Empowering Scientific Discovery

Beijing Hope Century Technology Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHE-480R
Sensor InputSingle-channel (electrode constant: 0.01 cm⁻¹)
Resistivity Range0–0.200 / 0–2.00 / 0–20.0 / 0–100.0 MΩ·cm
Alternate Range0–2.00 / 0–20.0 / 0–200.0 / 0–1000 kΩ·cm
Temperature Range0–100 °C (configurable 1- or 2-digit display)
Repeatability±0.5% (max, equivalent input)
Linearity±0.5% (max, equivalent input)
Output1 × isolated 4–20 mA DC (max load 900 Ω)
Relay Outputs2 × SPDT (240 V AC / 30 V DC, 3 A resistive)
Digital InterfaceRS-485 (2-wire, input/output isolated)
Temp. CompensationPt1000 sensor
Enclosure RatingIP65 (outdoor-rated)
ComplianceCE, FCC, RoHS, WEEE
Compatible SensorsHORIBA ERF-series resistivity sensors (0.01 cm⁻¹)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHE-200C
Enclosure RatingIP65
Power Supply90–264 V AC, 50/60 Hz
OutputThree isolated 4–20 mA analog outputs (configurable per parameter)
Relay OutputsThree SPDT dry-contact relays (R1–R3, user-assignable for high/low alarms, USP qualification, or fault indication)
CommunicationRS-485 (2-wire, galvanically isolated input/output circuitry)
Temperature CompensationPt1000 sensor (0 °C reference), 0–100 °C range
Measurement ParametersConductivity (0.000 µS/cm to 200.0 mS/cm, selectable cell constant: 0.01, 0.1, or 1.0 cm⁻¹), TDS (0.00–2000 mg/L, user-defined conversion factor 0.30–1.00), and temperature (0–100 °C, ±0.1 °C accuracy at reference point)
AccuracyConductivity ±0.5% FS (max), TDS ±1.5% FS (max), temperature ±0.5 °C linearity error
RepeatabilityConductivity ±0.5% FS (max), TDS ±1.5% FS (max)
Environmental RatingOutdoor-rated aluminum housing, CE/FCC/WEEE/RoHS compliant
MountingPole-mounted (50A pipe) or wall-mounted
Weight~4.5 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHP-100
Measurement PrincipleGlass Electrode Method
pH Range0–14
pH Repeatability±0.05 pH (max)
Output Signal4–20 mA DC (max load resistance: 500 Ω)
Power Supply22.5–30 V DC (nominal 24 V DC)
Power Consumption~0.48 W
Temperature Compensation Range0–100 °C
Ambient Operating Temperature−10 to +50 °C
Temperature SensorPt1000 (at 0 °C)
Enclosure RatingIP65
ComplianceCE, FCC, RoHS, WEEE
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHE-480C
EnclosureAluminum, IP65-rated
ComplianceCE, FCC, RoHS, WEEE
Input Channels1-channel conductivity sensor interface
Electrode Constants Supported0.01 cm⁻¹, 0.1 cm⁻¹, 1.0 cm⁻¹
Conductivity Range (selectable per electrode constant)2.000–20.00 µS/cm (0.01 cm⁻¹), 20.00–200.0 µS/cm (0.1 cm⁻¹), 200.0–2000 µS/cm / 0.2000–2.000 mS/cm / 2.000–20.00 mS/cm / 20.00–200.0 mS/cm (1.0 cm⁻¹)
TDS Range2.00–20.0 mg/L, 20.0–200 mg/L, 200–2000 mg/L (user-defined conversion factor 0.30–1.00)
Temperature Range0–100 °C (adjustable decimal places: 0/1/2)
Repeatability±0.5% of full scale (conductivity), ±1.5% FS (TDS)
Analog Output4–20 mA DC, isolated, max. load 900 Ω
Relay Outputs2 × SPDT (240 V AC / 30 V DC, 3 A resistive)
Temp. Compensation OptionsUltra-pure water mode (ref. 25 °C), user-defined reference temp. (5–95 °C) & temp. coefficient (±5 %/°C), NaCl curve, or none
Sensor CompatibilityHORIBA ESH-series and FS-series conductivity sensors
Power Supply90–264 V AC, 50/60 Hz, ≤10 VA
Operating Ambient−5 to +45 °C
Weight~4.5 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHE-200H
Measurement PrincipleFour-Electrode Conductivity
Range0.0–200.0 mS/cm (auto-ranging, dual-scale: 0.00–20.00 mS/cm & 0.0–200.0 mS/cm)
Temperature Compensation0–100 °C (Pt1000 sensor)
Accuracy±0.5% FS (0–20.00 mS/cm), ±1.0% FS (20.0–200.0 mS/cm)
Repeatability±0.1 °C (temperature), linearity ±0.5 °C
Enclosure RatingIP65
ComplianceCE, FCC, RoHS, WEEE
OutputThree isolated 4–20 mA analog outputs (max load 900 Ω), three SPDT relay contacts (R1/R2/failure), RS-485 Modbus RTU interface
Power Supply90–264 V AC, 50/60 Hz, ≤15 W
Operating Ambient−20 to +55 °C
Weight~4.5 kg
Housing MaterialAnodized aluminum alloy
MountingPole-mounted (50A pipe) or wall-mountable
Electrode Constant Options0.1 cm⁻¹ or 1.0 cm⁻¹
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHO-480
ORP Measurement Range−2000 to +2000 mV
Resolution1 mV
Repeatability±5 mV (max, at equivalent input)
Linearity±5 mV (max, at equivalent input)
Output Signal4–20 mA DC (galvanically isolated, max. load resistance: 900 Ω)
Relay Outputs2 SPDT relays (240 V AC / 30 V DC, 3 A resistive)
CalibrationSensitivity adjustment (−200 to +200 mV), span range adjustable from 50.0% to 150.0%
Operating Temperature−5 to +45 °C
Power Supply90–264 V AC, 50/60 Hz, ≤10 VA
ComplianceCE-marked, FCC-compliant, RoHS-conformant
Electrode TypeLead-free metal reference electrode and ToupH combination electrode
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHP-480
Measuring Range (pH)0–14
Resolution (pH)0.01 pH
Temperature Range0–100 °C
Resolution (Temperature)1 °C (selectable display)
Repeatability (pH)±0.05 pH (at equivalent input)
Output4–20 mA DC, isolated, max. load resistance 900 Ω
Relay Outputs2 points (SPDT, 240 V AC / 30 V DC, 3 A)
Calibration2-point auto-calibration (using standard buffers pH 2, 4, 7, 9, 10) and manual calibration
Temperature CompensationAutomatic, selectable sensor types (500 Ω, 6.8 kΩ, 350 Ω, 1 kΩ, 10 kΩ) or none
Enclosure RatingIP65 (panel mount)
Power Supply90–264 V AC, 50/60 Hz, ≤10 VA
Environmental Operating Range−5 to +45 °C
ComplianceCE, FCC, RoHS
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHE-300R
Enclosure RatingIP65
Measurement PrincipleTwo-Electrode Conductivity Method (Inverse Calculation for Resistivity)
Resistivity Range0.000–2.000 MΩ·cm (with 0.01 cm⁻¹ electrode), 0.00–20.00 kΩ·cm (with 0.1 cm⁻¹ electrode), 0.0–200.0 kΩ·cm (with 1.0 cm⁻¹ electrode)
Temperature Range0–100 °C
Temp. Resolution0.01 °C
Temp. Accuracy±0.02 °C (at reference point, Pt1000 sensor)
Resistivity Accuracy±0.01 MΩ·cm (max, calibrated per electrode constant)
Repeatability±0.1% (resistivity), ±0.1 °C (temperature)
Linearity±0.5% (resistivity), ±0.5 °C (temperature)
OutputIsolated 4–20 mA DC (max load 600 Ω)
Digital InterfaceRS-485 (2-wire, galvanically isolated I/O)
Power Supply24 V DC (21–32 V DC operating range), ≤0.6 W
Operating Ambient−20 to +60 °C
Housing MaterialAnodized Aluminum
MountingPole-mount (50A pipe) or wall-mount
Sensor CompatibilityHORIBA ERF-series resistivity sensors (e.g., ERF-001, electrode constant 0.01 cm⁻¹)
ComplianceCE, FCC, RoHS, WEEE
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHD-300
Measurement PrincipleClark-type Polarographic Membrane Sensor
DO Range0–20 mg/L (resolution: 0.01 mg/L)
Saturation Range0–200% (resolution: 1%)
Temperature Range0–50 °C (resolution: 0.1 °C)
Repeatability±0.5% FS (DO), ±0.5 °C (temperature)
Linearity±0.5% FS (DO), ±0.5 °C (temperature)
OutputIsolated 4–20 mA DC (max. load: 600 Ω)
Digital InterfaceRS-485 (2-wire, input/output isolated)
Power Supply24 V DC (21–32 V DC operating range), max. 0.6 W
Enclosure RatingIP65
Ambient Operating Temperature−20 to +60 °C
Housing MaterialAluminum alloy
MountingWall or 50A pipe column
Compatible ProbesDO-1100 (integrated Pt1000 temperature sensor)
Compliant StandardsCE, FCC, RoHS, WEEE
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHU-200TB
Measurement PrincipleDual-source 90° light scattering
Turbidity Range (Formazin)0–1000 NTU
Resolution0.1 NTU (10–100 NTU range)
Repeatability±2% or ±0.04 NTU (whichever is greater)
Linearity Deviation±2% or ±0.04 NTU at mid-range calibration point
OutputIsolated 4–20 mA DC (max load 900 Ω)
Relay Outputs3 SPDT dry contacts (configurable for alarm, cleaning activation, or fault indication)
Digital InterfaceRS-485 (2-wire, input/output isolated)
Enclosure RatingIP65
ComplianceCE, FCC, RoHS, WEEE
Operating Temperature0–55 °C
Power Supply90–264 V AC, 50/60 Hz, ≤35 VA
Weight~4.5 kg
Housing MaterialAluminum alloy
MountingPole-mounted (50A pipe) or wall-mount
Cleaning SystemMotor-driven wiper (integrated, converter-triggered)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHO-200
ORP Range−2000 to +2000 mV
ORP Resolution1 mV
Temperature Range0–100 °C
Temperature Resolution0.1 °C
ORP Repeatability±5 mV (max)
ORP Linearity±5 mV (max)
Temperature Accuracy±0.3 °C (equivalent input)
OutputTwo isolated 4–20 mA DC outputs (max load: 900 Ω)
Relay Outputs3 SPDT (1c) contact points
Digital InterfaceRS-485 (2-wire, input/output isolated)
Enclosure RatingIP65
Power Supply90–264 V AC, 50/60 Hz
Power Consumption≤15 VA
Operating Ambient Temperature−20 to +55 °C
Housing MaterialAluminum alloy
MountingWall or 50A pipe column
Weight~4.5 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHD-200
Protection RatingIP65
Measurement PrincipleClark-type Polarographic Membrane Sensor
DO Range0–20 mg/L (Resolution: 0.01 mg/L)
Saturation Range0–200% (Resolution: 1%)
Temperature Range0–50 °C (Resolution: 0.1 °C)
Repeatability±0.5% FS (DO), ±0.5 °C (Temp)
Linearity±0.5% FS (DO), ±0.5 °C (Temp)
OutputThree isolated 4–20 mA analog channels (max load: 900 Ω), three SPDT relay outputs, one dry-contact input, RS-485 Modbus RTU interface
Power Supply90–264 V AC, 50/60 Hz, ≤15 VA
Operating Ambient−20 to +55 °C
Enclosure MaterialAnodized Aluminum
Weight~4.5 kg
CompliantCE, FCC, RoHS, WEEE
Compatible ProbesDO-1100 (integrated Pt1000 temperature sensor), Electrodes: 5505, 5510
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHP-200
IP RatingIP65
Measurement PrincipleGlass Electrode Method
pH Range0–14
pH Resolution0.01
pH Repeatability±0.03
Temperature Range0–100 °C
Temperature Resolution0.1 °C
Temperature Accuracy±0.3 °C (equivalent input)
OutputTwo isolated 4–20 mA DC outputs (max load: 900 Ω)
Relay Outputs5 SPDT (1c) contacts, configurable for alarm/control
InputDry-contact hold input for electrode cleaning/operation status
CommunicationRS-485 (2-wire, input/output isolated)
Power Supply90–264 V AC, 50/60 Hz, ≤15 W
Enclosure MaterialAluminum alloy
MountingPole (Ø50 mm) or wall-mount
Ambient Operating Temperature−20 to +55 °C
Weight~4.5 kg
ComplianceCE, FCC, RoHS, WEEE
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHP-300
pH Range0–14
pH Resolution0.01
pH Repeatability±0.03
pH Linearity±0.03
Temperature Range0–100 °C
Temperature Resolution0.1 °C
Temperature Accuracy±0.3 °C (equivalent input)
Output4–20 mA DC, isolated, max. load resistance 600 Ω
Power Supply24 V DC (21–32 V DC), max. 0.6 W
Enclosure RatingIP65
Ambient Operating Temperature−20 to +60 °C
Housing MaterialAluminum alloy
MountingPipe (50A) or wall-mounted
Weight~4 kg
ComplianceCE, FCC, RoHS, WEEE
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHE-200R
Measurement PrincipleTwo-Electrode Conductivity Method (Inverse Calculation for Resistivity)
Resistivity Range0.000–2.000 MΩ·cm (with K=0.01/cm electrode), 0.00–20.00 kΩ·cm, 0.0–200.0 kΩ·cm
Temperature Range0–100 °C
Temperature CompensationPt1000 sensor (0 °C reference), 0–100 °C range
Accuracy (Resistivity)±0.01 MΩ·cm max (at reference temperature, standard units)
Accuracy (Temperature)±0.02 °C max
Repeatability (Resistivity)±0.1% FS
Linearity (Resistivity)±0.5% FS
OutputIsolated 4–20 mA (2 points), SPDT relay outputs (3 points), dry-contact input (1 point)
CommunicationRS-485 (2-wire, input/output isolated)
Enclosure RatingIP65
ComplianceCE, FCC, RoHS, WEEE
Power Supply90–264 V AC, 50/60 Hz, ≤15 VA
Operating Ambient−20 to +55 °C
Weight~4.5 kg
Sensor CompatibilityERF-series resistivity sensors (K=0.01 cm⁻¹)
CalibrationElectrode constant entry, single-point temperature calibration, atmospheric or saturated solution reference modes
Special FunctionsUltrapure water resistivity mode, trim adjustment, real-time self-diagnosis (electrode fault, analyzer error)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHR-480
Measurement PrincipleGalvanic Cell (Electrochemical)
MeasurandFree Residual Chlorine
Range0–2 mg/L
Resolution0.01 mg/L
Repeatability±0.05 mg/L (max)
Linearity±0.05 mg/L (max)
OutputIsolated 4–20 mA DC (max load: 900 Ω)
Relay Outputs2 × SPDT (240 V AC / 30 V DC, 3 A)
Enclosure RatingIP65
CertificationsCE, FCC, RoHS, WEEE
Operating Temperature−5 to +45 °C
Power Supply90–264 V AC, 50/60 Hz, ≤10 W
Compatible SensorHORIBA RS-1 Free Chlorine Electrode
CalibrationZero calibration (using zero solution), Span calibration (against DPD colorimetric reference)
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHR-200
Measurement PrincipleAmperometric (Polarographic)
Range0–3 mg/L Cl₂
Resolution0.01 mg/L
Repeatability±0.05 mg/L (max)
Linearity±0.05 mg/L (max)
Temperature Range0–50 °C
Temp. Resolution0.1 °C
Temp. Accuracy±0.5 °C
OutputTwo isolated 4–20 mA DC (max load 900 Ω)
Relay Outputs3 SPDT (1c), configurable for alarm/control/clean activation
Digital InterfaceRS-485 (2-wire, input/output isolated)
Enclosure RatingIP65
Power Supply90–264 V AC, 50/60 Hz, ≤35 VA
Operating Ambient0–55 °C
Weight~4.5 kg
ComplianceCE, FCC, RoHS, WEEE
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHD-480
Measurement PrincipleGalvanic Cell (Electrochemical)
DO Range0–20 mg/L (resolution: 0.01 mg/L)
Saturation Range0–200% (resolution: 1%)
Temperature Range0–40 °C (resolution: 0.1 °C)
Repeatability±0.5% FS (max)
Linearity±0.5% FS (max)
OutputIsolated 4–20 mA DC (max load: 900 Ω)
Relay Outputs2 × SPDT (240 V AC / 30 V DC, 3 A)
Calibration OptionsAir saturation, saturated aqueous solution, salinity (0.0–5.0%), dual-point temperature compensation
Sensor CompatibilityDP-1100 probe with built-in Pt1000 thermistor
Power Supply90–264 V AC, 50/60 Hz
Max Power Consumption10 W
Operating Ambient−5 to +45 °C
ComplianceCE, FCC, RoHS
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHU-200SS
Measurement PrincipleOptical Transmission (Infrared LED Light Path)
Measuring Range (MLSS)0–20,000 mg/L
Resolution1 mg/L (0–10,000 mg/L), 10 mg/L (10,000–22,000 mg/L)
Accuracy±3% of reading or ±10 mg/L (whichever is greater)
OutputIsolated 4–20 mA DC (max. load 900 Ω)
Relay Outputs3 SPDT dry contacts (configurable for alarm, cleaning, fault)
Digital InterfaceRS-485 (2-wire, input/output isolated)
Enclosure RatingIP65
Ambient Operating Temperature−20 °C to +55 °C
Power Supply90–264 V AC, 50/60 Hz, ≤35 VA
WeightApprox. 4.5 kg
ComplianceCE, FCC, RoHS, WEEE
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHO-300
ORP Range−2000 to +2000 mV
ORP Resolution1 mV
ORP Repeatability±5 mV (max)
ORP Linearity±5 mV (max)
Temperature Range0–100 °C
Temperature Resolution0.1 °C
Temperature Accuracy±0.3 °C (equivalent input)
Output4–20 mA DC, isolated, max load resistance 600 Ω
Digital InterfaceRS-485 (2-wire, isolated I/O)
Power Supply24 V DC (21–32 V DC), max 0.6 W
Enclosure RatingIP65
Ambient Operating Temperature−20 to +60 °C
Housing MaterialAluminum alloy
Mounting50A pipe or wall-mount
Weight~4 kg
Added to wishlistRemoved from wishlist 0
Add to compare
BrandHORIBA
OriginJapan
ModelHE-480H
Measurement PrincipleFour-Electrode Conductivity
Enclosure RatingIP65
ComplianceCE, FCC, RoHS, WEEE
InputSingle-Channel FES-Series Electrode (Cell Constant: 1.0 cm⁻¹)
Conductivity Range0.00–20.00 mS/cm
Temperature Range0–100 °C
Temp. CompensationAdjustable Reference Temp (5–95 °C) & Temp. Coefficient (±5 %/°C)
Output4–20 mA (Isolated, Max Load 900 Ω)
Relay Outputs2 × SPDT (240 V AC / 30 V DC, 3 A)
Repeatability±0.5 % FS (±1.0 % at 500 mS/cm or salinity mode)
Salinity Conversion0.00–4.00 % NaCl / 0.0–20.0 % Salt
Operating Ambient−5 to +45 °C
Show next
InstrumentHive
Logo
Compare items
  • Total (0)
Compare
0