- All
- Favorite
- Popular
- Most rated
| Brand | Thermo Fisher Scientific |
|---|---|
| Origin | USA |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Origin | Imported |
| Model | 410i |
| Instrument Type | Online Analyzer |
| Target Analyte | CO₂ (single-component gas) |
| Measurement Range (Standard) | 0–200 ppm |
| Resolution | 0.2 ppm |
| Repeatability | ±1.0% of reading |
| Zero Drift (24 h) | ±1.0 ppm |
| Span Drift | <0.5% of reading per 24 hours |
| Linearity Error | ±1.5% of full scale |
| Response Time (t₉₀) | 90 seconds (typical average: 30 seconds) |
| Zero Noise (300 s avg) | 0.1 ppm RMS |
| Detection Limit (Standard Range) | 1 ppm |
| Warm-up Time | <1 hour to operational stability |
| Brand | Thermo Fisher Scientific |
|---|---|
| Origin | USA |
| Manufacturer | Thermo Fisher Scientific |
| Type | Imported |
| Model | 48i |
| Configuration | Online Gas Analyzer |
| Target Analyte | Carbon Monoxide (CO), Single-Component |
| Measurement Range | 0–1 to 0–10,000 ppm (also available in mg/m³ equivalents) |
| Response Time (T90) | 60 s (with 30-s averaging) |
| Zero Drift (24 h) | < 0.1 ppm |
| Span Drift (24 h) | ±1% of full scale |
| Zero Noise | 0.02 ppm RMS (30-s average) |
| Limit of Detection | 0.04 ppm |
| Linearity Error | < ±1% of reading across full range |
| Data Storage | Flash memory with expanded capacity |
| Connectivity | Ethernet-enabled for remote access and control |
| Display | High-resolution backlit LCD with multi-tier menu interface |
| Soft-key Customization | User-definable function keys for rapid navigation |
| Compliance | Designed to meet EPA Performance Specification PS-1 (for ambient CO monitoring), ISO 9001-certified manufacturing, supports GLP/GMP data integrity requirements (audit trail, user authentication, electronic signature readiness per 21 CFR Part 11) |
| Brand | Thermo Scientific |
|---|---|
| Origin | USA |
| Manufacturer | Thermo Fisher Scientific |
| Type | Online Analyzer |
| Detection Principle | Non-Dispersive Infrared (NDIR) with Fixed Bandpass Interference Filter and Quantum Detector |
| Model | 410i |
| Configuration Options | 410i-D (Standard Range), 410i-E (High Range) |
| Measurement Range (D) | 0–10,000 ppm |
| Measurement Range (E) | 0–25% vol |
| Zero Noise (D) | 0.1 ppm RMS (300 s avg) |
| LOD (D) | 0.2 ppm |
| Zero Drift (7 days, D) | <0.20 ppm |
| Span Drift (7 days, D) | <1% of reading |
| Zero Noise (E) | 10 ppm RMS (60 s avg) |
| LOD (E) | 20 ppm |
| Zero Drift (24 h, E) | <40 ppm |
| Span Drift (7 days, E) | <2% of reading |
| Operating Temperature Range | –20 to +50 °C |
| Data Output Options | Direct Extraction or Dilution Mode Reporting |
| Compliance | Designed for EPA Method 3A, ASTM D6522, ISO 12039, and EN 15267-3 environments |
| Software Interface | i-Series Menu-Driven Firmware with Audit Trail Support (21 CFR Part 11 ready) |
| Brand | Thermo Fisher Scientific |
|---|---|
| Origin | USA |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Imported Instrument |
| Model | MIRAN SapphIRe |
| Instrument Type | Portable |
| Detection Target | Multi-component Gases |
| Wavelength Range | 7.7–14.1 µm (tunable filter) + seven fixed-bandpass filters at 1.8, 3.3, 3.6, 4.0, 4.2, 4.5, and 4.7 µm |
| Sample Cell Path Length | 0.5 m |
| Sample Cell Volume | 2.23 L |
| Flow Rate | 15 L/min |
| Analysis Time | 20 s (startup) to 3 min (full spectral acquisition) |
| Response Time | ≤18 s to 90% of final reading |
| Display | 8-line × 40-character LCD |
| Power | Internal rechargeable NiCd battery (7.2 V, 5.7 Ah) |
| Dimensions | 553 mm (W) × 365 mm (H) × 193 mm (D) |
| Weight | ~10 kg |
| Preloaded Calibration Curves | Up to 120 gases (configurable per model variant) |
| Brand | Thermo Fisher |
|---|---|
| Origin | USA |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Origin | Imported |
| Model Variants | 48i, 48i-HL, 48i-TLE |
| Instrument Type | Online Continuous Emission Monitor |
| Measurement Principle | Gas Filter Correlation Infrared (GFC-IR) Absorption |
| Detection Target | Single-component CO gas |
| Range Options | 0–1 to 0–10,000 ppm (48i) |
| Repeatability | ±0.1 ppm |
| Resolution | 0.1 ppm |
| Response Time (T90) | ≤60 s |
| Zero Drift (24 h) | <0.1 ppm (48i/48i-TLE), <4 ppm (48i-HL) |
| Span Drift (24 h) | ±1% FS (48i/48i-TLE), ±2% FS (48i-HL) |
| Linearity Error | ±1% FS (<1000 ppm), ±2.5% FS (>1000 ppm) |
| Zero Noise | 0.02 ppm RMS (30-s avg, 48i/48i-TLE), 2.0 ppm RMS (10-s avg, 48i-HL) |
| MDL | 0.04 ppm (48i/48i-TLE), 4.0 ppm (48i-HL) |
Show next