| Brand | Huipu |
|---|---|
| Origin | Hebei, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Pharmaceutical & Medical Device Residue Analysis |
| Detection Method | Headspace-GC/FID |
| Column | PEG-20M, 30 m × 0.32 mm × 0.5 µm |
| Oven Temp | 120 °C (isothermal) |
| Injector & Detector Temp | 150 °C |
| Carrier Gas | High-Purity N₂ (35 mL/min) |
| H₂ Flow | 35 mL/min |
| Air Flow | 350 mL/min |
| Headspace Temp | 80 °C (10 min equilibration) |
| Quantitation Range | EO: 0.1–100 µg/g |
| ECH | 3–100 ppm |
| Compliance Standards | GB/T 14233.1–2008, GB/T 16886.7–2015, ISO 10993-7:2008 |
| Brand | INFICON |
|---|---|
| Origin | USA |
| Model | Fusion |
| Instrument Type | Laboratory Micro Gas Chromatograph |
| Temperature Control Range | Ambient +5°C to 300°C |
| Maximum Ramp Rate | 300°C/min |
| Cooling Rate | 250°C to 50°C in <1 min |
| Carrier Gas Flow Range & Control | 3 mL/min |
| Carrier Gas Pressure Range & Control | 60 PSI |
| Injector Max Operating Temperature | 140°C |
| Injector Pressure Setting Range | 0–20 PSI |
| Injector Total Flow Setting Range | 0–10 mL/min |
| Brand | InFoMass |
|---|---|
| Origin | Beijing, China |
| Model | L10 |
| Cooling Options | Liquid Nitrogen (−189 °C) or Electronic (−90 °C) |
| Modulation Range | C₄–C₄₅ (LN₂ mode) |
| Column Configuration | Dual-column serial coupling via cryogenic loop modulator |
| Modulation Cycle | Programmable |
| Cold Jet Flow | Precisely controllable |
| Software Compatibility | MDCTM (GC Image LCC) |
| Structural Design | Fixed-angle column routing bracket eliminating manual alignment calibration |
| Moving Parts | None |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-Purpose with Baijiu (Chinese Liquor) Analysis Optimization |
| Temperature Control Zones | 6 Independent Channels (including Column Oven, Injector, Detector, AUX1–AUX3) |
| Column Oven Temp Range | +7 °C to 420 °C |
| Programmed Ramp Stages | 5-Stage Linear/Nonlinear |
| Oven Cooling Rate | Ambient air-cooled with motorized auto-backdoor |
| Carrier Gas Control | External modular gas control boxes (capillary, carrier, fuel/oxidant circuits) with manual flow/pressure adjustment |
| Injector Max Temp | 450 °C |
| Injector Pressure Range | 0–100 psi (adjustable) |
| Total Injector Flow Range | 0–200 mL/min |
| Detector Compatibility | Up to three simultaneous detectors (FID, TCD, FPD, NPD, ECD) |
| Safety Protections | Over-Temperature, Over-Current (TCD), and System Crash Protection |
| Display | Backlit LCD with color-coded membrane keyboard (green = temp, blue = detector, white = numeric, black = event) |
| Physical Design | Vertical column oven |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | GC1690 for DME in Fuels |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-Purpose with DME-Specific Configuration |
| Temperature Control Range | +7 °C to 420 °C |
| Programmable Ramp Stages | 5-step oven temperature programming |
| Carrier Gas Flow Control | External modular flow controller (capillary/twin-channel configuration) |
| Carrier Gas Pressure Control | Adjustable via external pressure regulator |
| Injector Maximum Operating Temperature | ≤420 °C |
| Injector Pressure Setting Range | 0–100 psi (adjustable) |
| Total Injector Flow Range | 0–100 mL/min (carrier + makeup gases) |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Authorized Distributor |
| Instrument Type | Laboratory Gas Chromatograph |
| Application | EO Residue Quantification in Medical Devices |
| Temperature Control Range | Up to 400 °C (Oven), 200 °C (Injector), 250 °C (Detector) |
| Heating Rate | 0.1–40 °C/min (programmable) |
| Cooling Rate | Ambient cooling with forced air assist |
| Carrier Gas Flow Control | Digital electronic pressure/flow control (EPC) for N₂, H₂, Air |
| Carrier Gas Pressure Range | 0–100 psi |
| Injector Max Temp | 200 °C |
| Injector Pressure Setting Range | 0–100 psi |
| Total Injector Flow Range | 0–100 mL/min |
| Compliance Standard | Aligned with GB/T 14233.1–2008, ISO 10993-7, USP <88>, ASTM F2889 |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Manufacturer Type | Authorized Distributor |
| Product Category | Domestic |
| Model | GC1690 for Mixed Inorganic & Organic Gases |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-Purpose |
| Temperature Control Range | +7 °C to 420 °C (Column Oven) |
| Programmable Ramp Rate | 5-Stage Programmed Temperature Ramping |
| Carrier Gas Flow Control | External Modular Flow Controller (Capillary, Carrier, Fuel/Auxiliary Gas Boxes) |
| Carrier Gas Pressure Control | Electronic Pressure Control (EPC)-Compatible Design |
| Injector Maximum Operating Temperature | ≤420 °C |
| Injector Pressure Setting Range | 0–100 psi (adjustable) |
| Total Injector Flow Setting Range | 0–500 mL/min (dependent on configuration) |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Model | GC1690 |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-purpose analysis of organic/inorganic compounds and gases (boiling point ≤ 399 °C) |
| Temperature Control Zones | 6 independent zones (including column oven, injector, detector, AUX1) |
| Column Oven Temp Range | +7 °C to 420 °C |
| Programmed Ramp | 5-step programmable heating |
| Injector Max Temp | 420 °C |
| Carrier Gas Pressure Control | External modular pressure controllers (single/dual channel) |
| Carrier Gas Flow Control | External modular flow controllers (single/dual channel) |
| Detector Options | FID, TCD, FPD, NPD, ECD (up to three simultaneously installed) |
| Injection Modes | Packed-column on-column, split/splitless, wide-bore capillary (WBC), packed-column vaporizing, 6-port gas sampling valve |
| Safety Protections | Over-temperature, over-current (TCD), and system crash protection with automatic power cutoff and audible/visual alarm |
| Data Interface | RS-232 or USB (optional) |
| Compliance | Designed for GLP-compliant laboratory environments |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-purpose with H₂ production process gas analysis focus |
| Temperature Control Zones | 6 independent zones (including column oven, injector, detector, AUX1) |
| Column Oven Temp Range | +7 °C to 420 °C |
| Programmed Ramp | 5-step temperature programming |
| Injector Max Temp | ≤420 °C |
| Carrier Gas Pressure Control | External modular pressure controllers (dual-channel optional) |
| Carrier Gas Flow Control | External modular mass flow controllers (dual-channel optional) |
| Detector Compatibility | FID, TCD, FPD, NPD, ECD (up to three simultaneously) |
| Injection Modes | Packed-column on-column, split/splitless, wide-bore capillary (WBC), packed-column vaporizing, 6-port gas sampling valve |
| Safety Protections | Over-temperature, over-current (TCD-specific), and system crash auto-shutdown with audible/visual alarm |
| Column Oven Architecture | Vertical design with active rear-door cooling |
| Display | Backlit LCD with color-coded membrane keyboard (green = temp, blue = detector, white = numeric, black = event control) |
| Compliance | Designed for GLP-compliant lab environments |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-Purpose with Pesticide Residue Focus |
| Temperature Control Zones | 6 independent zones (including column oven, injector, detector, AUX1–AUX3) |
| Column Oven Temp Range | +7 °C to 420 °C |
| Programmed Temperature Ramping | 5-step linear ramping |
| Injector Max Temp | 420 °C |
| Carrier Gas Pressure Control | Electronic pressure control (EPC) compatible |
| Carrier Gas Flow Control | Mass flow controller (MFC) optional |
| Detector Compatibility | FID, TCD, FPD, NPD, ECD (up to 3 simultaneously) |
| Injection Modes | Split/splitless, on-column, packed-column vaporizing, WBC for wide-bore capillary, 6-port gas sampling valve |
| Safety Protections | Over-temperature, over-current (TCD), and system crash protection circuits |
| Data Interface | RS-232 or USB (legacy PC connectivity) |
| Compliance | Designed to support GLP-compliant workflows |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | Power Industry SF₆ & Insulating Gas Analysis |
| Temperature Control Zones | 6 (including column oven, injector, detector, AUX1–AUX3) |
| Column Oven Temp Range | +7 °C to 420 °C |
| Programmed Ramp | 5-step ramp capability |
| Oven Cooling | Forced-air active cooling with auto rear door opening |
| Carrier Gas Control | External modular gas control boxes (capillary, carrier, fuel/oxidant circuits) |
| Injector Max Temp | 420 °C |
| Injector Pressure Range | 0–100 psi (adjustable) |
| Injector Total Flow Range | 0–100 mL/min |
| Detector Compatibility | FID, TCD, FPD, NPD, ECD (up to 3 simultaneously) |
| Safety Protections | Over-temperature, over-current (TCD), and system crash protection circuits |
| Compliance | Designed for GLP-compliant operation |
| Brand | Jiedao |
|---|---|
| Origin | Zhejiang, China |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Focus | Indoor Air VOC Analysis |
| Temperature Control Zones | 6 (including column oven, injector, detector, AUX1–AUX3) |
| Column Oven Temp Range | +7 °C to 420 °C |
| Programmable Ramp Stages | 5-step linear heating |
| Oven Cooling | Ambient air-cooled with motorized rear door |
| Carrier Gas Control | External modular gas control boxes (capillary, carrier, fuel/oxidant circuits) |
| Injector Options | Split/splitless, on-column, packed-column vaporizing, WBC for wide-bore capillaries, 6-port gas sampling valve |
| Detector Compatibility | Up to three simultaneous detectors (FID, TCD, FPD, NPD, ECD) |
| Safety Protections | Over-temperature cutoff, TCD over-current protection, system-level hardware watchdog and power-fail recovery |
| Brand | Jiedao/Kexiao |
|---|---|
| Origin | Zhejiang, China |
| Model | GC1690 NMHC |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | Environmental Air & Exhaust Monitoring |
| Detection Principle | Dual-Packed-Column Parallel Separation with Single FID Detection |
| Minimum Detectable Concentration | 0.05 mg/m³ (as methane equivalent) |
| Regulatory Compliance | GB 16297–1996, “Methods for Monitoring and Analysis of Air and Waste Gas” (4th Ed.) |
| Sample Introduction | 10-Port Valving System with 1 mL Gas Loop |
| Column Configuration | Two Parallel Packed Columns (Total Hydrocarbons Column + Methane-Specific Column) |
| Detector | Flame Ionization Detector (FID), 150 °C Operating Temperature |
| Injector Temperature | 150 °C |
| Oven Temperature Range | Ambient to 300 °C (programmable) |
| Carrier Gas | High-Purity Nitrogen or Helium |
| Flow Control | Mechanical or Electronic Pressure/Flow Regulation (model-dependent) |
| Origin | Beijing, China |
|---|---|
| Manufacturer Type | Authorized Distributor |
| Origin Category | Domestic (PRC) |
| Model | SP7800 BTX & Oxygenates GC System |
| Pricing | Available Upon Request |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC_8890 |
| Instrument Type | Online Gas Chromatograph |
| Application Field | Petrochemical Industry |
| Temperature Control Range | Ambient + 8 °C to 420 °C |
| Oven Ramp Rate | 0.1–40 °C/min |
| Oven Cool-down Rate | 450 °C to 50 °C in ~6 min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min (7 programmable segments) |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-2011 (Latest Revision) |
| Instrument Type | Benchtop Laboratory Gas Chromatograph |
| Application Scope | General-Purpose Chemical Analysis |
| Oven Temperature Range | 8–420 °C |
| Oven Ramp Rate | 0.1–30 °C/min (0.1 °C/min increments) |
| Temperature Stability | ±0.1 °C within 200 °C |
| Carrier Gas Flow Control | Electronic Pressure Control (EPC)-compatible design |
| Carrier Gas Pressure Control | Supports precision pressure regulation |
| Cooling Rate | Rapid cool-down to near-ambient (≤200 °C) via intelligent variable-frequency rear door system |
| Column Compatibility | Packed columns and capillary columns (split/splitless injection) |
| Standard Detectors | FID and/or TCD |
| Dimensions | 540 × 455 × 465 mm |
| Weight | 50 kg |
| Power Supply | AC 220 V ±10%, 50 Hz, 2 kW |
| Brand | Jinpu |
|---|---|
| Model | GC-2011 |
| Detector Options | FID or TCD |
| Temperature Control Range | 8 °C above ambient to 420 °C |
| Temperature Stability | ±0.1 °C (≤200 °C), ±0.2 °C (>200 °C) |
| Ramp Rate | 1–30 °C/min |
| Programmed Temperature | 5-step |
| FID Sensitivity | ≤2×10⁻¹¹ g/s (benzene), Linear Range: 10⁷ |
| TCD Sensitivity | ≥2500 mV·mL/mg (benzene) |
| Noise | ≤0.02 mV |
| Drift | ≤0.1 mV/h |
| Column Configuration | Single-column, dual-gas-path |
| Injector | Packed-column vaporizer + dedicated capillary inlet |
| Oven Cooling | Intelligent variable-frequency rear-door cooling to near-ambient |
| Safety Features | Overtemperature cutoff, gas failure/leak detection, power-fail parameter retention, anti-surge protection |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Manufacturer Type | Authorized Distributor |
| Instrument Type | Online Gas Chromatograph |
| Model | GC-2011 |
| Detection Modes | FID & TCD |
| Column Compatibility | Packed Columns & Capillary Columns (Split/Splitless) |
| Oven Temperature Range | Ambient +8°C to 420°C |
| Temperature Control Accuracy | ±0.1°C (≤200°C), ±0.2°C (>200°C) |
| Programmable Ramp Stages | 8-step linear ramp |
| Ramp Rate | 0.1–30.0°C/min (0.1°C increments) |
| Detector Operating Temperatures | Up to 420°C |
| Injector Max Temp | 420°C |
| Carrier Gas Control | Digital pressure/flow regulation (N₂, H₂, He) |
| Power Supply | AC 220 V ±10%, 50 Hz, ~2 kW |
| Dimensions (W×D×H) | 540 × 455 × 465 mm |
| Weight | 50 kg |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Manufacturer Type | Authorized Distributor |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-Purpose / EO Residual Testing |
| Temperature Control Range | 5°C above ambient to 400°C (adjustable in 0.1°C increments) |
| Oven Ramp Rate | 0.1–40°C/min (0.1°C increment) |
| Oven Cool-Down | 450°C → 50°C in ≤6 min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 400°C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Signal Outputs | Dual analog (A/B), RJ45 Ethernet interface |
| Gas Inlets | H₂, Air, Optional H₂ supply lines |
| Power Input | Standard AC mains with integrated fuse and charging circuitry |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-7890 |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Coal Mining Industry |
| Detection Targets | H₂, O₂, N₂, CO, CO₂, CH₄, C₂H₄, C₂H₆, C₂H₂ |
| Minimum Detectable Concentrations | H₂ ≤ 5 ppm |
| TCD Sensitivity | ≥ 2500 mV·mL/mg (n-hexadecane) |
| FID Detection Limit | 5 × 10⁻¹² g/s (n-hexadecane) |
| Column Oven Temp Range | 10–399 °C (1 °C increment) |
| Control Accuracy | ±0.1 °C |
| Detector Temp Range | 10–399 °C |
| TCD Control Accuracy | ±0.01 °C |
| Data Acquisition | 24-bit A/D, USB interface, 6–50 Hz sampling rate, dynamic range 10⁶ |
| Dimensions | 606 × 450 × 450 mm (W×H×D) |
| Weight | 45 kg |
| Power Supply | 220 V, 50 Hz, 2100 W |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-7890 |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-purpose |
| Oven Temperature Range | 5 °C above ambient to 400 °C (adjustable in 0.1 °C increments) |
| Ramp Rate | 0.1–40 °C/min (adjustable in 0.1 °C/min increments) |
| Cool-down Rate | 450 °C to 50 °C in ≤6 minutes |
| Carrier Gas Flow Range & Control | 0–1200 mL/min (7 flow segments) |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 400 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Temperature Stability (≤180 °C) | ±0.01 °C |
| Temperature Accuracy | ≤±0.1 °C |
| Temperature Gradient (100–360 °C, programmed ramp) | ±1 °C |
| Column Configuration | Dual packed columns + one capillary column, or dual capillary columns |
| Detector Capacity | Up to three detectors simultaneously (e.g., dual TCD, dual FID, or TCD+FID), plus optional methanizer |
| Inlet Options | Manual, auto-sampler compatible, valve injection, headspace, thermal desorption, liquid autosampler |
| Power Supply | 220 V ±10%, max. 2200 W |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-7890 (Black Chassis) |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-Purpose |
| Oven Temperature Range | 5°C above ambient to 450°C (adjustable in 0.1°C increments) |
| Oven Ramp Rate | 0.1–40°C/min (adjustable in 0.1°C increments) |
| Oven Cooling Rate | 450°C → 50°C in ≤6 min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 400°C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| FID Detection Limit | ≤3×10⁻¹² g/s (n-hexadecane) |
| FID Noise | ≤3×10⁻¹⁴ A |
| FID Drift | ≤5×10⁻¹³ A/30 min |
| FID Linear Dynamic Range | ≥10⁶ |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-7890 |
| Instrument Type | Benchtop Laboratory Gas Chromatograph |
| Temperature Range | -80 °C to 450 °C (with cryogenic cooling) |
| Column Compatibility | Dual packed columns or dual capillary columns |
| Detector Configurations | Dual TCD, Dual FID, or hybrid TCD/FID |
| Programmed Temperature | 7-step ramp, 6-zone independent temperature control |
| Injector Options | Manual, auto-sampler, headspace, thermal desorption, gas valve |
| Detection Sensitivity (FID) | ≤1×10⁻¹¹ g/s (benzene) |
| (ECD) | ≤1×10⁻¹³ g/mL (γ-HCH) |
| (NPD) | ≤5×10⁻¹² g/s (azobenzene) |
| Temperature Accuracy | ±0.1 °C |
| Heating Rate | 0.1–40 °C/min |
| Power Supply | AC 220 V ±10%, 50 Hz, 2.2 kW |
| Dimensions | 660 × 460 × 465 mm |
| Weight | 60 kg |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-7890SD |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Power Industry Dedicated |
| Oven Temperature Range | Ambient + 5°C to 420°C |
| Oven Ramp Rate | 0.1–40°C/min |
| Oven Cool-down | 450°C to 50°C in ~6 min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min (7-program segments) |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Max Operating Temperature | 450°C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-8890 (New Generation) |
| Instrument Type | Benchtop Laboratory Gas Chromatograph |
| Application Scope | General-Purpose Analysis |
| Oven Temperature Range | 5 °C above ambient to 400 °C (adjustable in 0.1 °C increments) |
| Oven Ramp Rate | 0.1–40 °C/min (0.1 °C/min increments) |
| Oven Cool-Down | 450 °C to 50 °C in ≤6 min |
| Carrier Gas Flow Control | 0–1200 mL/min, 7 programmable segments |
| Carrier Gas Pressure Control | 0–970 kPa |
| Injector Maximum Temperature | 400 °C |
| Injector Pressure Setpoint Range | 0–970 kPa |
| Injector Total Flow Setpoint Range | 0–1200 mL/min |
| Temperature Zones | 6 independently controlled zones |
| Max. Programmed Temperature Ramping | 16 steps |
| Electronic Pressure Control (EPC)/Electronic Flow Control (EFC) | 4-step pressure/flow programming |
| precision | ±0.01 kPa (pressure), ±0.01 sccm (flow) |
| Supported Detectors | FID, TCD, ECD, FPD, NPD |
| Display | 5.7-inch color LCD with dual-channel chromatogram overlay and optional touch interface |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-8890 |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-purpose |
| Temperature Control Range | 4 °C above ambient to 450 °C (1 °C increment |
| Oven Ramp Rate | 0.1–80 °C/min |
| Cooling Rate | 450 °C → 50 °C in ≤6 min |
| Carrier Gas Flow Range & Control | 0–1000 mL/min (4-program segments) |
| Carrier Gas Pressure Range & Control | 0–500 kPa (4-program segments) |
| Injector Max Temp | 450 °C |
| Injector Pressure Setting Range | 0–500 kPa |
| Injector Total Flow Setting Range | 0–1000 mL/min |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-8890PLUS |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Environmental Analysis |
| Temperature Control Range | 5°C above ambient to 450°C |
| Maximum Ramp Rate | 120°C/min |
| Cooling Rate | ≤3.5 min from 350°C to 100°C |
| Carrier Gas Flow Range & Control | 0–300 mL/min |
| Carrier Gas Pressure Range & Control | 0–100 psi |
| Injector Maximum Operating Temperature | 450°C |
| Injector Pressure Setting Range | 0–100 psi |
| Injector Total Flow Setting Range | 0–300 mL/min |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | GC-8890SD |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Power Industry Dedicated |
| Temperature Control Range | Ambient +6°C to 399°C (adjustable in 1°C increments) |
| Ramp Rate | 0–40°C/min (adjustable in 0.1°C increments) |
| Cooling Rate | ≤3 min (from 200°C to 100°C) |
| Carrier Gas Flow Range & Control | 0–1200 mL/min, 7-step programmable |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 450°C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Manufacturer Type | Authorized Distributor |
| Product Origin | Domestic (China) |
| Model | OG-2000V Online Gas Chromatograph for Hydrocarbon Show Evaluation |
| Instrument Type | Online Gas Chromatograph |
| Application Field | General-purpose Geochemical Logging & Rock Pyrolysis Analysis |
| Detector Configuration | Dual Flame Ionization Detectors (FID) |
| Cooling System | Integrated Forced-air Cooling Module |
| Interface | USB 2.0 Digital Output |
| Display | High-contrast Backlit LCD with Chinese Language Support |
| Brand | Jinpu |
|---|---|
| Origin | Shandong, China |
| Model | OG-3000 |
| Instrument Type | Laboratory Gas Chromatograph for Rock Pyrolysis |
| Application Field | Petroleum & Petrochemical Industry |
| Injector Maximum Temperature | 90–850 ± 0.5 °C (programmable) |
| FID Hydrocarbon Detection Limit | 0.001 mg HC/g rock |
| FID Baseline Drift | < 0.1 mV/30 min |
| FID Linear Range | 10⁶ |
| IR Detector Range | 0–100 % |
| IR Linearity Error | ≤ ±1 % F.S. |
| IR Repeatability | < 1 % |
| IR Zero Drift | ≤ ±2 % F.S./24 h |
| Pyrolysis Furnace Temp Control Accuracy | ±0.5 °C |
| Oxidation Furnace Temp Control Accuracy | ±0.5 °C |
| Programmable Ramp Segments | Up to 10 segments |