- All
- Favorite
- Popular
- Most rated
| Brand | Truelab |
|---|---|
| Origin | Tianjin, China |
| Model | MTN-2800D (12/24/40-position) |
| Heating Method | Dry-block (aluminum block) |
| Temperature Range | Ambient +5°C to 100°C (standard operation), up to 180°C with overtemperature protection |
| Temperature Uniformity | ±0.1°C |
| Temperature Stability | ±0.1°C |
| Sample Tube Diameter Compatibility | 10–15 mm |
| Nitrogen Flow Rate | 0–10 L/min (adjustable per channel) |
| Heating Power | 350 W |
| Block Dimensions | 150 W × 70 H × 95 D mm |
| Block Height | 70 mm |
| Timer Range | 10 min – 99 h 59 min |
| Overtemperature Cutoff | 185°C |
| Display | Quad-digit LED (set temp, actual temp, set time, countdown) |
| Safety Features | Auto cutoff at preset max temp, spill-resistant drip tray, independent gas channel control |
| Compliance | Designed for GLP-compliant sample prep workflows in analytical labs |
| Brand | Truelab |
|---|---|
| Origin | Shanghai, China |
| Model | R-1001 |
| Capacity | 0.9 L |
| Ultimate Vacuum | 1 mbar |
| Rotation Speed | 20–180 rpm |
| Bath Temperature Range | Ambient +5°C to 95°C |
| Temperature Control Accuracy | ±1.5°C (nominal), 0.1°C resolution |
| Condenser Surface Area | 0.126 m² |
| Sealing | Dual PTFE + FKM O-rings |
| Rotating Flask Tilt Angle | 15°–45° |
| Bath Material | 06Cr19Ni10 stainless steel (Φ254 mm × 130 mm) |
| Glassware | DIN ISO 3585 borosilicate 3.3 |
| Power Supply | 220–240 V~, 50/60 Hz |
| Heating Power | 1050 W |
| Total Power Consumption | 1080 W |
| IP Rating | IP20 |
| Max. Altitude | 2000 m |
| Operating Environment | 5–35°C, ≤70% RH |
| Brand | EYELA |
|---|---|
| Origin | Japan |
| Model | N-1200BT/BTF-W/WD |
| Capacity | 1 L (max 3 L) |
| Ultimate Vacuum | 399.9 Pa (≤3 mmHg) |
| Rotation Speed | 5–280 rpm |
| Bath Temperature Range | Water bath: RT+5 to 90°C |
| Oil/water bath | RT+5 to 180°C |
| Temperature Control Accuracy | ±1.5°C (water), ±3°C (oil) |
| Evaporation Rate | Max. 25 mL/min |
| Condenser Surface Area | 0.12 m² |
| Heating Power | 1.05 kW |
| Lift Stroke | 100 mm (motorized) + 130 mm (manual extension) |
| Dimensions (W×D×H) | 585 × 455 × 700 mm |
| Weight | 19.4 kg |
| Condenser Type | Wide-mouth vertical coil (Ø110 × 340 mm, internal Ø91 × 230 mm) |
| Flask Standard Taper | 29/38 (evaporating), S35/20 (receiving) |
| Sealing Material | PTFE vacuum seal |
| Brand | Truelab |
|---|---|
| Origin | Jiangsu, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | E2-TS / E2-TH / E2-TF / E2-TJ |
| Price Range | USD 1,400 – 4,200 (FOB) |
| Purified Water Grade | ASTM Type I (Ultra-Pure), GB/T 6682-2008 Grade I |
| Resistivity | 18.2 MΩ·cm @ 25 °C |
| Heavy Metals & Soluble Silica | <0.1 ppb |
| Total Organic Carbon (TOC) | <3 ppb (E2-TS/E2-TF), <5 ppb (E2-TH), <10 ppb (E2-TJ) |
| Endotoxin | <0.001 EU/mL (E2-TS/E2-TF), <0.002 EU/mL (E2-TH) |
| Microbial Count | <0.1 CFU/mL |
| Particles (>0.2 µm) | <1 CFU/mL |
| RNase | <0.01 ng/mL (E2-TS/E2-TF) |
| DNase | <4 pg/µL (E2-TS/E2-TF) |
| Feed Water Requirement | Municipal tap water, TDS <200 ppm, 5–45 °C, pressure 1.0–4.0 kgf/cm² |
| Flow Rate | 5–40 L/h (adjustable), instantaneous output 1.5–2.0 L/min (with pressurized tank) |
| Dimensions (W×D×H) | 500 × 400 × 580 mm |
| Weight | ~38 kg |
| Power | 220 VAC, 50 Hz, 50–80 W |
| Noise Level | <50 dB(A) |
| Certifications | NSF-certified tubing and fittings, RO membrane: Dow FilmTec™, UV lamp: dual-wavelength (185/254 nm), low-voltage DC control architecture (24 VDC), IPX1-rated enclosure |
| Brand | Truelab |
|---|---|
| Origin | Shanghai, China |
| Model | R-3001 |
| Capacity | 2 L |
| Ultimate Vacuum | 1 mbar |
| Rotation Speed | 10–280 rpm |
| Bath Temperature Range | 20–180 °C |
| Temperature Control Accuracy | ±1 °C |
| Condenser Cooling Area | 0.126 m² |
| Brand | Truelab |
|---|---|
| Origin | Jiangsu, China |
| Model | GREEN-Q2-5T / GREEN-Q2-10T |
| Output Capacity | 5 L/h / 10 L/h |
| Resistivity | 13–17.5 MΩ·cm @ 25 °C |
| Heavy Metals & Soluble Silica | < 0.1 ppb |
| Total Organic Carbon (TOC) | < 10 ppb |
| Microbial Count | < 1 CFU/mL |
| Particles (>0.2 µm) | < 1 particle/mL |
| Feed Water Requirement | Municipal tap water (TDS < 200 ppm, 5–45 °C, pressure 1.0–4.0 kgf/cm²) |
| RO Rejection Rate | >99% for ions, >99% for organics (MW > 200 Da), >99% for particles & bacteria |
| Power Supply | 220 V, 50 Hz, 50–80 W |
| Noise Level | <50 dB(A) |
| Dimensions (W×D×H) | 315 × 440 × 470 mm |
| Weight | ~18 kg |
| Certifications | NSF-certified tubing & fittings, food-grade PP resin cartridges, DOW RO membranes |
| Brand | Truelab |
|---|---|
| Origin | Tianjin, China |
| Model | MTN-2800W (12/24/40-position) |
| Heating Method | Water Bath |
| Temperature Range | RT + 5°C to 100°C |
| Temperature Stability | ±0.1°C |
| Tube Diameter Compatibility | 10–15 mm |
| Nitrogen Flow Rate | 0–10 L/min (adjustable per channel) |
| Heating Power | 350 W |
| Heating Block Height | 70 mm |
| Display | Dual digital displays for set/actual temperature + dual timers (set time & countdown) |
| Over-Temperature Protection | Activated at set temperature + 5°C |
| Timer Range | 10 min – 99 h 59 min |
| External Dimensions (W×H×D, excluding support rod) | 240 × 130 × 215 mm |
| Heating Block Dimensions (W×H×L) | 150 × 70 × 95 mm |
| Certifications | Compliant with GB/T 26572–2011 (Electrical Safety for Laboratory Equipment), RoHS-compliant construction |
| Brand | EYELA |
|---|---|
| Origin | Japan |
| Model | N-51 |
| Capacity | 50 L (evaporation flask), 20 L (recovery flask) |
| Ultimate Vacuum | ≤2.6 hPa (≤2 Torr) |
| Rotation Speed | 10–80 rpm |
| Bath Temperature Range | +5 to 80 °C |
| Temperature Control Accuracy | ±1–2 °C |
| Condenser Type | Vertical quadruple coil |
| Cooling Surface Area | 2.0 m² |
| Bath Dimensions | Ø560 × 300 mm (H) |
| Brand | Millipore |
|---|---|
| Origin | Germany |
| Model | D24 UV |
| Purification Grade | Type I (ASTM D1193-20) |
| Resistivity | 18.2 MΩ·cm @ 25 °C |
| Heavy Metals & Soluble Silica | < 4 pg/mL |
| Total Organic Carbon (TOC) | < 5 ppb |
| Microbial Contamination | < 0.1 CFU/mL |
| Endotoxin | < 0.001 EU/mL |
| Particles (>0.1 µm) | < 0.01 ng/mL |
| Pure Water Flow Rate | 24 L/h |
| Ultrapure Water Flow Rate | 2 L/min |
| Feed Water | Municipal tap water (1–6 bar, 5–35 °C, conductivity < 2000 µS/cm, TOC < 2000 ppb) |
| Brand | Truelab |
|---|---|
| Origin | Germany |
| Model Variants | S2-DS, S2-DH, S2-DF, S2-DJ |
| Feed Water Requirement | Deionized Water (≤3 µS/cm) |
| Product Water Grade | ASTM Type I / ISO 3696 Class 1 / GB/T 6682–2008 Grade I |
| Resistivity | 18.2 MΩ·cm @ 25 °C |
| Heavy Metals & Soluble Silica | <0.1 ppb |
| Total Organic Carbon (TOC) | <3 ppb (S2-DS/S2-DJ), <5 ppb (S2-DH), <10 ppb (S2-DF) |
| Endotoxin | <0.001 EU/mL (S2-DS/S2-DF), <0.002 EU/mL (S2-DH) |
| Bacteria | <0.1 CFU/mL |
| Particles (>0.2 µm) | <1 CFU/mL |
| RNase | <0.01 ng/mL (S2-DS/S2-DF) |
| DNase | <4 pg/µL (S2-DS/S2-DF) |
| Flow Rate | 10–90 L/h (adjustable), Instantaneous Output: 1.5–2.0 L/min |
| Dimensions (W×D×H) | 500 × 400 × 580 mm |
| Weight | ~38 kg |
| Power Supply | 220 V, 50 Hz, 50–80 W |
| Noise Level | <50 dB(A) |
| Conductivity Sensor Cell Constant | 0.01 cm⁻¹, Temp. Sensitivity: ±0.1 °C with Automatic Temperature Compensation |
| Safety | 24 VDC Low-Voltage Primary Power, NSF-Certified Tubing & Fittings, IPX4-Rated ABS Housing, Electrical/Water Separation Architecture |
Show next