- All
- Favorite
- Popular
- Most rated
| Brand | ALP |
|---|---|
| Origin | Japan |
| Model | CLG-32L / CLG-40M |
| Chamber Dimensions (Ø × Depth) | 320 × 676 mm (32L) / 400 × 672 mm (40M) |
| Chamber Volume | 54 L / 85 L |
| Sterilization Temp Range | 100–140 °C (CLG-32L), 100–137 °C (CLG-40M) |
| Dissolution/Heating Temp Range | 40–99 °C / 40–60 °C |
| Max Operating Pressure | 0.270 MPa / 0.250 MPa |
| Cycle Time | Sterilization/Dissolution: 0–10 h |
| Holding | 0–48 h |
| Programmable Cycles | 10 stored protocols with mode & data logging |
| Vacuum System | Triple-pulse pre-vacuum with high-capacity vacuum pump |
| Air Filtration | 0.2 µm hydrophobic filter for heated air drying |
| Drying Method | Vacuum-assisted hot-air drying post-sterilization |
| Safety Certification | PED 97/23/EC compliant |
| Construction | Inner chamber – SUS304 stainless steel |
| Power Supply | AC 220–240 V, 1-phase, 50/60 Hz |
| Rated Current | 12 A |
| Heating Power | 2.6 / 2.8 / 2.9 kW (model-dependent) |
| Dimensions (W×D×H) | 550 × 550 × 950 mm (CLG-32L) / 620 × 650 × 990 mm (CLG-40M) |
| Weight | 65 kg / 83 kg |
| Control Interface | Backlit LCD display with real-time monitoring of temperature, pressure, phase status, and elapsed time |
| Door Mechanism | Central lever lock with automatic interlock, motor-assisted hinge, and power-dependent electronic safety lock |
| Cooling | Active rapid cooling to <80 °C post-cycle |
| Drainage | Front-mounted manual drain valve and accessible front-positioned waste tank |
| Optional | Integrated product temperature probe port (probe sold separately) |
| Compliance | Meets IEC 61000-6-3 (EMC) and IEC 61000-6-4 (emission) standards |
| Brand | ALP |
|---|---|
| Origin | Japan |
| Model | CL-32L / CL-40M |
| Chamber Volume | 54 L (CL-32L), 85 L (CL-40M) |
| Max Operating Temperature | 140 °C (CL-32L), 137 °C (CL-40M) |
| Max Working Pressure | 0.27 MPa (CL-32L), 0.25 MPa (CL-40M) |
| Sterilization Temp Range | 40–140 °C (CL-32L), 40–137 °C (CL-40M) |
| Melting/Insulation Range | 40–99 °C / 40–60 °C |
| Cycle Time | Sterilization/Melting: 0–600 min or continuous |
| Insulation | 0–2880 min |
| Drying Temp Range | 60–180 °C |
| Drying Timer | 0–5940 min or continuous |
| Control System | Microprocessor-based with graphical LCD interface |
| Safety Compliance | ASME BPVC Section VIII Div. 1, PED 2014/68/EU, ISO 17665-1, GB/T 30690–2014, China Special Equipment Manufacturing License (Pressure Vessel) |