- All
- Favorite
- Popular
- Most rated
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XDLM |
| Measurement Principle | Energy-Dispersive X-ray Fluorescence (EDXRF) |
| Sample Positioning | Motorized Z-axis (standard) |
| Collimators | 4 interchangeable (diameters: 0.1, 0.2, 0.3, 0.5 mm) |
| Primary Filters | 3 interchangeable Al, Cu, Ti |
| Detector | Proportional counter with high-count-rate capability |
| Measurement Distance | 0–80 mm (adjustable via motorized Z-axis) |
| Optical Alignment | Integrated high-resolution color CCD camera with 10× digital zoom |
| Operating Temperature | 10–40 °C |
| Relative Humidity | ≤95% RH (non-condensing) |
| Instrument Weight | 100–120 kg |
| PC Interface | PCI/PCIe expansion slot required for spectrometer control and data acquisition |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | HM500 |
| Compliance | DIN EN ISO 14577-1, DIN EN ISO 14577-2, DIN 50359 |
| Measurement Principle | Continuous Stiffness Measurement (CSM) & Depth-Sensing Indentation |
| Load Range | 0.1 mN – 1000 mN |
| Depth Resolution | ≤ 0.05 nm (sub-pm interpolation capability) |
| Load Resolution | ≤ 10 nN |
| Indenter Types | Vickers (136° diamond pyramid), Berkovich (115.2° trihedral pyramid), spherical diamond (R = 1 µm to 100 µm) |
| Stiffness Measurement Mode | Enhanced Stiffness Profiling (ESP) |
| Data Acquisition Rate | Up to 1 kHz |
| Environmental Stability | Active temperature compensation (±0.1 °C) |
| Software Compliance | FDA 21 CFR Part 11 ready, GLP/GMP audit trail enabled |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | MP0/MP0R |
| Measurement Principle | Dual-mode electromagnetic induction (for ferrous substrates) and eddy current (for non-ferrous substrates) |
| Display | Dual independent pivotable LCD screens with backlight |
| Housing | IP65-rated ruggedized enclosure |
| Probe | Wear-resistant integrated probe |
| Compliance Modes | IMO PSPC, SSPC-PA2 |
| Operating Temperature | −10 °C to +50 °C |
| Battery Life | Up to 30,000 measurements per charge (standard alkaline AA batteries) |
| Data Output | USB interface for PC transfer |
| Calibration | Factory-calibrated with traceable NIST-equivalent reference standards |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XDLM Series |
| Measurement Principle | Energy-Dispersive X-Ray Fluorescence (ED-XRF) |
| Standards Compliance | DIN 50987, ISO 3497, ASTM B568 |
| X-Ray Tube | Microfocus, adjustable high voltage (30/40/50 kV) |
| Collimator Set | Circular (0.1 / 0.2 / 0.3 mm) and rectangular (0.05 × 0.3 mm) |
| Sample Chamber Dimensions (W×D×H) | 570 × 740 × 650 mm |
| Internal Chamber Volume (W×D×H) | 460 × 500 × 300 mm |
| Z-Axis Travel Range | 144 mm (motorized) |
| Camera System | Integrated color CCD camera with live imaging and on-screen crosshair overlay |
| Beam Spot Visualization | Real-time X-ray beam size indicator synchronized with focal distance |
| Distance Correction Method | DCM (Distance Control Measurement), effective up to 80 mm |
| Software Platform | WinFTM® v6 (Windows-based, 32-bit), multilingual UI (EN/DE/FR/IT/ES/ZH) |
| Data Interface | RS-232, Ethernet-compatible remote command protocol |
| Reporting & SPC | Embedded statistical engine with control charts (X̄/R, X̄/S), probability plots, histogram analysis, and customizable PDF/Excel report generation |
| Regulatory Support | Audit trail, user access levels, electronic signature readiness per FDA 21 CFR Part 11 and GLP/GMP requirements |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | RMP30-S |
| Measurement Principle | Micro-resistive Method per EN 14571:2004 |
| Display | Dual-line, 16-character LCD (2 × 16) |
| Memory Capacity | Up to 100 application setups |
| Data Storage | Up to 10,000 measurements across 1,000 data blocks with timestamp (date/time stamping) |
| Interface | Bidirectional RS232 port |
| Probe Recognition | Automatic probe identification |
| Audio Feedback | Audible limit alarm |
| Statistical Functions | Mean, Std. Dev., Min/Max, Cp/Cpk, histogram generation |
| Compatible Probes | ERCU-N, ERCU-D10 |
| Power Supply | Rechargeable battery or AC adapter |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XULM Series |
| Measurement Principle | Energy-Dispersive X-Ray Fluorescence (ED-XRF) |
| Standards Compliance | DIN 50987, ISO 3497, ASTM B568 |
| X-Ray Tube | Microfocus, adjustable high voltage (30/40/50 kV) |
| Collimator Options | Ø0.1 mm / Ø0.2 mm circular |
| Sample Chamber Dimensions (W×D×H) | 375 × 580 × 480 mm |
| Internal Chamber Dimensions (W×D×H) | 360 × 460 × 240 mm |
| Manual X-Y Stage Travel | 50 mm (X) × 50 mm (Y) |
| Stage Platform Size | 360 × 240 mm |
| Viewing System | Integrated color CCD camera with real-time zoom and auto-focus |
| Beam Alignment | On-screen calibrated crosshair + dynamic X-ray beam size indicator |
| Software Platform | WinFTM® v6 (Win32 native, multilingual UI) |
| Reporting | Embedded sample image capture, SPC charts (X̄/R, X̄/S), probability plots, histogram analysis |
| Regulatory Support | Audit trail, user access levels, electronic signature readiness per FDA 21 CFR Part 11 & GLP/GMP requirements |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | HM2000S |
| Instrument Type | Microhardness Tester |
| Compliance | DIN EN ISO 14577 |
| Measurement Principle | Instrumented Indentation Testing (IIT) |
| Key Output Parameters | Martens Hardness (HM), Indentation Hardness (HIT), Indentation Modulus (EIT), Indentation Creep (CIT), Elastic Work Ratio (ηIT), Depth- and Load-Dependent Property Mapping via ESP Mode |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XDV-SDD |
| Form Factor | Benchtop |
| Elemental Range | Al (Z=13) to U (Z=92) |
| Detection Limit | 1 ppm |
| Quantification Range | 1 ppm – 100 wt% |
| Energy Resolution | ≤140 eV (Mn Kα) |
| Repeatability | ≤5% RSD |
| Sample Stage | Motorized XYZ with Programmable XY Translation and Z-Axis Lift |
| X-ray Source | Microfocus Tungsten Anode Tube with Beryllium Window |
| High Voltage | 3-step Adjustable (up to 50 kV) |
| Filter Options | 6 Interchangeable Primary Filters |
| Collimators | Motorized, Switchable Apertures |
| Optical Alignment | Integrated HD Video System with 20× Digital Zoom, Crosshair Overlay, and Laser Spot Locator |
| Software Platform | WinFTM® v8.x (FDA 21 CFR Part 11 Compliant, Audit Trail Enabled, GLP/GMP-Ready) |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | FMP100 / FMP150 |
| Measurement Principle | Magnetic Induction (for ferrous substrates) & Eddy Current (for non-ferrous substrates) |
| Operating System | Windows CE |
| Display | High-Resolution Graphical Touchscreen |
| Onboard Memory | Supports >10,000 measurements |
| Compliance Modes | IMO PSPC, SSPC-PA2, QUALANOD, QUALICOAT |
| Probe Interface | Interchangeable probe design |
| Data Management | Compatible with DataCenter IP inspection planning software |
| Regulatory Context | Designed for GLP/GMP-aligned workflows |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | SIGMASCOPE SMP350 |
| Measurement Principle | Phase-Sensitive Eddy Current Method |
| Compliance | DIN EN 2004-1, ASTM E1004 |
| Frequency Range | 15 kHz to 1 MHz |
| Calibration Standards | Traceable to International Reference Standards (MS/m and %IACS) |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XUL220 |
| Measurement Principle | Energy-Dispersive X-Ray Fluorescence (ED-XRF) |
| Detector Type | Proportional Counter |
| Count Rate Capability | High |
| Imaging System | High-Resolution Color CCD Camera with Manual Focus and Adjustable LED Illumination |
| Sample Chamber Dimensions | Extra-Large (Enabling Large-Part Measurement) |
| Stage | Manual XY Translation Stage |
| Operating Temperature | 10–40 °C |
| Relative Humidity | ≤95% RH (Non-Condensing) |
| Compliance | Designed for ISO 3497, ASTM B568, DIN 50982, and RoHS/WEEE Screening Applications |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XDL-B Series |
| Measurement Principle | Energy-Dispersive X-Ray Fluorescence (ED-XRF) |
| Standards Compliance | DIN 50987, ISO 3497, ASTM B568 |
| X-ray Tube Voltage | Adjustable up to 50 kV |
| Collimator Options | 0.3 mm Ø standard |
| Measurement Chamber Dimensions (H×W×D) | 650 × 570 × 740 mm |
| Chamber Internal Dimensions (H×W×D) | 300 × 460 × 500 mm |
| Mass | ~105 kg |
| Viewing System | Integrated color CCD camera with real-time zoom and focus control |
| Distance Control Method | DCM (Distance Correction Method) for automated spectral intensity compensation across 3 fixed source-to-sample distances (standard: mid-position) |
| Sample Height Adjustment | Fixed-height X-ray head with removable support plate for oversized parts |
| Interface | RS-232, Ethernet-ready for remote command control |
| Software Platform | Native Win32 application for Windows 2000/XP |
| Language Support | English, German, French, Italian, Spanish, Chinese |
| Reporting & SPC | Embedded statistical engine with SPC charts, probability plots, histogram analysis, and customizable report generation including annotated sample images |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XAN500 |
| Detector Type | Silicon Drift Detector (SDD) |
| Measurement Principle | Energy-Dispersive X-ray Fluorescence (ED-XRF) |
| Portability | Handheld & Benchtop-Convertible Configuration |
| Compliance | CE, RoHS, ISO 3497, ASTM B568, DIN EN ISO 3497, ISO 2178 (for magnetic substrates), ISO 2360 (for non-magnetic substrates) |
| Software | WinFTM® v8.x with GLP/GMP-compliant audit trail, FDA 21 CFR Part 11 optional module |
| Price Range | USD 25,000 – 65,000 (FOB Hamburg) |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | MP10E |
| Measurement Principle | Electromagnetic Induction (DIN EN ISO 2178, ASTM B499) |
| Display | 60 × 30 mm LCD, dual-line alphanumeric (16 characters per line) |
| Probe Interface | Plug-in connector supporting up to 30 probe types (E- and VxE-series, e.g., EGAB1.3) |
| Auto-Probe Recognition | Yes |
| Data Storage | None |
| Connectivity | None |
| Power Supply | Replaceable or rechargeable batteries |
| Compliance | DIN EN ISO 2178, ASTM B499 |
| Included Accessories | Protective cap, carrying case, batteries, operating manual |
| Excluded | Probes, calibration standards |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XDVM-W |
| Measurement Principle | X-ray fluorescence (XRF) |
| Standards Compliance | DIN 50987, ISO 3497, ASTM B568 |
| Cabinet Dimensions (H×W×D) | 720 × 660 × 950 mm |
| Weight | ~135 kg |
| Internal Chamber (H×W×D) | 160 × 560 × 530 mm |
| X-ray Tube | Adjustable high voltage (30/40/50 kV), current up to 0.8 mA |
| Detector | Xenon-filled proportional counter with 4096-channel ADC (displayed as 256-channel spectrum) |
| Collimator Options | Standard Set I (0.1 mm, 0.2 mm, 0.05×0.05 mm², 0.03×0.2 mm² slot), optional Sets II & III |
| XY Stage Travel | 250 × 250 mm (XDVM-W standard configuration) |
| Z-axis Travel | 145 mm |
| Camera | High-resolution color CCD with dual magnification (×40 / ×80), crosshair + metric grid + spot size indicator |
| Software Platform | Native Win32 application for Windows 2000 and later |
| Language Support | English, German, French, Italian, Spanish, Chinese, Japanese |
| Data Interface | RS-232 serial port, remote command control via script files or network environment |
| Calibration | Automated beam centering, “Application Toolbox” with calibration standards and parameter disk |
| Reporting | User-defined report editor with embedded images (BMP), company logos, charts, font selection |
| Statistical Tools | SPC charts, probability plots, histogram analysis |
| Security | Customizable “short menu” lockout for critical system functions |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | FERITSCOPE FMP30 |
| Measurement Principle | Magnetic Induction (Ferrite-Specific Permeability Detection) |
| Display | Backlit LCD, 128 × 64 pixels |
| Probe Interface | Interchangeable modular probe connector |
| Calibration | Single-point calibration across full measurement range |
| Units | % Fe (ferrite number equivalent) and vol.% Fe |
| Compliance | Designed per ISO 8249:2017, ASTM A802, EN 17068 |
| Operating Temperature | 0 °C to 50 °C |
| IP Rating | IP65 (dust- and water-resistant) |
| Power Supply | Rechargeable Li-ion battery (≥10 h continuous operation) |
| Data Output | USB 2.0, optional Bluetooth 4.2 module |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | HM2000 |
| Instrument Type | Nanoindentation Tester |
| Max Indentation Depth | 500 µm |
| Effective Load Range | 0.1–2000 mN |
| Load Resolution | 150 nN |
| Displacement Range | 130 mm |
| Displacement Resolution | 0.01 nm |
| Max Friction Force | 50 N |
| Indenter Types | Vickers, Berkovich, Tungsten Carbide Sphere |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | HM500 LIGHT |
| Compliance | DIN EN ISO 14577-1, DIN EN ISO 14577-2, DIN 50359 |
| Measurement Principle | Continuous Stiffness Measurement (CSM) & Depth-Sensing Indentation |
| Load Range | 0.1 mN – 500 mN |
| Depth Resolution | ≤ 0.05 nm (50 pm) |
| Load Resolution | ≤ 0.01 mN |
| Indenter Types | Vickers (136° diamond pyramid), Berkovich (115.2° tri-pyramid), spherical WC or diamond |
| Stiffness Measurement Mode | Enhanced Stiffness Profiling (ESP) |
| Data Acquisition Rate | Up to 1 kHz |
| Zero-Point Detection | Capacitive contact sensing with sub-nm repeatability |
| Environmental Stability | Active thermal drift compensation (< 0.05 nm/s) |
| Software Platform | FISCHER FISCHERWIN v8.x (FDA 21 CFR Part 11 compliant audit trail, GLP/GMP-ready) |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XAN-DPP |
| Configuration | Benchtop/Free-standing |
| Excitation Source | Rhodium anode X-ray tube |
| Detector | Peltier-cooled Si-PIN diode (≈ −30 °C) |
| Energy Resolution | 180 eV (Mn Kα) |
| Elemental Range | Al (Z=13) to U (Z=92) |
| Collimators | Motorized, programmable: Ø0.2 mm, Ø0.6 mm, Ø1.0 mm, Ø2.0 mm |
| Measurement Chamber Internal Dimensions (incl. sloped door) | W 318 mm × D 307 mm × H 29–86 mm |
| Usable Height Zone (full-height region) | W 318 mm × D 203 mm × H 86 mm |
| Distance Compensation | Patented DCM (Distance Controlled Measurement) algorithm |
| Calibration | Standardless quantitative analysis enabled via fundamental parameters method |
| Simultaneous Output | Elemental concentration (ppm–wt%) and substrate thickness (µm) |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XDVM-u |
| Measurement Principle | X-ray Fluorescence (XRF) |
| Compliance | DIN 50987, ISO 3497, ASTM B568 |
| Max. Measurable Layers | 23 |
| Elements Analyzed | Cl (Z=17) to U (Z=92) |
| Element Quantification Range | x ppm to 100% |
| Spectral Channels | 1024 (raw: 4096) |
| Minimum Spot Size | 50 × 100 µm |
| Focal Depth Range | 2.5 mm |
| X-ray Tube | Adjustable HV (30/40/50 kV), Current up to 0.8 mA |
| Detector | Peltier-cooled semiconductor with high energy resolution |
| XY Stage Travel | X=250 mm, Y=220 mm, Positioning Accuracy: ±0.005 mm |
| Z-Axis Travel | 95 mm |
| Camera Magnification | 30×–1108× with crosshair & scale grid |
| Software | WinFTM® V6 with PDM (Product Data Management) |
| Interface | RS232, Ethernet, removable storage |
| Languages | English, German, French, Italian, Spanish, Chinese |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | SIGMASCOPE® GOLD |
| Measurement Principle | Eddy Current Conductivity Testing per ASTM E 1004 |
| Application | Non-Destructive Authentication of Gold, Silver, and Other Noble Metal Coins & Bullion |
| Penetration Depth Adjustable (M | up to ~8 mm |
| B | up to ~17 mm) |
| Compliance | ASTM E 1004, ISO 2360, EN 10045-1 |
| Sample Handling | Surface measurement through non-conductive coatings (e.g., plastic encapsulation films) |
| Display | High-resolution color touchscreen with real-time phase-angle analysis |
| Data Output | CSV export, GLP-compliant audit trail (optional), USB & Bluetooth connectivity |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | XUL Series |
| Measurement Principle | Energy-Dispersive X-Ray Fluorescence (ED-XRF) |
| Standards Compliance | DIN 50987, ISO 3497, ASTM B568 |
| X-ray Tube Voltage | Adjustable to 30 kV / 40 kV / 50 kV |
| Collimator Options | Ø0.3 mm circular or 0.05 × 0.3 mm rectangular (optional) |
| Sample Chamber Dimensions (W×D×H) | 455 × 580 × 510 mm |
| Internal Chamber Dimensions (W×D×H) | 360 × 380 × 240 mm |
| Manual X-Y Stage Travel | 50 mm (X) × 50 mm (Y) |
| Imaging System | Integrated color CCD camera with real-time magnified view and auto-focus |
| Beam Indicator | On-screen X-ray beam size overlay synchronized with focal distance |
| Software Platform | WinFTM® v6 (base) or Super WinFTM® (optional, order no. 602-950) |
| Language Support | English, German, French, Italian, Spanish, Chinese |
| Reporting | Graphical reports with embedded sample images, SPC charts, probability plots, histogram analysis |
| Regulatory Readiness | Supports GLP/GMP documentation workflows |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | PMP10 |
| Measurement Principle | Phase-Sensitive Eddy Current |
| Operating Frequencies | 60 kHz / 240 kHz / 1.25 MHz |
| Probe Compatibility | ESD20Cu, ESD20Ni, ESD20Zn, ESD2.4, ESL080 |
| Max Data Storage | 20,000 measurements across 4,000 data groups |
| Applications | 100 user-defined programs |
| Display Resolution | 3-level precision |
| Measurement Modes | Automatic, Continuous, External Trigger |
| Analysis Modes | Scientific, Statistical, Limit (Hi/Lo), Analog Simulation |
| Security | Password protection & key lock |
| Power Management | Auto-off |
| Brand | Fischer |
|---|---|
| Origin | Germany |
| Model | FMP10–FMP40 |
| Measurement Principles | Magnetic Induction (for ferrous substrates) & Eddy Current (for non-ferrous conductive substrates) |
| Compliance Modes | IMO PSPC, SSPC-PA2, QUALANOD, QUALICOAT |
| Memory Capacity | Thousands of measurement records |
| Probe Interface | Interchangeable probe system |
| Operating Principle | Dual-method substrate-agnostic thickness quantification via calibrated electromagnetic field perturbation analysis |
Show next