Empowering Scientific Discovery

Beijing Zolix Instruments Co., Ltd.

Categories
  • All
  • Favorite
  • Popular
  • Most rated
Added to wishlistRemoved from wishlist 0
Add to compare
BrandZOLIX
OriginBeijing, China
Manufacturer TypeDirect Manufacturer
Country of OriginChina
Instrument TypeVertical Cryostat
Base Temperature<3.5 K
Cooling Time (2nd Stage, Empty Load)<60 min to 4.2 K
Operating Temperature Range1.4 K – 300 K
Vacuum Chamber Pressure<1×10⁻³ Pa
Standard Optical Windows4 × quartz, Ø25 mm
Optional Window MaterialsCaF₂, sapphire, ZnSe
Sample Environment OptionsHigh-vacuum, He-gas (In-sealed), customizable sample space
Vibration LevelLow-noise mechanical design
Electrical Feedthroughs6-pin standard + optional 1–19-pin SMA/BNC/TRB connectors
Temperature Control Resolution±1 mK (with optional sensors)
Integrated Temperature ControllerDual-channel input, 1 active control loop, lower limit 300 mK
Compatible CryocoolerTwo-stage pulse tube cooler (PTC) with 1.3 W @ 4.2 K cooling power
Helium Compressor ModelKDC6000V
Water Cooling Requirement7–10 L/min @ 5–25 °C inlet, ΔT ≤ 19 °C
Total System Power Consumption6.5 kW (steady-state, 50 Hz)
Molecular PumpDN100 CF flange, 110 L/s (air), base pressure 6×10⁻⁷ Pa
Pre-pumpRV4 rotary vane pump (standard)
Weight (cryostat + cold head)20 kg
(compressor + chiller)118 kg