- All
- Favorite
- Popular
- Most rated
| Brand | Hach |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Portable |
| Measurement Principle | Ion-Selective Electrode (ISE) |
| Sensor Architecture | Integrated CARTRICAL Cartridge with pH, Temperature, and Selective Electrodes |
| Calibration | Factory-Calibrated Cartridge |
| Maintenance Interval | ~2x per year (typical operation) |
| Sample Handling | No pretreatment required |
| Controller Compatibility | SC100 / SC200 / SC1000 Digital Controllers |
| Housing Material | Stainless Steel Body with Ryton End Caps |
| Immersion Rating | Fully Submersible |
| Mounting Options | Rod, Chain, or Wall-Mount Kits |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer | HACH Company |
| Model | Amtax™ Compact |
| Instrument Type | Online Analyzer |
| Measurement Principle | Gas-Phase Separation with Colorimetric Detection (Nesslerization-based) |
| Range Options | 0.2–12.0 mg/L NH₄–N |
| Accuracy | ±2.5% of reading or ±0.2 mg/L (whichever is greater, per certified standard solution) |
| Detection Limit | 0.5 mg/L NH₄–N |
| Cycle Time | 13–120 min (user-selectable) |
| Output Signals | 4–20 mA (max. load 500 Ω) |
| Digital Interfaces | RS485 / RS232 with Modbus RTU or PROFIBUS DP protocol support |
| Alarm Outputs | Two programmable relay outputs (24 V DC, 1 A each) |
| Sample Channels | Dual-inlet capability (two independent sample streams) |
| Brand | Hach |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online Analyzer |
| Measurement Principle | Colorimetric Distillation Method |
| Measurement Range | 0.2–1200 mg/L NH₄–N |
| Accuracy | ≤ ±5% (varies by range |
| Detection Limit | 0.2 mg/L NH₄–N |
| Measurement Interval | User-definable (13 min to 24 h) |
| Batch Capacity | >10 samples per cycle |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Laboratory |
| Measurement Principle | Ion-Selective Electrode (ISE) Method |
| Measurement Range | 0.2–1000 mg/L NH₄–N |
| Accuracy | ±5% of reading or ±0.2 mg/L (with certified standard solutions) |
| Detection Limit | 0.2 mg/L NH₄–N |
| Response Time (T90) | <2 minutes |
| Sample Temperature Range | 0–40°C |
| Sample pH Range | 5–9 |
| Immersion Depth | 0.3–3.0 m |
| Max Sample Pressure | 0.3 bar |
| Sensor Housing Material | 316 Stainless Steel with Ryton® end caps |
| Dimensions | 48 × 361 mm |
| Cable Length | 10 m (standard) |
| Weight | 1.3 kg |
| Operating Certifications | CE compliant |
| Warranty | 2 years |
| Brand | Hach |
|---|---|
| Origin | Imported |
| Manufacturer | Hach Company |
| Model | NH6000sc |
| Instrument Type | Online Analyzer |
| Measurement Principle | Gas-Sensing Electrode (GSE) |
| Measurement Range | 0.02–5.0 mg/L NH₄–N (standard range), extendable to 10–1000 mg/L NH₄–N via range selection |
| Accuracy | ≤1 mg/L: ±0.02 mg/L + 3% of reading |
| >1 mg/L | ±0.02 mg/L + 5% of reading |
| Detection Limit | 0.02 mg/L NH₄–N |
| Response Time | <5 minutes (for concentrations >0.2 mg/L) |
| Measurement Interval | User-selectable (5, 10, 15, 20, or 30 min) |
| Sample Flow Rate | 1.0–20.0 L/h |
| Operating Temperature | –20 to 45 °C |
| Enclosure Rating | IP55 / NEMA 3R |
| Power Supply | 115 VAC or 230 VAC |
| Weight | ~45 kg |
| Dimensions | 575 × 991 × 425 mm |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Instrument Type | Online Analyzer |
| Measurement Principle | Ion-Selective Electrode (ISE) Method |
| Measurement Range | 0–1000 mg/L NH₄–N and 0–1000 mg/L K⁺ |
| Accuracy | ±5% of reading + 0.2 mg/L |
| Measurement Mode | Continuous Real-Time Monitoring |
| Calibration | Built-in Matrix Calibration with Auto-Compensation for Interfering Ions |
| Sensor Architecture | Integrated CARTRICAL™ Sensor Cartridge (K⁺-ISE, pH Reference Electrode, and Temperature Sensor in Single Housing) |
| Controller Compatibility | Digital SC Platform Integration |
| Optional | Automated Cleaning Module |
| Brand | Hach |
|---|---|
| Origin | Imported |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Model | Amtax NA8000 |
| Instrument Type | Online Analyzer |
| Measurement Principle | Salicylate–Indophenol Blue Method |
| Measurement Range | NA8000.01: 0.020–160.0 mg/L (auto-ranging) |
| NA8000.02 | 0.020–1000 mg/L (auto-ranging with dilution) |
| Accuracy | 0.020–15.00 mg/L: ±(0.06 mg/L or 3%) |
| 0.050–30.00 mg/L | ±(0.15 mg/L or 3%) |
| 12.00–160.0 mg/L | ±(0.90 mg/L or 3%) |
| 80.00–1000 mg/L | ±10% |
| Detection Limit | 0.02 mg/L |
| Measurement Cycle | 10–15 minutes |
| Repeatability | ≤2% (low range) to ≤3% (high range) |
| Display Resolution | 0.001 mg/L (≤9.999), 0.01 mg/L (10.00–99.99), 0.1 mg/L (100.0–999.9), 1 mg/L (≥1000) |
| Power Consumption | 100 W |
| Sampling Pressure | 0.04–1 bar (0.6–14.5 psi) |
| Sampling Flow Rate | 100–600 mL/min |
| Sampling Temperature | 5–40 °C |
| Operating Temperature | 5–40 °C |
| Relative Humidity | ≤95% RH (non-condensing) |
| Enclosure Rating | IP55 (indoor use only) |
| Dimensions (L×W×H) | 455 × 365 × 805 mm |
| Weight | ~28 kg (excl. reagents) |
| Power Supply | 100–240 VAC, 50/60 Hz |
| Outputs | 2 × 24 VDC/3 A SPDT relays |
| Data Storage | 20,000 records |
| Event Log Capacity | 20,000 entries (retained for ≥2 years) |
| Calibration | Automatic, manual, or remote-triggered |
| Sample Pretreatment Compatibility | CYQ-series water samplers (e.g., CYQ-106C, CYQ-104C) |
Show next