Beifen Sanpu
Filter
Showing 271–300 of 450 results
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Model | GC-9860 |
| Communication Interface | 10/100 Mbps Adaptive Ethernet |
| Max Concurrent Instruments | 5,000 |
| Data Import Format | AIA/CDF (ASTM E1947-compliant) |
| Analog Input Range | ±2.5 V |
| Integration Sensitivity | 0.05 µV·s |
| Minimum Resolution | 1 µV |
| Dynamic Range | 10⁷ |
| Linearity | ±0.005% |
| Repeatability | ±0.005% |
| Sampling Rate | Adjustable from 10 to 50 Hz |
| Peak Detection | Intelligent automated peak identification with baseline correction and deconvolution of co-eluting peaks |
| Quantitation Methods | Normalization, corrected normalization, single-point internal/external standard, multi-point internal/external calibration |
| Software Architecture | Multi-threaded, seat-scalable client-server system |
| Remote Data Export Protocols | HTTP/HTTPS, MODBUS TCP, CANopen, GPRS, 3G, private wireless network |
| Compliance | Designed for GLP/GMP environments |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Environmental Monitoring |
| Oven Temperature Range | 5 °C above ambient to 400 °C |
| Temperature Programming | 8-step linear ramping |
| Max. Heating Rate | 0.1–50 °C/min |
| Cooling Rate | ~6 min from 350 °C to 100 °C |
| Carrier Gas Flow Control Range | 0–1200 mL/min |
| Carrier Gas Pressure Control Range | 0–970 kPa |
| Injector Temperature Range | 100–350 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Total Injector Flow Setting Range | 0–1000 mL/min |
| Column Compatibility | Packed columns (3–5 mm OD), capillary columns (standard 6-inch mounting span) |
| Detectors Supported | Up to four simultaneous (FID, TCD, ECD, FPD, NPD) |
| Data Storage | 20 method/data sets onboard |
| Display | 5.7″ LCD (320×240), bilingual Chinese/English interface |
| Power Supply | 220 V ±10%, 2500 W |
| Dimensions (L×H×W) | 520×480×500 mm |
| Net Weight | 50 kg |
| [Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Portable Gas Chromatograph |
| Temperature Control Range | 100–350 °C |
| Heating Rate | 0.1–40 °C/min |
| Cooling Time | ~6 min (from 350 °C to 100 °C) |
| Carrier Gas Flow Range | 0–1200 mL/min |
| Carrier Gas Pressure Range | 0–970 kPa |
| Injector Max Operating Temperature | 100–350 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1000 mL/min] |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | GC-9860-5V |
| Pricing | Upon Request |
| Brand | Beifen Sanpu |
|---|---|
| Model | GC-9860-FID |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Alcoholic Beverage Chromatography |
| Oven Temperature Range | 450 °C |
| Oven Ramp Rate | 0–40 °C/min |
| Oven Cool-down Time | ~8 min |
| Carrier Gas Flow Control Range | 0–1200 mL/min |
| Carrier Gas Pressure Control Range | 0–1000 kPa |
| Injector Max Temp | 450 °C |
| Injector Pressure Setting Range | 0–1000 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Regional Classification | Domestic (China) |
| Model | GC-9860-IV-1 |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Environmental & Food Safety Testing |
| Oven Temperature Range | 450 °C |
| Maximum Ramp Rate | 80 °C/min |
| Maximum Cooling Rate | 40 °C/min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Temperature | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0.4–1200 mL/min |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Model | GC-9860-VI |
| Temperature Control Range | Ambient to 450 °C |
| Oven Ramp Rate | 1–40 °C/min |
| Cool-down Time | ~8 min (from 450 °C to 50 °C) |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Detection Method | Flame Ionization Detection (FID) |
| Sample Introduction | Automated Headspace Sampling (AHS-20A Plus) |
| Compliant Standard | GA/T 842–2019 (China Public Security Industry Standard for Blood Alcohol Testing) |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Food Industry-Specific |
| Temperature Control Range | Up to 450 °C |
| Maximum Ramp Rate | 50 °C/min |
| Cooling Rate | 40 °C/min |
| Carrier Gas Flow Range & Control | 0–1200 sccm |
| Carrier Gas Pressure Range & Control | 0–0.4 MPa |
| Injector Max Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–0.4 MPa |
| Total Injector Flow Setting Range | 0–1200 sccm |
| Key | Brand: Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Pharmaceutical & Cosmetic Quality Control |
| Max Oven Temperature | 450 °C |
| Heating Rate | 0.1–80 °C/min |
| Cooling Rate | 40 °C/min |
| Carrier Gas Flow Range | 0–1200 sccm |
| Carrier Gas Pressure Range | 0–0.4 MPa |
| Injector Max Temp | 450 °C |
| Injector Pressure Set Range | 0–0.4 MPa |
| Injector Total Flow Set Range | 0–1200 sccm |
| Brand | Beifen Sanpu |
|---|---|
| Model | GC-9860 with FID Detector + AHS-610 Headspace Sampler |
| Origin | Beijing, China |
| Manufacturer | Beifen Sanpu Instrument Co., Ltd. |
| Temperature Control Range | 450 °C |
| Oven Ramp Rate | 39 °C/min |
| Oven Cool-down Rate | 35 °C/min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–450 kPa |
| Injector Maximum Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–450 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Key | Brand: Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Environmental Monitoring |
| Temperature Control Range | Ambient to 400 °C |
| Maximum Ramp Rate | 39 °C/min |
| Cooling Rate | 50 °C/min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min (Electronic Flow Control) |
| Carrier Gas Pressure Range & Control | 0–970 kPa (Electronic Pressure Control) |
| Injector Max Temp | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Model | GC-9860HP |
| Temperature Control Range | 100–450 °C |
| Oven Ramp Rate | 0.1–80 °C/min |
| Oven Cool-down Time | ~7 min (from 450 °C to 100 °C) |
| Carrier Gas Flow Range | 0–1000 mL/min |
| Carrier Gas Pressure Range | 0–400 kPa |
| Injector Maximum Temperature | 450 °C |
| Injector Pressure Setting Range | 0–400 kPa |
| Injector Total Flow Setting Range | 0–100 mL/min |
| Detection Method | Flame Ionization Detection (FID) |
| Column Compatibility | Capillary columns (e.g., 30 m × 0.32 mm × 0.5 µm LZP-930) |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Model | GC-9860HP |
| Temperature Control Range | 100–450 °C |
| Oven Ramp Rate | 0.1–39 °C/min |
| Cool-down Time | ~7 min (from 450 °C to 100 °C) |
| Carrier Gas Flow Range | 0–1000 mL/min |
| Carrier Gas Pressure Range | 0–400 kPa |
| Injector Maximum Temperature | 450 °C |
| Injector Pressure Setting Range | 0–400 kPa |
| Injector Total Flow Setting Range | 0–100 mL/min |
| Number of Temperature-Controlled Zones | 6 |
| Max Programmed Temperature Ramps | 8 steps |
| Detection Modes | FID, TCD, ECD, FPD, NPD (configurable up to 3 detectors) |
| Data Interface | IEEE 802.3 Ethernet |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Scope | General-Purpose |
| Oven Temperature Range | Ambient to 400 °C |
| Maximum Ramp Rate | 40 °C/min |
| Maximum Cooling Rate | 40 °C/min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Model | GC-9860II |
| Temperature Control Range | Ambient to 400 °C |
| Oven Ramp Rate | Up to 40 °C/min |
| Oven Cool-down Rate | Up to 40 °C/min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–970 kPa |
| Injector Maximum Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Model | GC-9860II |
| Temperature Control Range | 4–450 °C |
| Oven Ramp Rate | 0.1–40 °C/min |
| Cooling Rate | 450 °C → 50 °C |
| Carrier Gas Pressure Control Range | 0–60 kPa |
| Carrier Gas Flow Control Range | 0–60 kPa |
| Injector Maximum Temperature | 450 °C |
| Injector Pressure Setting Range | 0–60 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Temperature Zones | 6 independent zones |
| Programmed Temperature Ramping | up to 8 steps |
| Detection Options | FID, TCD, ECD, FPD, NPD (up to 3 detectors simultaneously) |
| Interface | IEEE 802.3 Ethernet |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Model | GC-9860Ⅲ |
| Temperature Control Range | 8 °C above ambient to 450 °C |
| Oven Ramp Rate | 0.1–80 °C/min |
| Cool-down Time | ~7 min (450 °C → 50 °C) |
| Carrier Gas Flow Range & Control | 0–500 mL/min |
| Carrier Gas Pressure Control | 8 pressure steps |
| Injector Max Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–100 mL/min |
| Temperature Zones | 6 independent zones |
| Programmable Temperature Ramps | 16 stages |
| EPC/EFC Modes | Constant Flow, Constant Pressure, Split Flow |
| Supported Gases | N₂, H₂, Air, He, Ar |
| EPC/EFC Ramp Stages | 4 |
| Detector Options | FID, TCD, ECD (⁶³Ni), FPD, NPD |
| FID Detection Limit | ≤3×10⁻¹² g/s (n-hexadecane) |
| TCD Sensitivity | ≥4000 mV·mL/mg (n-hexadecane) |
| ECD Detection Limit | ≤1×10⁻¹⁴ g/s |
| Communication Interface | IEEE 802.3 Ethernet |
| [Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Category | Domestic |
| Model | GC-9860III/AHS-610 |
| Temperature Control Range | 450 °C |
| Oven Ramp Rate | 39 °C/min |
| Oven Cool-down Rate | 39 °C/min |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–450 kPa |
| Injector Maximum Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–450 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min] |
| [Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Benchtop Laboratory Gas Chromatograph |
| Model | GC-9860IV |
| Temperature Control Range | Up to 350 °C |
| Maximum Oven Ramp Rate | 22 °C/min |
| Maximum Cooling Rate | 410 °C/min |
| Carrier Gas Flow Range & Control | 0–562 mL/min (EPC) |
| Carrier Gas Pressure Range & Control | 0–485 kPa (EPC) |
| Injector Max Temp | 350 °C |
| Injector Pressure Setting Range | 0–512 kPa |
| Injector Total Flow Setting Range | 0–846 mL/min |
| Detectors Supported | FID, TCD, ECD, FPD, NPD |
| Temperature Zones | 6 independently controlled |
| Programmed Temperature Ramping | 16 steps |
| EPC/EFC Modes | Constant Flow / Constant Pressure |
| Supported Carrier Gases | N₂, H₂, He, Ar, Air |
| Communication Interface | Ethernet (IEEE 802.3) |
| Compliance | Designed for GLP/GMP-aligned workflows |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Laboratory Gas Chromatograph |
| Detection Method | Flame Ionization Detection (FID) |
| Column Configuration | Capillary Column (30 m × 0.25 mm × 0.25 µm) |
| Headspace Autosampler | AHS-20A (20-position) |
| Application Field | Food Industry–Specific Solvent Residue Testing |
| Compliance Standard | GB/T 5009.262–2016 |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Manufacturer |
| Product Category | Domestic |
| Model | GC2010+AHS-610 |
| Instrument Type | Laboratory Gas Chromatograph |
| Application Field | Pharmaceutical Analysis |
| Oven Temperature Range | 400 °C |
| Maximum Ramp Rate | 40 °C/min |
| Cool-down Time | ~10 min (from 400 °C to 50 °C) |
| Carrier Gas Flow Range & Control | 0–1000 mL/min |
| Carrier Gas Pressure Range & Control | 0–1000 kPa |
| Injector Maximum Temperature | 400 °C |
| Injector Pressure Setting Range | 0–1000 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Online Gas Chromatograph |
| Model | GC9860-5U |
| Price Range | USD 4,200 – 7,000 (FOB Beijing) |
| Temperature Control Range | 4 °C to 450 °C |
| Oven Ramp Rate | 0.1–39 °C/min |
| Cool-down Time | ~6 min (450 °C → 50 °C) |
| Carrier Gas Flow Range & Control | 0–1200 mL/min |
| Carrier Gas Pressure Range & Control | 0–0.6 MPa |
| Injector Max Operating Temperature | 450 °C |
| Injector Pressure Setting Range | 0–970 kPa |
| Injector Total Flow Setting Range | 0–1200 mL/min |
| Temperature Zones | 6 independently controlled |
| Programmed Temperature Ramping | Up to 16 steps |
| EPC/EFC Capabilities | Dual-mode (constant pressure / constant flow), supports N₂, H₂, Air, He, Ar |
| Pressure Accuracy | ±0.01 kPa |
| Flow Accuracy | ±0.01 sccm |
| Detector Options | FID, TCD, ECD, NPD (up to 3 simultaneously) |
| Data Interface | Ethernet (IEEE 802.3) |
| Display | 192×64 dot-matrix LCD or optional 7-inch industrial tablet |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Instrument Type | Online Gas Chromatograph |
| Model | GC9860-IV |
| Temperature Control Range | 4 °C to 450 °C |
| Oven Ramp Rate | 0.1–39 °C/min |
| Cooling Rate | ~6 min (450 °C → 50 °C) |
| Carrier Gas Flow Range | 0–1200 mL/min |
| Carrier Gas Pressure Control Accuracy | ±0.05 kPa |
| Injector Maximum Temperature | 450 °C |
| Injector Pressure Setting Range | ±0.05 kPa |
| Total Injector Flow Setting Range | 0–1200 mL/min |
| Temperature Zones | 6 independent zones |
| Temperature Increment | 1 °C |
| Temperature Stability | ±0.1 °C |
| Programmed Temperature Steps | 16 |
| External Events | 8 channels |
| Auxiliary Outputs | 4 channels |
| Supported Inlet Types | Packed column, capillary column, 6-port gas valve, headspace auto-sampler |
| Detector Capacity | Up to 3 simultaneous detectors |
| Compatible Detectors | FID, TCD, ECD, FPD, NPD |
| Gas Control Mode | Full Electronic Pressure Control (EPC) |
| EPC Operating Modes | Constant Flow, Constant Pressure |
| Supported Carrier Gases | N₂, H₂, Air, He, Ar |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | GDYS-101SE2 |
| Application | Field and Laboratory Quantitative Determination of Chlorine Dioxide in Water Samples |
| Light Source | 520 nm LED |
| Optical Stability | ±0.005 Abs/20 min |
| Detection Limit | 0.02 mg/L |
| Measurement Range | 0.00–3.00 mg/L |
| Measurement Time | 1 minute |
| Accuracy | ±5% |
| Power Supply | DC 9 V battery |
| Auto Power-Off | After 30 minutes idle |
| Battery Status Indicator | Integrated low-battery diagnostic circuit |
| Compliance | GB/T 5750.11–2006 (Standard Method for Examination of Disinfectants in Drinking Water) |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | GDYS-101SN |
| Application | Field-deployable photometric analyzer for residual and total chlorine in water |
| Detection Method | DPD colorimetric method per GB/T 5750.11–2006 |
| Detection Time | ≤3 min |
| Detection Range | 0.00–9.99 mg/L Cl₂ |
| LOD | 0.10 mg/L |
| Accuracy | ±5% of reading |
| Power Supply | 9 V alkaline battery (≥40 h continuous operation) |
| Reagent Format | Single-use, pre-packaged DPD reagent kits (50 tests per kit) |
| Included Accessories | Complete set of calibrated colorimetric cuvettes and reagents (50 tests) |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | GDYS-101SN2 |
| Application | Quantitative residual chlorine analysis in water samples |
| Detection Method | DPD colorimetric method per GB/T 5750.11–2006 |
| Detection Time | ≤3 minutes |
| Detection Range | 0.00–5.00 mg/L |
| Lower Detection Limit | 0.05 mg/L |
| Accuracy | ±5% of reading |
| Power Supply | 9 V alkaline battery (operating life: ≥40 hours) |
| Reagent Format | Single-use, pre-packaged DPD reagent kits (50 tests per kit) |
| Accessories Included | Complete set of calibrated colorimetric cuvettes and reagents (50 tests) |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Country of Origin | China |
| Model | GDYS-101SV |
| Application | Quantitative sulfide analysis in drinking water, surface water, domestic wastewater, and treated effluent |
| Detection Method | Colorimetric (methylene blue method per GB/T 5750.5–2006) |
| Detection Time | ≤10 min |
| Detection Limit | 0.02 mg/L |
| Measurement Range | 0.00–1.00 mg/L |
| Accuracy | ±5% |
| Power Supply | 9 V alkaline battery (operational life ≥40 h) |
| Reagent Format | Single-use, pre-packaged assay kits (50 tests per kit) |
| Included Accessories | Complete set of colorimetric glassware and reagents (50 assays) |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | GL-103A |
| Price | USD 650 (approx.) |
| Flow Measurement Principle | Volumetric Displacement (Soap Film Method) |
| Flow Range | 2 mL/min to 2 L/min |
| Accuracy | ±1% of reading |
| Temperature & Pressure Compensation | Yes |
| Display | LCD with real-time P, T, V, Q readouts |
| Timing Resolution | 0.01 s (range: 0.1–600.0 s) |
| Power | Rechargeable 4.8 V / 700 mAh battery (standby >300 h |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | GL-103B |
| Price | USD 685 (approx.) |
| Flow Measurement Principle | Volumetric (Soap Film Method) |
| Flow Range | 5 mL/min – 5 L/min |
| Accuracy | ±1% of reading |
| Display | LCD with real-time temperature, pressure, volume, and flow rate |
| Timing Resolution | 0.01 s (range: 0.1–600.0 s) |
| Power | Rechargeable 4.8 V / 700 mAh battery (standby >300 h |
| External Power Input | 9 V DC (simultaneous operation & charging) |
| Operating Interface | Three-button control with 5-point averaging function |
| Environmental Compensation | Built-in temperature and pressure correction for volumetric-to-mass flow conversion |
| Brand | Beifen Sanpu |
|---|---|
| Origin | Beijing, China |
| Manufacturer Type | Direct Manufacturer |
| Product Origin | Domestic (China) |
| Model | GL-105B |
| Price | USD 820 (FOB) |
| Flow Measurement Principle | Volumetric (Soap Film Method) |
| Flow Range | 30 mL/min – 30 L/min |
| Accuracy | ±1% of reading |
| Display | LCD with real-time temperature, pressure, volume, and flow rate |
| Timing Resolution | 0.01 s (range: 0.1–600.0 s) |
| Power | Rechargeable 4.8 V / 700 mAh battery (standby >300 h |
| Operating Interface | Three-button control with 5-point averaging function |
| Environmental Compensation | Built-in temperature and pressure correction for volumetric-to-mass flow conversion |