| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Import Status | Imported |
| Model | GA-370 |
| Measurement Principle | Dual-Beam Alternating Flow Modulated NDIR |
| Response Time | ≤ 180 s |
| Detection Limit | 10 ppb (for CO, CO₂, CH₄) |
| Gas Compatibility | N₂, O₂, He, Ar, H₂, Air |
| Compliance Notes | Designed for use in ISO/IEC 17025-accredited labs and GMP-aligned gas production facilities |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | GA-370 |
| Measurable Gases | CO, CO₂, CH₄ |
| Background Gases | N₂, O₂, He, Ar, H₂, Air |
| Measurement Principle | Dual-beam Non-Dispersive Infrared (NDIR) with Alternating Flow Modulation |
| Detection Limit (2σ) | 10 ppb |
| Range | 0–1/2/5/10 ppm |
| Repeatability | ≤ ±2% F.S. |
| Linearity | ≤ ±2% F.S. |
| Zero Drift | ≤ ±0.02 ppm/day, ≤ ±0.03 ppm/week |
| Span Drift | ≤ ±2% F.S./day, ≤ ±3% F.S./week |
| Response Time (T90) | ≤ 180 s |
| Sample & Reference Flow Rate | ~3.5 L/min each (at 50–100 kPa) |
| Analog Outputs | Up to two isolated outputs (0–1 V, 0–10 V, 0–16 mA, 4–20 mA, or 0–20 mA) |
| Operating Temperature | 0–40 °C |
| Humidity | ≤ 85% RH (non-condensing) |
| Vibration Tolerance | ≤ 0.29 m/s² at ≤ 100 Hz |
| Dimensions (W×H×D) | 430 × 221 × 555 mm (including protrusions) |
| Weight | ~18 kg |
| Power Supply | 100–240 V AC, ±10%, max. 250 V |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Import Status | Imported |
| Model | GD Profiler 2 |
| Pricing | Available Upon Request |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | H-1 Series |
| Measurement Parameters | pH, ORP, Dissolved Oxygen (DO), Resistivity, Conductivity, Residual Chlorine, Sludge Concentration, Turbidity |
| Compatible Electrodes | HP-100, HP-200, HP-300 pH Sensors |
| Compliance | Designed for continuous industrial process monitoring per ISO 5667-14, ASTM D1293, USP <791>, and EPA Method 150.1 |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | H-519 |
| Instrument Type | Portable Oil Content Analyzer |
| Measurement Principle | Infrared Absorption (at 3.4 µm) |
| Measurement Range | 0–200 mg/L |
| Detection Limit | < 0.2 mg/L |
| Accuracy | ±2% of reading |
| Resolution | 0.1 mg/L |
| Measurement Frequency | 1 test per 2 minutes (0.5 Hz equivalent sampling rate) |
| Compliance Context | Designed for use with OCMA-505H/555H analyzers per ASTM D7678, ISO 9377-2, and EPA Method 1664B workflows |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | 3200044462 |
| Application | Dedicated extraction reagent for OCMA-series oil analyzers |
| Compatibility | Designed for use with SR-305 solvent recovery system |
| Regulatory Context | Compliant with ISO 9377-2 (petroleum hydrocarbons in water) and EPA Method 1664A (n-hexane extractable material) |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Origin | Imported |
| Model | H1034 (Dual-Stage) |
| Grating Configuration | Dual Grating |
| Spectral Resolution | ≤5 nm (with 100 µm slit) |
| Wavelength Range | 200–1200 nm (optional blaze wavelengths: 250 nm, 350 nm, 500 nm) |
| F/# | F/2.8 |
| Stray Light | < 0.1% (typical) |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Origin | Imported |
| Model | H1034 (Single-Stage) |
| Grating Mount | Motorized 3-Grating Turret |
| Spectral Resolution | ≤5 nm (with 100 µm slit) |
| Wavelength Range | 200–1200 nm (optional blaze wavelengths: 250 nm, 350 nm, 500 nm) |
| F-number | F/2.8 |
| Stray Light | < 0.1% (typical) |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Model | H20-IHR320 |
| Dispersion Element | Grating |
| Slit Dimensions (Spectral Bandpass) | 10 µm × 1 mm |
| Spectral Resolution | 0.06 nm |
| Stray Light Level | 1.5 × 10⁻⁴ |
| Wavelength Accuracy | ±0.20 nm |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Manufacturer Type | Manufacturer |
| Import Status | Imported |
| Model | HCLUE |
| Pricing | Upon Request |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Manufacturer Type | Authorized Distributor |
| Import Status | Imported |
| Model | HCLUE |
| Pricing | Available Upon Request |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HD-200 |
| Protection Rating | IP65 |
| Measurement Principle | Clark-type Polarographic Membrane Sensor |
| DO Range | 0–20 mg/L (Resolution: 0.01 mg/L) |
| Saturation Range | 0–200% (Resolution: 1%) |
| Temperature Range | 0–50 °C (Resolution: 0.1 °C) |
| Repeatability | ±0.5% FS (DO), ±0.5 °C (Temp) |
| Linearity | ±0.5% FS (DO), ±0.5 °C (Temp) |
| Output | Three isolated 4–20 mA analog channels (max load: 900 Ω), three SPDT relay outputs, one dry-contact input, RS-485 Modbus RTU interface |
| Power Supply | 90–264 V AC, 50/60 Hz, ≤15 VA |
| Operating Ambient | −20 to +55 °C |
| Enclosure Material | Anodized Aluminum |
| Weight | ~4.5 kg |
| Compliant | CE, FCC, RoHS, WEEE |
| Compatible Probes | DO-1100 (integrated Pt1000 temperature sensor), Electrodes: 5505, 5510 |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HD-300 |
| Measurement Principle | Clark-type Polarographic Membrane Sensor |
| DO Range | 0–20 mg/L (resolution: 0.01 mg/L) |
| Saturation Range | 0–200% (resolution: 1%) |
| Temperature Range | 0–50 °C (resolution: 0.1 °C) |
| Repeatability | ±0.5% FS (DO), ±0.5 °C (temperature) |
| Linearity | ±0.5% FS (DO), ±0.5 °C (temperature) |
| Output | Isolated 4–20 mA DC (max. load: 600 Ω) |
| Digital Interface | RS-485 (2-wire, input/output isolated) |
| Power Supply | 24 V DC (21–32 V DC operating range), max. 0.6 W |
| Enclosure Rating | IP65 |
| Ambient Operating Temperature | −20 to +60 °C |
| Housing Material | Aluminum alloy |
| Mounting | Wall or 50A pipe column |
| Compatible Probes | DO-1100 (integrated Pt1000 temperature sensor) |
| Compliant Standards | CE, FCC, RoHS, WEEE |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HD-480 (Electrochemical Membrane-Type), HD-200FL (Optical Fluorescence-Based) |
| Measurement Principle | Clark-type membrane electrode (HD-480) |
| Output Signal | 4–20 mA, RS-485 (Modbus RTU) |
| Operating Temperature | 0–50 °C |
| IP Rating | IP66/IP67 |
| Power Supply | 12–24 VDC |
| Sensor Cable Length | Standard 5 m (extendable up to 30 m) |
| Compliance | CE, RoHS, ISO 5814:2012 (DO measurement), ASTM D888-22 (dissolved oxygen in water) |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HD-480 |
| Measurement Principle | Galvanic Cell (Electrochemical) |
| DO Range | 0–20 mg/L (resolution: 0.01 mg/L) |
| Saturation Range | 0–200% (resolution: 1%) |
| Temperature Range | 0–40 °C (resolution: 0.1 °C) |
| Repeatability | ±0.5% FS (max) |
| Linearity | ±0.5% FS (max) |
| Output | Isolated 4–20 mA DC (max load: 900 Ω) |
| Relay Outputs | 2 × SPDT (240 V AC / 30 V DC, 3 A) |
| Calibration Options | Air saturation, saturated aqueous solution, salinity (0.0–5.0%), dual-point temperature compensation |
| Sensor Compatibility | DP-1100 probe with built-in Pt1000 thermistor |
| Power Supply | 90–264 V AC, 50/60 Hz |
| Max Power Consumption | 10 W |
| Operating Ambient | −5 to +45 °C |
| Compliance | CE, FCC, RoHS |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Model | HE |
| Instrument Type | Confocal Micro-Raman Spectrometer |
| Excitation Wavelength Options | 532 nm / 633 nm / 785 nm |
| Spectral Range | 200–3200 cm⁻¹ |
| Detector | 1024 × 256 CCD array, MPP mode |
| Cooling | Peltier-cooled to −70 °C (air-cooled, no water required) |
| Read Noise | < 8 e⁻/pixel |
| Dark Current | < 0.01 e⁻/pixel/s |
| Laser Power Range | 80–300+ mW |
| Power Supply | 220 V, 15 A |
| Software | LabSpec 6 (Windows-compatible) |
| Brand | HORIBA |
|---|---|
| Origin | France |
| Manufacturer Type | Manufacturer |
| Import Status | Imported |
| Model | HE |
| Instrument Type | Portable/Handheld Raman Spectrometer |
| Spectral Range | 100–3200 cm⁻¹ |
| Spectral Resolution | 3 cm⁻¹ |
| Minimum Wavenumber | 100 cm⁻¹ |
| Spectral Reproducibility | < 0.25 cm⁻¹ |
| Detector | 1024 × 256-pixel CCD (MPP mode) |
| Cooling | Peltier-cooled to –70 °C (air-cooled, no water required) |
| Read Noise | < 8 e⁻/pixel |
| Dark Current | < 0.01 e⁻/pixel/s |
| Laser Options | 532 nm, 633 nm, or 785 nm (output power 80–300+ mW) |
| Fiber Probe Support | Up to 4 simultaneous SuperHead™ probes |
| Optical Configurations | Standard, confocal, immersion, or microscope-coupled via 40 mm, 10×, or 50× long-working-distance objectives |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HE-200C |
| Enclosure Rating | IP65 |
| Power Supply | 90–264 V AC, 50/60 Hz |
| Output | Three isolated 4–20 mA analog outputs (configurable per parameter) |
| Relay Outputs | Three SPDT dry-contact relays (R1–R3, user-assignable for high/low alarms, USP qualification, or fault indication) |
| Communication | RS-485 (2-wire, galvanically isolated input/output circuitry) |
| Temperature Compensation | Pt1000 sensor (0 °C reference), 0–100 °C range |
| Measurement Parameters | Conductivity (0.000 µS/cm to 200.0 mS/cm, selectable cell constant: 0.01, 0.1, or 1.0 cm⁻¹), TDS (0.00–2000 mg/L, user-defined conversion factor 0.30–1.00), and temperature (0–100 °C, ±0.1 °C accuracy at reference point) |
| Accuracy | Conductivity ±0.5% FS (max), TDS ±1.5% FS (max), temperature ±0.5 °C linearity error |
| Repeatability | Conductivity ±0.5% FS (max), TDS ±1.5% FS (max) |
| Environmental Rating | Outdoor-rated aluminum housing, CE/FCC/WEEE/RoHS compliant |
| Mounting | Pole-mounted (50A pipe) or wall-mounted |
| Weight | ~4.5 kg |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HE-200H |
| Measurement Principle | Four-Electrode Conductivity |
| Range | 0.0–200.0 mS/cm (auto-ranging, dual-scale: 0.00–20.00 mS/cm & 0.0–200.0 mS/cm) |
| Temperature Compensation | 0–100 °C (Pt1000 sensor) |
| Accuracy | ±0.5% FS (0–20.00 mS/cm), ±1.0% FS (20.0–200.0 mS/cm) |
| Repeatability | ±0.1 °C (temperature), linearity ±0.5 °C |
| Enclosure Rating | IP65 |
| Compliance | CE, FCC, RoHS, WEEE |
| Output | Three isolated 4–20 mA analog outputs (max load 900 Ω), three SPDT relay contacts (R1/R2/failure), RS-485 Modbus RTU interface |
| Power Supply | 90–264 V AC, 50/60 Hz, ≤15 W |
| Operating Ambient | −20 to +55 °C |
| Weight | ~4.5 kg |
| Housing Material | Anodized aluminum alloy |
| Mounting | Pole-mounted (50A pipe) or wall-mountable |
| Electrode Constant Options | 0.1 cm⁻¹ or 1.0 cm⁻¹ |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HE-200R |
| Measurement Principle | Two-Electrode Conductivity Method (Inverse Calculation for Resistivity) |
| Resistivity Range | 0.000–2.000 MΩ·cm (with K=0.01/cm electrode), 0.00–20.00 kΩ·cm, 0.0–200.0 kΩ·cm |
| Temperature Range | 0–100 °C |
| Temperature Compensation | Pt1000 sensor (0 °C reference), 0–100 °C range |
| Accuracy (Resistivity) | ±0.01 MΩ·cm max (at reference temperature, standard units) |
| Accuracy (Temperature) | ±0.02 °C max |
| Repeatability (Resistivity) | ±0.1% FS |
| Linearity (Resistivity) | ±0.5% FS |
| Output | Isolated 4–20 mA (2 points), SPDT relay outputs (3 points), dry-contact input (1 point) |
| Communication | RS-485 (2-wire, input/output isolated) |
| Enclosure Rating | IP65 |
| Compliance | CE, FCC, RoHS, WEEE |
| Power Supply | 90–264 V AC, 50/60 Hz, ≤15 VA |
| Operating Ambient | −20 to +55 °C |
| Weight | ~4.5 kg |
| Sensor Compatibility | ERF-series resistivity sensors (K=0.01 cm⁻¹) |
| Calibration | Electrode constant entry, single-point temperature calibration, atmospheric or saturated solution reference modes |
| Special Functions | Ultrapure water resistivity mode, trim adjustment, real-time self-diagnosis (electrode fault, analyzer error) |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HE-300R |
| Enclosure Rating | IP65 |
| Measurement Principle | Two-Electrode Conductivity Method (Inverse Calculation for Resistivity) |
| Resistivity Range | 0.000–2.000 MΩ·cm (with 0.01 cm⁻¹ electrode), 0.00–20.00 kΩ·cm (with 0.1 cm⁻¹ electrode), 0.0–200.0 kΩ·cm (with 1.0 cm⁻¹ electrode) |
| Temperature Range | 0–100 °C |
| Temp. Resolution | 0.01 °C |
| Temp. Accuracy | ±0.02 °C (at reference point, Pt1000 sensor) |
| Resistivity Accuracy | ±0.01 MΩ·cm (max, calibrated per electrode constant) |
| Repeatability | ±0.1% (resistivity), ±0.1 °C (temperature) |
| Linearity | ±0.5% (resistivity), ±0.5 °C (temperature) |
| Output | Isolated 4–20 mA DC (max load 600 Ω) |
| Digital Interface | RS-485 (2-wire, galvanically isolated I/O) |
| Power Supply | 24 V DC (21–32 V DC operating range), ≤0.6 W |
| Operating Ambient | −20 to +60 °C |
| Housing Material | Anodized Aluminum |
| Mounting | Pole-mount (50A pipe) or wall-mount |
| Sensor Compatibility | HORIBA ERF-series resistivity sensors (e.g., ERF-001, electrode constant 0.01 cm⁻¹) |
| Compliance | CE, FCC, RoHS, WEEE |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HE-480C |
| Enclosure | Aluminum, IP65-rated |
| Compliance | CE, FCC, RoHS, WEEE |
| Input Channels | 1-channel conductivity sensor interface |
| Electrode Constants Supported | 0.01 cm⁻¹, 0.1 cm⁻¹, 1.0 cm⁻¹ |
| Conductivity Range (selectable per electrode constant) | 2.000–20.00 µS/cm (0.01 cm⁻¹), 20.00–200.0 µS/cm (0.1 cm⁻¹), 200.0–2000 µS/cm / 0.2000–2.000 mS/cm / 2.000–20.00 mS/cm / 20.00–200.0 mS/cm (1.0 cm⁻¹) |
| TDS Range | 2.00–20.0 mg/L, 20.0–200 mg/L, 200–2000 mg/L (user-defined conversion factor 0.30–1.00) |
| Temperature Range | 0–100 °C (adjustable decimal places: 0/1/2) |
| Repeatability | ±0.5% of full scale (conductivity), ±1.5% FS (TDS) |
| Analog Output | 4–20 mA DC, isolated, max. load 900 Ω |
| Relay Outputs | 2 × SPDT (240 V AC / 30 V DC, 3 A resistive) |
| Temp. Compensation Options | Ultra-pure water mode (ref. 25 °C), user-defined reference temp. (5–95 °C) & temp. coefficient (±5 %/°C), NaCl curve, or none |
| Sensor Compatibility | HORIBA ESH-series and FS-series conductivity sensors |
| Power Supply | 90–264 V AC, 50/60 Hz, ≤10 VA |
| Operating Ambient | −5 to +45 °C |
| Weight | ~4.5 kg |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HE-480H |
| Measurement Principle | Four-Electrode Conductivity |
| Enclosure Rating | IP65 |
| Compliance | CE, FCC, RoHS, WEEE |
| Input | Single-Channel FES-Series Electrode (Cell Constant: 1.0 cm⁻¹) |
| Conductivity Range | 0.00–20.00 mS/cm |
| Temperature Range | 0–100 °C |
| Temp. Compensation | Adjustable Reference Temp (5–95 °C) & Temp. Coefficient (±5 %/°C) |
| Output | 4–20 mA (Isolated, Max Load 900 Ω) |
| Relay Outputs | 2 × SPDT (240 V AC / 30 V DC, 3 A) |
| Repeatability | ±0.5 % FS (±1.0 % at 500 mS/cm or salinity mode) |
| Salinity Conversion | 0.00–4.00 % NaCl / 0.0–20.0 % Salt |
| Operating Ambient | −5 to +45 °C |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HE-480R |
| Sensor Input | Single-channel (electrode constant: 0.01 cm⁻¹) |
| Resistivity Range | 0–0.200 / 0–2.00 / 0–20.0 / 0–100.0 MΩ·cm |
| Alternate Range | 0–2.00 / 0–20.0 / 0–200.0 / 0–1000 kΩ·cm |
| Temperature Range | 0–100 °C (configurable 1- or 2-digit display) |
| Repeatability | ±0.5% (max, equivalent input) |
| Linearity | ±0.5% (max, equivalent input) |
| Output | 1 × isolated 4–20 mA DC (max load 900 Ω) |
| Relay Outputs | 2 × SPDT (240 V AC / 30 V DC, 3 A resistive) |
| Digital Interface | RS-485 (2-wire, input/output isolated) |
| Temp. Compensation | Pt1000 sensor |
| Enclosure Rating | IP65 (outdoor-rated) |
| Compliance | CE, FCC, RoHS, WEEE |
| Compatible Sensors | HORIBA ERF-series resistivity sensors (0.01 cm⁻¹) |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HM-100 |
| Measurement Principle | Anodic Stripping Voltammetry (ASV) with Boron-Doped Diamond (BDD) Electrode |
| Detection Range | Zn: 0–2000 ppb |
| Cu | 0–2000 ppb (Semiconductor-grade Cu: 0–20.0 ppb) |
| As | 0–200 ppb |
| Cd | 0–500 ppb |
| Required Reagents | Acetate buffer solution, Deionized water (for rinsing), 3.3 M KCl solution (for reference electrode electrolyte exchange) |
| Portability | Benchtop with integrated carry handle for field deployment |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HO-200 |
| ORP Range | −2000 to +2000 mV |
| ORP Resolution | 1 mV |
| Temperature Range | 0–100 °C |
| Temperature Resolution | 0.1 °C |
| ORP Repeatability | ±5 mV (max) |
| ORP Linearity | ±5 mV (max) |
| Temperature Accuracy | ±0.3 °C (equivalent input) |
| Output | Two isolated 4–20 mA DC outputs (max load: 900 Ω) |
| Relay Outputs | 3 SPDT (1c) contact points |
| Digital Interface | RS-485 (2-wire, input/output isolated) |
| Enclosure Rating | IP65 |
| Power Supply | 90–264 V AC, 50/60 Hz |
| Power Consumption | ≤15 VA |
| Operating Ambient Temperature | −20 to +55 °C |
| Housing Material | Aluminum alloy |
| Mounting | Wall or 50A pipe column |
| Weight | ~4.5 kg |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HO-200 |
| Type | Industrial Online ORP Meter |
| ORP Range | −2000 to +2000 mV |
| Temperature Range | 0 to 100 °C |
| ORP Resolution | 1 mV |
| Temperature Resolution | 0.1 °C |
| ORP Repeatability | ±5 mV |
| Temperature Repeatability | ±0.3 °C (equivalent input) |
| Linearity | ±5 mV (ORP), ±0.3 °C (temperature) |
| Output | Two isolated 4–20 mA DC outputs (load ≤ 900 Ω) |
| Relay Outputs | Three SPDT dry-contact relays (R1, R2, FAIL) |
| Input | One opto-isolated contact input for external cleaning command |
| Communication | Isolated RS-485 (2-wire) |
| Power Supply | AC 100–240 V ±10%, 50/60 Hz |
| Power Consumption | ≤15 W |
| Enclosure | Aluminum, IP65-rated |
| Operating Ambient Temperature | −20 to +55 °C |
| Mounting | Wall or 50A pole mount |
| Weight | ~4.5 kg |
| Compliance | CE marking, FCC Rules, RoHS, WEEE |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HO-300 |
| ORP Range | −2000 to +2000 mV |
| ORP Resolution | 1 mV |
| ORP Repeatability | ±5 mV (max) |
| ORP Linearity | ±5 mV (max) |
| Temperature Range | 0–100 °C |
| Temperature Resolution | 0.1 °C |
| Temperature Accuracy | ±0.3 °C (equivalent input) |
| Output | 4–20 mA DC, isolated, max load resistance 600 Ω |
| Digital Interface | RS-485 (2-wire, isolated I/O) |
| Power Supply | 24 V DC (21–32 V DC), max 0.6 W |
| Enclosure Rating | IP65 |
| Ambient Operating Temperature | −20 to +60 °C |
| Housing Material | Aluminum alloy |
| Mounting | 50A pipe or wall-mount |
| Weight | ~4 kg |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HO-480 |
| ORP Measurement Range | −2000 to +2000 mV |
| Resolution | 1 mV |
| Repeatability | ±5 mV (max, at equivalent input) |
| Linearity | ±5 mV (max, at equivalent input) |
| Output Signal | 4–20 mA DC (galvanically isolated, max. load resistance: 900 Ω) |
| Relay Outputs | 2 SPDT relays (240 V AC / 30 V DC, 3 A resistive) |
| Calibration | Sensitivity adjustment (−200 to +200 mV), span range adjustable from 50.0% to 150.0% |
| Operating Temperature | −5 to +45 °C |
| Power Supply | 90–264 V AC, 50/60 Hz, ≤10 VA |
| Compliance | CE-marked, FCC-compliant, RoHS-conformant |
| Electrode Type | Lead-free metal reference electrode and ToupH combination electrode |
| Brand | HORIBA |
|---|---|
| Origin | Japan |
| Model | HP-100 |
| Measurement Principle | Glass Electrode Method |
| pH Range | 0–14 |
| pH Repeatability | ±0.05 pH (max) |
| Output Signal | 4–20 mA DC (max load resistance: 500 Ω) |
| Power Supply | 22.5–30 V DC (nominal 24 V DC) |
| Power Consumption | ~0.48 W |
| Temperature Compensation Range | 0–100 °C |
| Ambient Operating Temperature | −10 to +50 °C |
| Temperature Sensor | Pt1000 (at 0 °C) |
| Enclosure Rating | IP65 |
| Compliance | CE, FCC, RoHS, WEEE |