink™ PC software (Windows-compatible) |
| Compliance | Designed to meet ASTM D5210, ISO 5815-1, and EPA Method 405.1 requirements for BOD₅ determination |
| Brand | Hach |
|---|---|
| Origin | Imported |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Instrument Type | Laboratory BOD Analyzer |
| Model | BODTrak II |
| Measurement Principle | Manometric (Pressure-Difference) Method |
| Measurement Range | 0–700 mg/L O₂ |
| Accuracy | ±5% of reading |
| Sample Throughput | 6 samples per run (473 mL bottles) |
| Incubation Temperature | 20 °C (±0.5 °C) |
| Data Storage | Up to 480 data points per sample |
| Display | Graphical LCD with real-time BOD curve visualization |
| Sensor | Integrated microprocessor-controlled pressure transducer |
| Power Supply | AC-powered magnetic stirrer base |
| Dimensions | 28.9 × 26.0 × 9.8 cm |
| Compliance | Designed to meet ASTM D5210, ISO 5815-1, and EPA 405-B-16-001 requirements for BOD₅ testing |
| Brand | Hach |
|---|---|
| Origin | USA |
| Model | BODTrak II |
| Measurement Principle | Manometric (pressure-difference) respirometry |
| Sample Capacity | 6 simultaneous samples |
| BOD Range | 0–700 mg/L (four auto-ranging scales: 0–35, 0–70, 0–350, 0–700 mg/L) |
| Resolution | 0.1 mg/L |
| Repeatability | CV ≤ 5% |
| Data Logging | 480 points (15-min interval) |
| Display | Backlit LCD with real-time BOD curve and numeric readout |
| Connectivity | USB interface for HachLink™ software |
| Power Supply | AC 220 V / 50 Hz |
| Compliance | ASTM D5210, ISO 5815-1, US EPA 405.1 (respirometric BOD), GLP-compliant data storage with timestamped audit trail |
| Brand | Hach |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Laboratory BOD Analyzer |
| Principle | Manometric (Pressure-Difference) Method |
| Measurement Ranges | 0–35 mg/L, 0–70 mg/L, 0–350 mg/L, 0–700 mg/L |
| Detection Limit | 0.05 mg/L |
| Accuracy | ±0.1% of reading + 1 LSD |
| Single-Sample Analysis Time | 15 minutes |
| Repeatability | ≤5% |
| Sample Capacity | 6 simultaneous samples |
| Data Storage | 480 data points |
| Incubation Durations | 5, 7, or 10 days |
| Test Bottle Volume | 473 mL |
| Operating Temperature | 20 °C (68 °F) |
| Power Input | 230 V, 50/60 Hz |
| Output Voltage | 20.5 V DC |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | CA610 |
| Price | Upon Request |
| Measurement Range | 0.1–10 mg/L F⁻ |
| Accuracy | ±10% or ±0.10 mg/L (whichever is greater) |
| Detection Limit | 0.01 ppm |
| Cycle Time | 4.2 minutes |
| Analog Output | 1 × 4–20 mA (configurable within 0.1–10 mg/L range) |
| Alarm Outputs | 2 SPDT relays (5 A, 230 VAC), configurable for concentration alarm, system warning, or system shutdown |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Model | CA9400 Single-Channel / Dual-Channel Transmitter |
| Pricing | Available Upon Request |