BrandPri-eco
OriginBeijing, China
Manufacturer TypeManufacturer
ModelElement5 610
Detection ElementsC, H, N, S, O
PrincipleDumas Flash Combustion + Gas Chromatographic Separation
Furnace ConfigurationDual-Furnace (Oxidation + Reduction)
DetectorSelf-Calibrating TCD
Sample Mass Range0.01–100 mg
Detection Range200 ppm – 100%
Accuracy< 0.2% (vs. certified reference materials)
Precision< 0.1% RSD (vs. certified reference materials)
LOD/LOQ1–5 µg
Oxidation Temp800–1100 °C
Reduction Temp600–1100 °C
Analysis Time (CHNS)~15 min
GC Column2 m packed column (CN separation)
Calibration ModelsLinear, Quadratic, Cubic
SoftwareEAS CLARITY
Carrier GasHelium (99.999%, 3–5 bar)
Oxygen SupplyUltra-High-Purity O₂ (99.999%, 3–5 bar)
Power230 VAC ±10%, 1–10 A
Dimensions (L×W×H)760 × 350 × 700 mm
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPri-eco
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
ModelElement5 860
Detection PrincipleHigh-Temperature Catalytic Combustion (Dumas Method) with Gas Chromatographic Separation
Measured ElementsTotal Nitrogen, Crude Protein (calculated)
Sample FormsSolid and Liquid
Furnace ConfigurationDual-Furnace System (Oxidation + Reduction)
Carrier GasUltra-High-Purity Helium (99.999%, 3–5 bar)
Reactant GasUltra-High-Purity Oxygen (99.999%, 3–5 bar)
DetectorSelf-Calibrating Thermal Conductivity Detector (TCD)
Automation FeaturesAuto-Leak Detection, Intelligent Oxygen Dosage Control, Electronic Flow Regulation, Auto-Flow Setup
Sample CapacityUp to 147 positions (Pneumatic Autosampler)
Analysis Time per Sample4–8 min (N)
Nitrogen Measurement Range0.002–20 mg
Accuracy<0.2% RSD (Certified Reference Materials, >99.9% purity)
Precision<0.1% RSD (Same reference materials)
Furnace TemperatureUp to 1100 °C (each furnace)
Sample Mass Range0.1–400 mg
Power Supply230 V, 50/60 Hz, 5 A, 1100 W
Dimensions980 × 500 × 370 mm
Weight78 kg
SoftwareEAS Clarity (Compliant with FDA 21 CFR Part 11 Audit Trail & User Access Control)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPri-eco
OriginBeijing, China
Manufacturer TypeManufacturer
ModelElement5 624
Detection ElementsC, H, N, S, O
PrincipleDumas flash combustion coupled with gas chromatographic separation
炉 DesignSingle-furnace (oxidation and reduction in one furnace)
DetectorTCD with auto-calibration
Sample Mass Range0.01–100 mg
Detection Range200 ppm – 100%
Accuracy< 0.2% (vs. certified reference materials)
Precision< 0.1% RSD (vs. certified reference materials)
LOD1–5 µg
Combustion/Reduction Temperature800–1100 °C
Analysis Time (CHNS)~15 min
GC Column Length2 m
Calibration ModelsLinear, quadratic, cubic
SoftwareEAS CLARITY
Carrier GasHelium (99.999%, 3–5 bar)
OxidantOxygen (99.999%, 3–5 bar)
Power Supply230 VAC ±10%, 1–10 A
Dimensions (L×W×H)760 × 350 × 700 mm
Added to wishlistRemoved from wishlist 0
Add to compare
BrandNC Technologies
OriginItaly
ModelECS 8024
Measurable ElementsC, N
Sample TypesSolid (e.g., soil, sediment, plant tissue) and liquid
Detection PrincipleDumas combustion coupled with gas chromatographic separation and thermal conductivity detection (TCD)
Measurement RangeC/N: 0.002–20 mg
Accuracy<0.2% (certified reference materials, purity >99.9%)
Precision<0.1% RSD (n=10, CRM)
Analysis Time~5 min per CN determination
Furnace SystemDual-zone high-temperature combustion furnaces (up to 1100 °C each)
Separation Column3 m GC column
DetectorSelf-calibrating high-resolution TCD (no reference gas required)
Gas RequirementsHelium (99.999%, 3–5 bar), Oxygen (99.999%, 3–5 bar), oil-free compressed air
Auto-sampling OptionsPneumatic (147-position), Electronic (32/50/100-position), or manual
SoftwareEAS Clarity v5.x with GLP-compliant audit trail, calibration curve fitting (linear, quadratic, cubic), and method validation tools
ComplianceDesigned to support ISO 13879, ASTM D5373, EPA Method 9060A, and USP <231> for elemental carbon/nitrogen quantification in environmental matrices
Dimensions81 × 50 × 37 cm
Weight68 kg
Power230 V, 50/60 Hz, 5 A, 1100 W max
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPri-eco
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
ModelElement5 824
Detection ElementsC, N
Sample FormSolid (soil), liquid
Combustion SystemDual-furnace Dumas flash combustion
Separation MethodGC column (3 m)
DetectionHigh-sensitivity TCD (reference-gas-free self-calibration)
Autosampler OptionsPneumatic (147-position), electric (32/50/100-position), manual
Analysis Time (C/N)≤5 min
Measurement RangeC/N: 0.002–20 mg
Accuracy<0.2% (certified reference materials, purity >99.9%)
Precision<0.1% RSD (certified reference materials, purity >99.9%)
Furnace TempUp to 1100 °C (left & right), separation oven up to 110 °C
Carrier GasUltra-high-purity He (99.999%, 3–5 bar)
OxidantUltra-high-purity O₂ (99.999%, 3–5 bar)
Auxiliary GasOil-free compressed air
Power Supply230 V, 50/60 Hz, 5 A, 1100 W
Dimensions810 × 500 × 370 mm
Weight68 kg
SoftwareEAS Clarity (linear, quadratic, cubic calibration
ComplianceDesigned for GLP/GMP environments
Added to wishlistRemoved from wishlist 0
Add to compare
BrandNC Technologies
OriginItaly
ModelNCT ECS 4010
Detection PrincipleDumas Combustion Coupled with Gas Chromatographic Separation
Elements AnalyzedC, H, N, S, O
Sample Mass Range0.01–200 mg
Accuracy (std. materials)< 0.2%
Precision (std. materials)< 0.1%
LOQ1–5 µg
Oxidation Furnace Temp.800–1100 °C
Reduction Furnace Temp.600–1100 °C
GC Column Temp. Range30–110 °C
Analysis Time (CHNS)~15 min
GC Column Length2 m
DetectorPatented NCT Thermal Conductivity Detector (TCD), Self-Calibrating
Autosampler OptionsPneumatic (up to 147 positions), Electronic (32/50/100 positions), Manual
Gas RequirementsHigh-purity He (99.999%, 3–5 bar), O₂ (99.999%, 3–5 bar), Compressed Air
Power230 VAC ±10%, 1–10 A
Dimensions (L×W×H)760 × 350 × 700 mm
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPri-eco
OriginBeijing, China
Manufacturer TypeManufacturer
ModelElement5 822
Detection ElementsC, N
Sample FormsSolid and Liquid
Combustion ConfigurationDual-Furnace (Oxidation + Reduction)
Detection PrincipleDumas Flash Combustion Coupled with Gas Chromatographic Separation and Thermal Conductivity Detection (TCD)
Automation FeaturesAutomated Leak Detection, Intelligent Oxygen Dosage Control, Electronic Flow Regulation, Self-Calibrating TCD
Sample CapacityUp to 147 positions (Pneumatic Autosampler)
Analysis Time per CN Run~5 min
Measurement RangeC/N: 0.002–20 mg
Accuracy<0.2% (Certified Reference Materials, Purity >99.9%)
Precision (RSD)<0.1% (n ≥ 10, CRM)
Furnace TempUp to 1100 °C (Dual Independent Zones)
Separation Oven TempUp to 110 °C
Carrier GasUltra-High-Purity Helium (99.999%, 3–5 bar)
Reactant GasUltra-High-Purity Oxygen (99.999%, 3–5 bar)
Auxiliary GasOil-Free Compressed Air
GC Column3 m Packed Capillary Column
SoftwareEAS Clarity v3.x (Compliant with FDA 21 CFR Part 11 Audit Trail & User Access Control)
Dimensions810 × 500 × 370 mm
Weight68 kg
Power Supply230 V, 50/60 Hz, 5 A, 1100 W
Sample ContainerHigh-Purity Tin or Silver Cups
Optional AccessoriesAnalytical Balance Integration, Consumables Kits (Tubes, Cups, Reagents), Isotope Ratio Interface (IRMS-ready)
Added to wishlistRemoved from wishlist 0
Add to compare
OriginItaly
Manufacturer TypeAuthorized Distributor
Origin CategoryImported
ModelCostech ECS 4024
Price RangeUSD 13,500 – 27,000
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPri-eco
OriginBeijing, China
Manufacturer TypeManufacturer
ModelElement5 840
Detection ElementsC, H, N, S, O
Sample IntroductionSingle-furnace Dumas combustion with catalytic reduction for oxygen
Measurement PrincipleHigh-temperature flash combustion (Dumas method) coupled with gas chromatographic separation and thermal conductivity detection (TCD)
Automation FeaturesAuto leak detection, electronic flow control, intelligent O₂ dosing, self-calibrating TCD
Sample Capacity0.1–500 mg (up to 1000 mg for soil), solid/liquid
Analysis TimeC/N: 5 min
C/H/N8 min
C/H/N/S10–25 min
O4 min
Detection RangeC: 0.002–20 mg
H0.002–5 mg
N0.002–20 mg
S0.002–6 mg
O0.002–2 mg
Accuracy<0.2% (certified reference materials, purity >99.9%)
Precision<0.1% RSD (same reference materials)
Furnace Tempup to 1100 °C
Separation Oven≤110 °C
Carrier GasUltra-high-purity He (99.999%, 3–5 bar)
GC Column0.8–4 m packed capillary column
DetectorHigh-sensitivity, reference-gas-free TCD
SoftwareEAS Clarity (supports linear, quadratic, cubic calibration
Dimensions510 × 500 × 370 mm
Weight53 kg
Power230 V, 50/60 Hz, 4 A, 900 W
Optional AccessoriesAnalytical balance, consumables (Sn/Ag capsules), isotope interface kit for IRMS coupling
Added to wishlistRemoved from wishlist 0
Add to compare
BrandNC Technologies
OriginItaly
ModelECS 8022
Measured ElementsC, N
Sample FormsSolid and Liquid
Detection PrincipleDumas Combustion Coupled with Gas Chromatography–Thermal Conductivity Detection (GC-TCD)
Measurement RangeC/N: 0.002–20 mg
Accuracy<0.2% (vs. certified reference materials, purity >99.9%)
Precision<0.1% RSD (vs. certified reference materials, purity >99.9%)
Analysis Time~5 min per CN determination
Furnace SystemDual-zone high-temperature combustion furnaces (up to 1100 °C each)
Separation Column3 m GC column
DetectorSelf-calibrating high-resolution TCD (no reference gas required)
Carrier GasUltra-high-purity helium (99.999%, 3–5 bar)
OxidantUltra-high-purity oxygen (99.999%, 3–5 bar)
AutosamplersPneumatic (147-position), Electromechanical (32/50/100-position), Manual
SoftwareEAS Clarity™ with linear/quadratic/cubic calibration, real-time process monitoring, catalyst consumption tracking, and audit-ready data logging
ComplianceDesigned for GLP/GMP environments
Dimensions81 × 50 × 37 cm
Weight68 kg
Power230 V, 50/60 Hz, 5 A, 1100 W max
Added to wishlistRemoved from wishlist 0
Add to compare
BrandPri-eco
OriginBeijing, China
Manufacturer TypeManufacturer
ModelElement5 820
Detection ElementsC, H, N, S, O
Sample IntroductionPneumatic Auto-sampler (147 positions), Electric Auto-sampler (32/50/100 positions), Manual Sampler
Analysis TimeCN: 5 min
CHN8 min
CHNS10–25 min
O4 min
Measurement RangeC: 0.002–20 mg
H0.002–5 mg
N0.002–20 mg
S0.002–6 mg
O0.002–2 mg
Accuracy<0.2% (certified reference materials, purity >99.9%)
Precision<0.1% (certified reference materials, purity >99.9%)
Furnace ConfigurationDual independent high-temperature furnaces (max. 1100 °C each)
Separation Oven Temp.up to 110 °C
Carrier GasUltra-high-purity helium (99.999%, 3–5 bar)
Oxidant GasUltra-high-purity oxygen (99.999%, 3–5 bar)
Auxiliary GasOil-free compressed air
GC Column0.8–4 m capillary separation column
DetectorHigh-sensitivity TCD with auto-calibration (no reference gas required)
Flow ControlElectronic mass flow regulation
Leak DetectionFully automated self-diagnostic leak test
Oxygen DeliveryIntelligent, sample-mass-dependent dosing
SoftwareEAS Clarity (supports linear, quadratic, and cubic calibration models)
Sample Mass Range0.1–500 mg (up to 1000 mg for soil)
Sample CupsHigh-purity tin or silver capsules
Dimensions810 × 500 × 370 mm
Weight68 kg
Power Supply230 V, 50/60 Hz
Max. Power Consumption1100 W, 5 A
Touchscreen InterfaceIntegrated industrial-grade display
Added to wishlistRemoved from wishlist 0
Add to compare
BrandNC Technologies
OriginItaly
ModelECS 8060
Analysis MethodDumas Combustion
Detected ElementsNitrogen (N)
Protein CalculationBased on N × Conversion Factor (e.g., 6.25 for food)
Analysis Time2–8 min per sample
Sample Mass Range0.1–400 mg (up to 1000 mg for low-carbon samples)
Detection Limit (LOQ)5 µg N
DetectorSelf-Calibrating High-Sensitivity TCD
Carrier GasHelium (99.999%, 3–5 bar)
Oxidant GasOxygen (99.999%, 3–5 bar)
Auxiliary GasOil-free compressed air
Sample TypesSolid and liquid organic matrices
AutomationFully automated oxygen dosing, leak check, flow optimization, catalyst monitoring
InterfaceTouchscreen HMI
Isotope CompatibilityYes — direct coupling to IRMS systems (e.g., Thermo Fisher, Elementar, Sercon)
ComplianceDesigned for GLP/GMP environments