| Brand | METTLER TOLEDO |
|---|---|
| Origin | Switzerland |
| Model | DSC 1 |
| Temperature Range | –150 °C to 700 °C |
| Sensor Configuration | 56- or 120-pair star-arranged thermocouples |
| Optional Modules | UV-DSC, HP-DSC (up to 100 bar / down to 10 mbar), DSC-Microscopy System |
| Autosampler Capacity | Up to 34 samples |
| Brand | IKA |
|---|---|
| Origin | Germany |
| Model | C200 |
| Measurement Modes | Isothermal, Fast Dynamic, Manual, Time-Controlled |
| Standards Compliance | DIN 51900, ISO 1928, ASTM D240, ASTM D4809, ASTM D5865, ASTM D1989, ASTM D5468, ASTM E711 |
| Max Energy Range | 40,000 J |
| Operating Voltage | 24 V DC (instrument), 100–240 V AC / 50–60 Hz (external power supply) |
| Power Consumption | ≤100 W (calorimeter), ≤120 W (system) |
| Dimensions (W×D×H) | 400 × 400 × 400 mm |
| Weight | 21 kg |
| Protection Class | IP21, Class III |
| Ambient Conditions | 20–25 °C, ≤80 % RH, up to 2,000 m altitude |
| Interfaces | 1× RS-232, 1× Centronics parallel port |
| Software | Calwin C 5040 (real-time monitoring, data export to Excel/Word) |
| Brand | METTLER TOLEDO |
|---|---|
| Origin | Switzerland |
| Model | Flash DSC |
| Temperature Range | –95 to 450 °C |
| Heating/Cooling Rates | Up to 10⁶ K/s (heating), up to 10⁵ K/s (cooling) |
| Sensor Response Time | < 1 ms |
| Sample Mass Range | 1–500 ng |
| Compliance | ISO 11357, ASTM E794, ASTM E1269, USP <1151>, FDA 21 CFR Part 11 (with optional software module) |