- All
- Favorite
- Popular
- Most rated
| Brand | PSI (Czech Republic) |
|---|---|
| Origin | Czech Republic |
| Model | AquaPen AP110 |
| Detection Limit | 0.5 µg Chl/L |
| Excitation Wavelengths | 455 nm (blue), 620 nm (red) for AP110-C |
| Emission Detection Range | 667–750 nm |
| Optical Configurations | AP110-C (cuvette-based, 4 mL volume, OD680 & OD720 capability), AP110-P (submersible probe) |
| Time Resolution | 10 µs (100,000 measurements/s) for OJIP kinetics |
| Fluorescence Parameters | >50 standard and derived parameters including F₀, Fₘ, Fᵥ/Fₘ, ΦPSII, NPQ, Rfd, Area, Mo, Sm, PI_ABS, ABS/RC, TR₀/RC, ET₀/RC, DI₀/RC, and 26 OJIP-specific indices |
| Light Sources | Measuring light (0.09 µmol photons/m²·s, 10–100% adjustable), Actinic light (10–1000 µmol photons/m²·s), Saturating pulse (up to 3000 µmol photons/m²·s, 11–100% adjustable) |
| Data Storage | 16 MB flash memory (≈149,000 data points) |
| Power | Rechargeable 2000 mAh Li-ion battery (48 h continuous operation) |
| Dimensions | 165 × 65 × 55 mm (body only), Weight: 290 g |
| Environmental Rating | 0–55 °C operating, 0–95% RH (non-condensing) |
| Connectivity | Dual-mode Bluetooth 5.0 (20 m range, 3 Mbps max) + USB 2.0 |
| Integrated GPS | Yes (1.5 m horizontal accuracy) |
| Software | FluorPen v1.1 (Windows 7+, Excel export, kinetic curve visualization, remote instrument control, audit-ready data logging) |
| Origin | Italy |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Micromac 1000 |
| Instrument Type | Online Water Quality Analyzer |
| Measurement Principle | Colorimetric Flow Analysis (CFA) with Loop Flow Reactor (LFA) |
| Max Parameters per Cycle | 4 |
| Detection Limit | Sub-ppb to ppm range (method-dependent) |
| Sample Volume per Analysis | 20–30 mL |
| Reagent Consumption | < 500 µL per test |
| Dilution Ratio | Up to 1:100 (factory-calibrated) |
| Data Storage Capacity | 400 records (date, time, absorbance, concentration) |
| Output Signals | 4–20 mA (400 Ω load), 0–5 V, RS-232 (RS-485 optional) |
| Power Supply | 12 V DC |
| Power Consumption | 3 W (standby), 10 W (active) |
| Operating Temperature | 5–40 °C |
| IP Rating | IP55 (IP65 optional) |
| Dimensions (L×H×W) | 550 × 350 × 110 mm |
| Weight | 13 kg (excluding reagents) |
| Sample Temp. Range | 10–30 °C |
| Sample Turbidity/Color | No restriction (blank correction applied) |
| pH Range (method-dependent) | 3–12 |
| Reaction Cell Pathlength | 10–50 mm (50 mm for seawater) |
| Detector | Dual-beam silicon photodetector, multi-wavelength LED source |
| Flow System | Integrated peristaltic sampling pump + LFA reactor (10 mL volume) |
| Compliance | EPA, ISO, and DIN standard methods implemented |
| Origin | Imported |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Model | AquaFluo-Eutro Pro |
| Pricing | Available Upon Request |
| Origin | Denmark |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | BM-DAQ |
| Pricing | Upon Request |
| Brand | PSI |
|---|---|
| Origin | Czech Republic |
| Model | MC1000 |
| Channel Count | 8 |
| Culture Vessel Capacity | 100 mL (max. 85 mL working volume) |
| Optical Density Monitoring | OD₆₈₀ & OD₇₂₀ per channel |
| Light Source | Tunable LED array (0–2500 µmol/m²/s), wavelength options include 405 nm, 450 nm, 470/475 nm, 530/540 nm, 590 nm, 615 nm, 660 nm, 730 nm |
| Temperature Control | Water-bath heating (150 W), standard range: ambient +5°C to 60°C |
| optional refrigerated range | 15–60°C |
| Water Bath Volume | 5 L |
| Gas Mixing | Optional GMS150 precision gas blending system (N₂/air/CO₂) |
| Dissolved O₂ & CO₂ Monitoring | Optional 8-channel electrochemical/O₂ sensor and NDIR CO₂ analyzer (0–5% or 0–15% range, resolution ≤1 ppm) |
| Fluorescence Module | Optional probe-based or cuvette-based measurement of Ft, QY, OJIP, NPQ, and light-response curves |
| Software | PBR Control Suite (Basic or Advanced Edition), supporting real-time regression analysis, database logging, GUI-based protocol scheduling, email alerts, and GLP-compliant audit trail (Advanced Edition) |
| Power Supply | 110–240 V AC |
| Dimensions | 71 × 33 × 21 cm |
| Weight | 13 kg |
| Origin | Germany |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | AquapHOx RF-Series |
| Instrument Type | Online Water Quality Analyzer |
| Measured Parameters | Dissolved Oxygen (DO), pH, Temperature |
| Depth Rating | Up to 4000 m |
| Data Storage | 4 GB Industrial-Grade Flash Memory |
| Sensor Technology | RedFlash (RF) Optical Sensing |
| Compliance | Designed for ASTM D888, ISO 5814, ISO 10523, and EU Horizon 2020–validated protocols |
| Origin | Italy |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | EasyChem 200 |
| Pricing | Upon Request |
| Instrument Type | Multi-Parameter Ion Analyzer |
| Origin | Imported |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Model | AOM |
| Price | Available Upon Request |
| Detection Limit | Chlorophyll-a: 30 ng/L |
| Sensitivity | Green algae: 10 cells/mL |
| Cyanobacteria | 100 cells/mL |
| Measured Parameters | Fo, Ft, Fm, Fm′, OJIP transient, Turbidity |
| Actinic & Saturation Light Intensity | 0–3000 µE·m⁻²·s⁻¹ (adjustable) |
| Optical Detector | Photodiode with 660–750 nm bandpass filter |
| Data Interface | RS-232 or USB |
| Onboard Memory | 8 MB integrated data logger |
| IP Rating | IP65 |
| Operating Temperature | 0–45 °C |
| Dimensions | 198 × 60 × 295 mm |
| Weight | 1.8 kg |
| Origin | Imported |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Model | OmniCTRL Water Environment Automated Control System |
| Pricing | Available Upon Request |
| Origin | Italy |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | MICROMAC 1000 |
| Pricing | Upon Request |
| Brand | PSI (Czech Republic) |
|---|---|
| Origin | Czech Republic |
| Model | FMT150 |
| Volume Options | 400 mL / 1000 mL / 3000 mL |
| Temperature Control Range | 5–75 °C (dependent on vessel type) |
| PAR Intensity | Up to 3000 µmol photons·m⁻²·s⁻¹ (configurable per wavelength) |
| Light Sources | Red, Blue, White LEDs (single/dual/multi-spectral) |
| Fluorescence Detection | Dual-modulation chlorophyll fluorometer (F₀, Fₘ, Fₘ′, QY, ΦPSII) |
| Optical Density | OD₆₈₀ & OD₇₂₀ via PIN photodiode + bandpass filter (665–750 nm) |
| Gas Control | Optional GMS high-precision gas mixing system (N₂, air, CO₂, O₂) |
| pH/DO/CO₂ Sensing | Optional calibrated electrochemical sensors |
| Pump Capacity | Up to 8 per unit (PP600 peristaltic pumps) |
| Software | PSI FMT150 Control Suite (Basic & Advanced editions, FDA 21 CFR Part 11–compliant audit trail optional) |
| Compliance | CE-marked |