BrandKonica Minolta
OriginJapan
ModelCM-2300d
Instrument TypePortable
Optical Geometry0/45° integrating sphere
Light SourceDual pulsed xenon lamps
Spectral Range360 nm to 740 nm
Wavelength Interval10 nm
Spectral Bandwidth~10 nm
Reflectance Range0–175%
Display Resolution0.01%
Measurement Time~1.5 s (SCI & SCE simultaneously)
RepeatabilityΔE*ab ≤ 0.08 (after white calibration, 30 measurements at 10 s intervals)
Inter-instrument AgreementΔE*ab ≤ 0.4 (SCI, based on BCRA Series II standards)
Measuring/Aperture AreaØ8 mm / Ø11 mm
Weight670 g (without batteries)
Data Storage Capacity1,700 sets (each set includes paired SCI/SCE data)
ComplianceCIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033 Teil 7, JIS Z8722 Condition C
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
Instrument TypePortable
Optical Geometryd/8° integrating sphere
Light SourcePulsed Xenon Lamp
Spectral RangeY Reflectance 0.01–160.00%
RepeatabilityΔE*ab ≤ 0.07 (std. dev., white calibration, 30 measurements at 10 s intervals)
Measurement ApertureØ50 mm / Illumination Area: Ø53 mm
Measurement Time1 s
Minimum Interval3 s
Battery Life~800 measurements (AAA alkaline/Ni-MH, Konica Minolta specified conditions)
DisplayBacklit dot-matrix LCD (15×9 characters + icon line), 6 language options (EN/DE/FR/IT/ES/JP)
Data Storage1,000 measurements (instrument-only)
Standard CalibrationCR-A44 white calibration tile
ComplianceCIE 1931 2° standard observer, CIE No. 15:2004, ISO 7724-1, ASTM E308, JIS Z 8722
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCM-3500d
Illumination/Reception Geometry (Reflectance)d/8° (diffuse illumination, 8° viewing), switchable between SCI (Specular Component Included) and SCE (Specular Component Excluded)
Illumination/Reception Geometry (Transmittance)d/0° (diffuse illumination, 0° viewing)
DetectorDual 18-element silicon photodiode array with wedge-type continuous interference filter
Wavelength Range400–700 nm
Wavelength Interval20 nm
FWHM~20 nm (average)
Photometric Range0–175% (resolution: 0.01%)
Light SourcePulsed xenon arc lamp
Measurement TimeApprox. 2.5 s (to data output)
Minimum Measurement Interval3 s
Aperture Options (Reflectance)LAV (Φ30 mm measurement / Φ36 mm illumination), MAV (Φ8 mm measurement / Φ11 mm illumination)
Aperture (Transmittance)~Φ22 mm
Repeatability (after white calibration, 30 measurements at 10 s intervals on white calibration tile)Spectral reflectance SD ≤ 0.20%, ΔE*ad ≤ 0.05
Instrumental ErrorΔE*ad ≤ 0.15 (typical, BCRA Series II 12-tile average)
Temperature DriftSpectral reflectance ±0.30%/°C, ΔE*ab ≤ 0.05/°C
InterfaceRS-232C
Power SupplyAC 100/120/230 V, 50/60 Hz (with included AC adapter)
Operating Environment0–40°C, ≤85% RH (non-condensing at ≤35°C)
Storage Environment−20–45°C, ≤85% RH (non-condensing at ≤35°C)
Dimensions (W × H × D)413 × 190 × 240 mm
Weight10 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCM-3600d
Illumination/Viewing Geometryd/8° (diffuse illumination, 8° viewing) for reflectance
ComplianceCIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033-7, JIS Z8722 Condition C
Spectral Range360–740 nm
Spectral Interval10 nm
FWHM~10 nm
DetectorDual 40-element silicon photodiode arrays
Grating TypePlanar diffraction grating
Integrating Sphere DiameterΦ152 mm
Light SourceFour pulsed xenon lamps
Measurement Time~1.5 s (to data output)
AperturesLAV (Φ25.4 mm), MAV (Φ8 mm), SAV (Φ4 mm) for reflectance
RepeatabilityΔE*ab ≤ 0.02 (standard deviation), spectral reflectance SD ≤ 0.1%
Instrumental ErrorΔE*ab ≤ 0.15 (LAV/SCI, BCRA Series II tile average)
UV ControlDigital UV adjustment with selectable 400 nm / 420 nm cutoff filters
Transmittance Chamber133 mm (W) × ~50 mm (D), ~17 mm aperture
InterfaceRS-232C (D-sub 9-pin female)
PowerAC 120 V / 230 V, 50 / 60 Hz (via included adapter)
Operating Environment13–33 °C, ≤80% RH (non-condensing)
Dimensions (W×H×D)244 × 208 × 378 mm
Weight12 kg
Standard AccessoriesWhite calibration tile, aperture masks (Φ4/Φ8/Φ25.4 mm), zero calibration box, AC adapter, dust cover, accessory case, driver software, RS-232C cable (2 m)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCM-2500c
Illumination/Reception Geometry45°/0° (45° annular illumination, vertical reception)
DetectorDual 40-element silicon photodiode arrays
Spectral SeparationPlanar diffraction grating
Wavelength Range360–740 nm
Wavelength Interval10 nm
FWHMApprox. 10 nm
Reflectance Range0–175% (resolution: 0.01%)
Light SourceDual pulsed xenon lamps
Measurement Time~1.5 s
Minimum Measurement Interval4 s
Battery Life~1,000 measurements (alkaline AA, 23°C)
Aperture Size7 mm measuring / 11 mm illumination
RepeatabilityΔE*ab ≤ 0.05 (white calibration tile, 30 measurements at 10 s intervals)
Instrument ErrorΔE*ab ≤ 0.25 (BCRA Series II tiles, avg., vs. master unit, 23°C)
Measurement ModesSingle / auto-averaged (3/5/8 shots)
InterfaceRS-232C
Observer Angles2° and 10° (CIE 1931 & CIE 1964)
IlluminantsA, C, D50, D65, F2, F6, F7, F8, F10, F11, F12 (dual-illuminant simultaneous evaluation)
Color SpacesL*a*b*, L*C*h, CMC(1:1), CMC(2:1), CIE94, CIEDE2000, Yxy, XYZ, Hunter Lab, MI, WI(ASTM E313), WI(CIE), YI(ASTM E313/D1925), ISO Brightness (ISO 2470), Tint (CIE), A/T concentration
Data Storage700 measurements
Pass/Fail CriteriaBox or ellipse tolerance definition
Power Supply4 × AA batteries or AC adapter
Dimensions (W×H×D)69 × 96 × 193 mm
Weight~670 g (battery-free)
Operating Environment5–40°C, ≤80% RH (non-condensing at ≤35°C)
Storage Conditions0–45°C, ≤80% RH (non-condensing at ≤35°C)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCR-400 / CR-410
Illumination/Reception Geometryd/0 (diffuse illumination, 0° viewing
DetectorSilicon photodiodes (6 channels)
Light SourcePulsed xenon lamp
Measurement Time1 s
Minimum Interval Between Measurements3 s
RepeatabilityΔE*ab ≤ 0.07 (after white calibration, 30 measurements at 10 s intervals)
Instrument ErrorΔE*ab ≤ 0.6 (CR-400), ≤ 0.8 (CR-410) against BCRA Series II 12-color standard set
Aperture SizeCR-400: Ø8 mm measurement / Ø11 mm illumination
CR-410Ø50 mm measurement / Ø53 mm illumination
Reflectance RangeY = 0.01–160.00%
Observer AngleCIE 2° standard observer
IlluminantsC, D65
Color SpacesXYZ, Yxy, L*a*b*, L*C*h, CIELAB, CIELUV, CMC(l:c), CIE 1994, CIE 2000, Hunter Lab, CIE Whiteness (WI) & Yellowness (YI) per ASTM E313/D1925 (C or D65 dependent)
DisplayBacklit dot-matrix LCD (15 × 9 characters + icon line)
Memory Capacity1000 measurements (head-only mode)
InterfaceRS-232C (default baud rate: 9600 bps)
Power4 × AAA alkaline or Ni-MH batteries
Dimensions (W × H × D)102 × 244 × 63 mm
Weight~550 g (CR-400), ~570 g (CR-410), incl. batteries
Operating Environment0–40 °C, ≤85% RH (non-condensing)
Storage Environment−20–40 °C, ≤85% RH (non-condensing)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelMulti Gloss 268
Measurement Angles20°, 60°, 85°
Measurement Spot Size20°: 10 × 10 mm
60°9 × 15 mm
85°5 × 38 mm
Standards ComplianceISO 2813, ISO 7668, ASTM D523, ASTM D2457, DIN 67530, JIS Z 8741, BS 3900, BS 6161-12
Measurement Range20°: 0.0–2000 GU
60°0.0–1000 GU
85°0.0–160 GU
Resolution0.1 GU (0.0–99.9 GU), 1 GU (100–2000 GU)
Repeatability±0.2 GU (0.0–99.9 GU), ±0.2% of reading (100–2000 GU)
Instrument Error±0.5 GU (0.0–99.9 GU), ±0.5% of reading (100–2000 GU)
Data Storage999 measurements with timestamp
Standard Memory50 reference standards
Measurement Speed0.5 s per reading
Power Source1.5 V AA (R6) alkaline batteries (approx. 10,000 measurements)
Auto Power-OffAdjustable 10–99 s
Operating Environment15–40 °C, ≤85% RH (non-condensing at ≤35 °C)
Storage Conditions−10–60 °C, ≤85% RH (non-condensing at ≤35 °C)
Dimensions155 × 73 × 48 mm
Weight400 g
InterfaceRS-232C
Supported LanguagesEnglish, Spanish, German, French, Italian, Japanese
SoftwareGlossLink data exchange software
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelT-10
ComplianceJIS C 1609-1993 Class AA (T-10/T-10M), JIS Precision Grade (T-10P)
PhotodetectorSilicon photodiode
Spectral mismatch (fₛ)≤4% (T-10P), ≤8% (T-10/T-10M)
Cosine error±1% at 10°, ±2% at 30°, ±6% at 50°, ±7% at 60°, ±25% at 80°
Illuminance range0.01–299,900 lx
Integrated illuminance range0.01–999,900×10³ lx·h
Measurement functionsNORM (illuminance), Δ (difference/ratio), Σ (integrated illuminance, duration, hourly average)
Accuracy±2% ±1 digit
Temp./humidity drift±3% ±1 digit
OutputRS-232C digital, analog (1 mV/digit, max 3 V, 10 kΩ output impedance)
Display3-digit (some 4-digit) LCD with auto-backlight
PowerTwo AA alkaline batteries (≥72 h continuous use) or optional AC adapter
Dimensions69 × 174 × 35 mm
Weight (excl. batteries)200 g
Operating environment−10–40 °C, ≤85% RH (non-condensing), Pollution Degree 2, Equipment Class II
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKONICA MINOLTA
OriginJapan
ModelCS-100A
TypeSLR-type tristimulus spot chroma meter for luminance and colorimetric measurement
Luminance Range (Fast)0.01–299,000 cd/m²
Luminance Range (Slow)0.01–49,900 cd/m²
Luminance Accuracy±2% ±1 digit
Chromaticity Accuracy (x,y)±0.004 (CIE Illuminant A, 18–28°C)
Repeatability (Y)±0.2% ±1 digit
Repeatability (x,y)±0.001 (Y ≥100 cd/m², fast mode)
Measurement Angle
Field of View9° with 1° measurement area indicator
Optical System85 mm f/2.8 lens
Focus Distance1014 mm to infinity
SensorThree filtered silicon photodiodes approximating CIE 1931 color matching functions
Response TimeFast (sampling: 0.1 s, display: 0.8–1.0 s)
Slow (sampling0.4 s, display: 1.4–1.6 s)
DisplayExternal 3-digit LCD (Y, x, y)
InterfaceRS-232C (4800 bps)
Power9 V alkaline battery (included)
Dimensions (W×H×D)79 × 208 × 154 mm
Weight890 g (without battery)
Operating Environment0–40°C, ≤85% RH (non-condensing at ≤35°C)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCM-512m3
Illumination/Reception Geometry25°/0°, 45°/0°, 75°/0° (annular multi-angle)
Wavelength Range400–700 nm
Wavelength Interval20 nm
Reflectance Range25°: 0–300%
45°/75°0–200%
Photometric Resolution0.01%
Light SourcePulsed Xenon Arc Lamp
Minimum Measurement Interval7 s (with white calibration tile, at 23°C)
RepeatabilityΔE*ab ≤ 0.05 (n=30, 10-s interval), Spectral Reflectance SD ≤ 0.30%
ApertureMeasuring Ø12 mm / Illuminating Ø20 mm
Temperature SensingThermopile detector (8–13 µm), Range: −10 to +80°C (±2.5°C accuracy at 0–50°C, ε=0.93)
Display128×56-dot reflective LCD (21 chars × 7 lines)
InterfaceRS-232C (D-sub 9-pin female)
Power4× AA alkaline batteries or optional AC adapter
Dimensions (W×H×D)115×257×164 mm
Weight≈1.4 kg (battery-free)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCS-2000
Measurement Wavelength Range380–780 nm
Wavelength Resolution0.9 nm/pixel
Spectral Bandwidth (FWHM)<5 nm
Measurement Angles1°, 0.2°, 0.1° (interchangeable)
Luminance Range (A-source)0.003–5,000 cd/m² (1°), 0.075–125,000 cd/m² (0.2°), 0.3–500,000 cd/m² (0.1°)
Minimum Displayed Luminance0.00002 cd/m²
Luminance Accuracy±2% (A-source, ≥0.003 cd/m²)
Chromaticity Accuracyx,y ±0.003 to ±0.001 (dependent on luminance level)
Repeatability (2σ)Luminance 0.15–0.4%, Chromaticity 0.0004–0.002
Polarization Error≤2% (1°), ≤3% (0.1°/0.2°)
Integration TimeFast mode 0.005–16 s, Normal mode 0.005–120 s
Measurement Time2–243 s
Operating Environment5–35°C, RH ≤80% (non-condensing)
Dimensions158 × 200 × 300 mm (main unit), Ø70 × 95 mm (lens)
Weight6.2 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCM-700d / CM-600d
Illumination/Viewing Geometrydi:8°, de:8° (Diffuse illumination / 8° viewing)
Integrating Sphere DiameterΦ40 mm
DetectorSilicon photodiode array (dual 36-element)
Spectral DispersionDiffraction grating
Wavelength Range400–700 nm
Measurement Interval10 nm
FWHM~10 nm
Light SourcePulsed xenon lamp with UV-cut filter
Measurement TimeApprox. 1 sec
RepeatabilitySpectral reflectance SD ≤0.1%
Instrument ErrorΔE*ab ≤0.2 (MAV/SCI, vs. BCRA Series II 12-color tile set, 23°C)
AperturesMAV: Φ8 mm / Φ11 mm
SAVΦ3 mm / Φ6 mm (CM-700d only)
Display2.36″ TFT color LCD
InterfacesUSB 1.1, Bluetooth® v1.2 (Class 1)
Color SpacesCIE L*a*b*, L*C*h, Hunter Lab, Yxy, XYZ, Munsell (ΔE only for non-Munsell spaces)
Color Difference FormulasΔE*ab (CIE 1976), ΔE94 (CIE 1994), ΔE00 (CIE 2000), CMC (l:c)
Whiteness/Yellowness IndicesWI (ASTM E313), YI (ASTM E313-73 / D1925), ISO Brightness
Gloss8° gloss measurement capability
Data Storage4,000 sample measurements + 1,000 standard references
Power4 × AA alkaline or Ni-MH rechargeable (2300 mAh)
Dimensions73 (W) × 211.5 (H) × 107 (D) mm
Weight~550 g (excl. white calibration tile and batteries)
Operating Environment5–35°C, RH ≤80% (non-condensing)
Storage Environment0–45°C, RH ≤80% (non-condensing)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCA-2500
TypePortable Color Analyzer
Light SourceStandard White Reference Illuminant
Luminance Repeatability±0.5%
Chromaticity Repeatability±0.001
Measurement Range (with ND filter)0.05–100,000 cd/m²
Sensor2/3-inch Monochrome CCD, 1000 × 1000 Pixels, Integrated XYZ Tristimulus Filter (CIE 1931 Match)
Resolution Options980 × 980, 490 × 490, or 196 × 196 (via CA-S25w Software)
InterfaceUSB 2.0
Operating Temperature/Humidity10–30°C, ≤70% RH (non-condensing)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCR-10
Instrument TypePortable
Optical Geometryd/8° integrating sphere (with specular component included)
Light SourcePulsed Xenon Lamp
Spectral Range360 nm – 750 nm
RepeatabilityΔE*ab ≤ 0.1 (standard deviation, measured on certified white tile under D65/10°)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginShanghai, China
ModelCM-3700d
Illumination/Viewing Geometryd/8° (diffuse illumination / 8° viewing), switchable SCI (Specular Component Included) / SCE (Specular Component Excluded)
Transmission Moded/0°
Wavelength Range360–740 nm
Wavelength Interval10 nm
FWHM~14 nm (average)
Photometric Range0–200%
Resolution0.001%
Light SourcePulsed Xenon Arc Lamp
Measurement Time0.6–0.8 s (to data output)
Apertures (Reflectance)LAV (Φ28 mm illum. / Φ25.4 mm meas.), MAV (Φ11 mm / Φ8 mm), SAV (5×7 mm / 3×5 mm)
Aperture (Transmittance)~Φ20 mm
Repeatability (White Tile, 30 measurements @ 10 s intervals)ΔE*ad ≤ 0.005
Repeatability (Black Tile, BCRA Series II, 1% R)ΔE*ad ≤ 0.05
Instrument Error (LAV, BCRA Series II 12-tile avg.)ΔE*ad ≤ 0.08 (typical)
Temperature DriftSpectral Reflectance ±0.10%/°C
UV AdjustmentPC-controlled, continuously variable (1000-step)
Sample Thickness Limit (Transmittance)< 50 mm
InterfaceRS-232C (1200–9600 bps)
Power SupplyAC 100/120/230 V, 50/60 Hz
Operating Environment13–33 °C, ≤80% RH (non-condensing)
Storage Environment0–40 °C, ≤80% RH (non-condensing)
Dimensions (W×H×D)271×259×500 mm
Weight18 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelLS-150 / LS-160
Luminance Range0.001–999,900 cd/m² (LS-150), 0.01–9,999,000 cd/m² (LS-160)
Measurement Angle1° (LS-150), 1/3° (LS-160)
Optical SystemSingle-Lens Reflex (SLR), f=85 mm, F2.8
Spectral ResponseMatched to CIE V(λ) Photopic Luminosity Function
Accuracy±2% ±2 digits (≤1 cd/m²), ±2% ±1 digit (>1 cd/m²)
Repeatability0.2% + 1 digit
Minimum Measured AreaØ14.4 mm (1°), Ø4.5 mm (1/3°)
with optional macro lensØ1.3 mm (LS-150), Ø0.4 mm (LS-160)
Shortest Measurement Distance1012 mm (1°), 213 mm (1/3°)
Stray Light Suppression≤1%
Polarization Error≤1%
Temperature Drift≤0.2%/K
ComplianceDIN 5032-7 Class B (LS-150)
InterfaceUSB 2.0
PowerUSB bus power or AC adapter (optional)
Dimensions71 × 214 × 154 mm
Weight850 g (body only)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelUni-Gloss 60
Measurement Angle60°
Measurement Spot Size9 × 15 mm
Standards ComplianceISO 2813, ISO 7668, ASTM D523, ASTM D2457, DIN 67530, JIS Z 8741, BS 3900, BS 6161-12
Measurement Range (60°)0.0–1000 GU
Resolution0.1 GU (0.0–99.9 GU), 1 GU (100–1000 GU)
Repeatability±0.2 GU (0.0–99.9 GU), ±0.2% of reading (100–1000 GU)
Instrument Error±0.5 GU (0.0–99.9 GU), ±0.5% of reading (100–1000 GU)
Memory Capacity999 measurements with timestamp
Reference Standards Storage50 user-defined standards
Battery Life~10,000 measurements (with 1.5 V AA alkaline cells)
Measurement Interval0.5 s
Auto Power-OffAdjustable 10–99 s
Operating LanguageEnglish, Spanish, German, French, Italian, Japanese
Operating Environment15–40 °C, ≤85% RH (non-condensing at ≤35 °C)
Storage Conditions−10–60 °C, ≤85% RH (non-condensing at ≤35 °C)
Power Supply1.5 V AA (R6) alkaline batteries (4 pcs)
Dimensions155 × 73 × 48 mm (W × H × D)
Weight400 g
InterfaceRS-232C
Included Accessories60° calibration tile (high-gloss ceramic), RS-232C cable, carrying case, GlossLink data exchange software, 4 × AA batteries
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCR-400
TypePortable Color Difference Meter
Optical Geometryd/8° integrating sphere (diffuse illumination, 8° viewing, including specular component)
Light SourcePulsed Xenon Lamp
Spectral RangeY Reflectance: 0.01–160.00%
RepeatabilityΔE*ab ≤ 0.07 (std. dev., white calibration, 30 measurements at 10 s intervals)
Instrument ErrorΔE*ab ≤ 0.6 (BCRA Series II 12-color average, CIE D65 illuminant, 2° standard observer)
Measurement Aperture8 mm (measuring), 11 mm (illumination)
Measurement Time1 s
Minimum Interval3 s
Data Storage1,000 measurements (instrument only)
DisplayBacklit dot-matrix LCD (15×9 characters + icon line)
Power4× AAA alkaline/Ni-MH batteries or AC adapter AC-A17 (120 V / 230 V, 50–60 Hz)
Dimensions102 × 244 × 63 mm
Weight~550 g (incl. batteries, excl. cable)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCM-3700A
Instrument TypeDesktop
Optical Geometryd/8° integrating sphere
Light SourcePulsed Xenon Lamp
Spectral Range360 nm – 740 nm
Wavelength Interval10 nm
FWHM~14 nm
Reflectance Range0–200% (resolution: 0.001%)
RepeatabilityΔE*ab ≤ 0.005 (after white calibration, 30 measurements at 10 s intervals)
Transmittance Geometryd/0°
UV ControlMotorized filter, 1000-step adjustable
Measurement AperturesLAV Φ25.4 mm / Φ28 mm, MAV Φ8 mm / Φ11 mm, SAV 3 mm × 5 mm / 5 mm × 7 mm
Measurement Time0.6–0.8 s
Minimum Interval3 s
Inter-instrument AgreementAvg. ΔE*ab ≤ 0.08, Max. ΔE*ab ≤ 0.3 (BCRA Series II 12-color tile set)
Temperature DriftΔE*ab ≤ 0.05 / °C
Dimensions271 × 274 × 500 mm
Weight18 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCM-2500c / CM-2600d
Instrument TypePortable Spectrophotometer
Optical Geometry45°/0° (CM-2500c) / d/8° integrating sphere with SCI & SCE simultaneous measurement (CM-2600d)
Light SourcePulsed Xenon Lamp(s)
Spectral Range360–740 nm
Spectral Interval10 nm
FWHM~10 nm
RepeatabilityΔE*ab ≤ 0.04 (CM-2600d, MAV/SCI, 30 measurements) or ≤ 0.05 (CM-2500c)
Reflectance Range0–175%
Resolution0.01%
Measurement Time~1.5 s (standard), ~2 s (fluorescence mode)
Aperture OptionsMAV Φ8 mm / Φ11 mm
Data Storage1700 measurements (CM-2600d), 700 (CM-2500c)
ComplianceCIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033-7, JIS Z8722 Condition C
UV AdjustmentDigital real-time UV correction (CM-2600d only)
InterfaceRS-232C
Power4 × AA batteries or AC adapter
Dimensions (W×H×D)69 × 96 × 193 mm
Weight~670 g (battery-free)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCA-100Plus
Measurement ApertureΦ21 mm
Luminance Range (Standard Detector)0.05–1000 cd/m²
Luminance Range (High-Luminance Detector)0.05–2000 cd/m²
Chromaticity Accuracy (White, ≥0.50 cd/m²)±0.003
Luminance Accuracy±2% ±1 digit
Repeatability (Luminance)0.2% + 1 digit (2σ)
Measurement SpeedUp to 20 readings/sec (xyLv)
InterfaceUSB 1.1 or RS-232C (up to 38400 bps)
Storage Channels100
Operating Temperature10–28°C, RH ≤70% (non-condensing)
Power Supply100–240 VAC, 50–60 Hz, 50 VA
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelLS-100
Measurement Angle
Optical SystemSingle-lens reflex (SLR) with f=85 mm, F2.8 lens
Field of View (Finder)9° with diopter adjustment
Minimum Measurement Distance∞ to 1,014 mm (∞ to φ14.4 mm)
Close-up Lens OptionsNo.153 (φ18.7–φ8.0 mm), No.135 (φ8.7–φ5.2 mm), No.122 (φ4.3–φ3.2 mm), No.110 (φ1.5–φ1.3 mm)
DetectorSilicon photodiode
Spectral Match Error (fₛ) vs. CIE V(λ)≤8%
Response ModesFAST (display time: 0.8–1.0 s
sampling0.1 s), SLOW (display time: 1.4–1.6 s
sampling0.4 s)
Luminance RangeFAST: 0.001–299,900 cd/m²
SLOW0.001–49,990 cd/m²
Accuracy±2% ±1 digit (>1,000 cd/m²)
Repeatability±0.2% ±1 digit (>1,000 cd/m²)
Temperature Drift±3% ±1 digit (0–40°C, referenced to 20°C)
Calibration ModesKonica Minolta standard or user-defined reference (1 channel)
Luminance Ratio Mode1 channel, manual or measured reference input
Measurement FunctionsInstant luminance, peak luminance, luminance ratio (%)
Display4-digit LCD (external & eyepiece-mounted with LED backlight)
Digital InterfaceRS-232C compliant (4,800 bps), bidirectional control and data logging
Power SupplyOne 9 V alkaline battery (JIS 6F22) or optional DP-10 data printer
Power Consumption~16 mA (finder illumination ON, measurement active)
Operating Environment0–40°C, ≤85% RH (non-condensing), Pollution Degree 2, Overvoltage Category II
Storage−20–55°C, ≤85% RH (non-condensing)
Dimensions & Weight79 × 208 × 150 mm (W × H × D), 850 g (body only, no battery)
Standard AccessoriesLens cap, eyepiece ND filter, eyepiece cap, 9 V battery, carrying case
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelT-10 / T-10M
Illuminance Range0.01–299,900 lx
Spectral Matchf₁′ ≤ 8% (vs. CIE V(λ))
Cosine Error±1% at 10°, ±7% at 60°, ±25% at 80°
Accuracy±2% ±1 digit (Konica Minolta standard)
OutputRS-232C digital, 1 mV/digit analog
DisplayBacklit LCD, 3–4 digits
Operating Temp–10 to 40°C, RH ≤ 85% (non-condensing)
Power2× AA batteries (≥72 h continuous)
Dimensions (main unit)69 × 161.5 × 30 mm
Weight200 g (T-10), 205 g (T-10M)
ComplianceDesigned for ISO/CIE-compliant photometric measurement in industrial, architectural, and environmental applications
Added to wishlistRemoved from wishlist 0
Add to compare
BrandKonica Minolta
OriginJapan
ModelCS-200
TypeColor and Luminance Meter
Measurement PrincipleSpectral Matching Method (CIE 1931 2° Standard Observer)
PhotodetectorSilicon Photodiode
Luminance Range0.01 to 20,000,000 cd/m² (at 0.1° field of view)
Illuminance Range0.1 to 99,990 lx
Chromaticity Accuracy (xy)±0.002 (Std. Illuminant A, 800 lx)
Repeatability (xy)±0.0005
Temporal Stability2 measurements/sec (continuous)
Output InterfaceRS-232C
Display4-digit LCD with backlight
PowerFour AA alkaline batteries (≈72 h continuous use) or AC adapter AC-A10
Dimensions69 × 174 × 35 mm
Weight215 g (body only)
ComplianceJIS C 1609-1993 Class AA (fs ≤ 8%), ISO/CIE traceable calibration protocols