- All
- Favorite
- Popular
- Most rated
| Brand | HACH |
|---|---|
| Model | APA6000 |
| Measurement Principle | Automated titration pump coupled with colorimetric detection |
| Range (Low) | 0.05–2.0 mg/L Cu²⁺ |
| Range (High) | 1.0–10.0 mg/L Cu²⁺ |
| Accuracy (Low) | ±3% or ±0.03 mg/L (whichever is greater) |
| Accuracy (High) | ±5% or ±0.05 mg/L (whichever is greater) |
| Resolution | 0.001 ppm |
| Repeatability | ±3% of reading or <±0.03 mg/L (whichever is greater) |
| Cycle Time | ~4 minutes per analysis |
| Display | Digital and graphical output calibrated to Cu²⁺ concentration |
| Sample Flow Rate | 20–1000 mL/min |
| Operating Temperature | 5–50 °C |
| Relative Humidity | ≤95% RH, non-condensing |
| Inlet Pressure | 0.17–6.9 bar |
| Calibration Frequency | Configurable from twice daily to once weekly (calibration duration: ~35 min) |
| Analog Outputs | Two isolated 4–20 mA outputs |
| Alarm Outputs | Two SPDT relays |
| Power Supply | 95–240 VAC, 50/60 Hz |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting Options | Panel-mount, benchtop, or wall-mount (hardware included) |
| Data Storage | 30-day onboard history with trend and graphical analysis capability |
| Maintenance Interval | Up to 30 days unattended operation |
| Brand | HACH |
|---|---|
| Model | SP510 |
| Measurement Principle | Colorimetric (Titanium Yellow Method) |
| Range | 0.3–100 mg/L as CaCO₃ |
| Accuracy | ±10% of setpoint (0.3–2 mg/L), ±4% of setpoint (5–100 mg/L) |
| Cycle Time | ~2 minutes per measurement |
| Alarm Outputs | 1 SPDT relay |
| Power Supply | 115/230 VAC, 50/60 Hz |
| Certifications | UL, CSA, CE |
| Reagent Consumption | Up to 2 months per reagent pack |
| Maintenance Interval | Pump tube replacement every 6 months |
| Calibration | Automatic zero and span calibration supported |
| Compliance | Designed for continuous operation in industrial water treatment environments per ISO 9001-manufactured quality system |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | MS8000 Series Online Automatic Water Quality Analyzer |
| Pricing | Upon Request |
| Brand | HACH |
|---|---|
| Origin | USA |
| Manufacturer Type | Original Equipment Manufacturer (OEM) |
| Product Origin Category | Imported |
| Model | HACH Polymetron NA9600 sc |
| Pricing | Upon Request |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | APA6000 |
| Measurement Principle | Flow Injection Analysis (FIA) with Photometric Detection at 600 nm |
| Range | Total Alkalinity: 0–500 mg/L as CaCO₃ |
| Phenolphthalein Alkalinity | 0–250 mg/L as CaCO₃ |
| Accuracy | ±5% of reading or ±1.0 mg/L (whichever is greater) |
| Repeatability | ±3% of reading or ±0.6 mg/L (whichever is greater) |
| Response Time (90%) | <10 min |
| Analysis Cycle | 8 min |
| Auto-Calibration Interval | 60 min |
| Sample Temperature | 5–50 °C |
| Analog Outputs | Two isolated 4–20 mA outputs, fully configurable across full scale |
| Alarm Outputs | Two SPDT relays (230 VAC, 5 A) |
| Power Supply | 95–240 VAC, 50/60 Hz ±2 Hz |
| Enclosure Rating | NEMA 4X (indoor) / IEC 529 IP66 with air-purge interface |
| Drain Connection | Vertical orientation recommended |
| Sample Flow Rate | 100–2000 mL/min |
| Inlet/Outlet Fittings | 1/4" NPT inlet, 3/4" NPT drain |
| Sample Pressure | 0.03–2.04 bar |
| Dimensions (W×H×D) | 522 × 627 × 526 mm |
| Mounting Options | Wall-mounted, benchtop, or panel-mounted |
| Weight | 25.5 kg |
| Data Storage | 30 days of timestamped results with trend/graphing capability |
| Key Features | Fully automated calibration, reagent replenishment, and system cleaning |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | Filtrax |
| Sample Flow Rate | ~900 mL/h |
| Power Supply | 230 VAC ±10%, 50–60 Hz |
| Sample Temperature Range | 5–40 °C |
| Ambient Operating Temperature | −20 to +40 °C |
| Enclosure Rating | IP55 (Outdoor Installation) |
| Control Unit Dimensions | 430 × 530 × 220 mm |
| Filter Housing Dimensions | 92 × 500 × 340 mm |
| Weight | 41 kg |
| Heated Sample Intake Tubing Length | 5 m (standard) |
| Optional Sample Transport Tubing | 2 m (unheated), 10/20/30 m (heated) |
| Filtration Cut-off | 0.15 µm (ultrafiltration membrane) |
| Brand | HACH |
|---|---|
| Origin | USA |
| Manufacturer | HACH Company |
| Product Type | Imported |
| Model | GLI 3400 |
| Pricing | Available Upon Request |
| Electrode Constant Options | 0.05, 0.5, 1.0, 5.0, 10.0 |
| Conductivity Range | 0–200,000 µS/cm (dependent on electrode constant) |
| Resistivity Range | 0.001–20 MΩ·cm |
| Operating Temperature | –20 to 200 °C |
| Max Flow Rate | 3 m/s |
| Pressure Rating | 100 psig at 200 °C |
| Accuracy | ±0.01% of reading |
| Probe Weight | 0.45 kg |
| Brand | HACH |
|---|---|
| Model | APA6000 |
| Measurement Principle | Colorimetric EDTA Complexation (USEPA-Approved Method) |
| Low-Range Range | 0.05–10 mg/L as CaCO₃ |
| High-Range Range | 10–1000 mg/L as CaCO₃ |
| Accuracy (Low) | ±5% or ±0.05 mg/L as CaCO₃, whichever is greater |
| Accuracy (High) | ±5% or ±2 mg/L as CaCO₃, whichever is greater |
| Repeatability (Low) | ±3% or ±0.03 mg/L as CaCO₃ |
| Repeatability (High) | ±5% or ±2 mg/L as CaCO₃ |
| T90 Response Time (Low) | ≤5 min |
| T90 Response Time (High) | ≤17 min |
| Cycle Time (Low) | 4 min |
| Cycle Time (High) | 8.2 min |
| Sample Temp. | 5–50 °C |
| Sample Flow Rate | 100–2000 mL/min |
| Inlet Pressure | 0.03–2.04 bar |
| Drain Fitting | 3/4″ NPT |
| Analog Outputs | Dual 4–20 mA (isolated, configurable for hardness or status) |
| Alarm Outputs | Two SPDT relays |
| Certifications | NRTL Listed (UL/CSA), CE Marked |
| Power Supply | 95–240 VAC, 50/60 Hz ±2 Hz |
| Data Storage | 30-day internal logging with timestamped values, trend plotting, and export via serial interface |
| Brand | HACH |
|---|---|
| Model | GLI E53 / E33 / PRO-E3 |
| Measurement Principle | Inductive (Toroidal) Conductivity |
| Conductivity Range | 0.0–2,000,000 µS/cm |
| TDS Range | 0–9,999 ppm |
| Concentration Range | 0.0–200% w/w |
| Temperature Range | −20.0 to +200.0 °C |
| Accuracy (Conductivity) | ±0.2% FS (E53/E33), ±0.05% FS (PRO-E3) |
| Repeatability | ±0.1% FS (E53/E33), ±0.05% FS (PRO-E3) |
| Output | Dual 4–20 mA (E53/E33), Single 4–20 mA (PRO-E3) |
| Enclosure Rating | NEMA 4X / IP65 (E53/E33), NEMA 4X + Class I, Div 2, Groups A/B/C/D, T5 (PRO-E3) |
| Power Supply | 190–260 VAC (E53/E33), 16–30 VDC (PRO-E3) |
| Installation | Panel-mount, wall-mount, or pipe-mount |
| Compliance | CE, UL 61010-1, ATEX/IECEx (PRO-E3), FDA 21 CFR Part 11-ready software support |
| Brand | HACH |
|---|---|
| Model | UVASsc |
| Measurement Principle | UV Absorption at 254 nm |
| Optical Path Length Options | 1 mm, 2 mm, 5 mm, 50 mm |
| Measuring Range Options | 0.01–60 m⁻¹, 0.1–600 m⁻¹, 0–1500 m⁻¹, 2–3000 m⁻¹ |
| Measurement Cycle | 1 minute |
| Response Time | >1 min (adjustable) |
| Turbidity Compensation | Dual-beam optical system |
| Analog Outputs | Two isolated 0/4–20 mA outputs (max. load 500 Ω) |
| Communication Protocols | MODBUS RTU or PROFIBUS DP (configurable) |
| Enclosure Rating | IP65 |
| Operating Temperature | +2 °C to +40 °C |
| Probe Pressure Rating | Max. 0.5 bar (suitable for submersion up to 2 m depth) |
| Power Supply | 230 VAC ±10%, 50 Hz, 15 VA OR 24 VDC/AC ±25%, 800 mA |
| Probe Dimensions | Ø70 × 333 mm |
| Probe Weight | ~3.6 kg |
| Installation Options | Immersion or flow-through cell |
| Self-cleaning Mechanism | Mechanical wiper integrated |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer | HACH Company |
| Model | U53 |
| Measurement Principle | Ultrasonic Time-of-Flight (Pulse-Echo) |
| Range | 0.25–6.0 m |
| Depth Resolution | ±1 mm |
| Accuracy | ±0.5% of full scale |
| Repeatability | ±0.1% of full scale |
| Sensitivity | ±0.1% of full scale |
| Temperature Compensation Range | –40 to +80 °C |
| Operating Ambient Temperature | –20 to +60 °C |
| Output | Two independent 4–20 mA or 0–20 mA analog outputs |
| Relays | Four SPDT relays |
| Enclosure Rating | NEMA 4X (IP65) |
| Mounting Options | Panel, surface, and pipe mounting |
| Sensor Weight | 1.6 kg |
| Sensor Cable Length | 3.05 m (10 ft) |
| Instrument Weight | 0.5 kg |
| Flow Units | L/s, m³/s, gal/min, ft³/s, ML/d, Mgal/d |
| Depth Units | mm, m, in, ft |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | 3700 Series Electrodeless Conductivity Sensor |
| Connection Interface | Digital (SC100/SC1000 Compatible) |
| Sensor Body Materials | Polypropylene (PP), Polyvinylidene Fluoride (PVDF), Perfluoroalkoxy (PFA), Polyetheretherketone (PEEK) |
| Process Connections | 2" Sanitary Clamp or 2" NPT |
| Standard Cable Length | 6 m (with 1 m extension cable and digital adapter D6120800) |
| Compliance | Designed for continuous online water quality monitoring in compliance with ISO 7888, ASTM D1125, and USP <645> conductivity requirements |
| Brand | HACH |
|---|---|
| Model | pH/ORP Digital Sensor |
| Origin | Imported |
| Manufacturer | HACH Company |
| Sensor Interface | Digital (HACH SC Series Compatible) |
| Cable Options | 4.5 m integrated cable (standard), 10 m cable (6120600/6120500 variants) |
| Temperature Rating | Up to 70°C (PEEK/Ryton models), 60°C (stainless steel models) |
| Electrode Types | Glass pH (standard & HF-resistant), Pt or Au ORP |
| Materials | Ryton®, PVDF, 316SS, PEEK |
| Temperature Compensation | Integrated Pt1000 RTD |
| Compliance | Designed for continuous online water quality monitoring per ISO 5667, ASTM D1293, USP <791>, and EPA Method 150.1 |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Type | Manufacturer |
| Model | 3400 Series Digital Conductivity Sensor |
| Sensor Constants | 0.05, 0.5, 1.0, 5.0, 10 cm⁻¹ |
| Cell Materials | Titanium (D3422), Graphite (D3433), 316 Stainless Steel (D3444/D3455), Sanitary 316SS with Hygienic Flange (D3455) |
| Connection Interface | Digital Bus (HACH Smart Sensor Protocol) |
| Cable Length | 6–7 m standard (with 1 m extension kit included for D3433/D3444/D3455) |
| Compatible Controllers | SC100 and SC1000 Universal Digital Controllers |
| Mounting Types | Insertion (D3422), Submersible (requires analog 3422 analog variant), Sanitary Flanged (D3455) |
| Enclosure Material | PVDF or Ryton per model |
| Electrical Interface | Integrated digital bus connector with auto-ID and calibration memory |
| Brand | HACH |
|---|---|
| Model | OWF-100 |
| Measurement Principle | UV Fluorescence |
| Range | 0–2.0 mg/L and 0–10 mg/L (selectable) |
| Repeatability | ±7% FS |
| Stability | ±5% FS/day (in deionized water) |
| Response Time | T90 = 3 min |
| Sample Temperature | 0–30 °C |
| Sample Pressure | 0.02–0.3 MPa |
| Sample Flow Rate | 2–7 L/min |
| Ambient Operating Temperature | 0–45 °C |
| Analog Output | 4–20 mA, max. load 600 Ω |
| Power Supply | 220 VAC ±10%, 50/60 Hz |
| Power Consumption | ~50 VA |
| Dimensions (W×H×D) | 300 × 1500 × 500 mm |
| Weight | 40 kg |
| Installation | Indoor only |
| Optical Design | Windowless full-flow measurement cell |
| Maintenance | No moving parts, reagent-free operation |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer Status | Manufacturer |
| Measurement Principle | UV Absorption (Dual-Beam, 210 nm) |
| Optical Path Lengths | 1 mm, 2 mm, 5 mm |
| NO₃–N Range | 0.1–100.0 mg/L (1 mm), 0.1–50.0 mg/L (2 mm), 0.1–25.0 mg/L (5 mm) |
| LOD (NO₃–N) | 0.1 mg/L (typical) |
| Accuracy | ±3% ±0.5 mg/L (1 mm), ±5% ±1.0 mg/L (2 mm), ±5% ±0.5 mg/L (5 mm) |
| Resolution | 0.1 mg/L or 0.5 mg/L depending on configuration |
| Sludge Compensation | Yes |
| Measurement Interval | 1 min or 5 min |
| Response Time (T₁₀₀) | ≤1 min (1 mm path), ≤15 min (2 mm path) |
| Installation Options | Immersion & Flow-Through Cell |
| Housing Material | Stainless Steel (316L), Dual-Sealed or Single-Sealed variants |
| Calibration | Standard Solution-Based |
| Maintenance Interval | Every 6 months |
| Operating Temperature | 2–40 °C |
| Max Submersion Depth | 2 m (0.5 bar) |
| Probe Cable Length | 10 m (extendable) |
| Dimensions (L × D) | 33.3 × 7.0 cm, 32.7 × 7.0 cm, or 32.3 × 7.5 cm |
| Self-Cleaning Function | Integrated ultrasonic or mechanical wiper (model-dependent) |
| Brand | HACH |
|---|---|
| Model | TC2300 |
| Type | Online Optical Particle Counter |
| Origin | Imported (USA-manufactured) |
| Manufacturer | HACH Company, a Danaher Corporation business |
| Particle Detection Principle | Light Obscuration (LO) / Single-Particle Optical Sizing (SPOS) |
| Measurement Channels | 8 user-selectable size thresholds |
| Flow Control | Integrated precision peristaltic sampling pump with adjustable flow rate (typically 20–60 mL/min) |
| Display | 7-inch full-color capacitive touchscreen interface |
| Compliance | Designed to support ASTM D6978-05 (for clean water), ISO 21501-4, and USP <788> & <789> particulate matter testing guidelines (when configured per application requirements) |
| Data Output | Modbus RTU/TCP, 4–20 mA analog outputs, optional Ethernet/IP or OPC UA integration |
| Enclosure Rating | IP65-rated NEMA 4X housing for industrial environments |
| Brand | HACH |
|---|---|
| Model | Txpro-2 |
| Measurement Principle | Quadruple-beam Optical Transmission/Scattering |
| Measurement Range (RD240/242) | 0–10 g/L |
| Measurement Range (RD260/262) | 0–25 g/L |
| Accuracy (RD240/242) | ±2% of reading or ±0.1 g/L, whichever is greater |
| Accuracy (RD260/262) | ±2% of reading or ±0.2 g/L, whichever is greater |
| Repeatability (RD240/242) | ±2% of reading or ±0.05 g/L, whichever is greater |
| Repeatability (RD260/262) | ±2% of reading or ±0.1 g/L, whichever is greater |
| Operating Temperature | –20 °C to +55 °C |
| Analog Outputs | Two isolated 0/4–20 mA outputs, max. load 800 Ω |
| Relays | Three SPDT relays, 2 A @ 250 VAC |
| Signal Averaging | Adjustable from 1 second to 5 minutes |
| Power Supply | 90–265 VAC, 50/60 Hz |
| Max. Power Consumption | 25 VA |
| Enclosure Rating | NEMA 4X (IP66) |
| Mounting | Pipe clamp for up to 50 mm OD pipe |
| Compliance | CE, UL 61010-1, CSA C22.2 No. 61010-1 |
| Dimensions | 144 × 144 × 150 mm |
| Brand | HACH |
|---|---|
| Origin | Imported |
| Manufacturer | HACH Company |
| Model | AMTAX sc |
| Measurement Principle | Gas-Sensing Electrode (Ammonia-Selective Electrode) |
| Measurement Ranges (as NH₄-N) | Low Range: 0.05–20 mg/L |
| Medium Range | 1–100 mg/L |
| High Range | 10–1000 mg/L |
| LOD | 0.05 mg/L (low), 1 mg/L (medium), 10 mg/L (high) |
| Accuracy | ±3% ±0.05 mg/L (low), ±3% ±1.0 mg/L (medium), ±4.5% ±10 mg/L (high) |
| Repeatability | ±2% ±0.05 mg/L (low), ±2% ±1.0 mg/L (medium), ±2% ±10 mg/L (high) |
| Response Time (T₉₀) | ≤5 min |
| Measurement Interval | Adjustable from 5 to 120 min |
| Sample Temp. | 4–40 °C |
| Sample pH | 5–9 |
| Operating Temp. | −20–45 °C |
| Relative Humidity | ≤95% non-condensing |
| Power Supply | 230 VAC/50 Hz or 115 VAC/50–60 Hz |
| Power Consumption | 100 W nominal (peak: 1000 VA) |
| Enclosure Rating | IP66 |
| Sample Flow Rate | ≤3 m/s |
| Filtrate Flow | ≥5 mL/min for ≥4 min per 5-min cycle |
| External Cabinet Dimensions | 540 × 720 × 390 mm |
| Analyzer Weight | 35 kg (incl. reagents) |
| Cable Length | 2 m from enclosure |
| Optional Heated Sampling Hose | 5 m or 10 m |
Show next