- All
- Favorite
- Popular
- Most rated
| Brand | AIRTECH (Suzhou Antai) |
|---|---|
| Origin | Jiangsu, China |
| Instrument Type | Class II, Type B2 (Total Exhaust) Biosafety Cabinet |
| External Dimensions (L×W×H) | 1500 × 790 × 2050 mm |
| Internal Work Area Dimensions (L×W×H) | 1270 × 600 × 620 mm |
| HEPA Filter Efficiency | ≥99.995% @ 0.3 µm |
| ULPA Filter Efficiency | ≥99.9995% @ 0.12 µm |
| Downflow Velocity | 0.35 m/s |
| Inflow Velocity | 0.55 m/s |
| Noise Level | ≤65 dB(A) |
| Power Supply | AC 220 V, 1φ, 50 Hz |
| Rated Power | 1.65 kW |
| Weight | 270 kg |
| Air Cleanliness Class | ISO 4 (equivalent to Federal Standard 209E Class 10) |
| Personnel Protection | ≤5 CFU/plate (slit sampler) |
| Product Protection | ≤2 CFU/plate |
| Cross-Contamination Protection | ≤5 CFU/plate (impact sampler), ≤5 CFU/plate (slit sampler) |
| Lighting | 24.5 W LED, ≥900 lx uniformity |
| UV Lamp | 25 W, germicidal |
| Exhaust Direction | Top-mounted ducted exhaust |
| Brand | AIRTECH (Suzhou Jintan Airtech) |
|---|---|
| Origin | Jiangsu, China |
| Model | BSC-1600IIA2 |
| Instrument Type | Class II, Type A2 Biological Safety Cabinet |
| External Dimensions (L×W×H) | 1800 × 790 × 2050 mm |
| Internal Work Area Dimensions (L×W×H) | 1600 × 600 × 620 mm |
| Filtration Efficiency | HEPA ≥99.995% @ 0.3 µm |
| Downflow Velocity | 0.35 m/s |
| Inflow Velocity | 0.55 m/s |
| Noise Level | ≤62 dB(A) |
| Lighting Intensity | ≥800 lx |
| Power Consumption | 2.0 kW |
| Weight | 360 kg |
| Compliance | YY 0569–2011 (China), NSF/ANSI 49–2022 (USA) |