| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | DTA L61 |
| Instrument Type | Differential Thermal Analyzer (DTA) |
| Temperature Range | −150 °C to 2400 °C (configurable: −150–500 °C, −150–1000 °C, RT–1400 °C, RT–1500 °C, RT–1600 °C, RT–1750 °C, RT–2000 °C, RT–2400 °C) |
| Heating/Cooling Rate | 0.001–50 K/min |
| Vacuum Capability | down to 1×10⁻⁵ mbar |
| Sensor Types | E-, K-, S-, B-, and C-type thermocouples |
| Measurement Principle | Heat-Flux DTA |
| Atmosphere Control | Oxidizing, reducing, inert (static/dynamic) |
| Optional Coupling | Mass Spectrometry (MS), Fourier Transform Infrared Spectroscopy (FTIR) |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | TMA L71 (TMA PT1000) |
| Temperature Range | –260 to 1000 °C |
| Force Range | up to 1 / 5.7 / 20 N |
| Frequency Range | 1 Hz |
| Resolution | 0.125 nm |
| Atmosphere Options | Oxidizing, Reducing, Inert, Static or Dynamic |
| Sample Dimensions | up to 30 mm × 30 mm |
| Interface | USB |
| Optional Modules | DTA Analysis Software, Rate-Controlled Sintering (RCS) Software |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | STA/TGA L86 |
| Instrument Type | Universal Simultaneous Thermal Analyzer |
| Temperature Range | −150 °C to 2400 °C |
| Balance Sensitivity | 0.01 µg |
| Maximum Sample Capacity | 35 g / ±20000 mg |
| Vacuum Capability | 10⁻² mbar (pump-dependent) |
| Pressure Range (optional) | up to 5 bar |
| Heating Rate | 0.01–100 K/min |
| TG Resolution Options | 0.01 µg / 0.02 µg / 0.1 µg |
| DSC Sensor Types | E, K, S, B, C (C = DTA-only) |
| DSC Resolution | 0.3–1.2 µW |
| DTA Resolution | 0.03 nV |
| DTA Sensitivity | 1.5 µV/mW |
| DTA Range | ±250 µV / ±2500 µV |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | TEG-Tester (TEG L34) |
| Temperature Range | Ambient to 300 °C (hot side) / −20 °C to 300 °C (cold side) |
| Sample Diameter | Ø20–60 mm (circular) or 20×20–40×40 mm (rectangular) |
| Thickness Capacity | Up to 25 mm |
| Clamp Force | 2–5 kN (motor-driven) |
| Current Range | 0–3 A DC |
| Voltage Range | 0–12 V DC |
| Voltage Resolution | 1.6 µV |
| Current Resolution | 1 µA |
| Voltage Accuracy | ±0.3 % |
| Current Accuracy | ±0.3 % |
| Temperature Accuracy | ±0.1 °C |
| Thermocouple Type | E-type |
| Reference Block Materials | Aluminum, Brass, Copper (custom options available) |
| Heating Power | 1.0 kW |
| Cooling Capacity | 1.0 kW at 10 °C / 0.5 kW at −20 °C |
| Pump Flow Rate | 27 L/min at 0.7 bar |
| Reservoir Volume | 3.8–7.5 L |
| Refrigerant | R449A |
| Sampling Interval | <10 ms |
| Operating Modes | CC, CV, FOC, MPPT, P&O |
| Heat Dissipation Capacity | 36 W |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | STA L87 MSB |
| Temperature Range | –196 to 2400 °C |
| Max. Sample Mass | 10 g (standard balance) |
| Resolution | 1 µg |
| Pressure Range | UHV to 150 bar (metal version) |
| Measurement Principle | Magnetic suspension coupling with PID-controlled electromagnetic levitation |
| Compliance Options | ASTM E1131, ISO 11358, USP <731>, GLP/GMP-ready data logging |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | STA L82 |
| Temperature Range | RT to 1000 °C |
| Heating Rate | 0.01–100 K/min |
| Temperature Stability | ±0.001 °C |
| Maximum Sample Mass | 25 mg / 2500 mg |
| TG Resolution | 0.1 µg |
| DSC Sensitivity | ~4–17.6 µW |
| DTA Resolution | 0.03 nV |
| Vacuum Capability | 10⁻² mbar (≤800 °C) |
| Atmosphere Control | Inert, oxidative, reducing, humidified, or vacuum |
| Optional Autosampler | 42- or 90-position |
| Sensor Types | E/K/S thermocouples |
| Measurement Modes | TG, TG-DTA, TG-DSC, TG-DSC (Cp) |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Instrument Type | Differential Scanning Calorimeter |
| Model | DSC L63 |
| Temperature Range | −170 °C to 750 °C |
| Heating/Cooling Rate | 0.01–100 K/min |
| Temperature Accuracy | ±0.1 K |
| DSC Signal Range | ±750 mW |
| Sample Capacity | Up to 96 positions (with optional autosampler) |
| Data Acquisition Rate | 100 Hz |
| Enthalpy Accuracy | <1% (verified with In, Zn standards) |
| Sensor Type | Ceramic/metal hybrid high-resolution sensor |
| Gas Control | Optional MFC (3 gases) or integrated MFC |
| 炉 Body & Crucible Interchangeability | Yes |
| Compliance | ASTM E794, ISO 11357, USP <1163>, GLP-ready |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | LZT-Meter (LZT L33) |
| Thermal Conductivity Measurement Method | Laser Flash Analysis (LFA) |
| Thermal Diffusivity Range | 0.01 – 1000 mm²/s |
| Laser Source | Nd:YAG (25 J/pulse) |
| Pulse Width | 0.01 – 5 ms |
| Temperature Range | –100 °C to 1100 °C |
| Atmosphere Options | Inert, Reducing, Oxidizing, Vacuum |
| Seebeck Coefficient Range | 1 µV/K – 250 mV/K (Static DC Method) |
| Electrical Conductivity Range | 0.01 – 2×10⁵ S/cm |
| Sample Geometry | Disc-shaped (10, 12.7, 25.4 mm diameter) or Cylindrical/Rectangular (2–5 mm × 6–23 mm) |
| Detection | InSb / MCT Detectors |
| Furnace Options | IR Furnace (RT–800 °C / 1100 °C), Cryo Furnace (–100 °C to 500 °C), High-Temp Furnace (RT–1100 °C) |
| Measurement Principles | Laser Flash Analysis (ASTM E1461), Static DC Seebeck & Four-Terminal Resistivity, Harman Method (Direct ZT), AC Impedance Spectroscopy (TEG Modules) |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | TGA HP L85 |
| Temperature Range | RT to 1200 °C |
| Pressure Range | up to 150 bar |
| Balance Sensitivity | 0.1 µg |
| Maximum Sample Mass | 5 g |
| Vacuum Level | 10⁻⁴ mbar |
| Heating Rate | up to 300 K/min |
| Cooling Rate | up to 150 K/min |
| TG Sensor Types | E, K, S, B, C |
| TG-DSC Sensor Types | E, K, S, B, C |
| Interface | USB or Ethernet |
| Gas Configuration | 1–3 configurable gas channels (customizable) |
| Optional Accessories | Steam Generator, H₂/HC Auto-Vent & Gas Safety System |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | TGA L83 |
| Temperature Range | RT to 1100 °C |
| Heating/Cooling Rate | 0.001–250 °C/min |
| Maximum Sample Mass | 5 g |
| Cooling Time | <12 min (1100 °C → 100 °C) |
| Vacuum Capability | up to 10⁻³ mbar |
| Atmosphere Control | Inert, oxidative, reductive, vacuum |
| Gas Dosing | Integrated MFC module with 3 gas channels (1 purge + 2 reactive) |
| Sample Autosampler | Up to 90 positions |
| Crucible Materials | Pt, Al₂O₃, Au, Ag, Al, graphite, W, stainless steel (HP) |
| EGA Compatibility | FTIR, MS, GC-MS |
| Interface | USB |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | DIL L76 PT |
| Temperature Range | RT to 1600 °C |
| Measurement Range | ±2500 µm |
| Delta L Resolution | 0.05 nm |
| Contact Force | Manually adjustable up to 3 N |
| Sensor Type | High-precision LVDT (Linear Variable Differential Transformer) |
| Furnace Options | Quartz (up to 1100 °C), Al₂O₃ (up to 1600 °C) |
| Software | LINSEIS Thermal Analysis Suite with CTE calculation, softening point detection, and automated cooling trigger |
| Compliance | Designed for GLP-compliant thermal property characterization in industrial QC and R&D labs |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | STA HP L85 |
| Temperature Range | 170 °C to 1200 °C (optional 1600 °C / 1800 °C) |
| Temperature Stability | ±0.05 °C |
| Max Sample Mass | 35,000 mg |
| Operating Pressure | up to 150 bar |
| TGA Resolution | 0.1 μg |
| DSC Sensitivity | approx. 4–22.5 μW |
| DSC Resolution | 0.3–1 μW |
| DTA Resolution | 0.03 nV |
| Vacuum Level | ≤10⁻⁴ mbar (pump-dependent) |
| Sensor Types | E / K / S / B / C |
| Gas Control | Quantitative mass-flow-controlled inlet systems |
| Optional Accessories | Steam generator, top-loading modular sensor kits (TG-only or TG-DSC) |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | DIL L78 RITA |
| Temperature Range | −150 °C to 1600 °C |
| Heating/Cooling Rate | up to 2500 °C/s (induction furnace, hollow cylindrical sample) |
| Sample Geometry | Hollow cylinder, OD 3.5 mm / ID 3 mm / L 10 mm |
| Measurement Range | ±1.2 mm |
| Resolution | 5 nm |
| Data Acquisition Rate | up to 1 kHz |
| Compliance | ASTM A1033 |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | DIL L75 HP |
| Temperature Range | RT to 1100 °C / RT to 1400 °C / RT to 1800 °C |
| Maximum Pressure | 100 bar / 150 bar (configurable) |
| Vacuum Level | 1×10⁻⁴ mbar |
| Sample Length | up to 50 mm |
| Sample Diameter | φ7 / φ12 / φ20 mm |
| Measurement Range | 500 µm / 5000 µm |
| Resolution | 0.125 nm |
| Atmosphere Compatibility | Inert, Oxidizing, Reducing, Vacuum |
| Optional Accessories | Steam Generator, Mass Flow Controlled Gas Mixing System (MFC), QMS/FTIR Coupling Interface |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | LSR-1 (LSR L32) |
| Temperature Range (Base Unit) | RT to 200 °C |
| Extended Low-Temp Range (with LN₂ Accessory) | –160 °C to 200 °C |
| Extended High-Temp Resistivity Range (with HT Probe) | up to 600 °C |
| Seebeck Coefficient Range | 0–2.5 mV/K |
| Seebeck Voltage Range | ±8 mV |
| Electrical Resistance Range | 10 nΩ |
| Voltage Sensitivity | 0.5 nV/K |
| Temperature Accuracy | ±1.5 °C or 0.0040· |
| Heating Rate | 0.01–100 K/min |
| Atmosphere Options | Inert, Reducing, Oxidizing, Vacuum |
| Recommended Purge Gas | Low-pressure He |
| Sample Dimensions (Seebeck) | 8–25 mm (L) × 2–25 mm (W) × ≤2 mm (T) |
| Sample Dimensions (Resistivity) | 18–25 mm (L) × 18–25 mm (W) × ≤2 mm (T) |
| Measurement Principles | Van der Pauw method (resistivity), Static DC method & Slope-based Seebeck measurement |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | Chip-DSC 1 (Chip-DSC L66 Basic) |
| Instrument Type | DSC |
| Temperature Range | RT to 450 °C (no active cooling unit) |
| Heating/Cooling Rate | 0.001–100 K/min |
| Temperature Accuracy | ±0.2 K |
| DSC Signal Range | ±2.5–±1000 mW |
| Temperature Repeatability | ±0.02 K |
| Heat Flow Resolution | 0.03 µW |
| Atmosphere Options | Inert or oxidative (static/dynamic) |
| Calibration | Supplied with certified reference materials |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | LSR-3 (LSR L31) |
| Temperature Range | –100 °C to 1500 °C (configurable via interchangeable furnace modules) |
| Measurement Methods | Static DC Seebeck method / Slope method (Seebeck) |
| Seebeck Range | 1 µV/K to 5000 µV/K |
| Electrical Conductivity Range | 0.01 to 2 × 10⁵ S/cm |
| Furnace Options | IR furnace (RT–800 °C / RT–1100 °C), resistance furnace (RT–1500 °C), cryo furnace (–100 °C to 500 °C) |
| Atmosphere Compatibility | Inert, reducing, oxidizing, vacuum (recommended: low-pressure He) |
| Electrode Materials | Ni (–100 °C to 500 °C), Pt (–100 °C to 1500 °C) |
| Sample Geometry Support | Bulk rods (2–5 mm² cross-section, ≤23 mm length, ≤6 mm diameter), discs (10/12.7/25.4 mm Ø), thin films (with optional film adapter) |
| Thermocouple Types | K, S, C |
| Inter-thermocouple Spacing | 4 mm, 6 mm, 8 mm |
| Mandatory Cooling | Integrated water-cooling system |
| Optional Modules | Hammar Method ZT module (LSR-4), Impedance Spectroscopy for TEGs, Low-Temperature Cryo Kit (LN₂-cooled), High-Resolution Camera-Assisted Resistance Imaging, High-Resistance Adapter |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | TGA L81 |
| Temperature Range | –150 to 2400 °C |
| Balance Sensitivity | 0.1 µg |
| Maximum Sample Mass | 25 g |
| Vacuum Tightness | 10⁻² mbar |
| Optional DTA Calculation | Yes |
| Interface | USB |
| Compatibility | MS and/or FTIR for Evolved Gas Analysis (EGA) |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | DSC L62 |
| Instrument Type | DSC |
| Temperature Range | −150 °C to 1750 °C |
| Heating/Cooling Rate | 0.001–50 K/min |
| DSC Signal Range | ±2.5 mW to ±250 mW |
| Sensor Types | DSC, DTA, DSC–Cp |
| Thermocouple Options | E, K, S, B |
| Vacuum Capability | down to 10⁻⁵ mbar |
| Atmosphere Control | inert, oxidizing, reducing (static/dynamic) |
| Modulated Temperature Function | Yes |
| Interface | USB |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | TMA L72 (TMA PT1600) |
| Temperature Range | RT to 1600 °C |
| Force Range | up to 1 N or 5.7 N |
| Frequency Range | 1 Hz or 5 Hz |
| Resolution | 0.125 nm |
| Atmosphere Options | Oxidizing, Reducing, Inert, Static/Dynamic |
| Sample Max Height | 30 mm |
| Interface | USB |
| Optional Modules | DTA Analysis Software, Rate-Controlled Sintering (RCS) Software |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | LFA L52 |
| Measurement Principle | Laser Flash Analysis (LFA) |
| Thermal Conductivity Range | 0.1–4000 W/(m·K) |
| Thermal Diffusivity Range | 0.01–2000 mm²/s |
| Temperature Range | −125 °C to 2800 °C |
| Sample Forms | Solid, Powder, Liquid |
| Sample Geometry | Circular (Ø 6, 10, 12.7, 25.4 mm), Square (10×10 mm, 20×20 mm) |
| Sample Thickness | 0.1–6 mm |
| Pulse Source | User-replaceable Nd:YAG laser |
| Detection | Non-contact IR detection (InSb or MCT detector) |
| Data Acquisition Rate | 2 MHz |
| Atmosphere | Inert or reducing |
| Auto-sampler Options | 3-, 6-, or 18-position |
| Furnace Options | Multiple modular furnaces (e.g., cryo, graphite, SiC, high-temp metal) |
| Heating Rate | 0.01–100 °C/min (furnace-dependent) |
| Compliance | ASTM E1461, ISO 22007-4, DIN 30905 |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | HCS L36 |
| Instrument Type | Ambient-Temperature Hall Effect Tester |
| Constant Current Source Range | DC: 1 nA–125 mA (8 decades, ±12 V compatible) |
| AC | 16 µA–20 mA, input impedance > 100 GΩ, frequency range 1 mHz–100 kHz |
| Voltage Measurement | DC low-noise/low-offset: 1 µV–2500 mV, 4 gain ranges, digital resolution 300 pV |
| AC | 20 nV–1 V, adjustable integration time and gain |
| Mobility Range | 10⁻³–10⁷ cm²·V⁻¹·s⁻¹ |
| Resistivity Range | 10⁻⁴–10⁷ Ω·cm |
| Carrier Concentration Range | 10⁷–10²² cm⁻³ |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | TIM-Tester (TIM L58) |
| Measurement Principle | Laser Flash Method (ASTM D5470–17 compliant) |
| Sample Capacity | Single sample per test |
| Test Environment | Ambient temperature operation |
| Accuracy | ±0.1% at 50% stroke, ±0.25% at 100% stroke |
| Thermal Conductivity Range | 0.1–50 W/(m·K) (extendable upon configuration) |
| Sample Forms | Solid, powder, paste, liquid, adhesive, foil |
| Sample Dimensions | Circular Ø20/25/40 mm |
| Thermal Resistance Range | 0.01–8 K/W |
| Temperature Range | −30 °C to 300 °C (configurable) |
| Contact Pressure | 0–16 MPa (dependent on sample area) |
| Cooling System | External water-cooling unit |
| Brand | LINSEIS |
|---|---|
| Origin | Germany |
| Model | STA L81 Nuclear |
| Temperature Range | −150 °C to 500 / 700 / 1000 °C |
| Heating Rate | 0.01–100 K/min |
| Temperature Stability | 0.001 °C |
| Maximum Sample Mass | ±35,000 mg |
| Vacuum Capability | 10⁻² mbar (pump-dependent) |
| Pressure Range (optional) | up to 5 bar |
| TG Resolution | 0.01 / 0.02 / 0.1 µg |
| DSC Resolution | 0.3 / 0.4 / 1 / 1.2 µW |
| DTA Resolution | 0.03 nV |
| DTA Sensitivity | 1.5 µV/mW |
| DTA Range | 250 / 2500 µV |
| Sensor Types | E / K / S / B / C (C = DTA-only) |