- All
- Favorite
- Popular
- Most rated
| Brand | Sabio |
|---|---|
| Origin | Beijing, China |
| Model | 4010L |
| Diluent Flow Range | 0–10 SLM (optional up to 20 SLM) |
| Source Gas Flow Range | 0–100 SCCM / 0–10,000 SCCM |
| Flow Accuracy | ≤ ±1% FS |
| Flow Repeatability | ≤ ±0.15% FS |
| Linearity | ≤ ±0.5% FS |
| Inlet Pressure | 20–35 PSIG |
| Ozone Generator Output | 0.05–1.5 ppm at 5 SLPM |
| Ozone Generator Accuracy | ±1% of setpoint or ±2 ppb at 5 SLPM |
| UV Absorption Photometer Range | 0–0.5 / 1.0 / 2.0 ppm |
| Photometer Zero Drift | < ±1 ppb per 24 h (or 30 days) |
| Photometer Flow Rate | 1.0 LPM |
| Dilution Furnace Temperature Range | Ambient +5 °C to 65 °C |
| Temperature Stability | < ±0.1 °C |
| Calibration Programs | 20 user-configurable, time- and event-driven (7-day cycle support) |
| Input/Output | 24-bit digital I/O, 2× serial ports (RS232 standard), parallel printer port, external keyboard interface, air source control port |
| Display | High-brightness color LCD |
| Keypad | 55-key membrane panel |
| Gas Ports | 4 source gas inlets, 1 purge inlet, 1 diluent inlet, 5 calibrated outputs, 1 exhaust port |
| Brand | Environics |
|---|---|
| Origin | USA |
| Model | 4000 Series |
| Dilution Ratio | Up to 10,000:1 |
| Concentration Range | % to ppb |
| Flow Accuracy | ±1.0% of setpoint (concentration & flow) |
| Flow Repeatability | ±0.05% |
| Warm-up Time | 30 min |
| Gas Connection | 1/4" Swagelok® or compatible |
| Operating Pressure | 10–75 psig (0.67–5.04 bar) |
| UHP Leak Rate | ≤1×10⁻⁹ atm·cm³/sec He |
| Humidity Range (optional) | 0–95% RH (±5% accuracy, NIST-traceable) |
| Heating | Optional heated gas pathways (temperature-controlled) |
| Surface Treatment | Optional silanized stainless-steel tubing |
| Purge Pathway | Integrated electromagnetic valve purge system |
| Permeation Oven | Optional multi-zone permeation furnace with dual-permeation-tube capacity |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Model | 4010 |
| Flow Accuracy | ±1.0% of Full Scale |
| Flow Repeatability | ±0.15% of Full Scale |
| Flow Linearity | ±0.5% of Full Scale |
| Diluent MFC Range | 0–10 SLPM (optional 0–20 SLPM) |
| Source MFC Range | 0–100 SCCM (optional 0–50 / 0–200 / 0–500 SCCM / 0–1 SLM) |
| Input Pressure | 20–35 psig |
| Ozone Output Range | 0.05–1.5 ppm at 5 SLPM (extendable to 0.05–1.0 ppm at 15 SLPM) |
| Ozone Accuracy | ±2% of setpoint or ±3 ppb at 5 SLPM |
| Ozone Flow Control Precision | ±1 SCCM at 100 SCCM |
| 95% Setpoint Response Time | <60 s |
| Wetted Materials | 316 Stainless Steel, PTFE, PEEK, Halar |
| Seals | Viton, Kalrez |
| Optional Internal Permeation Oven | 94 cm³ Volume, 13.2 cm Length, 3.02 cm ID, Temperature Stability: ±0.1°C (setpoint +60°C above ambient), Preheat Time: <30 min |
| Brand | Environics |
|---|---|
| Origin | USA |
| Model | 4040 |
| Dilution Ratio | Up to 10,000:1 |
| Flow Accuracy | ±1.0% of setpoint |
| Concentration Accuracy | ±1.0% of setpoint |
| Flow Repeatability | ±0.05% |
| Dimensions (Portable) | 43.18 cm × 17.78 cm × 38.1 cm |
| Weight | 8.2 kg (2-MFC), 10.4 kg (4-MFC), 15.9 kg (8-MFC) |
| Brand | Sabio |
|---|---|
| Origin | Beijing, China |
| Model | Sabio2010 |
| Diluent Flow Range | 0–10 SLM (up to 20 SLM optional) |
| Source Gas Flow Range | 0–100 SCCM (up to 2 SLM optional) |
| Flow Accuracy | <1% FS |
| Flow Repeatability | ±0.15% FS |
| Linearity | ±0.5% FS |
| Inlet Pressure | 12–35 PSIG |
| Operating Temperature | 5–40°C |
| Dimensions | 156 mm (H) × 364 mm (W) × 316 mm (D) |
| Weight | 8.6 kg |
| Power Supply | 98–264 VAC, 100 VA, 50/60 Hz |
| Communication Interfaces | RS-232, Ethernet (TCP/IP optional) |
| Digital I/O | 8-bit TTL/contact-triggered |
| Calibration Paths | 20 preconfigured, each supporting up to 20 linear points |
| Gas Configuration | 1 diluent + 20 source gases |
| Output Ports | 3 calibrated outputs + 1 exhaust port (1 permeation port optional) |
| Input Interface | 42-key membrane keypad or external keyboard |
| Brand | VICI |
|---|---|
| Origin | USA |
| Model | VICI Cheminert M |
| Flow Range (M10) | 10 nL – 10 mL |
| Flow Range (M50) | 50 µL – 50 mL |
| Pump Type | Syringe-free, bidirectional, anodized displacement pump |
| Valve Configuration | Multi-position fluid path selection valve |
| Program Storage | Up to 100 user-defined methods |
| Physical Design | Compact benchtop footprint |
| Compliance Context | Designed for GLP-compliant liquid handling in regulated analytical laboratories |
| Brand | Sabio |
|---|---|
| Origin | USA |
| Manufacturer Type | Authorized Distributor |
| Origin Category | Imported |
| Model | Model 2010 |
| Pricing | Upon Request |
| Diluent Mass Flow Range | 0–10 SLM (optional up to 20 SLM) |
| Source Gas Mass Flow Range | 0–100 SCCM (optional up to 2 SLM) |
| Flow Accuracy | ≤1% FS |
| Flow Repeatability | ≤±0.15% FS |
| Linearity | ≤±0.5% FS |
| Response Time | <60 s |
| Control Interfaces | Keypad, Keyboard, RS-232 Serial, USB, Ethernet |
| Predefined Calibration Paths | 20 factory-configured flow paths, each supporting up to 20 linear calibration points |
| Supported Calibration Types | Diluent gas, Ozone (O₃), Gas Phase Titration (GPT) |
| Temperature Control System | Microprocessor-based proportional PID controller |
| Temperature Sensor | Platinum RTD (Pt100) |
| Heating Zone Dimensions | Internal Diameter 3.09 cm, Length 12.7 cm |
| Temperature Range | Ambient +2 °C to 65 °C |
| Front Panel Integration | Adjustable pressure regulator, analog pressure gauge, digital temperature controller, Teflon-lined dilution furnace inlet |
Show next