- All
- Favorite
- Popular
- Most rated
| Brand | Kyocera |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic |
| Model | LC3000A |
| Instrument Type | High-Pressure Preparative Liquid Chromatograph |
| Maximum Pressure (P3210) | 42 MPa |
| Maximum Pressure (P3220) | 30 MPa |
| Flow Rate Range (P3210) | 0.001–9.999 mL/min |
| Flow Rate Range (P3220) | 0.01–49.99 mL/min |
| Flow Accuracy | ≤0.4% |
| Flow Precision (RSD) | ≤0.1% |
| UV/VIS Detector (UV3000A) | Wavelength Range 190–700 nm |
| Light Source | Deuterium Lamp (standard), Tungsten Lamp (optional) |
| Wavelength Accuracy | ≤±2 nm |
| Wavelength Repeatability | 0.4 nm |
| Absorbance Range | 0–2 AU |
| Absorbance Error | <±5% |
| Baseline Noise (254 nm) | ≤1.5×10⁻⁵ AU (dynamic, 1 mL/min, MeOH/H₂O 80:20) |
| Baseline Drift (254 nm) | ≤4×10⁻⁴ AU/h (dynamic, 1 mL/min, MeOH/H₂O 80:20) |
| Time Constants | 0.1/0.2/0.5/1/2/5/10 s |
| Auto-Zero Function | Full-scale |
Show next