| Brand | Shinsei |
|---|---|
| Origin | Japan |
| Model | XO-326IIs |
| Sensor Type | Galvanic (Diffusion-Type Membrane) |
| Measurement Range | O₂: 0–25.0 vol% |
| Full-Scale Extended Range | 25.1–40.0 vol% |
| Accuracy | ±0.5 vol% |
| Response Time (t₉₀) | ≤20 s |
| Repeatability | ±0.5 vol% (typical) |
| Resolution | 0.1 vol% (3-digit LCD display) |
| Operating Temperature | 0–40 °C |
| Power Supply | Two AA alkaline batteries |
| Battery Life | Approx. 8,000 hours continuous operation |
| Enclosure Rating | Intrinsically Safe Ex ia IIB T3 |
| Dimensions (XO-326IIsB variant) | 66 × 120 × 29 mm |
| Weight (XO-326IIsB) | ~265 g |
| Brand | Hamilton |
|---|---|
| Origin | Switzerland |
| Model | H100 pH |
| Measurement Range | pH −2.00 to 16.00 |
| ORP | −1500 to +1500 mV |
| Temperature Input | Pt100 / Pt1000 / NTC 30 kΩ |
| Temperature Range | −20 to +150 °C |
| Temperature Resolution | 0.1 °C |
| Measurement Error | <1 mV |
| Analog Outputs | Dual 4–20 mA (floating, max 10 V) |
| Relay Output | Configurable (alarm/flush), 3 A / <250 VAC or 30 VDC |
| Housing Material | PBT |
| Enclosure Rating | IP65 / NEMA 4X |
| Power Supply | 24–230 V AC/DC |
| Display | 7-segment LCD with icons |
| Operating Humidity | 0–80% RH at ≤55 °C |
| FM Certification | Class I, Div. 2, Groups A/B/C/D & IIC, T4, Ta = 55 °C, Type 2 |