- All
- Favorite
- Popular
- Most rated
| Brand | PULL |
|---|---|
| Origin | Shaanxi, China |
| Model | PLD-11901 |
| Instrument Type | Benchtop Water Quality Analyzer |
| Detection Principle | Gravimetric Filtration (0.45 µm membrane) |
| Filter Membrane | Glass Fiber or Quantitative Filter Paper (φ50 mm or φ60 mm, pore size 0.45 µm) |
| Drying Temperature | 103–105 °C (controlled at 104 ± 1 °C) |
| Weighing Accuracy | ≥ 0.1 mg |
| Vacuum Pressure Range | 20–80 kPa |
| Maximum Vacuum Flow Rate | ≥ 20 L/min |
| Noise Level | ≤ 60 dB(A) |
| Power Supply | AC 220 V ± 10 %, 50 Hz |
| Detection Limit | 2 mg/L |
| Recommended Sample Volume | 100 mL (adjustable for 5–100 mg expected suspended solids mass) |
| Compliance Basis | Aligns with ISO 11923:1997, EPA Method 160.2, and GB/T 11901–1989 |
| Brand | PULUODY |
|---|---|
| Origin | Shaanxi, China |
| Model | PLD-11901 |
| Instrument Type | Portable Suspended Solids Analyzer |
| Detection Principle | Gravimetric Filtration (0.45 µm membrane) |
| Filtration Membrane | Glass fiber or quantitative filter paper (φ50 mm or φ60 mm, pore size 0.45 µm) |
| Drying Temperature | 104 ± 1 °C |
| Weighing Accuracy | ≥ 0.1 mg |
| Vacuum Pressure Range | 20–80 kPa |
| Maximum Vacuum | ≥ 80 kPa |
| Airflow Rate | ≥ 20 L/min |
| Noise Level | ≤ 60 dB(A) |
| Power Supply | AC 220 V ± 10%, 50 Hz |
| Detection Range | 2–1000 mg/L (practical optimal range: 5–100 mg suspended solids per sample) |
| Sample Volume | 100 mL (adjustable based on expected concentration) |
| Compliance Basis | ASTM D5907, ISO 11923, EPA Method 160.2, GB/T 11901–1989 |
| Brand | PULUODY |
|---|---|
| Origin | Shaanxi, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | PSD-351B |
| Instrument Type | Portable Water Quality Analyzer |
| Detection Targets | Surface Water, Groundwater, Domestic Wastewater, Industrial Effluent, Oilfield Injection Water |
| Applicable Standards | SY/T 5329–2012, GB 11901–1989, GB/T 7600, GB/T 11133, GB/T 606 |
| Gas Source | Nitrogen (optional or customer-supplied) |
| Filter Diameter | φ43 mm |
| Sample Volume Capacity | 3500 mL |
| Dimensions (L×W×H) | ≤500×500×800 mm |
| Construction Materials | Organic Glass (PMMA), Stainless Steel, Corrosion-Resistant Components |
| Pressure Regulation System | Integrated Inlet Pressure Control with Relief Valve, Pressure Gauge, Inlet/Outlet Valves |
| Temperature Measurement | Digital Thermometer |
| Timing Method | Precision Stopwatch |
| Filtration Module | Bubble-Elimination Enabled High-Molecular-Weight Membrane Holder |
| Brand | CALDEE |
|---|---|
| Origin | Shaanxi, China |
| Model | PLD-11901 |
| Instrument Type | Benchtop Water Quality Analyzer |
| Detection Principle | Gravimetric Filtration (0.45 µm membrane) |
| Filter Diameter | φ50 mm or φ60 mm |
| Drying Temperature | 104 ± 1 °C |
| Weighing Precision | ≥ 0.1 mg |
| Minimum Detectable Concentration | 2 mg/L |
| Maximum Negative Pressure | ≥ 80 kPa |
| Vacuum Adjustment Range | 20–80 kPa |
| Airflow Rate | ≥ 20 L/min |
| Noise Level | ≤ 60 dB(A) |
| Power Supply | AC 220 V ± 10 %, 50 Hz |
| Brand | PULUODY |
|---|---|
| Origin | Shaanxi, China |
| Manufacturer Type | Authorized Distributor |
| Country of Origin | China |
| Model | PLD-11901 |
| Instrument Type | Benchtop Laboratory Analyzer |
| Filtration Pore Size | 0.45 µm |
| Filter Diameter | φ50 mm or φ60 mm |
| Drying Temperature Control | 104 ± 1 °C |
| Gravimetric Accuracy | ≥0.1 mg |
| Maximum Vacuum Pressure | ≥80 kPa |
| Vacuum Adjustment Range | 20–80 kPa |
| Airflow Rate | ≥20 L/min |
| Acoustic Noise Level | ≤60 dB(A) |
| Power Supply | AC 220 V ±10%, 50 Hz |
| Detection Limit | 2 mg/L |
| Recommended Sample Mass Range | 5–100 mg (corresponding to 100 mL sample volume at typical concentrations) |
| Brand | PULUODY |
|---|---|
| Origin | Shaanxi, China |
| Manufacturer Type | Manufacturer |
| Country of Origin | China |
| Model | PLD-11901 |
| Instrument Type | Laboratory Water Quality Analyzer |
| Detection Parameter | Suspended Solids (SS) in Surface Water, Groundwater, Domestic Wastewater, and Industrial Effluents |
| Filtration Pore Size | 0.45 µm |
| Filter Diameter | φ50 mm or φ60 mm |
| Drying Temperature Range | 103–105 °C |
| Weight Precision | ≥0.1 mg |
| Maximum Vacuum Pressure | ≥80 kPa |
| Vacuum Adjustment Range | 20–80 kPa |
| Airflow Rate | ≥20 L/min |
| Noise Level | ≤60 dB(A) |
| Power Supply | AC 220 V ±10%, 50 Hz |
| Detection Limit | 2 mg/L |
| Gravimetric Accuracy Endpoint | ≤0.4 mg mass difference between consecutive weighings |
| Brand | PULUODY |
|---|---|
| Model | PSD-351B |
| Origin | Shaanxi, China |
| Instrument Type | Laboratory Water Quality Analyzer |
| Application Scope | Surface Water, Groundwater, Domestic Wastewater, Industrial Effluent, Oilfield Injection Water |
| Compliance Standards | SY/T 5329–2012, GB 11901–1989, GB/T 606–2006, GB/T 11133–2015, GB 7600–2014 |
| Gas Source | Nitrogen (externally supplied or optional integrated regulator) |
| Filter Diameter | φ43 mm |
| Housing Materials | Corrosion-resistant stainless steel + optical-grade acrylic |
| Pressure Regulation | Integrated pressure-reducing valve, precision pressure gauge, and dual-stage flow control valves |
| Temperature Monitoring | Digital electronic thermometer (±0.5 °C accuracy) |
| Timing | Manual stopwatch interface with synchronized logging capability |
| Membrane De-airing Function | Integrated venting mechanism to eliminate air bubbles on membrane surface |
| Filtration Mode | Constant-pressure gravimetric filtration |